K230195 · Comerge AG · OLO · Mar 8, 2024 · Neurology
Device Facts
Record ID
K230195
Device Name
Neo ADVISE software
Applicant
Comerge AG
Product Code
OLO · Neurology
Decision Date
Mar 8, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The Neo ADVISE™ System is indicated to calculate the relative positions and distances between implanted Neo Pedicle Screw System™ screws, as estimated by visually identifying screw guides, to estimate rod length, visualize rod positioning, and create a rod template that can be used to visually guide the manually creation a contoured rod by the surgeon, with a mechanical rod bender. The Neo ADVISE™ System can be used in posterior open or percutancous spine surgery in the thoracic, lumbar, and/or sacral spine when using the Neo Pedicle Screw System™.
Device Story
System supports spinal surgery via iPad-based Augmented Reality (AR). Surgeon scans pedicle screw guides intraoperatively using either 3D scanning (overlaying virtual guides) or Marker Detection (using Neo Markers/Base). Software calculates screw head distances, estimates rod length, and displays rod placement relative to screws in coronal/sagittal planes without X-ray. Generates 1:1 scale rod template on screen for manual contouring with mechanical bender. Used in sterile field (iPad in sterile bag) by surgeons. Benefits include reduced intraoperative imaging and improved rod fit accuracy. Final X-ray confirmation required.
Clinical Evidence
No clinical data. Evidence includes two cadaver studies and usability testing. First cadaver study (n=11 surgeons) evaluated percutaneous screw insertion. Second cadaver study (n=4 surgeons) assessed accuracy of screw-rod construction using CBCT, measuring translation of screw heads and angular change (Cobb angle). Usability studies confirmed interface design and identified/mitigated potential user errors.
Technological Characteristics
Software-based stereotaxic instrument running on Apple iPad. Utilizes AR functionality, camera, and sensors for spatial tracking. Employs optical marker detection (Neo Markers/Base) for guide recognition. Standalone system; no external connectivity mentioned. Sterile barrier provided by iPad sterile bag. No specific materials of construction listed for software components.
Indications for Use
Indicated for use in posterior open or percutaneous spine surgery (thoracic, lumbar, sacral) to assist surgeons in calculating screw positions, estimating rod length, and creating rod templates for manual bending when using the Neo Pedicle Screw System.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
March 8, 2024
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left and features a stylized image of a human figure. The FDA logo is on the right and features the letters "FDA" in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
Comerge AG % Frederike Bruhschwein-Mandic Senior Consultant Confinis Ag Hauptrasse 16 Dudingen, 3186 Switzerland
Re: K230195
Trade/Device Name: Neo ADVISE software Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 4, 2024 Received: March 4, 2024
Dear Frederike Bruhschwein-Mandic:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali -S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230195
Device Name Neo ADVISE System
#### Indications for Use (Describe)
The Neo ADVISE™ System is indicated to calculate the relative positions and distances between implanted Neo Pedicle Screw System™ screws, as estimated by visually identifying screw guides, to estimate rod length, visualize rod positioning, and create a rod template that can be used to visually guide the manually creation a contoured rod by the surgeon, with a mechanical rod bender. The Neo ADVISE™ System can be used in posterior open or percutancous spine surgery in the thoracic, lumbar, and/or sacral spine when using the Neo Pedicle Screw System™.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K230195
Neo ADVISE System
## I. SUBMITTER
Applicant Name: Comerge AG Applicant Adress: Bubenbergstrasse 1 Zurich 8045 Switzerland Applicant Contact Telephone: +41445525264 Applicant Contact: Mr. Beat Hugi Applicant Contact Email: beat.huqi@comerge.net Correspondent Name: confinis ag Correspondent Address: Hauptstrasse 16 Dudingen 3186 Switzerland Correspondent Contact Telephone: +4915201892430 Correspondent Contact: Mrs. Frederike Brühschwein-Mandic Correspondent Contact Email: frederike.bruehschwein@confinis.com Date prepared: March 07, 2024
## II. DEVICE
Device Trade Name: Neo ADIVSE system Common Name: Stereotaxic instrument Classification Name Orthopedic Stereotaxic Instrument Regulation Number: 882.4560 Product Code: OLO
## III. PREDICATE DEVICE
Predicate #: K122742
Predicate Trade Name: NuVasive Bendini Product Code: OLO
## IV. DEVICE DESCRIPTION
The Neo ADVISE system is a tool that supports the surgeon in open and percutaneous spinal surgeries performed with the Neo Medical Pedicle Screw System (K222256). After the placement of the pedicle screws, the software scans the the screw head positions and shows them in
{4}------------------------------------------------
Augmented Reality (AR) on an iPad. Based on the scanned screw positions, the software displays the distance between the screw heads and calculates the minimum rod length. The surgeon can select any of the pre-bent or straight rods of the Neo PSS and check the placement of the rod in the patient's body without touching the body or taking X-Ray images. The software calculates the distance between the screw head and the rod in the coronal and sagittal plane for the selected rod, indicating possible screw head adjustments.
Again, based on the scanned screw position, the software can generate a custom bent rod template that follows the path through all screw heads. This provides the surgeon with a rod template that assists the surgeon in rod bending. The rod template is displayed in a 1-1 scale on the device in the sagittal and coronal plane, allowing the surgeon to copy the rod shape according to the template.
The scanning of the screw heads can be performed using two distinct scanning methods, 3D scanning and Marker detection. 3D Scanning does not require any additional hardware. The surgeon scans each guide separately by overlaying the physical guide with a virtual guide in AR. 3D Scanning of the guides is limited to 10 guides. The Marker Detection scanning method allows the scanning of up to 20 guides. This is achieved by the use of the Neo Markers and the Neo Marker Base. The Neo Markers are attached to each guide, providing the software with a defined pattern, similar to QR codes, that can be recognized by the software.
The Neo ADVISE System consists of several components. The software runs on an Apple iPad, which is put into a sterile bag to allow use in the sterile environment of the patient's area. The Neo Markers and the Neo Marker Base are needed to allow the scanning of the guides using the Marker Detection scanning method.
#### V. INDICATIONS FOR USE
The Neo ADVISE™ System is indicated to calculate the relative positions and distances between implanted Neo Pedicle Screw System™ screws, as estimated by visually identifying screw quides, to estimate rod length, visualize rod positioning, and create a rod template that can be used to visually quide the manually creation a contoured rod by the surqeon, with a mechanical rod bender. The Neo ADVISE™ System can be used in posterior open or percutaneous spine surgery in the thoracic, lumbar, and/or sacral spine when using the Neo Pedicle Screw System™.
## VI. INDICATIONS FOR USE COMPARISON
Both the Neo ADVISE System and the predicate Bendini Spinal Rod Bending System are used to locate the screw position during spinal surgery and relative to the identified positions, generate bend instructions to shape a spinal rod.
The Neo ADVISE System is intended to be used with off-the-shelf rod benders while the Bendini system has a dedicated rod bending instrument.
The Bendini system can be used with any pedicle screw system while the Neo ADVISE System is dedicated for use with the Neo Pedicle Screw System (K222256) only.
Like the NuVasive Bendini Spinal Rod Bending System the Neo ADVISE System is intraoperatively digitizing the relative location of anatomical landmarks (i.e. the position of the pedicle screws), through the use of a camera and markers that are pre-fixed on the screw guides (same as the pointer stylus with NuVasive Bendini).
The differences in the indications for use are not critical to the intended surgical use of the devices as in both cases, a final intra-operative X-ray must be taken to confirm the spinal rod placement at the end of the procedure.
## VII. TECHNOLOGICAL COMPARISON
Augmented reality and optical/electromagnetic localizer systems share similarities in their use during orthopedic surgery. The Neo ADVISE System's software utilizes the augmented reality functionality of an Apple iPad to scan the position of guides and pedicle screws, without the need for traditional tracking arrays or pre-operative data planning. Instead, markers may be used to
{5}------------------------------------------------
enhance object recognition. The Apple iPad's camera and sensors provide the interface for communication between the user and the software.
Similarly, the predicate device cleared under K122742 also uses a localized pointer or stylus to intraoperatively digitize the relative location of anatomical landmarks, which are then interpreted by proprietary algorithms, in place of the stereotaxic principle.
#### VIII. PERFORMANCE DATA
Non-clinical testing included two cadaver studies to evaluate the Neo ADVISE System, a usability study including a follow-up study to evaluate the human factors/usability aspects of the Neo ADVISE System, and bench testing.
The first cadaver study, conducted by 11 German orthopedic and trauma surgeons, utilized a biomechanical pilot study design with a single human cadayeric spine specimen. One investigator inserted 6 polyaxial pedicle screws (Neo Pedicle Screw SystemTM) percutaneously at levels L1-S1 on one side of the spine, resulting in instrumentation across 5 levels for fusion.
The second cadaver study aimed to assess the accuracy and effectiveness of the Neo ADVISE™ System in assembling and fixing posterior screw-rod constructions in thoraco-lumbo-sacral spine fusion surgery. Using CBCT (Cone-beam computed tomography systems) scans, primary endpoints measured translation of screw heads along sagittal and transverse axes, as well as angular change (Cobb angle) along sagittal and coronal planes. Four German spine surgeons conducted the study, demonstrating that AR-guided surgery maintained anatomy during rod placement and fixation.
The usability studies of the Neo ADVISE System's device-user interface, focused on identifying preventable errors attributable to device design or labeling and affirmed the usability of the Neo ADVISE™ system.
No clinical tests have been submitted.
#### IX. CONCLUSION
Based on equivalence of indications for use, technology characteristics and operational principle the substantial equivalence between the new and the predicate device has been demonstrated. The Neo ADVISE System, is at least as safe and effective as the legally marketed predicate Nuvasive Bendini Spinal Rod Bending System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.