K162627 · Ensodata, Inc. · OLZ · Mar 31, 2017 · Neurology
Device Facts
Record ID
K162627
Device Name
EnsoSleep
Applicant
Ensodata, Inc.
Product Code
OLZ · Neurology
Decision Date
Mar 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K162627 · Mar 31, 2017
EnsoSleep
Ensodata, Inc.
Retrospective clinical polysomnography (PSG) records
Retrospective clinical PSG data were used to evaluate the performance of the EnsoSleep software in staging sleep, detecting sleep disordered breathing, arousals, and leg movements, and to establish substantial equivalence to predicate devices.
EnsoSleep is intended for use for the diagnostic evaluation by a physician to assess sleep quality and as an aid for the diagnosis of sleep and respiratory related sleep disorders in adults only. EnsoSleep is a software-only medical device to be used under the supervision of a clinician to analyze physiological signals and automatically score sleep study results, including the staging of sleep, detection of arousals, leg movements, and sleep disordered breathing events including obstructive apneas. All automatically scored events are subject to verification by a qualified clinician. Central apneas, mixed apneas, and hypopneas must be manually marked within records.
Device Story
EnsoSleep is a software-only medical device that analyzes previously recorded physiological signals (EDF/EDF+ files) from sleep studies. It uses automated algorithms to score sleep stages (Wake, N1-N3, REM), respiratory events (obstructive apneas, hypopneas), arousals, and periodic leg movements. The software operates as a standalone application on Windows 7/8, with processing, scoring, and analysis performed on cloud servers. Clinicians use the output to assist in diagnosing sleep disorders; all automated scores require manual verification by a qualified clinician. The device improves workflow efficiency through automated study initiation, upload, and scoring, and includes cybersecurity controls like end-to-end encryption and authentication. It benefits patients by providing rapid, standardized analysis of sleep data to support clinical decision-making.
Clinical Evidence
Clinical performance evaluated using a retrospective cross-sectional study of 72 subjects (59,719 epochs). Performance compared against a 2/3 majority scoring reference. Endpoints included sleep staging, sleep apnea diagnostic agreement (AHI thresholds), and event detection (SDB, arousal, leg movements). Results showed EnsoSleep achieved positive, negative, and overall agreement substantially equivalent to the Sleep Profiler (K153412) and MICHELE (K112102) benchmarks. 95% percentile bootstrap confidence intervals were calculated for all metrics.
Technological Characteristics
Software-only device; processes EDF/EDF+ files. Cloud-based processing architecture. Compatible with Windows 7/8. Features automated signal quality rejection, end-to-end encryption, authentication, and vulnerability scanning. Scoring based on AASM Manual for Scoring and Associated Events. No physical hardware components.
Indications for Use
Indicated for adult patients undergoing diagnostic evaluation for sleep quality and sleep/respiratory-related sleep disorders. Used under clinician supervision to assist in scoring sleep studies.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
Predicate Devices
Advanced Brain Monitoring, Inc. Sleep Profiler (K153412)
Reference Devices
Younes Sleep Technologies MICHELE Sleep Scoring System (K112102)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem that resembles an abstract caduceus, with three stylized human profiles forming the staff and a serpent-like shape winding around them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2017
EnsoData, Inc. % Seth Mailhot Partner Michael Best & Friedrich, LLP 601 Pennsylvania Ave, NW Suite 700 South Washington, District of Columbia 20004
Re: K162627
Trade/Device Name: EnsoSleep Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: OLZ Dated: March 27, 2017 Received: March 29, 2017
Dear Mr. Mailhot:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162627
Device Name EnsoSleep
### Indications for Use (Describe)
EnsoSleep is intended for use for the diagnostic evaluation by a physician to assess sleep quality and as an aid for the diagnosis of sleep and respiratory related sleep disorders in adults only. EnsoSleep is a software-only medical device to be used under the supervision of a clinician to analyze physiological signals and automatically score sleep study results, including the staging of sleep, detection of arousals, leg movements, and sleep disordered breathing events including obstructive apneas. All automatically scored events are subject to verification by a qualified clinician. Central apneas, mixed apneas, and hypopneas must be manually marked within records.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|---------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitted by: | EnsoData, Inc. |
|-------------------|----------------------------------------------------------------------------------|
| Address: | 111 N. Fairchild Street, Suite 240 |
| | Madison, WI, 53703 |
| Telephone: | (608) 509-4704 |
| Contact Name: | Chris Fernandez, Co-founder and CEO |
| Date Submitted: | August 13, 2016 |
| Trade Name: | EnsoSleep |
| Common Name: | Automatic Event Detection Software for Polysomnograph with Electroencephalograph |
| Product Code: | OLZ |
| Regulatory Class: | II (21 C.F.R. 882.1400) |
| Review Panel: | Neurology |
| Predicate Device: | Advanced Brain Monitoring, Inc. Sleep Profiler (K153412) |
| Reference Device: | Younes Sleep Technologies MICHELE Sleep Scoring System (K112102) |
#### Device Description:
EnsoSleep is a software application that analyzes previously recorded physiological signals obtained during sleep. The EnsoSleep software can analyze any EDF or EDF+ files.
Automated algorithms are applied to the raw signals in order to derive additional signals and interpret the raw and derived signal information. The software automates recognition of:
Sleep Stage Events
- Wake
- Stage N1
- Stage N2
- · Stage N3
- Stage REM
Respiratory Events
- · Sleep disordered breathing (apneas and hypopneas)
- · Apneas detected with airflow signal are classified as obstructive apnea (OSA), and can be edited to be central or mixed appeas
- · Sleep disordered breathing events not detected to be apneas are marked as hypopnea
- · Central apneas, mixed apneas, and hypopneas must be manually marked within records
Arousal Events
· Arousals
- Movement Events
- · Periodic Leg Movements during Sleep (PLMS)
The EnsoSleep software can be used as a stand-alone application for use on Microsoft Windows 7 & 8 operating system platforms. All processing, scoring, and analysis of signal data occurs on the EnsoSleep cloud servers.
#### Indications for Use:
EnsoSleep is intended for use for the diagnostic evaluation by a physician to assess sleep quality and as an aid for the diagnosis of sleep and respiratory related sleep disorders in adults only. EnsoSleep is a software-only medical device to be used under the supervision of a clinician to analyze physiological signals and automatically score study results, including the staging of arousals, leg movements, and sleep disordered breathing events including obstructive apneas. All automatically scored events are subject to verfication by a qualified clinician. Central apneas, and hypopneas must be manually marked within records.
#### Determination of Substantial Equivalence:
#### Summary of Technology:
The EnsoSleep software-only device is similar in inctionality to the Sleep Profiler (K153412) by Advanced Brain Monitoring, Inc. and Younes Sleep Technologies MCHELE Sleep Scoring System (K112102) electroencephalograph analysis software programs. The EnsoSleep device is similar with respect to indications for use and physical characteristics to the predicate device and
{4}------------------------------------------------
reference device in terms of 510(k) substantial equivalency.
Based on analysis and comparison of technological characteristics and features between EnsoSleep and reference device, EnsoSleep is determined to have the same technological characteristics as the predicate and reference devices, and does not raise different questions of safety or efficacy as demonstrated by the device design. EnsoSleep uses the same fundamental technology as the legally marketed predicate device; automated algorithms are applied to raw signals in order to derive additional signals and interpret raw and derived signal information. Each of the EnsoSleep, Sleep Profiler (K153412), and MCHELE Sleep Scoring System (K112102) devices include as features the ability for full disclosure recording of derived signals and automated analyses to be visually inspected and edited prior to the results being integrated into one of several sleep study report data formats. Each of the EnsoSleep, Sleep Profiler (K153412), and MCHELE Sleep Scoring System (K112102) devices base the automatic scoring of physiological events on the American Academy of Sleep Medicines, definitions, and procedures. Additionally, the EnsoSleep predicate device Sleep Profiler (K153412, K13007, K120450), in an earlier cleared version of the device, utilized MCHELE Sleep Scoring (K112102) as its own predicate device for establishing substantial equivalence validation testing comparisons.
Both the predicate Sleep Profiler (K153412) and EnsoSleep devices can be used as a stand-alone software application with a user interface delivered on the Microsoft Windows 7 or 8 operating system platforms, and automatically reject periods of poor EEG signal quality. The following features were not included in the EnsoSleep application as they were deemed unnecessary based on end-user feedback and/or the fact that they are not requirement AASM guidelines: detection of heart rate, head position, snoring levels, head movements, and respiratory event related apneas (RERAs); disease management comments; two-night reports. The chosen EnsoSleep design and implementation features scoing, efficiency, reliability, and security improvements over Sleep Profiler (K153412) and MICHELE Sleep Scoring System (K112102). EnsoSleep provides more comprehensive coverage of the event types specified for scoring by the AASM Manual for Scoring and Associated Events recommendations than the Sleep Profiler (K153412) device by automatically scoring leg movement events in addition to the sleep disordered breathing, obstructive apnea, arousal, and sleep staging functionality supported by both the predicate and reference devices. Both the Sleep Profiler (K153412) predicate device and EnsoSleep classify each apnea as obstructive, each non-apnea sleep disordered breathing event as a hypopnea (excluding RERAs), and enable apneas to be edited and manually marked to be central or mixed apneas within records.
EnsoSleep differs from Sleep Profiler (K153412) and MICHELE Sleep Scoring System (K112102) in that EnsoSleep automates the initiation, upload, scoring, and download of studies, enabling a user experience that is optimized for efficiency and fast analysis prior to the required user over-read of EnsoSleep scring, Futhermore, the high-performance specifications of the EnsoSleep Processing Platform enables throughput efficiency and scalability, as the speed by which a study processes is dependent on the number of distributed servers, parallel computer processor(s), and the amount of RAM. EnsoSleep improves network protocols with automatic connection-recovery that are robust to poor or interrupted network conditions. Finally, EnsoSleep provides cybersecurity improvements over the predicate Sleep Profiler (K153412) and MCHELE Sleep Scoring System (K112102) with verified authentication, authorization, and access controls and checksum controls including end-to-end encryption; secure software distribution mechanisms and controls, intrusion systems and vulnerability scanning; and other network, systems, and database controls.
#### Summary of Non-Clinical Tests:
Support for the substantial equivalence of EnsoSleep was provided by risk management and software testing. Both EnsoSleep and Sleep Profiler (K153412) conduct and document verification, and performance testing as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The EnsoSleep software has been thorough verification of specifications. One or more verification tests are provided for each requirement specified with detailed protocols, objective passfial criteria, and clearly documented test executions with results. Detailed plans and protocols were developed prospectively for system level, performance, and usability validation testing was conducted with qualified clinical and non-clinical users, with objective passffall criteria, ability to provide comments, and with all testing reports and results documented for review.
#### Summary of Clinical Tests:
Substantial equivalence was also established through a testing protocol that used clinical polysomnography (PSG) data to evaluate the performance of EnsoSleep. Clinical performance testing was completed by evaluating EnsoSleep device performance using a cross-sectional experimental design on a representative N=72 subject sample of retrospective clinical PSG data. First, the intended use population, study population, conditions of interest, designated comparative reference, designated comparative benchmarks, and experimental endpoints were defined. Upon applying predeined selection controls, a statistically representative sample of the defined intended use and user population sample consisting of N=22 archived collection of retrospective diagnostic clinical PSG data collected from an AASM Accredity. Same as the predicate and reference devices respectively, the study population was then sent to a clinical testing laboratory where each PSG was manually scored by three (3) independent registered sleep technologists (RPSGT) that met all acquisition, sooring-bind, and rater controls. A designated comparative reference was constructed using 2/3 Majority Scoring to evaluate the EnsoSleep device performance versus the predicate Sleep Profiler and MICHELE device performance benchmarks and acceptance criteria of positive percent agreement (PA), negative percent agreement (NA), and overall percent (OA) were predefined competitively based on analysis of sleep staging event detection and diagnostic agreement performance reported in the predicate device 510(k) documentation. The predicate device did not report dinical testing results for shing, arousal, and leg movement event detection agreement performance, and therefore the reference device clinical performance and 510(k) documentation were used to facilitate a valid comparison. EnsoSleep device performance was evaluated using the definental design, statistical methodology, and controls, across the following three (3) experimental endpoints:
- 1. Endpoint 1: As EnsoSleep is intended to assist clinicians with the assessment of sleep scoring must be validated. For Endpoint 1, EnsoData evaluated a performance goal comparing the predicated ovice Sleep Profiler (K153412) PA, NA, and OA sleep staging performance, and the bootstrapped point estimate of EnsoSleep sleep staging PA, NA, and OA versus a 2/3 Majority Scoring reference.
{5}------------------------------------------------
- 2. Endpoint 2: As EnsoSleep is intended to assist clinicians with the scoring sleep disordered breathing events used in diagnostic evaluation, device performance for diagnosing sleep apnea must be validated. For Endpoint 2, EnsoData evaluated a performance goal comparing the predicate device Sleep Profiler (K153412) PA. NA, and OA diagnostic agreement performance, and the bootstrapped point estimate of median performance for EnsoSleep diagnostic agreement PA, NA, and OA versus a 2/3 Majority Scoring reference.
- 3. Endpoint 3: As EnsoSleep is intended to analyze physiological signals and automatically score sleep study results, including detection of sleep disordered breathing events, arousal events, and leg movement events, device performance for detecting each event type must be validated. For Endpoint 3, EnsoData evaluated a performance goal comparing the reference device MICHELE Sleep Scoring System (K112102) PA, NA, and OA event detection performance, and the bootstrapped point estimate of median performance for EnsoSleep event detection PA, NA, and OA versus a 2/3 Majority Scoring reference.
The final experimental results and statistical and ont-on-NA, OA performance with 95% percentile bootstrap confidence intervals (R=1000 resamples) were calculated by overall-pochs versus 2/3 Majority Scoring in event detection experiments evaluating Wake, N1, N2, N3, REM, SDB, Apneal, and Leg Movement event detection performance respectively. Furthermore. EnsoSleep device diagnostic agreement was evaluated versus 2/3 Majority on both mild and moderate sleep apnea diagnostic thresholds by computing the overall and REM-only apneal. bootstrapped point-estimates for median PA, NA, OA, performance with 95% percentile bootstrap confidence intervals (Re1000 resamples) and likelihood ratio pairs were computed in diagnosic agreement experiments evaluating overall-mild AHI, REM-mid AHI, and REM-moderate AHI. The final experimental results are summarized for each endpoint in Table 1, Table 2, and Table 3 below.
Table 1 shows the EnsoSleep and Sleep Profiler (K153412) clinical performance results compared for Endboint 1, sleep staging. For all sleep staging event types evaluated in EnsoSleep PA, NA, and OA performance was observed to show no statistically significant differences or was observed to be significantly greater in all comparisons relative to the predicate device. There were no cases where predicate device Wake, N1, N2, N3, REM, or Total Staging performance was statistically greater than EnsoSleep event detection performance (i.e. the precicate device point estimate for PA. NA, and OA were never higher than EnsoSleep PA, NA, and OA 95% CI upper bounds). The results confirm EnsoSleep achieves clinical performance for sleep staging positive, negative, and overall agreement that is substantially equivalent to the Sleep Profiler positive, and overall agreement across all sleep stages.
| Table 1: Sleep Staging Clinical<br>Performance Comparisons | | | Overall-Epochs EnsoSleep<br>vs 2/3 Majority Sleep Staging Performance | | | Overall-Epochs Sleep Profiler (K153412)<br>vs 2/3 Majority Sleep Staging Performance | | | |
|------------------------------------------------------------|-------|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|---------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--|--|--|
| | | (N=72 subjects,<br>59719 epochs) | Bootstrapped point-estimate of median<br>Percent Agreement (%) with 95%<br>percentile bootstrap confidence interval<br>(R=1000 resamples) | (N=43 subjects,<br>31361 epochs) | Point-estimate of Percent Agreement (%) | | | | |
| | | Total Epochs | Positive Agreement<br>(PA) Negative Agreement<br>(NA) Overall Agreement<br>(OA) | Total Epochs | Positive Agreement<br>(PA) Negative Agreement<br>(NA) Overall Agreement<br>(OA) | | | | |
| | Wake | 17459 | 86%<br>(82%, 88%) 97%<br>(95%, 98%) 94%<br>(92%, 95%) | 7424 | 73% 94% 89% | | | | |
| Overall-epochs<br>assigned by 2/3<br>Majority Scoring | N1 | 3293 | 41%<br>(33%, 48%) 94%<br>(93%, 96%) 91%<br>(90%, 93%) | 1752 | 25% 93% 89% | | | | |
| | N2 | 26839 | 77%<br>(73%, 81%) 87%<br>(85%, 90%) 83%<br>(80%, 85%) | 12582 | 77% 84% 81% | | | | |
| | N3 | 5587 | 81%<br>(74%, 88%) 93%<br>(91%, 95%) 92%<br>(90%, 94%) | 4704 | 76% 94% 91% | | | | |
| | REM | 6541 | 79%<br>(72%, 84%) 99%<br>(98%, 99%) 96%<br>(96%, 97%) | 3749 | 74% 97% 95% | | | | |
| | Total | 59719 | 78%<br>(77%, 80%) 95%<br>(94%, 95%) 91%<br>(91%, 92%) | 31361 | 73% 93% 87% | | | | |
| | None | 1432 | - - - | 1150 | - - - | | | | |
Table 2 shows the EnsoSleep and Sleep Profiler (K153412) dinical performance results compared for Endposition agreement. For all diagnostic agreements evaluated in Endpoint 2, mild-REM, moderate-overall, and moderate-REM, EnsoSleep PA and NA performance were observed to show no statistically significant differences compared to predicate device performance (e.g. the two-sided 95% percentle bounds contained the predicate device point estimates for PA and NA in all cases). The only statistically significant difference observed was for overall-mild PA, with a 2% difference in the EnsoSleep Cl upper bound and Sleep Profiler point estimate (91% (82%,98%) vs. 100%) and no statistically significant differences in overall-mild NA. Furthermore, the point estimate of EnsoSleep PA, NA, and OA performance exceeded, were equivalent to, or were within 10% of the predicate device PA and NA point estimates for all comparisons. The EnsoSleep positive likelihood ratios were observed to be above 3.5 for overall/REM-mild and above 5.0 for overall/REM-moderate in all diagnostic agreement experiments, similarly exceeding the performance goal targeted by Sleep Profiler (K153412) in the predicate device 510(k) documentation. The resoSleep achieves clinical performance for positive sleep apnea diagnostic agreement that is substantially equivalent to the Sleep Profiler positive and negative agreement across all mild-REM, moderate-overall, and moderate-REM comparisons.
{6}------------------------------------------------
| | Per-Patient EnsoSleep vs 2/3 Majority<br>Sleep Apnea Diagnostic Agreement | | Per-Patient Sleep Profiler (K153412) vs 2/3<br>Majority<br>Sleep Apnea Diagnostic Agreement | | | | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|------------------------|---------------------------------------------------------------------------------------------|----------------------------|----------------------------|-----------------------------|--------------------------------|---------------------------------|
| Table 2: Sleep Apnea Diagnostic Agreement<br>Clinical Performance Comparisons<br><br>EnsoSleep bootstrapped point-estimate of<br>median Percent Agreement (%) with 95%<br>percentile bootstrap confidence interval<br>(R=1000 resamples), Sleep Profiler<br>point-estimate of Percent Agreement (%), and<br>likelihood ratio pairs | EnsoSleep<br>AHI >= 5 | EnsoSleep<br>AHI >= 15 | EnsoSleep REM<br>AHI >= 5 | EnsoSleep REM<br>AHI >= 15 | Sleep Profiler<br>AHI >= 5 | Sleep Profiler<br>AHI >= 15 | Sleep Profiler REM<br>AHI >= 5 | Sleep Profiler REM<br>AHI >= 15 |
| Sample size (N) | 72 | 72 | 72 | 72 | 60 | 60 | 40 | 40 |
| Positive Agreement (PA) | 91%<br>(82%, 98%) | 95%<br>(83%, 100%) | 83%<br>(72%, 94%) | 79%<br>(56%, 94%) | 100% | 100% | 84% | 73% |
| Negative Agreement (NA) | 76%<br>(61%, 90%) | 98%<br>(94%, 100%) | 89%<br>(79%, 97%) | 96%<br>(90%, 100%) | 85% | 97% | 90% | 96% |
| Overall Agreement (OA) | 85%<br>(77%, 92%) | 97%<br>(93%, 100%) | 86%<br>(79%, 93%) | 92%<br>(85%, 97%) | N/A | N/A | N/A | N/A |
| Likelihood ratio (+) | 3.76 | 52.25 | 7.71 | 22.00 | 6.67 | 9.50 | 8.84 | 18.33 |
| Liklihood ratio (-) | 0.12 | 0.05 | 0.19 | 0.22 | 0.00 | 0.00 | 0.00 | 0.28 |
Table 3 shows the EnsoSleep and MICHELE Sleep Scoring (K112102) clinical performance results compared for Endpoint 3, event detection agreement. The MCHELE Sleep Scoring performance testing did not calculate OA for individual event types in the same way as EnsoSleep, and as such only PA and NA comparisons were made to avoid biased performance evaluation. For all event detection experiments including SDB, OSA, Arousal, and Leg Movement event types, the point-estimates of EnsoSleep PA and NA event detection performance exceeded, were equivalent to, or were within 10% of the refermance, with statistically significant differences observed in a minority of cases. On the basis that EnsoSleep met or exceeded objective PA and NA performance goals for these event types in all comparisons, and the reference device did not provide information to adequately compare the statistical significance of results (i.e. two-sided confidence intervals for point-estimates of agreement were not reported), EnsoSleep is considered substantially equivalent to the reference device SDB, Arousal, and Leg Movement event detection performance.
| Table 3: Event Detection Clinical<br>Performance Comparisons | | | | Overall-Epochs MICHELE (K112102)<br>vs 2/3 Majority Event Detection Performance | | | | | |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|----------------|-------------------------|-------------------------|------------------------|-------|
| Overall-Epochs EnsoSleep vs 2/3 Majority Event Detection Performance | | | | | | | | | |
| (N=72 subjects, 59719 epochs) | Bootstrapped point-estimate of median Percent Agreement (%) with 95% percentile bootstrap confidence interval (R=1000 resamples) | (N=30 subjects, 24987 epochs) | Weighted average of Positive and Negative Percent Agreement (%) point-estimates and Overall Percent Agreement (%) point-estimate | | | | | | |
| | Total Epochs | Positive Agreement (PA) | Negative Agreement (NA) | Overall Agreement (OA) | Total Epochs | Positive Agreement (PA) | Negative Agreement (NA) | Overall Agreement (OA) | |
| Overall-epochs assigned by 2/3 Majority Scoring | Sleep Disordered Breathing Events (apneas and hypopneas) | 4705 | 67% (58%, 75%) | 93% (92%, 94%) | 91% (90%, 92%) | 2439 | 75.5% | 98.1% | 93.0% |
| | Apnea Events | 1690 | 56% (41%, 70%) | 96% (96%, 97%) | 95% (95%, 96%) | 750 | N/A | N/A | N/A |
| | Obstructive Apnea Events | 1066 | 53% (35%, 71%) | 97% (96%, 97%) | 96% (95%, 97%) | 359 | 55.9% | 99.3% | N/A |
| | Arousal Events | 7686 | 66% (61%, 71%) | 90% (88%, 91%) | 87% (85%, 88%) | 2278 | 60.0% | 94.1% | 89.9% |
| | Leg Movement Events | 5796 | 71% (60%, 80%) | 90% (89%, 92%) | 89% (87%, 90%) | 1714 | 78.4% | 97.6% | 95.7% |
In summary, performance test results demonstrated that EnsoSleep staging event detection, sleep apnea diagnostic agreement, and sleep disordered breathing, apnea, arousal, and leg movement event detection agreement that is substantially equivalent to the predicate device Sleep Profiler (K153412) performance and reference device MICHELE (K153412) performance for all comparisons in all endpoints analyzed respectively. The EnsoSleep performance the safety and effectiveness of EnsoSleep when used for the defined indications for use and demonstrates that the device performs as well as the legally marketed predicate Sleep Profiler (K153412) and reference MICHELE Sleep Scoring (K112102).
#### Conclusion:
Nor-Clinical and Clinical verification, and performance testing was conducted in accordance with FDA guidance recommendations to confirm the device design met all specifications, user needs, and was acceptable to qualified clinical and non-clinical users. EnsoSleep has passed all of the aforementioned Verification and Validation tests and provided Clinical Performance testing results with a library clinical dataset in order to demonstrate safety or effectiveness. It is therefore concluded that EnsoSleep is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.