K223453 · Promisemed Hangzhou Meditech Co., Ltd. · FMF · Mar 2, 2023 · General Hospital
Device Facts
Record ID
K223453
Device Name
Insulin Syringe
Applicant
Promisemed Hangzhou Meditech Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Mar 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.
Device Story
Insulin Syringe is a sterile, single-use, manual piston syringe used for subcutaneous insulin delivery. Device consists of a calibrated plastic barrel, plunger with polyisoprene piston, and integrated non-retractable stainless steel needle. Operated by patient or clinician via manual pushing force on plunger rod to deliver insulin. Device features include 0.3ml, 0.5ml, and 1.0ml volumes and needle gauges from 28G to 32G. Lubricated with aminofunctional siloxane for smooth movement. Used in home or clinical settings. Output is the controlled delivery of insulin to subcutaneous tissue, facilitating diabetes management.
Clinical Evidence
No clinical data. Bench testing only. Performance verified against ISO 8537 requirements including needle dimensions, bond strength, leakage, dead space, and graduation accuracy. Biocompatibility confirmed per ISO 10993-1, -4, -5, -10, -11, and ASTM F 756-17. Sterilization validated per ISO 11135.
Indicated for subcutaneous injection of U-40 and U-100 insulin in patients with diabetes.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 2, 2023
Promisemed Hangzhou Meditech Co., Ltd. Zearou Yang Regulatory affairs manager No. 1388 Cangxing Street, Cangqian Community, Yuhang District Hangzhou City, Zhejiang 311121 China
Re: K223453
Trade/Device Name: Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF Dated: January 30, 2023 Received: January 30, 2023
Dear Zearou Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sincerely,
Alan St
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K223453
Device Name Insulin Syringe
Indications for Use (Describe)
Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### 1. Date Prepared
Feb. 24, 2023
### 2. Submitter's Information
## Name of Sponsor:
Promisemed Hangzhou Meditech Co., Ltd.
### Address:
No. 1388 Cangxing Street, Cangqian Community, Yuhang District, Hangzhou City, 311121
Zhejiang, China
Contact Name: Zearou Yang
### Telephone No.:
+86 571 88772985
Fax No.:
+86 571 88772985
Email Address:
zearou.yang@promisemed.ca
### 3. Trade Name, Common Name, Classification
Trade/Product Name: Insulin Syringe
Common Name: Insulin Syringe
Classification name: Piston syringe
Regulation Number: 880.5680
Device Class: Class II
Product Code: FMF
### 4. Identification of Predicate Device K110421; Disposable Insulin Syringe
## 5. Description of the Device
The proposed device Insulin Syringe, a sterile device consisting of a calibrated barrel with plunger, is intended to be used to administer an injection of insulin to a patient subcutaneously. A non-retractable integrated needle is included. The syringe is made of plastic and silicone materials and allowing smooth plunger movement. This is a single-use device. The device is the same as K193273. The purpose of the submission is to add OTC labeling.
{4}------------------------------------------------
### 6. Indications for use
Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.
### Similarities and Differences of the Proposed Devices to the Predicate Devices 7.
A detailed comparison to the predicate is provided in Table 1.
| Items | Subject Device<br>(K223453) | Predicate Device<br>(K110421) | Comments |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Insulin Syringe | Disposable Insulin<br>Syringe | |
| Manufacturer | Promisemed Hangzhou<br>Meditech Co., Ltd | Wen Zhou Wuzhou<br>Import & Export Co., Ltd. | |
| Device Class | Class II | Class II | Same |
| Product Code | FMF | FMF | Same |
| Regulation number | 880.5680 | 880.5680 | Same |
| Regulation Name | Piston syringe | Piston syringe | Same |
| Intended Use/<br>Indications for Use | Insulin Syringe is intended<br>for subcutaneous injection<br>of U-40 and U-100 insulin<br>in the treatment of<br>diabetes. | Disposable insulin syringe is a<br>device intended for medical<br>purposes for the manual<br>aspiration of insulin, and for<br>the injection of insulin into<br>parts of the body below the<br>surface skin. | Same |
| Type of use | Prescription use and over-<br>the-counter use | Over-the-counter use | Different.<br>1. "Prescription use" of<br>subjective device has<br>been cleared with<br>K193273;<br>2. "Over-the-counter<br>use" is same with<br>Predicate device;<br>The difference of use type<br>does not impact safety<br>and effectiveness. |
| Operating<br>Principle | The insulin is injected to<br>subcutaneous tissue by<br>pushing force generated<br>through<br>pushing plunger rod of the<br>insulin syringe. | The insulin is injected to<br>subcutaneous tissue by<br>pushing force generated<br>through<br>pushing plunger rod of the<br>insulin syringe. | Same |
| Specific drug<br>use | Insulin | Insulin | Same |
| Length | 120mm | 120mm | Same |
| Volume | 0.3ml, 0.5ml, 1.0ml | 0.3ml, 0.5ml, 1.0ml | Same |
| Needle length | 6mm, 8mm,12mm | 5mm, 6mm, 8mm, 12mm | Different.<br>The needle length of<br>proposed device is covered<br>by the predicate device. |
| Needle gauge | 32G, 31G, 30G, 29G, 28G | 31G, 30G, 29G, 28G | Different.<br>Compared to the predicate<br>device, the proposed<br>device has an additional<br>32G model needle. 32G<br>model for insulin syringe |
| | | | 510(K) 223453<br>needle safety and<br>performance has been<br>confirmed by testing. This<br>difference does not raise<br>new questions of safety<br>and effectiveness. |
| Needle tip<br>configuration | 3 bevels | 3 bevels | Same |
| Nozzle type | Not applicable | Not applicable | Same |
| Numbering of<br>scale | At every five units for the<br>0.3mL and 0.5mL syringes,<br>and at every 10units for<br>1.0mL | At every five units for the<br>0.3mL and 0.5mL syringes,<br>and at every 10units for<br>1.0mL | Same |
| Gradations<br>legibility | Legible | Legible | Same |
| Needle cover<br>dimensions | Length:25mm,<br>Diameter: 6mm | Length:25mm,<br>Diameter: 6mm | Same |
| Needle cover<br>color | Red (U-40) and orange<br>(U-100) | Red (U-40) and orange<br>(U-100) | Same |
| Lubricant<br>composition | Aminofunctional siloxane | Aminofunctional siloxane | Same |
| Lubricant<br>amount/cm² | The lubricant is not form<br>pools of fluid on the interior<br>surface of the syringe or<br>outside surfaces of the<br>needle tube. | The lubricant is not form<br>pools of fluid on the interior<br>surface of the syringe or<br>outside surfaces of the<br>needle tube. | Same |
| Barrel transparency | Transparent | Transparent | Same |
| Needle cover<br>strength | <15N | <15N | Same |
| Hub/needle bond<br>strength | >22N | >22N | Same |
| Biocompatibility | No cytotoxicity<br>No irritation reactivity<br>No significant evidence<br>of skin sensitization<br>No significant evidence<br>of systemic toxicity<br>No evidence of Hemolysis<br>No evidence of pyrogens | No cytotoxicity<br>No irritation reactivity<br>No significant evidence<br>of skin sensitization<br>No significant evidence<br>of systemic toxicity<br>No evidence of Hemolysis<br>No evidence of pyrogens | Same |
| Configuration<br>and Materials | Needle: Stainless Steel<br>(SUS304)<br>Barrel: Polypropylene<br>Plunger: Polypropylene<br>Piston: Polyisoprene<br>rubber<br>Needle cap: Polyethylene<br>Protective end cap (only<br>type 8): Polyethylene | Needle: Stainless Steel<br>(SUS304)<br>Barrel: Polypropylene (PP)<br>Plunger: Polypropylene (PP)<br>Piston: Polyisoprene Rubber<br>Needle cover: Polypropylene<br>(PP)<br>Protective end cap:<br>Polypropylene (PP) | Different.<br>The materials of needle<br>cap and protective end<br>cap are different<br>between the subject<br>device and predicate<br>device. The<br>biocompatibility test of<br>the subject device was<br>conducted to<br>demonstrate that the<br>subject device met the<br>biocompatibility<br>requirements. This<br>difference does not raise<br>any new safety and |
| | | | effectiveness questions. |
| Label | Device name,<br>indication, instruction<br>for use, precaution,<br>warning, shelf life,<br>manufacturer | Device name,<br>indication, instruction<br>for use, precaution,<br>warning, shelf life,<br>manufacturer | Same |
| Sterilization<br>method and SAL | Sterilized by ethylene<br>oxide gas SAL = 10-6 | Sterilized by ethylene<br>oxide gas SAL = 10-6 | Same |
| Sterilization method | EO Sterilization | EO Sterilization | Same |
| EO and ECH residues<br>testing | Conform ISO 10993-7 | Conform ISO 10993-7 | Same |
{5}------------------------------------------------
## Insulin Syringe
510(k) 223453
{6}------------------------------------------------
### Insulin Syringe
510(k) 223453
#### 8. Performance Testing Summary
The bench testing performed verifies that the performance of the subject device is substantially equivalent in terms of critical performance characteristics to the predicate device. The data are provided in submission, K193273.
Performance Testing:
The Insulin Syringes have been designed and successfully tested (relying on K193273 submission) to meet the applicable requirements outlined in ISO 8537, such as:
- Needle tube length; -
- -Diameter of needle tube;
- Bond between hub and needle tube; -
- -Needle points;
- -Freedom from defects;
- -Lubrication;
- -Fit of plunger stopper in barrel;
- -Dead space;
- -Freedom from leakage at needle;
- -Freedom from leakage past plunger stopper;
- -Tolerance on graduated capacity;
- -Limits for acidity or alkalinity;
- -Limits for extractable metals;
- -Fragmentation;
- -Penetration force ;
- -Particulate contamination;
### Biocompatibility testing:
The material of the Insulin Syringes have successfully passed testing as outlined in ISO 10993-1 for devices categorized as External communicating devices, Limited exposure.
ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro
{7}------------------------------------------------
cytotoxicity
ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Test for Irritation and
Sensitization
ISO 10993-11: 2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
ISO 10993-4:2017 Biological evaluation of medical devices -Part 4: Selection of tests for interactions with blood.
ASTM F 756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
The United States Pharmacopeia <151> (Pyrogen test), method of limulus amebocyte lysate [LAL].
Sterilization and Shelf-life Testing:
Sterilization of the Insulin Syringes has been validated according to ISO 11135 with half-cycle method.
Testing demonstrated maximum levels of residues of ethylene oxide (≤9mg/d) and ethylene chlorohydrins (≤4mg/d) met with the limits presented in ISO 10993-7. Shelf-life testing supports a shelf life of 5-years after sterilization.
#### Clinical study 9.
No prospective clinical trials were conducted in support of this 510(K).
### 10. Colorants
The colorant used in the needle cap was listed as below:
| Component | Color code | Colorant |
|--------------------|------------|----------|
| Needle cap (U-100) | Orange | PE-M2766 |
| Needle cap (U-40) | Red | PE-M2282 |
### 11. Conclusion
Based on the information provided within this 510(k) submission, the proposed subject device is substantially equivalent to the predicate device and is as safe, as effective and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.