K211733 · Lunit, Inc. · QFM · Nov 10, 2021 · Radiology
Device Facts
Record ID
K211733
Device Name
Lunit INSIGHT CXR Triage
Applicant
Lunit, Inc.
Product Code
QFM · Radiology
Decision Date
Nov 10, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2080
Device Class
Class 2
Attributes
AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K211733 · Nov 10, 2021
Lunit INSIGHT CXR Triage
Lunit, Inc.
NIH chest X-ray dataset; Multi-institutional clinical dataset from India
The clinical dataset was used to evaluate the standalone performance (sensitivity, specificity, and ROC AUC) of the Lunit INSIGHT CXR Triage algorithm in identifying pleural effusion and pneumothorax.
1,385 images classified into positive and negative groups.
>1 (board-certified radiologists)
1,708 anonymized chest radiographs from NIH dataset and multiple institutions in India.
—
Indications for Use
Lunit INSIGHT CXR Triage is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). Lunit INSIGHT CXR Triage uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage. As a passive notification for prioritization-only software tool within standard of care workflow, Lunit INSIGHT CXR Triage does not send a proactive alert directly to the appropriately trained medical specialists. Lunit INSIGHT CXR Triage is not intended to direct attention to specific portions of an image. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
Device Story
Software tool for radiological triage; analyzes frontal chest X-ray images for pleural effusion and pneumothorax. Uses AI algorithm to identify features suggestive of critical findings; provides case-level output to PACS/workstation for worklist prioritization. Operates in parallel to standard clinical workflow; does not send proactive alerts to specialists. De-identifies DICOM images; flags suspicious cases in worklist. Does not provide diagnostic information or direct attention to specific image regions. Used by radiologists/trained specialists to prioritize review of time-sensitive cases. Generates contextual technologist notification 15 minutes post-interpretation. Benefits patient by enabling earlier review of critical findings.
Clinical Evidence
Clinical pivotal studies using NIH and India datasets (1,708 radiographs). Pleural effusion: AUC 0.9686, 89.86% sensitivity, 93.48% specificity. Pneumothorax: AUC 0.9630, 88.92% sensitivity, 90.51% specificity. Performance time averaged ~20 seconds. Results demonstrate effective triage capabilities comparable to predicates.
Technological Characteristics
Software-only device; AI-based image analysis algorithm. Compatible with DICOM frontal chest X-ray images. Deployed on PACS, radiological imaging equipment, or cloud. Passive notification system; no image markup. Vendor-agnostic.
Indications for Use
Indicated for adult patients to analyze frontal chest X-ray images for suspected pleural effusion and/or pneumothorax to enable worklist prioritization. Not for standalone clinical decision-making; not for directing attention to specific image portions.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Lunit Inc. % Colin Jacob 15th Floor, 27 Teheran-ro 2-gil Gangnam-gu, Seoul REPUBLIC OF KOREA
Re: K211733
November 10, 2021
Trade/Device Name: Lunit INSIGHT CXR Triage Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QFM Dated: October 1, 2021 Received: October 1, 2021
Dear Colin Jacob:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211733
Device Name Lunit INSIGHT CXR Triage
#### Indications for Use (Describe)
Lunit INSIGHT CXR Triage is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). Lunit INSIGHT CXR Triage uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage.
As a passive notification for prioritization-only software tool within standard of care workflow, Lunit INSIGHT CXR Triage does not send a proactive alert directly to the appropriately trained medical specialists. Lunit INSIGHT CXR Triage is not intended to direct attention to specific portions of an image. Its results are not intended to be used on a standalone basis for clinical decision-making.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
× | Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## K211733
# This summary of safety and effectiveness information is submitted in accordance with 21 CFR § 807.92.
# Date Prepared: November 3, 2021
Submitter Lunit Inc., 15th Floor, 27 Teheran-ro 2-gil, Gangnam-gu, Seoul, Republic of Korea Phone: +82 2 2138 0827 Facsimile: +82 2 6969 5116 Contact Person: Colin S. Jacob
| Name of Device: | Lunit INSIGHT CXR Triage |
|----------------------------|--------------------------------------------------------------------|
| Common or Usual Name: | Lunit INSIGHT CXR Triage |
| Classification Name: | Radiological Computer-Assisted Prioritization Software for Lesions |
| Classification Regulation: | 21 CFR 892.2080 |
| Regulatory Class: | Class II |
| Product Code: | QFM |
#### Predicate Device
The Lunit INSIGHT CXR Triage is substantially equivalent to the following devices:
#### Primary Predicate Device
| Manufacturer Name | Zebra Medical Vision, Ltd. |
|-------------------|----------------------------|
| Device Trade Name | HealthCXR |
| 510(k) Number | K192320 |
#### Secondary Predicate Device
| Manufacturer Name | GE Medical Systems, LLC |
|-------------------|-------------------------|
| Device Trade Name | Critical Care Suite |
| 510(k) Number | K183182 |
{4}------------------------------------------------
# Device Description
Lunit INSIGHT CXR Triage is a radiological computer-assisted prioritization software that utilizes AI-based image analysis algorithms to identify pre-specified critical findings (pleural effusion and/or pneumothorax) on frontal chest X-ray images and flag the images in the PACS/workstation to enable prioritized review by the appropriately trained medical specialists who are qualified to interpret chest radiographs. The software does not alter the order or remove cases from the reading queue.
Chest radiographs are automatically received from the user's image system (e.g. Picture Archiving and Communication System (PACS)) or other radiological imaging equipment (e.g. X-ray systems) and processed by the Lunit INSIGHT CXR Triage for analysis. Following receipt of chest radiographs, the software device de-identifies a copy of each chest radiographs in DICOM format (.dcm) and automatically analyzes each image to identify features suggestive of pleural effusion and/or pneumothorax. Based on the analysis result, the software notifies PACS/workstation for the presence of the critical findings as indicating either "flag" or "(blank)". This would allow the appropriately trained medical specialists to group suspicious exams together that may potentially benefit for their prioritization. Chest radiographs without an identified anomaly are placed in the worklist for routine review, which is the current standard of care. Lunit INSIGHT CXR Triage can flag more than one critical finding per radiograph and the user may select the option to turn on and off notification of critical findings (pleural effusion and pneumothorax).
When deployed on other radiological imaging equipment, Lunit INSIGHT CXR Triage automatically runs after image acquisition. It prioritizes and displays the analysis result through the worklist interface of PACS/workstation. Moreover, the analysis result can also be provided in the form of DICOM files containing information on the presence of suspicious radiologic findings. In parallel, the algorithms produce an on-device notification indicating which cases were prioritized by Lunit INSIGHT CXR Triage in PACS. The on-device notification does not provide any diagnostic information and it is not intended to inform any clinical decision, prioritization, or action to the technologist.
Lunit INSIGHT CXR Triage works in parallel to and in conjunction with the standard care of workflow; therefore, the user enables to review the study containing critical findings earlier than others. As a passive notification for prioritization-only software tool within standard of care workflow, the software does not send a proactive alert directly to the appropriately trained medical specialists who are qualified to interpret chest radiographs. Lunit INSIGHT CXR Triage is not intended to direct attention to specific portions or anomalies of an image and it should not be used on a stand-alone basis for clinical decision-making.
In parallel, an on-device, technologist notification is generated 15 minutes after interpretation by the user, indicating which cases were prioritized by Lunit INSIGHT CXR Triage in PACS. The technologist notification is contextual and does not provides any diagnostic information. The ondevice, technologist notification is not intended to inform any clinical decision, prioritization, or action.
{5}------------------------------------------------
# Intended Use / Indications for Use
Lunit INSIGHT CXR Triage is a radiological computer-assisted triage and notification software that analyzes adult chest X-ray images for the presence of pre-specified suspected critical findings (pleural effusion and/or pneumothorax). Lunit INSIGHT CXR Triage uses an artificial intelligence algorithm to analyze images for features suggestive of critical findings and provides case-level output available in the PACS/workstation for worklist prioritization or triage.
As a passive notification for prioritization-only software tool within standard of care workflow, Lunit INSIGHT CXR Triage does not send a proactive alert directly to the appropriately trained medical specialists. Lunit INSIGHT CXR Triage is not intended to direct attention to specific portions of an image. Its results are not intended to be used on a stand-alone basis for clinical decision-making.
There is a minor difference between the subject and primary predicate device HealthCXR (K192320) in their indications for use. The primary predicate device is indicated for use to analyze chest X-ray images for the presence of pleural effusion, while Lunit INSIGHT CXR Triage detects pleural effusion as well as pneumothorax. While the specific clinical finding differs, the intended use of the device which is to provide passive notification for triage and prioritization for time sensitive radiologic findings in the chest X-ray is the same. Additionally, the time sensitivity of additional potential finding of pneumothorax has been already established in a secondary predicate device; Critical Care Suite (K183182) and the detection of such potential finding in Lunit INSIGHT CXR Triage is supported by performance testing.
## Summary of Technological Characteristics
Both the Lunit INSIGHT CXR Triage and the predicate devices are software only devices that use Artificial intelligence (AI) algorithms and are intended to aid in triage and prioritization of radiological images. At a high level, the subject and the predicate devices are based on the following same technological elements:
- · Artificial Intelligence Algorithm(s)
- · Triage and prioritization software
The following technological differences exist between the subject and predicate devices:
- · Clinical condition(s)
Lunit INSIGHT CXR Triage detects suspected findings, pleural effusion as well as pneumothorax. In comparison, the primary predicate device detects suspected findings of pleural effusion alone. However, both devices identify time sensitive findings and provide passive notification and pleural effusion as well as pneumothorax have been accepted by FDA as appropriate for regulation as CAD t devices.
Comparison of the key features of the subject, and predicate devices is provided in Table 1.
{6}------------------------------------------------
# Performance Data
The performance of Lunit INSIGHT CXR Triage was validated by clinical and nonclinical tests. All verification testing met the acceptance criteria (passed), demonstrating that the software fulfills its requirement specifications.
# Nonclinical Tests
The company conducted an internal validation test to assess the standalone performance of Lunit INSIGHT CXR Triage. The validation was completed to determine whether a distinction between target findings is properly completed by the subject device. A total of 1,385 images were collected according to the inclusion and exclusion criteria and classified into positive and negative groups by highly experienced board-certified radiologists. For each target finding, positive includes any case that contains at least one target finding in the image and negative represents the case that does not have the target finding, which includes cases without any findings and cases that contain findings other than the target radiologic finding.
Summary of results: ROC AUC was 0.9864 (95% CI: 0.9815 - 0.9913) with 94.29% sensitivity and 95.72% specificity for pleural effusion. For pneumothorax, 0.9973 ROC AUC (95% CI: 0.9955 -0.9992), 96.08% sensitivity, and 99.14% specificity were reported. By confirming that the lower bound of ROC AUC exceeds 0.95 and the sensitivity and specificity for both target radiologic findings are above 0.80, the performance of the algorithm of Lunit INSIGHT CXR Triage is validated to demonstrate that the prespecified target performance is satisfied.
# Clinical Tests
Two individual clinical pivotal studies were conducted to evaluate the effectiveness of the Lunit INSIGHT CXR Triage. The studies were conducted with NIH chest X-ray dataset that represents the US population and India dataset collected respectively from multiple institutions in India (6 sites for pleural effusion and 3 sites for pneumothorax).
A total of 1,708 anonymized chest radiographs (754 cases for pleural effusion and 954 cases for pneumothorax) collected from NIH chest X-ray dataset and multiple institutions in India were enrolled in the validation dataset for the standalone performance test and device performance time estimation. For pleural effusion, the results are as follow: ROC AUC 0.9686 (95% CI: 0.9547 - 0.9824), sensitivity 89.86% (95% CI: 86.72 - 93.00) and specificity 93.48% (95% CI: 91.06 - 95.91). For pneumothorax, the results are as follows: ROC AUC 0.9630 (95% CI: 0.9521 - 0.9739), Sensitivity 88.92% (95% CI: 85.60 - 92.24) and Specificity 90.51% (95% CI: 88.18 - 92.83). The performance for the primary predicate device indicated for pleural effusion (Zebra Medical Vision, HealthCXR) are as follows: ROC AUC was 0.9885 (95% CI: 0.9815 - 0.9956) with sensitivity 96.74% (95% CI: 92.79 -96.48) and specificity 93.17% (95% CI: 89.57 – 95.58). The performance data show that the lower bound of ROC AUC exceeds 0.95 and the lower bounds of both sensitivity and specificity are above
Image /page/6/Picture/11 description: The image shows the word "Lunit" in a simple, sans-serif font. To the left of the word is a small, stylized graphic consisting of two circles connected by a diagonal line. The graphic is in a light blue color, while the text is in a darker shade of blue or black. The overall design is clean and modern.
{7}------------------------------------------------
0.85 for both pleural effusion and pneumothorax. Accordingly, the Lunit INSIGHT CXR Triage is demonstrated to achieve effective image analysis and triage capabilities.
With regards to the device performance time, the company assessed the performance time of the Lunit INSIGHT CXR Triage that reflects the time it takes for the device to analyze the study and send a notification to the worklist. The performance time of the Lunit INSIGHT CXR Triage was 20.76 seconds (95% CI: 20.23 - 21.28) for pleural effusion and 20.45 seconds (95% CI: 19.99 - 20.92) for pneumothorax. As compared the device performance time of the Lunit INSIGHT CXR Triage with cleared commercial products (HealthCXR (Zebra, K192320) and red dot™ (Behold.AI, K161556)), the result was comparable with the primary predicate device and other 510(k) cleared products.
In summary, the Lunit INSIGHT CXR Triage performed successfully in the clinical pivotal studies. The Lunit INSIGHT CXR Triage established standalone adequate detection performance and device performance time for pleural effusion and pneumothorax as compared to the primary predicate device.
# Conclusions
The Lunit INSIGHT CXR Triage is as safe and effective as the predicate devices. The Lunit INSIGHT CXR Triage has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor differences in indications do not alter the intended use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the Lunit INSIGHT CXR Triage and its predicate devices raise no new issues of safety or effectiveness. The clinical and non-clinical performance data demonstrates that the Lunit INSIGHT CXR Triage is as safe and effective as the predicate devices. Thus, the Lunit INSIGHT CXR Triage support a decision is substantially equivalent to its predicate devices for triage and notification.
{8}------------------------------------------------
Lunit, Inc.
15 Floor, 27 Teheran-ro 2-gil, Gangnam-gu, Seoul, 06241, Republic of Korea
#### Traditional 510(k) Lunit INSIGHT CXR Triage
Page 6 of 10
# Table 1 Substantial Equivalence Table
| Technological<br>Characteristics | Proposed Device:<br>Lunit INSIGHT CXR Triage | Primary Predicate Device:<br>HealthCXR (K192320) | Secondary Predicate Device:<br>Critical Care Suite (K183182) |
|------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>classification | Radiological Computer Assisted<br>Prioritization Software Class II,<br>QFM | Radiological Computer Assisted<br>Prioritization Software Class II,<br>QFM | Radiological Computer Assisted<br>Prioritization Software Class II,<br>QFM |
| Indication for<br>Use/Intended<br>Use | Lunit INSIGHT CXR Triage is a<br>radiological computer-assisted triage<br>and notification software that<br>analyzes adult chest X-ray images for<br>the presence of pre-specified<br>suspected critical findings (pleural<br>effusion and/or pneumothorax).<br>Lunit INSIGHT CXR Triage uses an<br>artificial intelligence algorithm to<br>analyze images for features<br>suggestive of critical findings and<br>provides case-level output available<br>in the PACS/workstation for<br>worklist prioritization or triage.<br>As a passive notification for<br>prioritization-only software tool<br>within standard of care workflow,<br>Lunit INSIGHT CXR Triage does<br>not send a proactive alert directly to<br>the appropriately trained medical<br>specialists. Lunit INSIGHT CXR<br>Triage is not intended to direct<br>attention to specific portions of an<br>image or to anomalies other than | The Zebra HealthCXR device is a<br>software workflow tool designed to<br>aid the clinical assessment of adult<br>Chest X-Ray cases with features<br>suggestive of pleural effusion in the<br>medical care environment.<br>HealthCXR analyzes cases using an<br>artificial intelligence algorithm to<br>identify suspected findings. It makes<br>case-level output available to a<br>PACS/workstation for worklist<br>prioritization or triage. HealthCXR is<br>not intended to direct attention to<br>specific portions or anomalies of an<br>image. Its results are not intended to<br>be used on a stand-alone basis for<br>clinical decision-making nor is it<br>intended to rule out Pleural Effusion<br>or otherwise preclude clinical<br>assessment of X-Ray cases. | Critical Care Suite is a computer<br>aided triage and notification device<br>that analyzes frontal chest x-ray<br>images for the presence of pre-<br>specified critical findings<br>(pneumothorax). Critical Care Suite<br>identifies images with critical finding<br>to enable case prioritization or triage<br>in the PACS/workstation.<br>Critical Care Suite is intended for<br>notification only and does not<br>provide diagnostic information<br>beyond the notification. Critical Care<br>Suite should not be used in-lieu of<br>full patient evaluation or solely relied<br>upon to make or confirm a diagnosis.<br>It is not intended to replace the<br>review of the x-ray image by a<br>qualified physician.<br>Critical Care Suite is indicated for<br>adult-size patients. |
| Technological<br>Characteristics | Proposed Device:<br>Lunit INSIGHT CXR Triage | Primary Predicate Device:<br>HealthCXR (K192320) | Secondary Predicate Device:<br>Critical Care Suite (K183182) |
| | pleural effusion and/or<br>pneumothorax. Its results are not<br>intended to be used on a stand-alone<br>basis for clinical decision-making. | | |
| Notification-<br>only, parallel<br>workflow tool | Yes | Yes | Yes |
| User | Appropriately trained medical<br>specialists who are qualified to<br>interpret chest radiographs. | Radiologist | Radiologist |
| Targeted<br>clinical<br>condition,<br>anatomy, and<br>modality | Pleural effusion, pneumothorax<br>Chest/Lung<br>Frontal Chest X-ray | Pleural effusion<br>Chest/Lung<br>Frontal Chest X-ray | Pneumothorax<br>Chest/Lung<br>Frontal Chest X-ray |
| Algorithm for<br>pre-specified<br>critical findings<br>detection | AI algorithm designed to detect<br>pleural effusion and pneumothorax<br>in chest X-ray images.<br>Lunit INSIGHT CXR Triage uses a<br>vendor agnostic algorithm<br>compatible with DICOM chest X-ray<br>images. | AI algorithm designed to detect<br>pleural effusion in chest X-ray<br>images.<br>HealthCXR employs a vendor<br>agnostic algorithm compatible with<br>DICOM chest X-ray images. | AI algorithm designed to detect<br>pneumothorax in frontal chest X-ray<br>images.<br>Critical Care Suite uses a vendor<br>agnostic algorithm compatible with<br>DICOM frontal chest X-ray images. |
| Radiological<br>images format | DICOM | DICOM | DICOM |
| Computational<br>Platform | Lunit INSIGHT CXR Triage is<br>designed as a software module that<br>can be deployed on several | HealthCXR is designed as a software<br>module that can be deployed on | Critical Care Suite is designed as a<br>software module that can be<br>deployed on several computing and |
| Technological<br>Characteristics | Proposed Device:<br>Lunit INSIGHT CXR Triage | Primary Predicate Device:<br>HealthCXR (K192320) | Secondary Predicate Device:<br>Critical Care Suite (K183182) |
| | computing and X-ray imaging<br>platforms such as radiological<br>imaging equipment, PACS, On<br>Premise or On Cloud. | PACS and Standalone desktop<br>application, Zebra Worklist. | X-ray imaging platforms such as<br>Digital Projection Radiographic<br>Systems, PACS, On Premise or On<br>Cloud. |
| | When deployed on other radiological<br>imaging equipment, Lunit INSIGHT<br>CXR Triage automatically runs after<br>image acquisition and prioritizes and<br>displays the analysis result through<br>the worklist interface of<br>PACS/workstation. | Integration module notifies the<br>PACS/workstation for prioritization<br>through the worklist interface. | Critical Care Suite enables case<br>prioritization or triage through direct<br>communication of the Critical Care<br>Suite notification during image<br>transfer to the PACS. |
| Device output<br>in case of<br>positive<br>detection | No markup on original image. | No markup on original image. | No markup on original image |
| | Secondary capture of the finding. | | |
| | Upon image acquisition from other<br>radiological imaging equipment (e.g.<br>X-ray systems), an on-device,<br>technologist notification indicating<br>which cases were flagged by Lunit<br>INSIGHT CXR Triage in PACS, is<br>generated 15 minutes after<br>interpretation by the user. The on-<br>device notification is contextual and<br>does not provide any diagnostic<br>information. It is not intended to | | Upon image acquisition on a Digital<br>Projection Radiographic System, an<br>on-device, technologist notification is<br>generated 15 minutes after exam<br>closure, indicating which cases were<br>prioritized by Critical Care Suite in<br>PACS. The technologist notification<br>is contextual and does not provides<br>any diagnostic information. The on-<br>device, technologist notification is<br>not intended to inform any clinical<br>decision, prioritization, or action. |
| Republic of Korea | | | |
| Technological<br>Characteristics | Proposed Device:<br>Lunit INSIGHT CXR Triage | Primary Predicate Device:<br>HealthCXR (K192320) | Secondary Predicate Device:<br>Critical Care Suite (K183182) |
| | inform any clinical decision,<br>prioritization, or action to the<br>technologist. | | |
| Notification<br>(i.e., recipient,<br>timing and<br>means of<br>notification) | Passive notification.<br>Images with suspicion of pleural<br>effusion and/or pneumothorax are<br>flagged in PACS/workstation. | Passive notification.<br>Images with suspicion of pleural<br>effusion are flagged in<br>PACS/workstation. | Passive notification.<br>Images with suspicion of<br>pneumothorax are flagged in<br>PACS/workstation. |
| Where<br>generated<br>results (i.e.,<br>DICOM files)<br>are stored | PACS/Workstation | PACS/Workstation | PACS/Workstation |
| Performance<br>level –<br>Timing of<br>notification | The average time taken for the<br>notification to travel from the Lunit<br>INSIGHT CXR Triage to the point<br>at which the result is displayed in the<br>destination PACS/RIS/EPR worklist<br>is 14.66 seconds. | Passive notification is visible upon<br>transfer to the PACS with a delay of<br>about 22 seconds for image transfer<br>to the cloud, computation, and<br>results transfer. | The average time to acquire,<br>annotate, process and transfer an<br>image from the x-ray system to<br>PACS was measured and found to<br>take 42 seconds on average. Exams<br>arrive on PACS with the passive<br>notification already incorporated,<br>therefore there is no delay for image<br>transfer or computation. The<br>worklist prioritization happens<br>immediately once the exam is<br>received on the PACS.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.