SOMNUM (V.1.1.2.)

K223922 · Honeynaps Co., Ltd. · OLZ · Aug 16, 2023 · Neurology

Device Facts

Record IDK223922
Device NameSOMNUM (V.1.1.2.)
ApplicantHoneynaps Co., Ltd.
Product CodeOLZ · Neurology
Decision DateAug 16, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2
AttributesSoftware as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K223922 · Aug 16, 2023SOMNUM (V.1.1.2.)Honeynaps Co., Ltd.Retrospective polysomnography (PSG) records from 48 subjectsThe sponsor used a retrospective cohort of 48 clinical sleep study records to validate the software's automated scoring performance (sleep staging, respiratory events, arousals, and leg movements) by comparing it against a consensus of three human experts.Retrospective data; Polysomnography; Performance validation; Sleep disorders

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Cross-sectional experimental design using retrospective data; Follow-up/Duration: Full night studiesAdult patients undergoing sleep studies; Sample Size: 48 subjectsThree human expert scorers (2/3 majority rule)Epoch-by-epoch agreement (PPA, NPA, OPA, Cohen's kappa) for sleep staging, respiratory events, arousals, and PLMs; summary variables (e.g., TST, AHI)

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Sleep Stage Scoring——87.5% (OPA)——N=48 subjects of data recorded in the sleep laboratory3 (expert scoring)
Arousal Detection——82.5% (OPA)——N=48 subjects of data recorded in the sleep laboratory3 (expert scoring)
Periodic Leg Movement Detection——94.1% (OPA)——N=48 subjects of data recorded in the sleep laboratory3 (expert scoring)
Respiratory Event Detection——92.3% (OPA)——N=48 subjects of data recorded in the sleep laboratory3 (expert scoring)

Indications for Use

SOMNUM is a computer program (software) intended for use as an aid for the diagnosis of sleep and respiratory related sleep disorders. SOMNUM is intended to be used for analysis (automatic scoring and manual re-scoring), display, redisplay(retrieve), summarize and reports generation of digital data collected by monitoring devices typically used to evaluate sleep and respiratory related sleep disorders. The device is to be used under the supervision of a physician. Use is restricted to files obtained from adult patients. For respiratory events - Sleep Disordered Breathing (Apneas)- obstructive, central, mixed apneas, and hypopneas must be manually scored by physician. The device does not output specific apnea or hypopnea events and therefore should not be used for management decisions.

Device Story

Standalone software application (Windows 10) for analysis of polysomnography (PSG) data (EDF files). Inputs: EEG, EOG, EMG (chin/leg), ECG, and respiratory signals (chest/abdomen movement, airflow, thermistor). Algorithms automatically score sleep stages (Wake, N1-N3, REM), arousals, respiratory events, and periodic leg movements (PLMs). Output: Automated scoring, summary reports, and data visualization for physician review. Used in healthcare facilities; operated by clinicians. Software assists in diagnostic evaluation; physician performs manual re-scoring and final verification of respiratory events. Benefits: Streamlines sleep study analysis, reduces manual scoring burden, and provides standardized reporting to support clinical diagnosis of sleep disorders.

Clinical Evidence

Bench testing only. Performance evaluated using a cross-sectional design on 48 subjects (43,971 epochs). Primary endpoints: epoch-by-epoch agreement for sleep staging, arousals, SDB, and PLMs compared to a 2/3 majority rule of three expert scorers. Results: Sleep staging OPA 87.5%, Arousal OPA 82.5%, SDB PPA 94.18%, PLMS PPA 92.89%. Performance metrics were compared against predicate devices and literature-based targets, demonstrating substantial equivalence.

Technological Characteristics

Standalone software for Windows 10. Processes EDF, XML, and JSON files. Inputs: EEG, EOG, EMG, ECG, respiratory bands, thermistor. Scoring based on AASM 2016 manual. Performs cardiac artifact removal. No internet access required. Software verification and validation conducted per FDA guidance.

Indications for Use

Indicated for adult patients undergoing sleep studies to aid in the diagnosis of sleep and respiratory-related sleep disorders. Used under physician supervision for analysis, display, and reporting of polysomnography data. Contraindicated for automated management decisions regarding specific apnea or hypopnea events, which require manual physician scoring.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ August 16, 2023 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Honeynaps Co., Ltd Tony Lee Official Correspondent 4F. Marine Tech B/D. 529. Nonhyeon-ro. Gangnam-gu Seoul. 06126 Korea, South #### Re: K223922 Trade/Device Name: Somnum (v.1.1.2.) Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLZ Dated: July 17, 2023 Received: July 17, 2023 #### Dear Tony Lee: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Patrick Antkowiak -S for Jay Gupta Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K223922 Device Name SOMNUM #### Indications for Use (Describe) SOMNUM is a computer program (software) intended for use as an aid for the diagnosis of sleep and respiratory related sleep disorders. SOMNUM is intended to be used for analysis (automatic scoring and manual re-scoring), display, redisplay(retrieve), summarize and reports generation of digital data collected by monitoring devices typically used to evaluate sleep and respiratory related sleep disorders. The device is to be used under the supervision of a physician. Use is restricted to files obtained from adult patients. For respiratory events - Sleep Disordered Breathing (Apneas)- obstructive, central, mixed apneas, and hypopneas must be manually scored by physician. The device does not output specific apnea or hypopnea events and therefore should not be used for management decisions. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Honey Naps. The logo consists of a stylized letter "H" in two shades of blue, with the left side being a lighter blue and the right side being a darker blue. To the right of the "H" is the text "HONEY NAPS" in a dark blue sans-serif font. The word "HONEY" is above the word "NAPS". Traditional 510(k) Premarket Submission ## 510(k) Summary Device Name : SOMNUM (V.1.1.2. Final finished device) {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Honey Naps, a company that specializes in sleep products. The logo features a stylized "H" in blue, with the words "HONEY NAPS" in blue text to the right of the "H". Below the logo, the words "510(k) Summary" are written in a large, bold font. In accordance with 21 CFR 807.92 the following summary of information is provided: Date : September 30, 2022 ## I. SUBMITTER | SUBMITTER : | Honeynaps | |-------------------------|--------------------------------| | | 529, Nonhyeon-ro, Gangnam-gu | | | Seoul, Republic of Korea 06126 | | | Tel : +82-2-567-0134 | | CONTACT PERSON: | Honeynaps | | | YoungJun Lee | | | Chief Executive Officer | | | tony.lee@honeynaps.com | | PRIMARY CONTACT PERSON: | Honeynaps | | | YoungJun Lee | | | Chief Executive Officer | | | tony.lee@honeynaps.com | ## II. DEVICE | Trade Name | SOMNUM | |-----------------------------|-------------------------------------------------------------------------------------| | Common Name | Sleep Analysis System | | Regulation Number | 21 CFR 882.1400 | | Regulation Name | Electroencephalograph | | Regulation Class | II | | Classification Product Code | OLZ | | Device Classification Name | Automatic Event Detection Software For Polysomnograph<br>with Electroencephalograph | | Review Panel | Neurology | | Ⅲ. PREDICATE DEVICE | | |-----------------------------|-------------------------------------------------------------------------------------| | 510(k) Number | K162627 | | Applicant | ENSODATA | | Device Name | EnsoSleep | | Regulation Number | 21 CFR 882.1400 | | Regulation Name | Electroencephalograph | | Regulation Class | II | | Classification Product Code | OLZ | | Device Classification Name | Automatic Event Detection Software For Polysomnograph<br>with Electroencephalograph | {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains a logo with the text "HONEY NAPS". To the left of the text is a stylized letter "H" made up of two vertical bars connected by a horizontal bar. The left vertical bar is a lighter shade of blue than the right vertical bar. The logo appears to be for a company or product called "Honey Naps". | 510(k) Number | K112102 | |-----------------------------|------------------------------| | Applicant | YOUNES SLEEP TECHNOLOGIES | | Device Name | MICHELE SLEEP SCORING SYSTEM | | Regulation Number | 21 CFR 868.2375 | | Regulation Name | Breathing Frequency Monitor | | Regulation Class | II | | Classification Product Code | MNR | | Device Classification Name | Ventilatory Effort Recorder | {6}------------------------------------------------ SOMNUM is a standalone software application that analyze previously recorded physiological data obtained during level 1 sleep studies, referred to as polysomnography (PSG) records. The SOMNUM software can analyze any EDF files. Automated algorithms are applied to the raw signals in order to identify the occurrence of certain events. The software automates recognition of: - · Sleep Stage Events : Wake, Stage N1, Stage N2, Stage N3, Stage REM - Respiratory Events : Sleep Disordered Breathing (device output does not distinguish between Apneas and Hypopneas. Obstructive, central, mixed apneas, and hypopneas must be manually scored by physician) - · Arousal Events - · Leg Movement Events : Periodic Leg Movements during Sleep (PLMs) The SOMNUM software can be used as a stand-alone application for use on Windows 10 operating system platform. All processing, scoring, and analysis of signal data occurs on local desktop PC. ### V. INDICATIONS FOR USE SOMNUM is a computer program (software) intended for use as an aid for the diagnosis of sleep and respiratory related sleep disorders. SOMNUM is intended to be used for analysis(automatic scoring and manual re-scoring), display, redisplay(retrieve), summarize, reports generation of digital data collected by monitoring devices typically used to evaluate sleep and respiratory related sleep disorders. The device is to be used under the supervision of a physician Use is restricted to files obtained from adults patients. For respiratory events - Sleep Disordered Breathing (Apneas and Hypopneas)- obstructive, central, mixed apneas, and hypopneas must be manually scored by physician. The device does not output specific apnea or hypopnea events and therefore should not be used for management decisions. | Description | Subject Device | Predicate Device | |---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | SOMNUM | ENSOSLEEP<br>(K162627) | | Indications for Use | SOMNUM is a computer program<br>(software) intended for use as an<br>aid for the diagnosis of sleep and<br>respiratory related sleep disorders.<br>SOMNUM is intended to be used<br>for analysis(automatic scoring and<br>manual re-scoring), display,<br>redisplay(retrieve), summarize,<br>reports generation of digital data<br>collected by monitoring devices<br>typically used to evaluate sleep and<br>respiratory related sleep disorders. | EnsoSleep is intended for use for the<br>diagnostic evaluation by a physician to<br>assess sleep quality and as an aid for the<br>diagnosis of sleep and respiratory related<br>sleep disorders in adults only. EnsoSleep<br>is a software-only medical device to be<br>used under the supervision of a clinician<br>to analyze physiological signals and<br>automatically score sleep study results,<br>including the staging of sleep, detection of<br>arousals, leg movements, and sleep<br>disordered breathing events including<br>obstructive apneas. All automatically | #### Table 1. Comparison of Indications for Use {7}------------------------------------------------ | | The device is to be used under the<br>supervision of a physician Use is<br>restricted to files obtained from<br>adults patients.<br><br>For respiratory events - Sleep<br>Disordered Breathing (Apneas and<br>Hypopneas)- obstructive, central, mixed<br>apneas, and hypopneas must be<br>manually scored by physician. The<br>device does not output specific apnea<br>or hypopnea events and therefore<br>should not be used for management<br>decisions | scored events are subject to verification<br>by a qualified clinician. Central apneas,<br>mixed apneas, and hypopneas must be<br>manually marked within records. | |--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Analyze physiological data previously<br>recorded during sleep and present<br>a report. | Analyze physiological data previously<br>recorded during sleep and present a<br>report. | | Environment of Use | Healthcare Facility | Healthcare Facility | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the logo for Honey Naps. The logo consists of a stylized letter "H" on the left and the words "HONEY NAPS" on the right. The "H" is made up of two vertical bars connected by a horizontal bar, with circles at the top and bottom of each vertical bar. The word "HONEY" is above the word "NAPS". The colors of the logo are shades of blue. ## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE There are no significant differences in the technological characteristics of this device compared to the predicate device which adversely affect safety or effectiveness. The below table is summarized and compared with the technological characteristics between the SOMNUM and the predicate device : #### Table 2. Comparison of Technological Characteristics | Description | Subject Device<br>SOMNUM | Predicate Device<br>ENSOSLEEP<br>(K162627) | |-------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------------------------| | <b>Clinical Criteria :</b> | | | | Clinical condition or purpose:<br>Diagnosis of sleep and<br>respiratory disorders | YES | YES | | Population: Human subjects<br>undergoing sleep studies | YES | YES | | Five-stage Sleep Stage Scoring<br>(Wake, REM, Three non-REM<br>stages) | YES | YES | | Arousal Scoring | YES | YES | | Respiratory Events Scoring | YES | YES | | Leg Movements Scoring | YES | YES | | Performance assessed by<br>percent agreement(and Cohen's<br>kappa) between automatic and<br>human scoring | YES | YES | | Basic operation: processing of<br>polysomnography data<br>recorded from patients in<br>sleep laboratories and<br>polysomnography report<br>generation | YES | YES | | <b>Data inputs for Sleep Stage and Arousal Scoring:</b> | | | | Central<br>electroencephalogram(EEG) | YES | YES | | Left and right eye<br>electroocculogram(EOG) | YES | YES | | Chin electromyogram(EMG) | YES | YES | | Electrocardiogram (ECG) | YES | YES | | <b>Data inputs for Respiratory Events Scoring :</b> | | | | Chest and abdomen<br>movements measured by<br>respiratory bands | YES | YES | | Oxygen saturation | YES | YES | {9}------------------------------------------------ Image /page/9/Picture/0 description: The image shows the logo for Honey Naps. The logo consists of a stylized letter H on the left and the words "HONEY NAPS" on the right. The letter H is made up of two vertical bars connected by a horizontal bar in the middle. The words "HONEY NAPS" are written in a sans-serif font. | Respiratory airflow | YES | YES | |-------------------------------------------------------------|-------------------------------------------------------------------------|-------------------------------------------------------------------------| | Thermister | YES | YES | | Audio | NO | YES | | Body position | NO | YES | | Airway CO2 | NO | YES | | Data inputs for Leg Movement Scoring : | | | | EMG recorded from right and<br>left legs | YES | YES | | Additional Technical Criteria : | | | | Polysomnography records<br>scored per 30 second epoch | YES | YES | | Cardiac artifacts removed from<br>EEG, EMG and EOG channels | YES | YES | | Others : | | | | Standard of scoring manual | The AASM Manual for the Scoring of<br>Sleep and Associated Events, 2016 | The AASM Manual for the Scoring of<br>Sleep and Associated Events, 2007 | | Re-scoring | YES | YES | | File Format | .EDF, .XML, .JSON format | .EDF, . | | Network | Access to Internet is not required. | Access to Internet is required | | Operating System | Windows 10 | Windows XP, Windows 7, Windows 10 | #### VII. PERFORMANCE DATA #### Software Verification & Validation Support for the substantial equivalence of the SOMNUM Software was provided as a result of risk management and software testing. Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guideline for Industry and FDA Staff, "Guidance for the Content of Premarket Submission for Software Contained in Medical Device."(Document issued on : May 11, 2005) - · Verification of each independent software subsystem against defined requirements - · Verification of interfaces between software subsystems against defined interface requirements - · Validation of fully integrated system including all subsystems against overall system requirements The results of the verification and validation activities that have been performed demonstrate that the software meets requirements to support substantial equivalence. #### Performance Test (Non-clinical test) Performance testing was completed by evaluating SOMNUM device performance using a cross-sectional experimental design on a representative N=48 subjects of data recorded in the sleep laboratory. Substantial Equivalence was established through a testing protocol that used full night studies data to evaluate the performance of SOMNUM. The goal of the test reported here is to establish that SOMNUM performance is equivalent to the performance of the predicate device. We confirm that SOMNUM does not present any new concerns about safety or effectiveness compared to the predicate device, MICHELE Sleep Scoring System(K112102). The performance was measured by epoch by epoch agreement between SOMNUM's {10}------------------------------------------------ Image /page/10/Picture/0 description: The image contains the logo for Honey Naps. The logo consists of a stylized "H" made of rounded rectangles and circles, with the left side in a lighter blue and the right side in a darker blue. To the right of the "H" are the words "HONEY" stacked on top of "NAPS", both in the same darker blue color as the right side of the "H". ing and the scoring of three technologists using 2/3 majority rule, that is, at least two out of three export scoring(medical professionals certified on PSG recording and analysis) agree on the presenceof an event within an epoch. SOMNUM device performance was evaluated using the defined test design, statistical methodology, and controls across the following two endpoints: Endpoint 1. As SOMNUM is intended to analyze physiological signals and automatically score sleep study results, including sleep staging and detection of respiratory events, PLMs events, device performance for detecting each event type must be validated. Endpoint 2. As SOMNUM is intended to present summary variables that appear in the clinical report used by physician to assess sleep disorders. #### Test Result Summary #### Data for End point 1 #### Sleep Stage | | | SOMNUM | | | | | Reference[2] | | | | | |---------|----|--------------|--------------|--------------|--------------|--------------|--------------|-------------|-------------|-------------|-------------| | | | W | N1 | N2 | N3 | R | W | N1 | N2 | N3 | R | | | W | <b>89.96</b> | 6.93 | 0.86 | 0.14 | 4.13 | <b>82.2</b> | 35.9 | 4.0 | 0.5 | 5.8 | | | | (89.22- | (6.35- | (0.73- | (0.00- | (3.69- | | | | | | | | | 90.67) | 7.51) | 0.98) | 0.32) | 4.61) | | | | | | | | N1 | 6.15 | <b>77.93</b> | 5.46 | 0.28 | 5.88 | 9.6 | <b>26.1</b> | 4.7 | 0.4 | 8.7 | | | | (5.63- | (76.95- | (5.16- | (0.09- | (5.38- | | | | | | | | | 6.73) | 78.86) | 5.78) | 0.52) | 6.43) | | | | | | | Scorers | N2 | 2.06 | 13.21 | <b>91.13</b> | 15.52 | 5.37 | 3.8 | 33.2 | <b>85.0</b> | 16.0 | 12.3 | | | | (1.75- | (12.42- | (90.71- | (14.02- | (4.84- | | | | | | | | | 2.41) | 13.96) | 91.47) | 16.95) | 5.84) | | | | | | | | N3 | 0.01 | 0.07 | 1.84 | <b>84.04</b> | 0.00 | 0.2 | 0.2 | 5.4 | <b>82.3</b> | 0.0 | | | R | 1.81 | 1.86 | 0.72 | 0.05 | <b>84.62</b> | 4.2 | 4.7 | 0.8 | 0.8 | <b>73.2</b> | | | | (1.49- | (1.53- | (0.60- | (0.00- | (83.81- | | | | | | | | | 2.14) | 2.19) | 0.84) | 0.14) | 85.48) | | | | | | Table 3. Performance Comparison of sleep stage. #### Arousal #### Table 4. Performance Comparison of sleep stage. | SOMNUM | | | | Reference[3] | | | | |-------------------------|------------------------|------------------------|---------------------|-------------------------|-------|-------|-----| | Event<br>Total<br>Epoch | PPA | NPA | OPA | Event<br>Total<br>Epoch | PPA | NPA | OPA | | 13108 | 82.02<br>(81.42-82.64) | 82.78<br>(82.41-83.17) | 82.5<br>(82.2-82.9) | 6468 | 76.82 | 82.48 | - | {11}------------------------------------------------ Image /page/11/Picture/0 description: The image contains the logo for Honey Naps. The logo consists of a stylized letter "H" made of two vertical bars connected by a horizontal bar in the middle. The left vertical bar is light blue, while the right vertical bar is dark blue. To the right of the "H" are the words "HONEY" and "NAPS" stacked on top of each other, both in dark blue. #### SDB & PLMS | | SOMNUM | | | | Reference[4] | | | Reference[1] | | | | | |------|-------------------------|------------------------|------------------------|---------------------|-------------------------|------|------|--------------|-------------------------|----------------|----------------|----------------| | | Event<br>Total<br>Epoch | PPA | NPA | OPA | Event<br>Total<br>Epoch | PPA | NPA | OPA | Event<br>Total<br>Epoch | PPA | NPA | OPA | | SDB | 15984 | 94.18<br>(93.79-94.52) | 91.29<br>(90.98-91.61) | 92.3<br>(92.1-92.6) | 2439 | 75.5 | 98.1 | 93.0 | 4705 | 67<br>(58, 75) | 93<br>(92, 94) | 91<br>(90, 92) | | PLMS | 5215 | 92.89<br>(92.18-93.55) | 94.30<br>(94.08-94.51) | 94.1<br>(93.9-94.4) | 1741 | 78.4 | 97.6 | 95.7 | 5796 | 71<br>(60, 80) | 90<br>(89, 92) | 89<br>(87, 90) | #### Table 5. Performance Comparison of sleep stage. Please note that Event total Epoch means the total number of epochs with events. The total epoch is 43971 epochs (event epoch + non-event epoch). #### Discussion: For Sleep Stage and Arousal, SOMNUM exceeded all performance targets, while for SDB and PLMS, SOMNUM exceeded the performance targets for PPA. Comparing NPA of SOMNUM with that of predicate device, the NPA of SOMNUM' is 3.3% lower. however, it is unknown whether a statistically significant comparison can be made due to lack of Cl. Therefore, the NPA of K162627 was compared as the reference device, and in this case, SOMNUM showed good performance by 4.3%. In conclusion, the NPA performance for PLMS is considered to be within clinically acceptable range. The NPA for SDB showed a performance that was 1.7% lower than K162627 and about 6.8% lower than K112102. However, considering both PPA and OPA of SOMNUM, the overall performance of SOMNUM is better than that of K162627. In addition, considering the situation where there is no standard reference, it cannot be said that this difference shows the low performance of SOMNUM. Furthernore, the manual scorer agreement[8], which is generally acceptable in clinical practice, is considered to have an error range of around 15%, these performance differences are considered to be within the clinically acceptable error range. In conclusion, SOMNUM passed all pass/fail criteria for Endpoint 1. Data for End point 2: Table 6. Performance Comparison of Relevant Scoring Variables Between SOMNUM and Performance Target for Endpoint | Variable | Type Of<br>Limit | SOMNUM | abs. MAX | Target | abs. MAX | Unit | |----------|------------------|--------|----------|--------|----------|-------| | TST | U | 10 | 20 | 60 | 120 | [min] | | | L | -20 | | -120 | | | {12}------------------------------------------------ | | NAPS | | | | | | Variable | LOA | SOMNUM | References | Ref. No. | Unit | |-------------------------------------|------|-------|-----|------|-------------|-------------|--------------------------------------------------------------|-----|--------|------------|----------|-------------| | SE | U | 2.5 | 5 | 10 | 13 | % | TST<br>(Total Sleep Time) | U | 7.8 | 35 | [5] | [min] | | | L | -5 | | -13 | | | | L | -14.21 | -70 | | | | SOL | U | 8 | 23 | 40 | 40 | [min] | SE<br>(Sleep Efficiency) | U | 1.69 | 10 | | % | | | L | -23 | | -5 | | | | L | -3.31 | -12 | [7] | | | ROL | U | 120 | 120 | 170 | 170 | [min] | SOL<br>(Sleep On Latency) | U | 2.93 | 15 | [7] | [min] | | | L | -70 | | -130 | | | | L | -4.21 | -11 | | | | Wake | U | 22 | 22 | 60 | 60 | [min] | ROL<br>(REM On Latency) | U | 62.01 | 70 | [5] | [min] | | | L | -10 | | -45 | | | | L | -60.69 | -90 | | | | N1 | U | 30 | 45 | 80 | 80 | [min] | Wake | U | 14.21 | 70 | [7] | [min] | | | L | -45 | | -60 | | | | L | -7.80 | -45 | | | | N2 | U | 65 | 65 | 70 | 120 | [min] | N1 | U | 25.30 | 30 | [5] | [min] | | | L | -50 | | -120 | | | | L | -17.29 | -30 | | | | N1_N2 | U | 55 | 55 | 70 | 70 | [min] | N2 | U | 31.61 | 10 | [5] | [min] | | | L | -35 | | -60 | | | | L | -29.10 | -75 | | | | N3 | U | 30 | 40 | 140 | 140 | [min] | N1_N2 | U | 31.58 | 30 | [5] | [min] | | | L | -40 | | -10 | | | | L | -19.17 | -75 | | | | REM | U | 20 | 40 | 50 | 80 | [min] | N3 | U | 18.83 | 65 | [5] | [min] | | | L | -40 | | -80 | | | | L | -30.45 | -5 | | | | Arousal Index | U | 30 | 30 | - | | events/hour | REM | U | 6.07 | 18 | [5] | [min] | | | L | -20 | | - | - | | | L | -31.95 | -55 | | | | PLMS Index | U | 43 | | 7 | | events/hour | Arousal Index | U | 23.91 | - | - | events/hour | | | L | -2 | 43 | -43 | 43 | | | L | -15.21 | - | | | | AHI Index | U | 18 | | 7 | | events/hour | PLMS Index<br>(Periodic Limb<br>Movements of Sleep<br>Index) | U | 12.24 | 13 | | events/hour | | | L | -7 | 18 | -20 | 20 | | | L | -0.37 | -15 | [6] | | | AHI Index<br>(Apnea-Hypopnea Index) | U | 8.79 | 4 | [5] | events/hour | | | | | | | | | | L | -3.28 | -2 | | | | | | | | | | Discussion: TST, SE, ROL, Wake, N1, N2, N1_N2, and REM satisfy the provided Upper and Lower LOA. However, SOL has a target of -5 to 40, while SOMNUM has a range of -23 to 8, and N3 has a target of -10 to 140, while SOMNUM has a range of -40 to 30. The PLMS index has a target of -40 to 10, but SOMNUM has a range of -2 to 43, and the AHI index has a target of -20 to 7, but SOMNUM has a range of -7 to 18. When considering the actual range of absolute errors, SOL has a target of 40 minutes, while SOMNUM has 23 minutes, and N3 has a target of 140, while SOMNUM has 40 minutes. The AHI index has a target of 22, but SOMNUM has 18, resulting in smaller absolute errors. PLMS index has a target of 43, while SOMNUM has 43, it shows the almost same value. However, considering the 15% error among scorers presented in Reference [8], it cannot be said that it affects the performance. In addition, we compared the LOA of Relevant Sleep Scoring Variables of SOMNUM and those of References. The table 12 showed the LOA of SOMNUM and References. {13}------------------------------------------------ Image /page/13/Picture/0 description: The image shows the logo for Honey Naps. The logo consists of a stylized letter "H" on the left, with the word "HONEY" above the word "NAPS" on the right. The letter "H" is made up of two vertical bars connected by a horizontal bar, with circles at the top and bottom of each vertical bar. The left vertical bar is light blue, and the right vertical bar is dark blue. The words "HONEY NAPS" are also dark blue. #### Table 7. Performance Comparison of LOA between SOMNUM and References for Endpoint 2 When comparing the LOA ranges of SOMNUM and References, the CI ranges of SOMNUM is narrower than those of the references except for AHI. In conclusion, SOMNUM passed all pass/fail criteria for Endpoint 2. ## Output Comparison Data {14}------------------------------------------------ Image /page/14/Picture/0 description: The image contains the logo for Honey Naps. The logo consists of a stylized letter "H" on the left, with the words "HONEY" and "NAPS" stacked on top of each other to the right of the "H". The "H" is made up of two vertical bars connected by a horizontal bar in the middle, with circles at the top and bottom of each vertical bar. The logo is colored in shades of blue. | SCORING<br>FUNCTION | SOMNUM | | | | | ENSOSLEEP(K162627) | | | | | |-----------------------|-----------------|---------------------|----------------------|---------------------|---------------------|--------------------|---------------|----------------|---------------|--------------| | | Total<br>Epochs | PPA<br>(%) | NPA<br>(%) | OPA<br>(%) | kappa<br>(%) | Total<br>Epochs | PPA<br>(%) | NPA<br>(%) | OPA<br>(%) | kappa<br>(%) | | SLEEP<br>STAGING | 43971 | | | 87.5<br>(87.2,87.8) | 82.1<br>(81.6,82.5) | 59719 | 78<br>(77,80) | 95<br>(94,95) | 91<br>(91,92) | | | Awake | 6719 | 89.9<br>(89.2,90.6) | 97.4<br>(97.3,97.6) | | | 17459 | 86<br>(82,88) | 97<br>(95,98) | 94<br>(92,95) | | | N1 | 6522 | 77.9<br>(77.0,78.9) | 94.6<br>(94.4,94.9) | | | 3293 | 41<br>(33,48) | 94<br>(93, 96) | 91<br>(90,93) | | | N2 | 20448 | 91.1<br>(90.7,91.5) | 92.4<br>(92.0,92.7) | | | 26839 | 77<br>(73,81) | 87<br>(85,90) | 83<br>(80,85) | | | N3 | 1012 | 84.1<br>(82.6,85.6) | 99.1<br>(99.0,99.2) | | | 5587 | 81<br>(74,88) | 93<br>(91,95) | 92<br>(90,94) | | | REM | 7364 | 84.7<br>(83.8,85.4) | 98.9<br>(98.8,99.0) | | | 6541 | 79<br>(72,84) | 99<br>(98,99) | 96<br>(96,97) | | | No Consensus | 756 | | | | | 1432 | | | | | | AROUSAL | 43971 | | | 82.5<br>(82.2,82.9) | 60.8<br>(60.1,61.5) | 59719 | | | 87<br>(85,88) | - | | Yes | 13108 | 82<br>(81.4,82.6) | 82.8<br>(82.4,83.2) | | | 7686 | 66<br>(61,71) | 90<br>(88,91) | | | | None | 30863 | | | | | | | | | | | No Consensus | 0 | | | | | 0 | | | | | | PLMS | 43971 | | | 94.1<br>(93.9,94.4) | 75.7<br>(74.7,76.5) | 59719 | | | 89<br>(87,90) | - | | Yes | 8509 | 92.9<br>(92.1,93.6) | 94.3<br>(94.1, 94.5) | | | 5796 | 71<br>(60,80) | 90<br>(89,92) | | | | None | 35462 | | | | | | | | | | | No Consensus | 0 | | | | | 0 | | | | | | RESPIRATORY<br>EVENTS | 43971 | | | 92.3<br>(92.1,92.6) | 83.8<br>(83.2,84.3) | 59719 | | | 91<br>(90,92) | - | | YES | 15984 | 94.2<br>(93.8,94.5) | 91.3<br>(91.0,91.6) | | | 4705 | 67<br>(58,75) | 93<br>(92,94) | | | | None | 27987 | | | | | | | | | | | No Consensus | 0 | | | | | 0 | | | | | Discussion : Except for the NPA of Arousal, SOMNUM is superior to the predicate device in all PPA and NPA. However, the NPA of arousal is about 8% lower, but this is within the limit that can be seen between general scorers. The manual scorer agreement[8], which is generally acceptable in clinical practice, is considered to have an error range of around 15%, these performance differences are considered to be within the clinically acceptable error range. {15}------------------------------------------------ Image /page/15/Picture/0 description: The image contains the logo for Honey Naps. The logo consists of a stylized "H" made of two vertical bars connected by a horizontal bar in the middle. The "H" is colored in shades of blue. To the right of the "H" is the text "HONEY NAPS" in a sans-serif font, with "HONEY" in a lighter blue and "NAPS" in a darker blue. #### Clinical Test Summarv No clinical studies were considered necessary and performed. #### Summary In summary, performance test results demonstrated that SOMNUM achieved sleep disordered breathing, arousal, PLMs events detection agreement that is substantially equivalent to the predicate device Ensosleep (K162627) and MICHELE Sleep Scoring(K112102) performance for all comparisons in all endpoints analyzed respectively. The SOMNUM performance validation testing demonstrates the safety and effectiveness of SOMNUM whenused for the defined indications for use. The performance data demonstrates that the SOMNUM device performs comparably to the predicate device that is currently marketed for the same intended use. #### References [1] FDA, 510(k) Summary(MICHELE Sleep Scoring System, K112102) [2] DJ Levendowski, et. al. "The Accuracy, Night-to-Night Variability, and Stability of Frontopolar Sleep Electroencephalography Biomarkers" Journal of Clinical Sleep Medicine, Vol. 13, No. 6, pp.791 - 803, 2017 [3] F De Carli, et. al, "A Method for the Automatic Detection of Arousals During Sleep, Vol. 22, No. 5, pp. 561~572, 1999 [4] FDA, K162627 510K summary report [5] Atul Malhotra, et. al., "Performance of an Automated Polysomnography Scoring System Versus Computer-Assisted Manual Scoring" SLEEP, Vol. 36, No. 4, pp. 573-582, 2013 (6) S. D. Pittman, et. al., "Assessment of Automated Scoring of Polysomnographic Recordings in a Population with Suspected Sleep-disordered Breathing", SLEEP, Vol. 27, No. 7, 2004 pp.1394-1403 [7] C. Stepnowsky, et. al, "Scoring accuracy of automated sleep staging from a bipolar recording compared to manual scoring by multiple raters, Sleep Medicine 14 (2013) 1199-1207 [8] T. Penzel, "Sleep scoring moving from visual scoring", SLEEP, 2022, Vol. 45, No. 10. pp. 1 – 3. {16}------------------------------------------------ Image /page/16/Picture/0 description: The image shows the logo for Honey Naps, which is a blue stylized H with two circles on the left, and the words "HONEY NAPS" in blue on the right. Below the logo, the words "VIII. CONCLUSIONS" are written in black. The text is in a simple, sans-serif font. Based on a comparison of the intended use and technological characteristics, performance test, the SOMNUM software is substantially equivalent to the identified predicate device. Minor differences in technological and performance characteristics did not raise new or different questions of safety and effectiveness. Additionally, the non-clinical testing supports that the system performs in accordance with its intended use and is as safe, as effective, and performs as well as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...