K230069 · S.I.N. – Sistema DE Implante Nacional S.A. · NHA · Apr 6, 2023 · Dental
Device Facts
Record ID
K230069
Device Name
S.I.N. Dental Implant System
Applicant
S.I.N. – Sistema DE Implante Nacional S.A.
Product Code
NHA · Dental
Decision Date
Apr 6, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
S.I.N. Dental Implant System is intended for placement in the maxillary or mandbular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading. All digitally-designed custom abutments for use with Pre-Milled CAD-CAM Abutments are to be sent to a S.I.N.-validated milling center for manufacture.
Device Story
S.I.N. Dental Implant System adds Pre-Milled CAD-CAM Abutments to existing implant line; components feature 11.5° or 16° CM connections. Abutments are titanium alloy (Ti-6Al-4V) blanks; clinicians or technicians design patient-specific abutments using CAD-CAM software; designs must be sent to S.I.N.-validated milling center for final manufacture. Device used in dental clinics; operated by dentists/oral surgeons. Output is custom-milled abutment providing support for prosthetic restorations. Benefits include patient-specific fit for dental rehabilitation. Abutments provided non-sterile; require end-user moist heat sterilization.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing: moist heat sterilization validation (ISO 17665), biocompatibility (ISO 10993-5), MR environment safety (ASTM F2052, F2213, F2182, F2119), and engineering analysis of mechanical performance (ISO 14801).
Technological Characteristics
Materials: Titanium alloy (Ti-6Al-4V) per ASTM F136. Connection: 11.5° or 16° CM conical interface. Design limits: 0.55 mm min wall/gingival height, 4.0 mm min post height, 30° max angulation. Sterilization: Moist heat (autoclave) by end user. Packaging: PET blister pack with Tyvek cover.
Indications for Use
Indicated for patients requiring functional and esthetic rehabilitation of the edentulous mandible or maxilla via single-unit or multi-unit restorations supported by dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
S.I.N. - Sistema de Implante Nacional S.A. % Kevin Thomas Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K230069
Trade/Device Name: S.I.N. Dental Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: January 9, 2023 Received: January 10, 2023
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
## Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K230069
Device Name
S.I.N. Dental Implant System
Indications for Use (Describe)
S.I.N. Dental Implant System is intended for placement in the maxillary or mandbular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
All digitally-designed custom abutments for use with Pre-Milled CAD-CAM Abutments are to be sent to a S.I.N.-validated milling center for manufacture.
Type of Use (Select one or both, as applicable)
| <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------------|
| <span style="font-size: large;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary K230069 S.I.N. - Sistema de Implante Nacional S.A. S.I.N Dental Implant System
April 6, 2023
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | S.I.N. - Sistema de Implante Nacional S.A.<br>Avenida Vereador Abel Ferreira, 1100<br>São Paulo, São Paulo<br>03340-000 Brazil<br>Telephone +55-11-21693000 ext 3236 | | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Official Contact | Denise Domiciano, Quality and Regulatory Manager | | |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email kthomas@paxmed.com<br>flarson@paxmed.com | | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | S.I.N. Dental Implant System |
|------------------------|------------------------------------------------------------------------------------------------|
| Common Names | Dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory,<br>ENT and Dental Devices) |
| Reviewing Division | Division of Dental and ENT Devices |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K 193096, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A.
Reference Devices K221453, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A. K222231, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A.
{4}------------------------------------------------
K211921, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A. K200992, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A. K 170392, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A.
#### INDICATIONS FOR USE STATEMENT
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
All digitally-designed custom abutments for use with Pre-Milled CAD-CAM Abutments are to be sent to a S.I.N.-validated milling center for manufacture.
#### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to add components to the S.I.N. Dental Implant System, which includes components cleared previously in K221453, K222231, K211921, K193096, and K170392.
This submission adds Pre-Milled CAD-CAM Abutments to the S.I.N. Dental Implant System. The subject device abutments are provided with 11.5° CM and 16° CM connections compatible with the above S.I.N. dental implants.
The design parameters for the CAD-CAM fabrication of patient-specific abutments from Pre-Milled CAD-CAM Abutments are the same for the 11.5° CM implant connection and the 16° CM implant connection. The design limit parameters are:
Minimum wall thickness - 0.55 mm Minimum post height for single-unit restoration - 4.0 mm Minimum gingival height - 0.55 mm Maximum gingival height - 3.5 mm Minimum prosthetic platform diameter - 3.5 mm Maximum prosthetic post height - 6 mm Maximum angulation - 30°
The subject device abutments are compatible with previously cleared S.I.N. Dental Implant bodies summarized in the following table.
| Compatible Previously Cleared Implants, 11.5° CM | 510(k) Number |
|------------------------------------------------------------------------------------|---------------|
| Unitite Implant, body Ø 3.5 mm, 4.3 mm, 5.0 mm | K170392 |
| Implant Strong SW, body Ø 3.5 mm, 3.8 mm, 4.5 mm, 5.0 mm | |
| Implant Strong SW Plus, body Ø 3.5 mm, 3.8 mm, 4.5 mm, 5.0 mm | |
| Implant Tryon CM Conical, body Ø 3.5 mm, 4.0 mm, 5.0 mm | K200992 |
| Implant Tryon CM Cylindrical Body and Conical Apex, body Ø 3.5 mm, 3.75 mm, 4.0 mm | |
| Implant Tryon CM Cylindrical, body Ø 3.5 mm, 3.75 mm, 4.0 mm, 5.0 mm | |
| Implant Epikut CM, body Ø 3.5 mm, 3.8 mm, 4.5 mm, 5.0 mm | |
| Implant Epikut Plus CM, body Ø 3.5 mm, 3.8 mm, 4.5 mm, 5.0 mm | K211921 |
| Implant Epikut CM, body Ø 3.8 mm, 4.0 mm, 4.5 mm | |
| Implant Epikut Plus CM, body Ø 3.8 mm, 4.0 mm, 4.5 mm | K221453 |
{5}------------------------------------------------
| Compatible Previously Cleared Implants, 16° CM | 510(k) Number |
|----------------------------------------------------------------------|---------------|
| Strong SW CM Implant, body Ø 3.5 mm, 3.8 mm, 4.5 mm, 5.0 mm | K170392 |
| Strong SW CM Plus Implant, body Ø 3.5 mm, 3.8 mm, 4.5 mm, 5.0 mm | K193096 |
| Implant Epikut S, body Ø 3.5 mm, 3.8 mm, 4.0 mm, 4.5 mm, 5.0 mm | K222231 |
| Implant Epikut S Plus, body Ø 3.5 mm, 3.8 mm, 4.0 mm, 4.5 mm, 5.0 mm | |
All subject device abutments are manufactured from titanium alloy (Ti-6Al-4V) conforming to ASTM F136, and are to be used with compatible abutment screws previously cleared in K170392. All subject device abutments are provided non-sterile and are to be moist heat sterilized by the end user.
#### PERFORMANCE DATA
Non-clinical data submitted or referenced to demonstrate substantial equivalence included:
- . provided in this submission was moist heat sterilization for subject devices provided non-sterile to the end user, validated to a sterility assurance level of 105 by the overkill method according to ANSI/AAMI/ISO 17665-1 and ANSI/AAMI/ISO TIR 17665-2;
- . referenced from K170398 was biocompatibility testing according to ISO 10993-5 (cytotoxicity) for the abutment material ASTM F136;
- . referenced from K221453 was non-clinical analysis and testing to evaluate the metallic subject devices and compatible dental implants in the MR environment according to ASTM F2052 (magnetically induced displacement force), ASTM F2213 (magnetically induced torque), ASTM F2182 (RF induced heating), and ASTM F2119 (image artifact), and the FDA guidance document Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued May 2021);
- . provided in this submission was engineering analysis to demonstrate that the subject device abutments, in combination with compatible previously cleared dental implants, do not create a new worst-case construct in terms of mechanical testing according to ISO 14801; and
- referenced from K200992 was mechanical testing conducted according to ISO 14801 to support the ● performance of the subject device abutments, in conjunction with the engineering analysis provided in this submission.
No clinical data were included in this submission.
#### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the primary predicate device and the reference devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the reference devices.
The primary predicate device K193096 is in support of substantial equivalence of the Indications for Use including CAD-CAM abutments, abutment designs, materials, manufacturing, sterilization, and compatible implants. The reference devices K221453, K222231, K211921, K200992, and K170392 are in support of substantial equivalence of the Indications for Use including CAD-CAM abutments (K221453), abutment designs, materials, manufacturing, sterilization, and compatible implants.
{6}------------------------------------------------
The Indications for Use Statement (IFUS) for the subject device includes language concerning placement in the maxillary or mandibular arches and regarding immediate loading that is identical to the language in K193096. The IFUS for the subject device also includes language concerning CAD-CAM abutments and validated milling centers that is identical, besides the trade name, to that included in K193096.
Differences between the IFUS for the subject device and the reference devices include: language in K221453 and K222231 regarding indications for implants with lengths of 18-24 mm that is not relevant to the subject device; language in K170392 regarding implant lengths less than 7 mm that is not relevant to the subject device; and the IFUS for K222231, K211921, K200992, and K170392 do not contain language regarding CAD-CAM abutments and validated milling centers.
The subject device abutments with the 11.5° CM interface are compatible with implants cleared in K200992, K221453, K211921, and K170392 in terms of the abutment-implant connection. The subject device abutments with the 11.5° CM interface in final finished form are substantially equivalent to the Abutment Cemented Angled Indexed SIT cleared in K200992 in terms of material and design limit parameters including prosthetic platform diameter, maximum gingival height, and maximum angulation.
The subject device abutments with the 16° CM interface are compatible with implants cleared in K222231, K193096, and K170392 in terms of the abutment-implant connection. The subject device abutments with the 16° CM interface in final finished form are substantially equivalent to the Abutment Cemented Morse Angled cleared in K200992 in terms of material and design limit parameters including prosthetic platform diameter, maximum gingival height, and maximum angulation.
The subject device abutments are to be end user sterilization instructions are provided in the Instructions for Use. Similar components to be end-user sterilized (abutments manufactured from Ti-6Al-4V alloy material) were cleared in K221453, K193096, and K170392.
The subject device abutments are packaged in a polyethylene terephthalate blister pack with a Tyvek cover. The Tyvek cover for the blister pack is printed with the appropriate label for the corresponding component. This is the same packaging used for similar abutments (provided non-sterile) in K221453.
An engineering rationale was provided to demonstrate that the use of the subject device abutments in combination with the compatible implants raise no new risks and that mechanical testing data provided in a prior submission (K200992) are applicable to the subject abutments.
#### CONCLUSION
The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are made of identical or similar materials. The subject device, the primary predicate, and the reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the primary predicate device and the reference devices listed above.
{7}------------------------------------------------
### Table of Substantial Equivalence
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K230069<br>S.I.N. Dental Implant System | K193096<br>S.I.N. Dental Implant System | K200992<br>S.I.N. Dental Implant System | K221453<br>S.I.N. Dental Implant System | K170392<br>S.I.N. Dental Implant System |
| | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. |
| Indications for Use Statement<br>(IFUS) | S.I.N. Dental Implant System is intended for placement<br>in the maxillary or mandibular arch to provide support<br>for single-unit or multi-unit restorations. When a one-<br>stage surgical approach is applied, the S.I.N. Dental<br>Implant System is intended for immediate loading when<br>good primary stability is achieved and with appropriate<br>occlusal loading.<br>All digitally-designed custom abutments for use with<br>Pre-Milled CAD-CAM abutments are to be sent to a<br>S.I.N .- validated milling center for manufacture. | S.I.N. Dental Implant System is intended for placement<br>in the maxillary or mandibular arch to provide support<br>for single-unit or multi-unit restorations. When a one-<br>stage surgical approach is applied, the S.I.N. Dental<br>Implant System is intended for immediate loading when<br>good primary stability is achieved and with appropriate<br>occlusal loading.<br>All digitally-designed custom abutments for use with<br>Interface CAD-CAM abutments are to be sent to a<br>S.I.N.-validated milling center for manufacture. | S.I.N. Dental Implant System is intended for placement<br>in the maxillary or mandibular arch to provide support<br>for single-unit or multi-unit restorations. When a one-<br>stage surgical approach is applied, the S.I.N. Dental<br>Implant System is intended for immediate loading when<br>good primary stability is achieved and with appropriate<br>occlusal loading. | S.I.N. Dental Implant System is intended for placement<br>in the maxillary or mandibular arch to provide support<br>for single-unit or multi-unit restorations. When a one-<br>stage surgical approach is applied, the S.I.N. Dental<br>Implant System is intended for immediate loading when<br>good primary stability is achieved and with appropriate<br>occlusal loading.<br>S.I.N. Dental Implant System implants with lengths of<br>18, 20, 22, or 24 mm may be tilted up to 30°. When used<br>in the mandible or maxilla with implants with lengths of<br>18, 20, 22, or 24 mm at an angulation of 30°, a<br>minimum of four implants must be used and must be<br>splinted. When placed in the maxilla with lengths of 18,<br>20, 22, or 24 mm at angulations between 0° and less<br>than 30°, the S.I.N. Dental Implant System implants are<br>only indicated for multiple unit restorations in splinted<br>applications that utilize at least two implants.<br>All digitally-designed custom abutments for use with<br>Interface CAD-CAM abutments are to be sent to a<br>S.I.N.-validated milling center for manufacture. | S.I.N. Dental Implant System is intended for placement<br>in the maxillary or mandibular arch to provide support<br>for single-unit or multi-unit restorations. When a one-<br>stage surgical approach is applied, the S.I.N. Dental<br>Implant System is intended for immediate loading when<br>good primary stability is achieved and with appropriate<br>occlusal loading. Implants with lengths less than 7 mm<br>are intended for delayed loading only. |
| Reason for<br>Predicate Device | Not applicable | IFUS including CAD-CAM abutments and validated<br>milling centers;<br>Compatible implants | Abutment designs; materials; manufacturing;<br>sterilization;<br>Compatible implants | IFUS including CAD-CAM abutments and validated<br>milling centers;<br>Compatible implants | Abutment designs; materials; manufacturing;<br>sterilization<br>Compatible implants |
| Product Codes | NHA | DZE, NHA | DZE, NHA | DZE, NHA | DZE, NHA |
| Intended Use | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla |
| Abutment Designs | | | | | |
| CM 11.5° interface | Pre-Milled Blank Abutments<br>CM 11.5° interface<br>Minimum post height for single-unit restoration: 4 mm<br>Prosthetic platform Ø: 3.5 mm (min)<br>Gingival height: 3.5 mm (max), 0.55 mm (min)<br>Angle: 30° (max)<br>Indexed | | Abutment Cemented Angled Indexed SIT<br>CM 11.5° interface<br>Prosthetic platform Ø: 3.3, 4.5 mm<br>Gingival height: 1.5-3.5 mm<br>Angle: 17°, 30°<br>Indexed | | |
| CM 16° interface | Pre-Milled Blank Abutments<br>CM 16° interface<br>Minimum post height for single-unit restoration: 4 mm<br>Prosthetic platform Ø: 3.5 mm (min)<br>Gingival height: 3.5 mm (max), 0.55 mm (min)<br>Angle: 30° (max)<br>Indexed | | Abutment Cemented Morse Angled<br>CM 16° interface<br>Prosthetic platform Ø: 3.3, 4.5 mm<br>Gingival height: 1.5-3.5 mm<br>Angle: 17°, 30<br>Indexed | | |
| Abutment Materials | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 | Titanium alloy, ASTM F136 |
| | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Reference Device |
| | K230069<br>S.I.N. Dental Implant System | K193096<br>S.I.N. Dental Implant System | K200992<br>S.I.N. Dental Implant System | K221453<br>S.I.N. Dental Implant System | K170392<br>S.I.N. Dental Implant System |
| | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. | S.I.N. - Sistema de Implante Nacional S.A. |
| Compatible Implants | | | | | |
| CM 11.5° Connection | | | Strong SW and<br>Strong SW Plus<br>Body/platform $Ø$ 3.5/3.5; 3.8/3.8; 4.5/4.5; 5.0/5.0 mm<br>Lengths 8.5-15 mm | Epikut CM<br>Epikut Plus CM<br>Body/platform $Ø$ 3.8/3.8; 4.0/4.0; 4.5/4.5 mm<br>Lengths 18-24 mm | Unitite<br>Body/platform $Ø$ 3.5/3.5; 4.3/4.3, 5.0/5.0 mm<br>Lengths 8.5-15 mm |
| | | | Tryon CM Conical<br>Body/platform $Ø$ 3.5/3.5, 4.0/4.0; 5.0/5.0 mm<br>Lengths 8.5-15 mm | | |
| | | | Tryon CM Cylindrical Body and Conical Apex<br>Body/platform $Ø$ 3.5/3.5, 3.8/3.8, 4.5/4.5; 5.0/5.0 mm<br>Lengths 8.5-15 mm | | |
| | | | Tryon CM Cylindrical<br>Body/platform $Ø$ 3.5/3.5, 3.75/3.75, 4.0/4.0; 5.0/5.0 mm<br>Lengths 7-15 mm | | |
| CM 16° Connection |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.