Femur Reconstruction Interlocking Nail System, Femur Retrograde Interlocking Nail System, Humerus Interlocking Nail System, Tibia Interlocking Nail System, Compression Hip Nail System
Femur Reconstruction Interlocking Nail System, Femur Retrograde Interlocking Nail System, Humerus Interlocking Nail System, Tibia Interlocking Nail System, Compression Hip Nail System
Applicant
Tdm Co., Ltd.
Product Code
HSB · Orthopedic
Decision Date
Aug 22, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Femur Reconstruction Interlocking Nail System is intended to stabilize femoral shaft fractures, subtrochanteric fractures, ipsilateral neck/shaft fractures, impending pathologic fractures, non-unions and malunions. The Femur Retrograde Interlocking Nail System is indicated for simple long bone fractures: severely comminuted, spiral, large oblique and segmental fractures; nonunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grating; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of long bones being treated. The Humerus Interlocking Nail System indicated for fractures of the proximal humerus, including 2-part surgical neck fractures, 3-part fractures, and 4-part fractures with diaphyseal extension, diaphyseal extension, diaphyseal fractures of the humeral shaft, and impending pathologic humeral fractures. The Tibia Interlocking Nail System is intended to stabilize fractures of the proximal and distal thatt; open and closed tibial shaft fractures; certain pre- and post-isthmic fractures; and tibial malunions and non- unions. The Compression Hip Nail System is intended to treat stable proximal femoral fracture including Pertrochanteric fractures, Intertrochanteric fractures and High sub trochanteric fractures and combination of these fractures.
Device Story
Intramedullary fixation system for stabilization of humeral, femoral, and tibial fractures; consists of various interlocking nails, lag screws, and supplemental screws. Device implanted into medullary cavity of long bones; fracture stabilized by locking screws at proximal and distal ends. Used in clinical/surgical settings by orthopedic surgeons. Provides mechanical fixation to facilitate bone healing and alignment. Benefits include stabilization of complex fractures, non-unions, and pathologic fractures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including biocompatibility (ISO 10993 series: genotoxicity, cytotoxicity, implantation, sensitization, reactivity, systemic/subchronic toxicity, pyrogens) and mechanical testing (ASTM F1264, ASTM F543, ASTM F384).
Technological Characteristics
Intramedullary fixation rod system; materials: titanium alloy; components: nails, lag screws, supplemental screws; dimensions: similar range to predicate; sterilization: not specified; connectivity: none; software: none; mechanical fixation principle.
Indications for Use
Indicated for patients with long bone fractures (femur, humerus, tibia) including shaft, subtrochanteric, supracondylar, proximal, distal, comminuted, spiral, oblique, segmental, and open/closed fractures; non-unions; malunions; polytrauma; impending pathologic fractures; and tumor resection reconstruction. Compression Hip Nail indicated for stable/unstable proximal femoral fractures (Pertrochanteric, Intertrochanteric, high subtrochanteric).
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
{0}------------------------------------------------
August 22, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Tdm Co., Ltd. % Dave Kim Medical Regulatory Affairs Mtech Group 7505 Fannin St. Suite 610 Houston, Texas 77054
Re: K230071
Trade/Device Name: Femur Reconstruction Interlocking Nail System, Femur Retrograde Interlocking Nail System. Humerus Interlocking Nail System. Tibia Interlocking Nail System. Compression Hip Nail System
Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: August 15, 2024 Received: August 15, 2024
Dear Dave Kim:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Farzana Sharmin -S | Digitally signed by Farzana Sharmin -S<br>Date: 2024.08.22 14:16:28 -04'00' |
|--------------------|-----------------------------------------------------------------------------|
|--------------------|-----------------------------------------------------------------------------|
Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230071
#### Device Name
Femur Reconstruction Interlocking Nail System. Femur Retrograde Interlocking Nail System, Nail System, Tibia Interlocking Nail System, Compression Hip Nail System
### Indications for Use (Describe)
### (1) Femur Reconstruction Interlocking Nail System
The Femur Reconstruction Interlocking Nail System is intended to stabilize femoral shaft fractures, subtrochanteric fractures, ipsilateral neck/shaft fractures, impending pathologic fractures, non-unions and malunions.
### (2) Femur Retrograde Interlocking Nail System
The Femur Retrograde Interlocking Nail System is indicated for simple long bone fractures: severely comminuted, spiral, large oblique and segmental fractures; nonunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grating; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of long bones being treated.
### (3) Humerus Interlocking Nail System
The Humerus Interlocking Nail System indicated for fractures of the proximal humerus, including 2-part surgical neck fractures, 3-part fractures, and 4-part fractures with diaphyseal extension, diaphyseal extension, diaphyseal fractures of the humeral shaft, and impending pathologic humeral fractures.
### (4) Tibia Interlocking Nail System
The Tibia Interlocking Nail System is intended to stabilize fractures of the proximal and distal thatt; open and closed tibial shaft fractures; certain pre- and post-isthmic fractures; and tibial malunions and non- unions.
### (5) Compression Hip Nail System
The Compression Hip Nail System is intended to treat stable proximal femoral fracture including Pertrochanteric fractures, Intertrochanteric fractures and High sub trochanteric fractures and combination of these fractures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K230071
The following 510(k) summary is being submitted as required by 21 CFR Part 807.92;
| 5.1 | Submitter: | TDM CO., LTD.<br>69, Cheomdan Venture So-Ro 37beon-Gil<br>Gwangju, KR 61003<br>Tel: +82-62-971-7460<br>Establishment registration number: 3014257776 |
|-----|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Contact Person:<br>(Official Correspondent) | Mr. Dave Kim<br>7505 Fannin St. Suite 610, Houston, TX 77054<br>Tel: 713-467-2607<br>Email: davekim@mtechgroupllc.com |
| | Date Prepared: | 01/25/2024 |
| 5.2 | Device Identification | |
| | Device Trade Name | Femur Reconstruction Interlocking Nail System, Femu<br>Retrograde Interlocking Nail System, Humerus Interlocking<br>Nail System, Tibia Interlocking Nail System, Compression<br>Hip Nail System |
| | Common Name | Intramedullary Fixation Nailing System |
| | Classification Name<br>(Regulation Number) | Intramedullary Fixation Rod<br>(21 CFR 888.3020) |
| | Device Classification | II |
| | Product Code | HSB |
#### 5.3 Predicate and Reference devices:
Predicate device: K192003, "Auxein Nailing System", manufactured by "Auxein Medical Private Limited" Reference device #01: K120807, "Synthes Multiloc Humeral Nailing System", manufactured by "Synthes"
Reference device #02: K040762, "Synthes(USA) Tibial Nail System EX", manufactured by "Synthes" Reference device #03: K200869, "Gamma3 System Lag Screws", manufactured by "Stryker
Trauma GmbH"
Reference device #04: K200880, "T2 Femur Nail", manufactured by "Stryker GmbH"
#### 5.4 Device Description
This intramedullary fixation rod is used for stabilization of shaft fracture in humeral bone, femoral bone and tibial bone. It is implanted into medullary cavity inside of the long bone (humeral bone, femoral bone and tibial bone) with a fracture and the fractured long bone (humeral bone, femoral bone and tibial bone) is fixed to the nail by using the locking screw at the upper and lower parts of the fractured part.
{4}------------------------------------------------
The nailing system consists of the Femur Reconstruction Interlock Nail System, Femur Retrograde Interlocking Nail System, Humerus Interlocking Nail System, Tibia Interlocking Nail System, and Compression Hip Nailing System.
#### . Indications for use
- Femur Reconstruction Interlocking Nail System ●
The Femur Reconstruction Interlocking Nail System is intended to stabilize femoral shaft fractures, subtrochanteric fractures, ipsilateral neck/shaft fractures, impending pathologic fractures, non-unions and malunions.
- . Femur Retrograde Interlocking Nail System
The Femur Retrograde Interlocking Nail System is indicated for simple long bone fractures; severely comminuted, spiral, large oblique and segmental fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; supracondylar fractures; bone lengthening and shortening. Interlocking intramedullary nails are indicated for fixation of fractures that occur in and between the proximal and distal third of long bones being treated.
- Humerus Interlocking Nail System .
The Humerus Interlocking Nail System indicated for fractures of the proximal humerus, including 2-part surgical neck fractures, 3-part fractures, and 4-part fractures, proximal humeral fractures with diaphyseal extension, diaphyseal fractures of the humeral shaft, and impending pathologic humeral fractures.
. Tibia Interlocking Nail System The Tibia Interlocking Nail System is intended to stabilize fractures of the proximal and distal tibia and the tibial shaft; open and closed tibial shaft fractures; certain pre- and post-isthmic fractures; and tibial malunions and non- unions.
. Compression Hip Nail System
The Compression Hip Nail System is intended to treat stable and unstable proximal femoral fracture including Pertrochanteric fractures, Intertrochanteric fractures and High sub trochanteric fractures and combination of these fractures.
#### ● Non-clinical Test Conclusion
The following properties were tested based on the referenced standards. All the test results support substantial equivalence to the predicate devices.
- = ISO 10993-3- Genotocixity (Bacterial reverse mutation) & Genotocixity (chromosomal aberration)
{5}------------------------------------------------
- ISO 10993-5- Cytotoxicity
- ISO 10993-6- Implantation
- = ISO 10993-10 Skin Sensitization & Intracutatneous Reactivity
- = ISO 10993-11 Acute systemic toxicity & Subchronic Toxicity & Pyrogens
- = testing per ASTM F1264, ASTM F543, and ASTM F384
#### ● Clinical Test Conclusion
Clinical testing was not required for this submission.
#### Technical Characteristics and Substantial Equivalence ●
The subject device is substantially equivalent to "Auxein Nailing System (K192003)", the predicate device, based on the following:
- . The subject device does not have a new intended use. It shows equivalent specifications with the predicate device in most of the parameters.
- The subject device is intramedullary fixation nailing system which employs Femur Reconstruction ● Interlocking Nail, Femur Retrograde Interlocking Nail System Humerus Interlocking Nail Tibia Interlocking Nail System, and Compression Hip Nail System, designed similar to the predicate device.
- . Both the subject device and the predicate device are made of titanium alloy.
- The nails, lag screws and supplemental screws are in the similar range of lengths and diameters as the predicate device.
## 6. Conclusion:
Based on the testing results, TDM CO., LTD. concludes that the subject devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.