Retrospective clinical datasets were used to conduct an external independent performance validation study to assess the sensitivity, specificity, and ROC AUC of the BraveCX algorithm for detecting pleural effusion and pneumothorax.
External independent testing; Retrospective performance validation study
Adult patients (≥18 years old) undergoing chest X-ray imaging; Sample Size: 2,509 images (Pleural Effusion); 3,245 images (Pneumothorax); Number of Sites: 3 (Beth Israel Deaconess Medical Center, NIH Clinical Center, Stanford Hospital)
Training: 80%, Validation: 10%, Internal Test: 10% of 1,387 cases
3 (board-certified Radiologists)
MIMIC-CXR, NIH Chest X-Ray, and CheXpert datasets: 2,114 cases
3 (board-certified Radiologists)
Indications for Use
BraveCX is a radiological computer-assisted triage and notification software that analyzes adult (≥18 years old) chest Xray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). BraveCX uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, BraveCX does not send a proactive alert directly to the appropriately trained medical specialists. BraveCX is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
Device Story
BraveCX is a deep learning AI software for radiological triage of adult chest X-rays. It ingests DICOM images (AP/PA views) from X-ray equipment or PACS. The algorithm analyzes images for features of pleural effusion and/or pneumothorax. It outputs a 'flag' or 'blank' status to the PACS/workstation worklist to enable prioritization by radiologists or clinicians. It operates as a passive, parallel tool within the standard of care; it does not provide diagnostic alerts directly to specialists or highlight specific image regions. It is deployed via API as a cloud-based or on-premises service. By prioritizing suspicious exams in the reading queue, it aims to expedite review of critical findings, potentially improving patient outcomes through faster diagnosis.
Clinical Evidence
Bench testing only. Performance validated using an internal set (n=1,209 pleural effusion; n=1,387 pneumothorax) and an external independent study using MIMIC-CXR, NIH, and CheXpert datasets (n=2,509 for pleural effusion; n=3,245 for pneumothorax). For pleural effusion: ROC AUC 0.988, 92.62% sensitivity, 98.11% specificity. For pneumothorax: ROC AUC 0.972, 93.38% sensitivity, 97.27% specificity. Performance was consistent across demographics (age, gender, BMI, ethnicity).
Technological Characteristics
Deep learning AI software; vendor-agnostic; compatible with DICOM chest X-ray images. Deployed as an API (cloud or on-premises). Integrates with PACS/workstation. No hardware components. Software verification/validation per IEC 62304.
Indications for Use
Indicated for adult patients (≥18 years old) undergoing chest X-ray imaging to assist in the triage and prioritization of cases with suspected pleural effusion and/or pneumothorax. Not for stand-alone clinical decision-making.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Bering Ltd % Stephan Toupin - Official Correspondent Dawa Medical LLC 7320 NW 12th Street Suite 103 Miami, FL 33126
November 9, 2023
### Re: K223754
Trade/Device Name: BraveCX Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological Computer Aided Triage And Notification Software Regulatory Class: Class II Product Code: QFM Dated: October 11, 2023 Received: October 11, 2023
Dear Stephan Toupin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to. among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100,
{1}------------------------------------------------
Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223754
Device Name BraveCX
#### Indications for Use (Describe)
BraveCX is a radiological computer-assisted triage and notification software that analyzes adult (≥18 years old) chest Xray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). BraveCX uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides caselevel output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, BraveCX does not send a proactive alert directly to the appropriately trained medical specialists. BraveCX is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# BraveCX
# K223754
# 1. Submission Sponsor
Bering Ltd
54 Portland Place, 2nd Floor
London
W1B 1DY
United Kingdom
Contact: DROZDOV Ignat Title: Managing Director
# 2. Submission Correspondent
Stéphan Toupin, Msc
Dawa Medical LLC
7320 NW 12th Street Suite 103 Miami,
Florida, 33126, USA
Email: stoupin@dawamedical.com
Cellphone number: (786) 731-1159
# 3. Date Prepared
March 14, 2023
# 4. Device Identification
Trade/Proprietary Name: BraveCX
Common/Usual Name: BraveCX
Classification Name: Radiological computer aided triage and notification software
{4}------------------------------------------------
| Regulation Number: | 892.2080 |
|-----------------------|-------------------------------------------------------------------|
| Product Code: | QFM, Radiological computer aided triage and notification software |
| Device Class: | Class II |
| Classification Panel: | Radiology |
### 5. Legally Marketed Predicate Device(s)
#### Primary Predicate
| 510(k) Number: | K211733 |
|----------------|--------------------------|
| DEVICE NAME: | Lunit INSIGHT CXR Triage |
| MANUFACTURER: | Lunit Inc. |
### 6. Indication for Use Statement
BraveCX is a radiological computer-assisted triage and notification software that analyzes adult (≥18 years old) chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). BraveCX uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, BraveCX does not send a proactive alert directly to the appropriately trained medical specialists. BraveCX is not intended to direct attention to specific portions of an image or to anomalies other than pleural effusion and/or pneumothorax. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
#### 7. Device Description
The company conducted an internal independent testing set to assess the performance of BraveCX. The validation was completed to determine whether a distinction between target findings is properly completed by the subject device. The internal independent testing set contained n=1,209 cases for pleural effusion and n=1,387 cases of pneumothorax, obtained between June 2007 and August 2019 across 14 acute sites in NHS Greater Glasgow and Clyde. Each image corresponded to a single patient. Images were obtained by following industry standards using both mobile and departmental X-Ray equipment. Table 5A lists X-Ray equipment manufacturers included in the training and internal independent testing of the BraveCX device.
{5}------------------------------------------------
Images used in the training, validation, and testing of the subject device were all manually-curated ground truths provided by three board-certified Radiologists with at least 10 years in specialist radiology training. Model training, validation, and testing sets were generated by stratified random partitions of 80%, 10%, and 10% respectively. Each partition was stratified according to the frequency of abnormalities, gender, and View Position. To avoid data leakage, each stratified split contained non-overlapping patient identifiers.
Table 5A - Distribution of device manufacturers used for training and internal validation of the BraveCX device.
| Manufacturer | Proportion of DICOMs |
|--------------------------|----------------------|
| GE Healthcare | 2% |
| AGFA | 1% |
| Fujifilm Corporation | 70% |
| KODAK | 8% |
| Phillips Medical Systems | 1% |
| Samsung Electronics | 18% |
## Summary of results:
ROC AUC was 0.96 (95% CI: 0.95 - 0.97) with 82% sensitivity and 95% specificity for pleural effusion.
For pneumothorax, 0.98 ROC AUC (95% CI: 0.98-0.99), 89% sensitivity, and 97% specificity were reported.
By confirming that the lower bound of ROC AUC exceeds 0.95 and the sensitivity and specificity for both target radiologic findings are above 80%, the performance of the algorithm of BraveCX is validated to demonstrate that the prespecified target performance is satisfied.
## Product deployment:
BraveCX is supplied as a licensed Application Programming Interface (API) that can be deployed either as a cloud-based service, directly on premises, or integrated with third-party systems. The system can be configured to work with multiple DICOM storage platforms, including Picture Archiving and Communications (PACS) or other persistent data storage systems.
BraveCX is a Deep Learning Artificial Intelligence (AI) software that analyzes adult (≥18 years old) chest X-ray images for the presence of pre-specified suspected critical findings (pleural
{6}------------------------------------------------
effusion and/or pneumothorax. It uses deep learning to analyze each image to identify features suggestive of pleural effusion and/or pneumothorax.
Upon image acquisition from other radiological imaging equipment (e.g. X-ray systems), Anteroposterior (AP) and Posteroanterior (PA) chest X-Rays are received and processed by BraveCX. Following receipt of an image, BraveCX de-identifies a copy of each DICOM file and analyses it for features suggestive of pleural effusion and/or pneumothorax. Based on the analysis result, the software notifies PACS/workstation for the presence of the critical findings, indicated by "flag" or "(blank)". This allows the appropriately trained medical specialists to group suspicious exams together with potential for prioritization. Chest radiographs without an identified anomaly are placed in the worklist for routine review, which is the current standard of care.
The intended user of the BraveCX software is a health care professional such as radiologist or another appropriately trained clinician. The software does not alter the order or remove cases from the reading queue.
The software output to the user is a label of "flag" or "(blank)" that relates to the likelihood of presence of pneumothorax and/or pleural effusion.
BraveCX platform ingests prediction requests with either attached DICOM images or DICOM UIDs referencing images already uploaded to DICOM storage. The results will be made available via a newly generated DICOM that is stored in DICOM storage or as a JSON file. The DICOM storage component may be a Picture Archiving and Communications (PACS) system or some other local storage platform.
BraveCX works in parallel to and in conjunction with the standard of care workflow to enable prioritized review by the appropriately trained medical specialists who are qualified to interpret chest radiographs. As a passive notification for prioritization-only software tool within standard of care workflow, BraveCX does not send a proactive alert directly to the appropriately trained medical specialists who are qualified to interpret chest radiographs. BraveCX is not intended to direct attention to specific portions or anomalies of an image and it should not be used on a standalone basis for clinical decision-making.
BraveCX automatically runs after image acquisition. It prioritises and displays the analysis results through the worklist interface of PACS/workstation. An on-device, technologist notification is generated within 15 minutes after interpretation by the user, indicating which cases were prioritized by BraveCX in PACS. The technologist notification is contextual and does not provide
{7}------------------------------------------------
any diagnostic information. The on-device, technologist notification is not intended to inform any clinical decision, prioritization, or action.
## 8. Substantial Equivalence Discussion
The following table compares BraveCX to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
| Manufacturer | Bering Ltd | Lunit Inc. |
|------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | BraveCX | Lunit INSIGHT CXR Triage |
| 510(k) Number | NA | K211733 |
| Product Code | QFM | QFM |
| Regulation<br>Number | 892.2080 | 892.2080 |
| Regulation<br>Name | Radiology | Radiology |
| Indications for<br>Use | BraveCX is a radiological<br>computer-assisted triage and<br>notification software that analyzes<br>adult (≥18 years old) chest X-ray<br>images for the presence of pre-<br>specified suspected critical findings<br>(pleural effusion and/or<br>pneumothorax). BraveCX uses an<br>artificial intelligence algorithm to<br>analyze images for features<br>suggestive of critical findings and<br>provides case-level output available<br>in the PACS/workstation for<br>worklist prioritization or triage. As<br>a passive notification for<br>prioritization-only software tool<br>within standard of care workflow,<br>BraveCX does not send a proactive | Lunit INSIGHT CXR Triage is<br>a radiological computer-<br>assisted triage and notification<br>software that analyzes adult<br>chest X-ray images for the<br>presence of pre-specified<br>suspected critical findings<br>(pleural effusion and/or<br>pneumothorax). Lunit<br>INSIGHT CXR Triage uses an<br>artificial intelligence algorithm<br>to analyze images for features<br>suggestive of critical findings<br>and provides case-level output<br>available in the<br>PACS/workstation for worklist<br>prioritization or triage. As a<br>passive notification for |
| Manufacturer | Bering Ltd | Lunit Inc. |
| Trade Name | BraveCX | Lunit INSIGHT CXR Triage |
| | alert directly to the appropriately<br>trained medical specialists.<br>BraveCX is not intended to direct<br>attention to specific portions of an<br>image or to anomalies other than<br>pleural effusion and/or<br>pneumothorax. Its results are not<br>intended to be used on a stand-<br>alone basis for clinical decision-<br>making. | prioritization-only software<br>tool within standard of care<br>workflow, Lunit INSIGHT<br>CXR Triage does not send a<br>proactive alert directly to the<br>appropriately trained medical<br>specialists. Lunit INSIGHT<br>CXR Triage is not intended to<br>direct attention to specific<br>portions of an image or to<br>anomalies other than pleural<br>effusion and/or pneumothorax.<br>Its results are not intended to<br>be used on a stand-alone basis<br>for clinical decision-making. |
| Notification-<br>only, parallel<br>workflow tool | Yes | Yes |
| User | Appropriately trained medical<br>specialists who are qualified to<br>interpret chest radiographs | Appropriately trained medical<br>specialists who are qualified to<br>interpret chest radiographs |
| Targeted<br>clinical<br>condition,<br>anatomy, and<br>modality | Pleural effusion, pneumothorax<br>Chest/Lung Frontal Chest X-ray | Pleural effusion, pneumothorax<br>Chest/Lung Frontal Chest X-<br>ray |
| Algorithm for<br>pre-specified<br>critical findings<br>detection | BraveCX is a Deep Learning<br>Artificial Intelligence (AI) software<br>that was trained to detect pleural<br>effusion and pneumothorax in chest<br>X-Ray images. BraveCX uses a<br>vendor agnostic algorithm<br>compatible with DICOM chest X-<br>Ray images | Lunit INSIGHT CXR is deep<br>learning based software that<br>assists radiologists or clinicians<br>in the interpretation of chest x-<br>ray. AI algorithm designed to<br>detect pleural effusion and<br>pneumothorax in chest X-ray<br>images. Lunit INSIGHT CXR<br>Triage uses a vendor agnostic<br>algorithm compatible with<br>DICOM chest X-ray images |
Table 5B - Comparison of Characteristics
{8}------------------------------------------------
{9}------------------------------------------------
| Manufacturer | Bering Ltd | Lunit Inc. |
|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | BraveCX | Lunit INSIGHT CXR Triage |
| Radiological<br>images format | DICOM | DICOM |
| Computational<br>Platform | BraveCX is supplied as a licensed<br>Application Programming Interface<br>(API) that can be deployed either as<br>a cloud-based service, directly on<br>premises, or integrated with third-<br>party systems. | Lunit INSIGHT CXR Triage is<br>designed as a software module<br>that can be deployed on several<br>computing and X-ray imaging<br>platforms such as radiological<br>imaging equipment, PACS, On<br>Premise or On Cloud. |
| Device output in<br>case of positive<br>detection | BraveCX automatically runs after<br>image acquisition.<br>The user may prioritize reporting<br>tasks by grouping images flagged by<br>BraveCX together. Results are<br>displayed through the worklist<br>interface of PACS/workstation | Lunit INSIGHT CXR Triage<br>automatically runs after image<br>acquisition and prioritizes and<br>displays the analysis result<br>through the worklist interface<br>of PACS/workstation. |
| | No markup on original image.<br>Secondary capture of the finding. | No markup on original image.<br>Secondary capture of the finding. |
| | Upon image acquisition from other<br>radiological imaging equipment<br>(e.g. X-ray systems), an on-device,<br>technologist notification indicating<br>which cases were flagged in the<br>Secondary Capture image by Brave<br>CX in PACS, is generated 15<br>minutes after interpretation by the<br>user. | Upon image acquisition from<br>other radiological imaging<br>equipment (e.g. X-ray systems),<br>an on-device, technologist<br>notification indicating which<br>cases were flagged in the<br>Secondary Capture image by<br>Lunit INSIGHT CXR Triage in<br>PACS, is generated 15 minutes<br>after interpretation by the user. |
| | The on-device notification is<br>contextual and does not provide any<br>diagnostic information. It is not<br>intended to inform any clinical<br>decision, prioritization, or action to<br>the technologist. | The on-device notification is<br>contextual and does not provide<br>any diagnostic information. It is<br>not intended to inform any<br>clinical decision, prioritization,<br>or action to the technologist. |
| Manufacturer | Bering Ltd | Lunit Inc. |
| Trade Name | BraveCX | Lunit INSIGHT CXR Triage |
| Notification<br>(i.e., recipient,<br>timing and<br>means of<br>notification) | Passive notification.<br>Images with suspicion are flagged<br>in PACS/workstation. | Passive notification.<br>Images with suspicion<br>are<br>flagged in PACS/workstation. |
| Where<br>generated<br>results (i.e.,<br>DICOM files)<br>are stored | Picture Archiving and<br>Communications (PACS) system or<br>some other local storage platform | PACS/Workstation |
| Performance<br>level – Timing<br>of notification | The average time taken for the<br>notification to travel from the<br>BraveCX API to the point at which<br>the result is displayed in the<br>destination PACS/RIS/EPR<br>worklist is 10.4 seconds. | The average time taken for the<br>notification to travel from the<br>Lunit INSIGHT CXR Triage to<br>the point at which the result is<br>displayed in the destination<br>PACS/RIS/EPR worklist is<br>14.66 seconds. |
| Performance<br>level -<br>accuracy of<br>classification | Pleural Effusion<br>ROC AUC > 0.95<br>AUC: 0.988 (95% CI: [0.988,<br>0.9887]<br>Sensitivity 92.62% (95% CI :<br>[90.67%, 94.27%])<br>Specificity 98.11% (95% CI<br>[97.33%, 98.71%])<br>Pneumothorax<br>ROC AUC > 0.95<br>AUC : 0.972 (95% CI: [0.9727,<br>0.9729])<br>Sensitivity 93.38% (95% CI:<br>[92.23%, 94.40%])<br>Specificity 97.27% (95%CI:<br>[96.49%-97.92%]) | Pleural Effusion<br>ROC AUC > 0.95<br>AUC: 0.9686 (95% CI:<br>[0.9547, 0.9824])<br>Sensitivity 89.86% (95% CI:<br>[86.72, 93.00])<br>Specificity 93.48% (95% CI:<br>[91.06, 95.91])<br>Pneumothorax<br>ROC AUC > 0.95<br>AUC: 0.9630 (95% CI:<br>[0.9521, 0.9739])<br>Sensitivity 88.92% (95% CI:<br>[85.60, 92.24])<br>Specificity 90.51% (95% CI:<br>[88.18, 92.83]) |
{10}------------------------------------------------
# 9. Non-Clinical Performance Data
{11}------------------------------------------------
The performance of BraveCX was validated by non-clinical tests. All verification testing met the (passed), demonstrating that the software fulfills its requirement acceptance criteria specifications.
As part of demonstrating safety and effectiveness of BraveCX and in showing substantial equivalence to the predicate devices that are subject to these 510(k) submissions, Bering Ltd completed non-clinical performance tests.
The BraveCX meets all the requirements confirming that the design output meets the design inputs and specifications for the device. BraveCX passed all the testing of the subject device according to Software verification and validation testing per IEC 62304/FDA Guidance.
The company conducted an external independent testing to assess the performance of BraveCX. The studies were conducted with MIMIC Chest X-ray (MIMIC-CXR) Database v2.0.020, NIH Chest X-Ray dataset (NIH-CXR), and CheXpert dataset (Stanford Hospital) that represent the US population.
The datasets contained 867 cases for pleural effusion and 2,114 cases for pneumothorax obtained from Beth Israel Deaconess Medical Center in Boston, MA, NIH Clinical Center, and Stanford Hospital. In all cases, each image corresponded to a single patient. Patients with multiple studies were excluded from the performance validation process. All images were manually labelled by three board-certified Radiologists with at least 10 years of experience in specialty radiology training.
Demographic characteristics of images used in clinical validation studies of BraveCX are shown in the tables (Table 5C, 5D) and reflect a diverse range of ages, gender, and ethnicities.
| Characteristic | MIMIC (Pleural Effusion) | NIH (Pleural Effusion | CheXpert (Pleural Effusion) |
|----------------|--------------------------|-----------------------|-----------------------------|
| Gender | | | |
| Male | 54% | 58% | 55% |
| Female | 46% | 42% | 45% |
| Ethnicity | | | |
| Asian | 3% | N/A | 12% |
Table 5C - Demographic characteristics of patients in the clinical validation cohort of the Pleural Effusion classifier.
{12}------------------------------------------------
| Black | 10% | N/A | 6% |
|---------------|-----|-----|-----|
| Hispanic | 3% | N/A | 4% |
| Other/Unknown | 30% | N/A | 25% |
| White | 54% | N/A | 53% |
| Age | | | |
| 18-25 | 2% | 9% | 9% |
| 25-35 | 4% | 16% | 8% |
| 35-65 | 33% | 63% | 47% |
| >65 | 12% | 12% | 36% |
Table 5D - Demographic characteristics of patients in the clinical validation cohort of the Pneumothorax classifier.
| Characteristic | MIMIC<br>(Pneumothorax) | NIH<br>(Pneumothorax) | CheXpert<br>(Pneumothorax) |
|----------------|-------------------------|-----------------------|----------------------------|
| Gender | | | |
| Male | 60% | 58% | 53% |
| Female | 40% | 42% | 47% |
| Ethnicity | | | |
| Asian | 5% | N/A | 10% |
| Black | 10% | N/A | 7% |
| Hispanic | 3% | N/A | 2% |
| Other/Unknown | 5% | N/A | 25% |
| White | 77% | N/A | 56% |
| Age | | | |
| 18-25 | 4% | 9% | 8% |
{13}------------------------------------------------
| 25-35 | 7% | 16% | 7% |
|-------|-----|-----|-----|
| 35-65 | 48% | 63% | 43% |
| >65 | 41% | 12% | 42% |
Summary of results:
For pleural effusion, the results are as follows:
A total of n=2,509 images were included in the analysis (n=867 Pleural Effusion). ROC AUC 0.988 (95% CI:0.9885-0.9887) Sensitivity 92.62% (95% CI:90.67%-94.27%) Specificity 98.11% (97.33%-98.71%).
Model performance was unaffected by chest X-ray View Position (anteroposterior or posteroanterior), patient sex, age quartiles (18-53, 53-65, 65-75, ≥75), BMI quartiles (8-25, 25-30, 30-35, and >35), and ethnicities (Asian, Black, Hispanic, Other, and White) DeLong's p-values 0.89 - 0.98
The performance for the predicate device indicated for pleural effusion (Lunit INSIGHT CXR Triage, K211733) are as follows: ROC AUC 0.9686 (95% CI: 0.9547 - 0.9824), sensitivity 89.86% (95% CI: 86.72 - 93.00) and specificity 93.48% (95% CI: 91.06 - 95.91).
For pneumothorax, the results are as follows:
A total of n=3,245 images were included in the analysis (n=2,114 Pneumothorax). ROC AUC 0.972 (95% CI:0.9727-0.9729) Sensitivity 93.38% (95% CI:92.23%-94.40%) Specificity 97.27% (96.49%-97.92%).
Model performance was unaffected by chest X-ray View Position (anteroposterior or posteroanterior), patient sex, age quartiles (18-53, 53-65, 65-75, ≥75), BMI quartiles (8-25, 25-30, 30-35, and ≥35), and ethnicities (Asian, Black, Hispanic, Other, and White) DeLong's pvalues 0.07 - 0.98.
As compared the device performance time of the BraveCX with the predicate device (Lunit INSIGHT CXR Triage, K211733), the result was comparable with the predicate device.
The performance data show that the lower bound of ROC AUC exceeds 0.95 and the lower bounds of both sensitivity and specificity are above 0.85 for both pleural effusion and pneumothorax. Accordingly, the BraveCX is demonstrated to achieve effective image analysis and triage capabilities.
{14}------------------------------------------------
With regards to the device performance time, the company assessed the performance time of the BraveCX that reflects the time it takes for the device to analyze the study and send a notification to the worklist.
Time-to-notification of BraveCX was 4.8 seconds-10.4 seconds (95% CI: 4.2-10.41s) for simultaneous prediction of Pleural Effusion and Pneumothorax.
The performance for the predicate device (Lunit INSIGHT CXR Triage, K211733) indicated are as follows: The performance time was 20.76 seconds (95% CI: 20.23 - 21.28) for pleural effusion and 20.45 seconds (95% CI: 19.99 - 20.92) for pneumothorax.
As compared the device performance time of the BraveCX with predicate device (Lunit INSIGHT CXR Triage, K211733), the result was comparable with the predicate device.
In summary, the BraveCX performed successfully in the external standalone study. The BraveCX established standalone adequate detection performance and device performance time for pleural effusion and pneumothorax as compared to the predicate device.
Results of the clinical investigation support the indications for use of BraveCX. External standalone study conclusion confirms that BraveCX is safe and effective as used according to the instructions for use.
## 10. Statement of Substantial Equivalence
The BraveCX is as safe and effective as the predicate devices. The BraveCX has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices.
The minor technological differences between the BraveCX and its predicate device raise no new issues of safety or effectiveness. The clinical and non-clinical performance data demonstrates that the BraveCX is as safe and effective as the predicate device. Thus, the BraveCX support a decision is substantially equivalent to its predicate devices for triage and notification.
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise additional questions regarding its safety and effectiveness as compared to the predicate device.
BraveCX, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.