BonOs® Inject Bone Cement; NEO Pedicle Screw System™
K222256 · Neo Medical SA · PML · Aug 26, 2022 · Orthopedic
Device Facts
Record ID
K222256
Device Name
BonOs® Inject Bone Cement; NEO Pedicle Screw System™
Applicant
Neo Medical SA
Product Code
PML · Orthopedic
Decision Date
Aug 26, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NEO Pedicle Screw System™, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. The system is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (scolosis, and/or lordosis), turnor, pseudarthrosis, and/ or failed previous fusion. When used in conjunction with BonOs® Inject Cement, the NEO Pedicle Screw System™ is intented to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NEO Pedicle Screws augmented with BonOs® Inject Cement are for use with 5 mm to 8 mm screw diameters at spinal levels where the structural integrity of the spine is not severely compromised.
Device Story
System comprises pedicle screws, rods, and connectors for posterior spinal fixation; includes BonOs® Inject bone cement for vertebral augmentation. Screws/rods/connectors implanted by surgeons in clinical settings to stabilize spinal segments or restore integrity in tumor patients. Cement (PMMA powder/liquid) mixed to form injectable paste; injected into vertebral body; cures via exothermic polymerization to stabilize lesions/fractures. Screws (titanium alloy) and rods (titanium/cobalt chrome) provide mechanical fixation. System includes single-use sterile instruments. Output is mechanical stabilization of spine; aids clinical decision-making by providing structural support in degenerative or oncological cases. Benefits include pain relief, fracture stabilization, and spinal column integrity restoration.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including static and dynamic mechanical testing and axial gripping capacity in accordance with ASTM 1798.
Technological Characteristics
Materials: Titanium alloy (screws, connectors), titanium or cobalt chrome alloy (rods), PMMA/Zirconium dioxide (bone cement). Dimensions: Pedicle screws 4.5–8.0 mm diameter, 25–55 mm length; Iliac screws 8.0 mm diameter, 70–100 mm length; rods 5.5 mm diameter. Energy: Mechanical fixation. Sterilization: Validated process (cleared under K212489). Connectivity: None. Software: None.
Indications for Use
Indicated for skeletally mature patients requiring posterior, non-cervical spinal immobilization/stabilization as an adjunct to fusion for degenerative disc disease, spondylolisthesis, trauma, spinal stenosis, curvatures, tumor, pseudarthrosis, or failed fusion. Also indicated for use with BonOs® Inject Cement to restore spinal integrity in patients with advanced stage thoracic/lumbar tumors where fusion is not feasible due to limited life expectancy.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Predicate Devices
BonOs® Inject Bone Cement, NEO Pedicle Screw System™ (K212489)
Submission Summary (Full Text)
{0}------------------------------------------------
August 26, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Neo Medical SA % Sandra Soniec Managing Director meditec Consulting GmbH Obermoosstrasse 23 Boll, Berne 3067 Switzerland
Re: K222256
Trade/Device Name: BonOs® Inject Bone Cement; NEO Pedicle Screw System™ Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) Bone Cement Regulatory Class: Class II Product Code: PML, NDN, NKB Dated: July 15, 2022 Received: July 27, 2022
Dear Sandra Soniec:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222256
Device Name NEO Pedicle Screw System™
#### Indications for Use (Describe)
The NEO Pedicle Screw System™, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. The system is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (scolosis, and/or lordosis), turnor, pseudarthrosis, and/ or failed previous fusion.
When used in conjunction with BonOs® Inject Cement, the NEO Pedicle Screw System™ is intented to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NEO Pedicle Screws augmented with BonOs® Inject Cement are for use with 5 mm to 8 mm screw diameters at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:flex; align-items:center;"><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K22256
Device Name BonOs® Inject
Indications for Use (Describe)
BonOs® Inject bone cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using vertebroplasty or balloon kyphoplasty procedure.
When used in conjunction with NEO Pedicle Screw System™ BonOs® Inject is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NEO Pedicle Screws augmented with BonOs® Inject Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following information is provided for the BonOs® Inject Bone Cement, NEO Pedicle Screw System™.
### ADMINISTRATIVE INFORMATION
| Date prepared | July 22, 2022 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Submission type | Special 510(k) |
| Purpose of 510(k) | Modification to currently marketed NEO Pedicle Screw<br>System™, additional Iliac Connector |
| Submitter | Neo Medical S.A.<br>Route de Lausanne 157 A<br>1096 Villette (Lavaux), Switzerland |
| Official Correspondent | Sandra Soniec<br>Senior consultant, meditec Consulting GmbH<br>Phone +41 31 535 3193<br>Email soniec@meditec-consulting.ch |
| US agent | Vicky Verna, confinis corporation<br>Phone +1 240 252 0891<br>Email vicky.verna@confinis.com |
# DEVICE NAME AND CLASSIFICATION
| Trade name | 1) BonOs® Inject Bone Cement,<br>2) NEO Pedicle Screw System™ |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name | 1) PMMA bone cement for vertebroplasty<br>2) Pedicle Screw System |
| Regulation number,<br>Regulation name,<br>Regulatory class, Product<br>Codes | 1) BonOs® Inject Bone Cement<br>21 CFR 888.3027<br>Polymethylmethacrylate (PMMA) bone cement<br>Class II<br>PML, NDN<br><br>2) NEO Pedicle Screw System™<br>21 CFR 888.3070<br>Thoracolumbosacral pedicle screw system<br>Class II<br>NKB |
{5}------------------------------------------------
# PREDICATE DEVICES
BonOs® Inject Bone Cement, NEO Pedicle Screw System™ (K212489)
### INDICATIONS FOR USE
### 1) BonOs® Inject
BonOs® Inject bone cement is indicated for the treatment of pathological fractures of the vertebral body due to osteoporosis, cancer, or benign lesions using vertebroplasty or balloon kvphoplastv procedure.
When used in coniunction with NEO Pedicle Screw System™ BonOs® Inject is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NEO Pedicle Screws augmented with BonOs® Inject Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
#### 2) NEO Pedicle Screw System™
The NEO Pedicle Screw System™, when used as a posterior pedicle screw system, is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion. The system is intended for posterior, non-cervical fixation for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (scoliosis, kyphosis, and/or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.
When used in conjunction with BonOs® Inject Cement, the NEO Pedicle Screw System™ is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. NEO Pedicle Screws augmented with BonOs® Inject Cement are of the spine is not severely compromised. for use with 5 mm to 8 mm screw diameters at spinal levels where the structural integrity of the spine is not severely compromised.
# DEVICE DESCRITION
#### 1) BonOs® Inject
BonOs® Inject is a radiopaque, injectable bone cement for use in spine surgery like percutaneous vertebral augmentation during vertebroplasty or kyphoplasty. It is a twocomponent system consisting of a powder and a liquid. Methylmethacrylate polymer is the primary constituent of the powder component. Zirconium dioxide is added as radiopacifier. Methylmethacrylate monomer is the primary constituent of the liquid component. Mixing the two separate sterile components, initially an injectable paste is produced which can be transferred into a svringe and which then can be iniected under slight pressure into the vertebral body. After curing of the bone cement by exothermic polymerization it stabilizes the vertebral lesions and vertebral compression fractures.
{6}------------------------------------------------
# 2) NEO Pedicle Screw System™
The NEO Pedicle Screw System™ consists of screws, rods and connectors which are available in different sizes. The system includes the relevant instruments which are mainly single use, disposable and delivered sterile, just few optional instruments are reusable and delivered non-sterile.
All the system components are made of materials compliant with ASTM and/or ISO standards. The screws are made out of a titanium alloy and delivered pre-mounted to a screw extender including a tissue dilator and sterile. The rods are made out of a titanium alloy or cobalt chrome alloy and delivered sterile. Connectors are made out of titanium allov and delivered sterile. Iliac connectors are being added in this special 510(k).
The pedicle screws are offered in diameters of 4.5 – 8.0 mm and lengths of 25 - 55 mm. Illiac screws are offered in diameters of 8.0 mm and lengths of 70 – 100 mm. Three different types of rods are available: pre-bent, straight or special bent rod for S1/L5. All rods have a diameter of 5.5 mm. Pre-bent rods are offered in lengths of 40 - 100 mm, straight rods in lengths from 30 - 500 mm and the special-bent rod in either 30 or 40 mm length.
### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS AND COMPARISON TO PREDICATE DEVICE
The modified NEO Pedicle Screw System™ is substantially equivalent to the currently cleared device in terms of intended use, material, design, mechanical properties and function.
Non-clinical performance testing demonstrate that the NEO Pedicle Screw System™ meets the requirements for Pedicle screw spinal systems according to Spinal System 510(k)s Guidance for Industry and FDA Staff Document issued on: May 3, 2004 and is substantially equivalent to its predicate devices.
# SUMMARY OF PERFORMANCE DATA
To demonstrate that the modified NEO Pedicle Screw System™ is substantially equivalent to a cleared predicate, static and dynamic testing and axial gripping capacity in accordance with ASTM 1798.
The new system component has been evaluated in context of K212489 with regards to Usability, Biocompatibility testing, MRI safety and compatibility and is sterilized using the currently validated sterilization process cleared under K212489. No clinical studies were conducted.
# CONCLUSION
Based on the indications for use, technological characteristics, and comparison to predicate device, the subject modified BonOs® Inject Bone Cement, NEO Pedicle Screw System™ has been shown to be substantially equivalent to currently cleared device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.