Terran NM-101 is a post-processing software medical device intended for use in visualization of the brain in older adults between the age of 51 to 83 years. Terran NM-101 analyzes input data acquired from Siemens 3T MR imaging systems, using a specified protocol (i.e., Turbo Spin Echo, TSE). Terran NM-101 can generate qualitative parametric maps from non-contrast T1-weighted image MR acquisition. Terran NM-101 is also intended for automatic labeling, visualization, volumetric quantification and contrast quantification of segmentable brain tissues from a set of Siemens 3T acquired MR images. Brain tissue characterization and volumes of neuromelanin associated signal contrasts are determined based on analysis of qualitative parametric maps. When interpreted by a neuroradiologist, Terran NM-101 images can provide information useful in determining neuromelanin association as an adjunct to diagnosis. Terran NM-101 must always be used in combination with a T1-weighted image MR acquisition.
Device Story
Terran NM-101 is a fully automated post-processing SaMD; analyzes non-contrast T1-weighted MRI data from Siemens 3T scanners (TSE protocol). Device generates qualitative parametric maps; performs automatic labeling, volumetric quantification, and contrast quantification of segmentable brain tissues. Specifically measures neuromelanin-associated contrast-to-noise ratio (CNR) in substantia nigra (SN) and locus coeruleus (LC) regions. Used by neuroradiologists in clinical settings; output provided as DICOM images/reports for display on PACS/workstations. Neuroradiologist interprets output to assist in clinical diagnosis. Benefits include standardized, automated quantification of neuromelanin-associated signals to support diagnostic decision-making.
Clinical Evidence
No clinical studies were required. Evidence consists of non-clinical design verification, system integration testing, performance testing (V&V), and simulated use testing. Clinical performance reviews by neuroradiologists confirmed that user needs and intended use requirements were fulfilled and all acceptance criteria were met.
Technological Characteristics
Post-processing SaMD; operates on cloud-based, Windows, or Linux platforms. Inputs: Siemens 3T MRI (TSE protocol). Outputs: DICOM parametric maps, volumetric/contrast quantification reports. Features automated quality control (input check, protocol verification, brain alignment, NM-associated CNR check, PHI check, and cybersecurity threat assessment).
Indications for Use
Indicated for visualization of the brain in adults aged 51-83 years. Used as an adjunct to diagnosis for determining neuromelanin association when interpreted by a neuroradiologist. Requires use with T1-weighted MR acquisition.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
September 29, 2023
Terran Biosciences, Inc. % Kay Fuller Principal Regulatory & Clinical Research Consultant Medical Device Regulatory Solutions, LLC 230 Collingwood Dr., Suite 260 Ann Arbor, Michigan 48103
Re: K230187
Trade/Device Name: Terran NM-101 Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH, LLZ Dated: August 26, 2023 Received: August 29, 2023
Dear Kay Fuller:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Daniel M. Krainak, Ph.D. Assistant Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230187
Device Name Terran NM-101
#### Indications for Use (Describe)
Terran NM-101 is a post-processing software medical device in visualization of the brain in older adults between the age of 51 to 83 years. Terran NM-101 analyzes input data acquired from Siemens 3T MR imaging systems, using a specified protocol (i.e., Turbo Spin Echo, TSE). Terran NM-101 can generate qualitative parametric maps from non-contrast T1-weighted image MR acquisition.
Terran NM-101 is also intended for automatic labeling, volumetric quantification and contrast quantification of segmentable brain tissues from a set of Siemens 37 acquired MR images. Brain tissue characterization and volumes of neuromelanin associated signal contrasts are determined based on analysis of qualitative parametric maps.
When interpreted by a neuroradiologist. Terran NM-101 images can provide information useful in determining neuromelanin association as an adjunct to diagnosis.
Terran NM-101 must always be used in combination with a T1-weighted image MR acquisition.
Type of Use (*Select one or both, as applicable*)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
## 510(k) SUMMARY
### Terran Biosciences, Inc. Terran NM-101
#### September 19, 2023
The following summary is provided pursuant to Section 513 (I) (3) (A) of the Federal Food Drug and Cosmetic Act:
## 1. GENERAL INFORMATION
| Submitter Information: | Terran Biosciences, Inc.<br>2457 Collins Ave., Suite 504<br>Miami Beach, FL 33140 |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Information: | Kay Fuller, RAC<br>Principal Regulatory & Clinical Research Consultant<br>Medical Device Regulatory Solutions, LLC<br>734-846-7852 |
| 2. DEVICE INFORMATION | |
| Device Name: | Terran NM-101 |
| Proprietary Name: | Terran NM-101 |
| Common Name: | System, Nuclear Magnetic Resonance Imaging |
| Classification Name: | Magnetic Resonance Diagnostic Device |
| Classification Code: | LNH, LLZ |
| Regulation Number: | 21 CFR §892.1000 |
| 3. PREDICATE DEVICE(S) | The Terran NM-101 is similar to the primary predicate device<br>SyMRI cleared for US commercialization via K191036, on<br>6/13/2019 and predicate device NeuroQuant cleared for US<br>commercialization via K170981, on 9/7/2017. |
| 4. DEVICE DESCRIPTION | Terran NM-101 is a fully automated post-acquisition software as a<br>medical device (SaMD) that measures neuromelanin associated<br>contrast to noise ratio (CNR) signal in the substantia nigra (SN) and<br>locus coeruleus (LC) regions of non-contrast brain magnetic<br>resonance imaging (MRI) images from Siemens 3 tesla (3T) MRI<br>scanners. Terran NM-101 can generate qualitative parametric<br>maps from non-contrast T1-weighted image MR acquisition. |
{4}------------------------------------------------
5. INDICATIONS FOR USE Terran NM-101 is a post-processing software medical device intended for use in visualization of the brain in older adults between the age of 51 to 83 years. Terran NM-101 analyzes input data acquired from Siemens 3T MR imaging systems, using a specified protocol (i.e., Turbo Spin Echo, TSE). Terran NM-101 can generate qualitative parametric maps from non-contrast T1-weighted image MR acquisition.
> Terran NM-101 is also intended for automatic labeling, visualization, volumetric quantification and contrast quantification of seqmentable brain tissues from a set of Siemens 3T acquired MR images. Brain tissue characterization and volumes of neuromelanin associated signal contrasts are determined based on analysis of qualitative parametric maps. When interpreted by a neuroradiologist, Terran NM-101 images can provide information useful in determining neuromelanin association as an adjunct to diagnosis.
> Terran NM-101 must always be used in combination with a T1weighted image MR acquisition.
The Terran NM-101 fundamental technological characteristics are similar to those of the predicate devices as described herein, and as described in the following table.
| Feature<br>Comparison<br>Criteria | Subject Device<br>Terran NM-101<br>K230187 | Primary Predicate Device A<br>K191036<br>SyMRI | Subject Device<br>SE to K191036? | Predicate Device B<br>K170981<br>NeuroQuant | Subject Device<br>SE to K173224? |
|---------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| 21 CFR Reg #,<br>Product Code &<br>Classification | 21 CFR §892.1000<br>21 CFR §892.2050<br>LNH / LLZ<br>Class II | 21 CFR §892.1000<br>LNH<br>Class II | Yes | 21 CFR §892.2050<br>LLLZ<br>Class II | Yes |
| Regulation Name | Magnetic resonance diagnostic device;<br>Picture archiving and communication system | Magnetic resonance diagnostic device | Yes | Picture archiving and<br>communications<br>system | Yes |
| Prescription Device<br>Rx Only | Yes | Yes | Yes | Yes | Yes |
| Indications for Use | Terran NM-101 is a post-processing software<br>medical device intended for use in visualization<br>of the brain in older adults between the age of<br>51 to 83 years. Terran NM-101 analyzes input<br>data acquired from Siemens 3T MR imaging<br>systems, using a specified protocol (i.e., Turbo<br>Spin Echo, TSE). Terran NM-101 can generate<br>qualitative parametric maps from non-contrast<br>T1-weighted image MR acquisition. Terran NM-<br>101 is also intended for automatic labeling,<br>visualization, volumetric quantification and<br>contrast quantification of segmentable brain<br>tissues from a set of Siemens 3T acquired MR<br>images. Brain tissue characterization and<br>volumes of neuromelanin associated signal<br>contrasts are determined based on analysis of<br>qualitative parametric maps. When interpreted<br>by a neuroradiologist, Terran NM-101 images<br>can provide information useful in determining<br>neuromelanin association as an adjunct to<br>diagnosis. Terran NM-101 must always be used<br>in combination with a T1-weighted image MR<br>acquisition. | SyMRI is a post-processing software medical<br>device intended for use in visualization of the<br>brain. SyMRI analyzes input data from MR<br>imaging systems. SyMRI utilizes data from a<br>multi-delay, multi-echo acquisition (MDME)<br>to generate parametric maps of R1, R2<br>relaxation rates, and proton density (PD).<br>SyMRI can generate multiple image<br>contrasts from the parametric maps.<br>SyMRI enables post-acquisition image<br>contrast adjustment. SyMRI is indicated for<br>head imaging. SyMRI is also intended for<br>automatic labeling, visualization and<br>volumetric quantification of segmentable<br>brain tissues from a set of MR images. Brain<br>tissue volumes are determined based on<br>modeling of parametric maps from MDME.<br>When interpreted by a trained physician,<br>SyMRI images can provide information useful<br>in determining diagnosis. SyMRI should<br>always be used in combination with at least<br>one other MR acquisition (e.g., T2-FLAIR). | Yes | NeuroQuant is intended<br>for automatic labeling,<br>visualization and<br>volumetric quantification<br>of segmentable brain<br>structures and lesions<br>from a set of MR<br>images. Volumetric<br>measurements may be<br>compared to reference<br>percentile data. | Yes |
| Intended Users | Qualified Radiologist | Qualified Radiologist | Yes | Qualified Radiologist | Yes |
| Feature | Subject Device | Primary Predicate Device A | Subject Device | Predicate Device B | Subject Device |
| Comparison<br>Criteria | Terran NM-101<br>K230187 | K191036<br>SyMRI | SE to<br>K191036? | K170981<br>NeuroQuant | SE to<br>K173224? |
| Type of Imaging Scans | MRI | MRI | Yes | MRI | Yes |
| Target Organ/System | MR Brain | MR Brain | Yes | MR Brain | Yes |
| Handle Multiple Studies | Yes | Yes | Yes | Yes | Yes |
| Field Strength | No : 1.5T<br>Yes : 3.0T | No: 1.5T<br>Yes: 3.0T | Yes | Yes: 1.5T<br>No: 3.0T | Yes |
| Mode | 3D | 3D | Yes | 3D | |
| Plane | Sagittal | Sagittal | Yes | Sagittal | Yes |
| Contrast Enhancement | T1 | T1 | Yes | T1 | Yes |
| Sequence | MPRAGE | MPRAGE | Yes | MPRAGE | Yes |
| Slice Thickness | 1.8 mm | Unknown | No | 1.2 mm | Yes |
| TR | 600 ms | 650 ms | Yes | 2300 ms | Yes |
| TE | 10 ms | 10 ms | Yes | Minimum | Yes |
| TI | ~ | ~ | Yes | 900 ms | Yes |
| Acquisition Time | 8 mins | 6 mins | Yes | 5-7 mins | Yes |
| Flip Angle | 120° | Unknown | No | 9° | Yes |
| Filter | Non | Non | Yes | Non | Yes |
| Data Source | Siemens 3T MRI scanner: T1 MRI<br>scans acquired with specified protocols<br>(i.e., TSE) | MRI scanner: 3D T1 MRI scans acquired<br>with specified protocols (i.e., MDME - GE<br>MAGiC, Philips SyntAc; Siemens 3 T<br>TSE_MDME) | Yes | MRI scanner: 3D T1 MRI scans<br>acquired with specified protocols | Yes |
| | | | | NeuroQuant Supports DICOM<br>format as input | |
| Operating System | Terran NM-101 Supports DICOM<br>format as input<br>Cloud-Based, Windows, Linux…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.