T2 Tibial Nailing System, T2 Femoral Nail System, T2 Supracondylar Nail System, T2 Recon Nail System, T2 Greater Trochanter Nail (GTN),T2 Ankle Arthrodesis Nail,T2 Arthrodesis Nail System
T2 Tibial Nailing System, T2 Femoral Nail System, T2 Supracondylar Nail System, T2 Recon Nail System, T2 Greater Trochanter Nail (GTN),T2 Ankle Arthrodesis Nail,T2 Arthrodesis Nail System
Applicant
Stryker GmbH
Product Code
HSB · Orthopedic
Decision Date
Jul 1, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The T2 Tibial Nailing System is intended to provide temporary stabilization of various types of fractures, malunions and nonunion of the tibia. The nails are inserted or closed technique and can be statically, dynamically and compressed locked.
Device Story
T2 Nail Systems are titanium alloy intramedullary fixation devices used for stabilization of long bone fractures and arthrodesis. The system includes various nail designs (tibial, femoral, supracondylar, recon, greater trochanter, ankle arthrodesis, and arthrodesis nails) and associated locking screws, compression screws, and end caps. Devices are inserted into the medullary canal via open or closed surgical techniques by orthopedic surgeons. The nails provide mechanical stabilization of bone segments to facilitate healing. The current submission establishes compatibility between existing T2 nails and the IMN Screws System (K172774). The device is used in clinical settings (OR) to treat fractures, malunions, and nonunions, benefiting patients by providing rigid internal fixation to restore bone alignment and stability.
Clinical Evidence
No clinical data. Bench testing only, including functional interface analysis and computational assessment of construct strength comparing T2 nails with IMN Locking Screws versus original T2 locking screws.
Technological Characteristics
Titanium alloy intramedullary nails; cannulated and solid designs; various diameters and lengths; locking/compression screws and end caps. Mechanical fixation via intramedullary rod. Single-use. No software or electronic components.
Indications for Use
Indicated for long bone fracture fixation, including open/closed fractures, pseudoarthrosis, correction osteotomy, pathologic fractures, tumor resections, nonunion, and malunion in the tibia, femur, and hindfoot/ankle (for arthrodesis).
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Stryker GmbH Jackie Perri Senior Regulatory Affairs Specialist 325 Corporate Drive MAHWAH, NJ 07430
July 01, 2020
Re: K200880
Trade/Device Name: T2 Tibial Nailing System, T2 Femoral Nail System, T2 Supracondylar Nail System. T2 Recon Nail System, T2 Greater Trochanter Nail (GTN), T2 Ankle Arthrodesis Nail, T2 Arthrodesis Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: April 1, 2020 Received: April 2, 2020
Dear Jackie Perri:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorywebsite the DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael Owens Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K200880
Device Name T2 Tibial Nailing System
Indications for Use (Describe)
The T2 Tibial Nailing System is intended to provide temporary stabilization of various types of fractures, malunions and nonunion of the tibia. The nails are inserted or closed technique and can be statically, dynamically and compressed locked.
The T2 Tibial Nailing System is indicated for long bone fracture fixation, specifically tibial fracture fixation, which may include the following:
- · Open and closed tibial fractures
- · Pseudoarthrosis and correction osteotomy
- · Pathologic fractures, impending pathologic fractures, and tumor resections
- · Nonunion and malunion
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
{3}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K200880
Device Name T2 Femoral Nail System
Indications for Use (Describe)
The T2 Femoral Nail is indicated for long bone fracture fixation specifically femoral fracture fixation which may include the following:
- · Open and closed femoral fractures
- · Pseudoarthrosis and correction osteotomy
- · Pathologic fractures, impending pathologic fractures, and tumor resections
- · Supracondylar fractures, including those with intra-articular extension
- Ipsilateral femur fractures
- · Fractures proximal to a total knee arthroplasty
- · Fractures distal to hip joint
- Nonunions and malunions
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
Expiration Date: 06/30/2020
See PRA Statement below.
Form Approved: OMB No. 0910-0120
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K200880
Device Name T2 Supracondylar Nail System
Indications for Use (Describe)
The T2 Supracondylar Nail System is indicated for:
- · Open and closed femoral fractures
- · Pseudoarthrosis and correction osteotomy
- · Pathologic fractures, impending pathologic fractures, and tumor resections
- · Supracondylar fractures including those with intra-articular extension
- · Fractures distal to total hip prosthesis
- Nonunions and malunions
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden. to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
Expiration Date: 06/30/2020
See PRA Statement below.
Form Approved: OMB No. 0910-0120
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K200880
#### Device Name T2 Recon Nail System
## Indications for Use (Describe)
The T2 Recon Nail is indicated for:
- Subtrochanteric fractures
- Intertrochanteric fractures ●
- Ipsilateral neck/shaft fractures
- Comminuted proximal femoral shaft fractures
- Femoral fixation required as a result of pathological disease
- Temporary stabilization of fractures of the femoral shaft ranging from the femoral neck to the ● supracondylar regions of the femur.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120 |
|-------------------------------------------------------------------------|---------------------------------------------------------|
| Indications for Use | Expiration Date: 06/30/2020<br>See PRA Statement below. |
| 510(k) Number ( <i>if known</i> ) | K200880 |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Device Name | T2 Greater Trochanter Nail (GTN) |
| Indications for Use ( <i>Describe</i> ) | The T2 Greater Trochanter Nail is indicated for long bone fracture fixation, which may include the following: |
- Open and closed femoral fractures
- Pseudarthrosis and correction osteotomy
- · Pathologic fractures, impending pathologic fractures, and tumor resections
- · Ipsilateral femur fractures
- · Fractures proximal to a total knee arthroplasty
- · Nonunions and malunions.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart
# C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{7}------------------------------------------------
#### FORM FDA 3881 (7/17)
443-6740
EF
PSC Publishing Services (301)
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K200880
#### Device Name
T2 Ankle Arthrodesis Nail
Indications for Use (Describe)
The T2 Ankle Arthrodesis Nail is intended for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include:
- · Post-traumatic or primary Arthrosis
- · Previously infected arthrosis (second degree)
- · Revision of Failed Ankle Arthrodesis
- · Failed Total Ankle Replacement
- · Avascular Necrosis of the Talus (requiring tibiocalcaneal arthrodesis)
- Neuroarthropathy or Neuromuscular Deformity or other neuromuscular disease with severe deformity or instability of the ankle
- Rheumatoid Arthritis with severe deformity such as rheumatoid hindfoot
- · Osteoarthritis
- Nonunions or Pseudarthrosis of hindfoot and distal tibia
- · Malunited tibial pilon fracture
- · Charcot foot
- · Severe endstage degenerative arthritis
- · Severe defects after tumor resection
- Pantalar arthrodesis
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
Over-The-Counter Use (21 CFR 801
Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
> > Page 6 of 7
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{8}------------------------------------------------
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber." Page 1 of 1 FORM FDA 3881 (7/17) PSC Publishing Services (301)443-6740 PI Form Approved: OMB No. 0910-0120 DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Expiration Date: 06/30/2020 Indications for Use See PRA Statement below. 510(k) Number (if known) K200880 Device Name T2 Arthrodesis Nail System Indications for Use (Describe) The T2 Arthrodesis Nail is intended for long bone internal fixation, which may include the following: Aseptic failed total knee arthroplasty ● Open and closed femoral fractures ● ● Pseudoarthrosis and correction osteotomy Pathological fractures, impending pathological fractures, and tumor resections ● Ipsilateral femur fractures ●
- Failed external fixation, nonunions and malunions
- Periarticular fractures where repair is not possible ●
- Knee arthrodesis
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
]Over-The-Counter Use (21 CFR 801
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An aqency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber."
FORM FDA 3881 (7/17)
Page 7 of 7 EF
PSC Publishing Services (301)
443-6740
{9}------------------------------------------------
Stryker GmbH T2 Nail Systems
Traditional 510(k) Premarket Notification
# 510(k) Summary
| Sponsor: | Stryker GmbH |
|-------------------------|-----------------------------------------------------------------------------------------|
| | Bohnackerweg 1 |
| | 2545 Selzach / Switzerland |
| Contact Person: | Jackie Perri |
| | Senior Regulatory Affairs Specialist |
| | 325 Corporate Drive |
| | Mahwah, NJ 07430 |
| | Phone: (732) 770-5616 |
| Date Prepared: | March 31, 2020 |
| Proprietary Name: | T2 Tibial Nailing System |
| | T2 Femoral Nail System |
| | T2 Supracondylar Nail System |
| | T2 Recon Nail System |
| | T2 Greater Trochanter Nail (GTN) |
| | T2 Ankle Arthrodesis Nail |
| | T2 Arthrodesis Nail System |
| Common Name: | T2 Tibial Nailing System |
| | Tibial Nail |
| | T2 Femoral Nail System |
| | Intramedullary Nail, Femoral Nail |
| | T2 Supracondylar Nail System |
| | Intramedullary Nail |
| | T2 Recon Nail System |
| | Intramedullary Nail |
| | T2 Greater Trochanter Nail (GTN) |
| | Intramedullary Nail, Femoral Nail |
| | T2 Ankle Arthrodesis Nail |
| | Intramedullary Nail |
| | T2 Arthrodesis Nail System |
| | Intramedullary Nail |
| Regulation Number/Name: | T2 Tibial Nailing System |
| | 21CFR 888.3020 (Intramedullary Fixation Rod) |
| | T2 Femoral Nail System |
| | 21CFR 888.3020 (Intramedullary Fixation Rod) |
| | T2 Supracondylar Nail System |
| | 21CFR 888.3020 (Intramedullary Fixation Rod) |
| | T2 Recon Nail System |
| | 21CFR 888.3020 (Intramedullary Fixation Rod) |
| | T2 Greater Trochanter Nail (GTN)<br>21CFR 888.3020 (Intramedullary Fixation Rod) |
| | T2 Ankle Arthrodesis Nail<br>21CFR 888.3020 (Intramedullary Fixation Rod) |
| | T2 Arthrodesis Nail System<br>21CFR 888.3020 (Intramedullary Fixation Rod) |
| Product Code: | T2 Tibial Nailing System<br>HSB (Rod, fixation, intramedullary and accessories) |
| | T2 Femoral Nail System<br>HSB (Rod, fixation, intramedullary and accessories) |
| | T2 Supracondylar Nail System<br>HSB (Rod, fixation, intramedullary and accessories) |
| | T2 Recon Nail System<br>HSB (Rod, fixation, intramedullary and accessories) |
| | T2 Greater Trochanter Nail (GTN)<br>HSB (Rod, fixation, intramedullary and accessories) |
| | T2 Ankle Arthrodesis Nail<br>HSB (Rod, fixation, intramedullary and accessories) |
| | T2 Arthrodesis Nail System<br>HSB (Rod, fixation, intramedullary and accessories) |
| Device Class: | Class II |
| Predicate Device: | T2 Tibial Nailing System – K003018 and K131365 |
| | T2 Femoral Nail System - K081152, K021744 and K01080 |
| | Synthes Solid Femoral Nail - K923580 |
| | T2 Supracondylar Nail System – K023267 |
| | T2 Recon Nail System - K102992 and K032898 |
| | T2 Greater Trochanter Nail (GTN) – K101438 |
| | T2 Ankle Arthrodesis Nail - K051590 |
| | T2 Arthrodesis Nail System - K020384 |
{10}------------------------------------------------
# Description
#### T2 Tibial Nailing System
The T2 Tibial Nailing System is a cylindrical tube manufactured from titanium alloy and slightly bowed to accommodate the shape of the tibia. Locking screws, compression screws and an end cap are manufactured from titanium alloy and are used with the nails. The T2 Tibial Nailing System is available in three versions, each differing from the other only in diameter, length and number and orientation of screw holes.
#### T2 Femoral Nail System
The T2 Femoral Nail is a cylindrical, cannulated titanium alloy tube, slightly bowed to accommodate the shape of the femur. The T2 Femoral Nail may be inserted into the femoral canal using either a retrograde or antegrade surgical approach.
{11}------------------------------------------------
# T2 Supracondylar Nail System
The T2 Supracondylar Nails are retrograde nails with a one-piece round profiled shaft design. The nails are cannulated and have a closed-section design with proximal rounded end. The T2 Supracondylar Nail is available in two versions: Short and Long. The T2 Supracondylar nails are available in lengths from 170 mm to 440 mm and in diameters from 9 mm to 14 mm. The T2 Supracondylar Nail System offers nails in varying lengths, a combination of locking screws, condyle screws, nuts and end caps.
## T2 Recon Nail System
The T2 Recon Nail System is a family of nails for various types of femoral fractures. The system includes Recon Nails in various lengths and diameters (both left and right versions), Lag Screws, Locking Screws, end caps, an Antegrade Set Screw and other accessories for use with the nails.
## T2 Greater Trochanter Nail (GTN)
The T2 Greater Trochanter Nail (GTN) is a cylindrical, cannulated titanium alloy tube, slightly bowed to accommodate the shape of the femur. The T2 GTN may be inserted into the femoral canal using either a retrograde or antegrade surgical approach.
## T2 Ankle Arthrodesis Nail
The T2 Ankle Arthrodesis Nail is a fluted, cannulated, titanium alloy nail and has a compression screw to provide internal compression of the upper ankle joint. The nail is inserted using an open or closed technique and can be locked in static, dynamic or compression mode. The T2 Ankle Arthrodesis Nail and accessories are intended for single use only. The T2 Ankle Arthrodesis Nail is available in left and right version in diameters 10 mm to 13 mm with the length ranging from 150 mm to 480 mm in increments of either 20mm or 50 mm. It has a 5 "lateral bend.
## T2 Arthrodesis Nail System
The T2 Arthrodesis Nails have a one-piece round profiled shaft design. The nails are cannulated and have a closed-section design with distal rounded end. The T2 Arthrodesis Nail is available in lengths from 540 mm to 780 mm in 40 mm increments, and in diameters from 10 mm to 15 mm.
## Indications for Use
## T2 Tibial Nailing System
The T2 Tibial Nailing System is intended to provide temporary stabilization of various types of fractures, malunions and nonunion of the tibia. The nails are inserted using an opened or closed technique and can be statically, dynamically and compressed locked.
The T2 Tibial Nailing System is indicated for long bone fracture fixation, specifically tibial fracture fixation, which may include the following:
- · Open and closed tibial fractures
- · Pseudoarthrosis and correction osteotomy
- · Pathologic fractures, impending pathologic fractures, and tumor resections
- Nonunion and malunion
{12}------------------------------------------------
# T2 Femoral Nail System
The T2 Femoral Nail is indicated for long bone fracture fixation specifically femoral fraction, which may include the following:
- Open and closed femoral fractures
- Pseudoarthrosis and correction osteotomy ●
- Pathologic fractures, impending pathologic fractures, and tumor restrictions ●
- . Supracondylar fractures, including those with intra-articular extension
- . Ipsilateral femur fractures
- . Fractures proximal to a total knee arthroplasty
- Fractures distal to hip joint
- Nonunions and malunions ●
## T2 Supracondylar Nail System
The T2 Supracondylar Nail System is indicated for:
- Open and closed femoral fractures
- Pseudoarthrosis and correction osteotomy
- . Pathologic fractures, impending pathologic fractures, and tumor resections
- . Supracondylar fractures including those with intra-articular extension
- Fractures distal to total hip prosthesis ●
- . Nonunions and malunions
## T2 Recon Nail System
The T2 Recon Nail is indicated for:
- Subtrochanteric fractures ●
- Intertrochanteric fractures ●
- Ipsilateral neck/shaft fractures
- Comminuted proximal femoral shaft fractures ●
- Femoral fixation required as a result of pathological disease
- . Temporary stabilization of fractures of the femoral shaft – ranging from the femoral neck to the supracondylar regions of the femur.
## T2 Greater Trochanter Nail (GTN)
The T2 Greater Trochanter Nail is indicated for long bone fixation, which may include the following:
- Open and closed femoral fractures
- Pseudarthrosis and correction osteotomy ●
- Pathologic fractures, impending pathologic fractures, and tumor resections
- Ipsilateral femur fractures
- Fractures proximal to a total knee arthroplasty
- Nonunions and malunions.
{13}------------------------------------------------
### T2 Ankle Arthrodesis Nail
The T2 Ankle Arthrodesis Nail is intended for tibiotalocalcaneal arthrodesis (fusion) and to provide stabilization of the hindfoot and ankle including the transverse tarsal joints coupling the mid-foot to the hindfoot. Examples of specific indications include:
- Post-traumatic or primary Arthrosis
- Previously infected arthrosis (second degree) ●
- . Revision of Failed Ankle Arthrodesis
- Failed Total Ankle Replacement .
- Avascular Necrosis of the Talus (requiring tibiocalcaneal arthrodesis) .
- . Neuroarthropathy or Neuromuscular Deformity or other neuromuscular disease with severe deformity or instability of the ankle
- Rheumatoid Arthritis with severe deformity such as rheumatoid hindfoot
- Osteoarthritis
- Nonunions or Pseudarthrosis of hindfoot and distal tibia ●
- Malunited tibial pilon fracture ●
- Charcot foot
- Severe endstage degenerative arthritis ●
- . Severe defects after tumor resection
- Pantalar arthrodesis
#### T2 Arthrodesis Nail System
The T2 Arthrodesis Nail is indicated for long bone internal fixation, which may include the following:
- Aseptic failed total knee arthroplasty
- Open and closed femoral fractures
- Pseudoarthrosis and correction osteotomy
- Pathologic fractures, impending pathologic fractures, and tumor resections ●
- Ipsilateral femur fractures
- . Failed external fixation, nonunions and malunions
- . Periarticular fractures where repair is not possible
- Knee arthrodesis ●
## Summary of Technologies
The purpose of this 510(k) is to gain clearance to market T2 Nails cleared in the systems listed below as compatible with the Locking Screws of the IMN Screws System, previously cleared in K172774, and the IMN Instrumentation.
A comparison of the systems demonstrated that the subject T2 Tibial Nailing System is substantially equivalent to the predicate T2 Tibial Nailing System (K003018 and K131365) regarding intended use, material, design and operational principals.
A comparison of the systems demonstrated that the subject T2 Femoral Nailing System is substantially equivalent to the predicate T2 Femoral Nailing System (K081152, K021744 and K010801) and Synthes Solid Femoral Nail (K923580) regarding intended use, material, design and operational principals.
{14}------------------------------------------------
A comparison of the systems demonstrated that the subject T2 Supracondylar Nail System is substantially equivalent to the predicate T2 Supracondylar Nail System (K023267) regarding intended use, material, design and operational principals.
A comparison of the systems demonstrated that the subject T2 Recon Nail System is substantially equivalent to the predicate T2 Recon Nail System (K102992 and K032898) regarding intended use, material, design and operational principals.
A comparison of the systems demonstrated that the subject T2 Greater Trochanter Nail System (GTN) is substantially equivalent to the predicate T2 Greater Trochanter Nail System (GTN) (K101438) regarding intended use, material, design and operational principals.
A comparison of the systems demonstrated that the subject T2 Ankle Arthrodesis Nail System is substantially equivalent to the predicate T2 Ankle Arthrodesis Nail System (K051590) regarding intended use, material, design and operational principals.
A comparison of the systems demonstrated that the subject T2 Arthrodesis Nail System is substantially equivalent to the predicate T2 Arthrodesis Nail System (K020384) regarding intended use, material, design and operational principals.
# Non-Clinical Testing
The Locking Screws (05mm, 25mm - 120mm length) of the IMN Screws System (IMN LS, recently cleared in K172774) were developed for locking of the T2 Alpha nails (recently cleared in K172774 and K180436). Further, compatibility of the IMN LS was proven to the existing T2 nails formerly cleared in:
- K003018 and K131365 [T2 Tibial Nailing System]
- K081152, K021744 and K010801 [T2 Femoral Nail System]
- K023267 [T2 Supracondylar Nail System]
- K102992 and K032898 [T2 Recon Nail System]
- K101438 [T2 Greater Trochanter Nail (GTN)]
- K051590 [T2 Ankle Arthrodesis Nail] ●
- K020384 [T2 Arthrodesis Nail System]
In a computational assessment, constructs of T2 nails and IMN Locking Screws were compared to constructs of T2 nails used with T2 Locking Screw, Fully Threaded (T2 Locking Screw, formerly cleared in K003018) as defined in the respective operative techniques.
The interface compatibility of the IMN Locking Screws (Ø 5 mm) with all T2 nails was shown in IMN Screws DOF 25-039 Functional Interface Analysis-Locking Screws. The performance of the IMN Locking Screw (Ø5 mm) was shown to be equal or higher than the performance of the T2 Locking
{15}------------------------------------------------
Screw (05 mm) in various test disciplines. If the same nail is used in a construct test with IMN Locking Screw (05 mm) and T2 Locking Screw (05 mm), the IMN Locking Screw construct will provide an equal or higher performance. Based on this, equal or higher construct strength of the IMN Locking Screw (Ø5 mm) in combination with the T2 nails is ensured and all acceptance criteria were fulfilled.
# Updated Report data for the previously cleared T2 Femoral Nail System
Following reevaluation of the dynamic fatigue strength testing of the T2 Femoral Nails, new Student t-tests were performed with the corrected data input. An additional predicate device was identified, the 09mm Synthes Solid Femoral Nail (K923580), which was tested in the same manner and statistically compared to the T2 Femoral Nails. The T2 Femoral Nail System was shown to be substantially equivalent to the additional predicate regarding intended use, material, design and operational principals.
# Clinical Testing
Clinical testing was not required for this submission.
## Conclusion
The subject T2 Tibial Nailing System , T2 Femoral Nail System, T2 Supracondylar Nail System, T2 Recon Nail System, T2 Greater Trochanter Nail (GTN), T2 Ankle Arthrodesis Nail, and T2 Arthrodesis Nail System are substantially equivalent to the predicate T2 Tibial Nailing System , T2 Femoral Nail System, T2 Supracondylar Nail System, T2 Recon Nail System, T2 Greater Trochanter Nail GTN), T2 Ankle Arthrodesis Nail, and T2 Arthrodesis Nail System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.