K193273 · Promisemed Hangzhou Meditech Co., Ltd. · FMF · Jan 5, 2021 · General Hospital
Device Facts
Record ID
K193273
Device Name
Insulin Syringe
Applicant
Promisemed Hangzhou Meditech Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Jan 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.
Device Story
Insulin Syringe is a sterile, single-use piston syringe consisting of a calibrated barrel, plunger, and non-retractable integrated needle. It is used by patients or clinicians to administer subcutaneous insulin injections. The device operates via manual pushing force applied to the plunger rod to deliver insulin from the barrel through the needle into subcutaneous tissue. It is available in 0.3ml, 0.5ml, and 1.0ml volumes with various needle gauges (28G-32G) and lengths (6mm-12mm). The device facilitates accurate insulin dosing through graduated scales on the barrel. It is intended for home or clinical use to manage diabetes by enabling precise delivery of insulin concentrations.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including performance testing per ISO 8537:2016, biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, systemic toxicity, hemolysis), pyrogen testing (USP <151>), particulate matter testing (USP <788>), and sterilization validation (ISO 11135).
Indicated for subcutaneous injection of U-40 and U-100 insulin in patients with diabetes.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
February 26, 2021
Promisemed Hangzhou Meditech Co., Ltd. % Christy Young Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town. Nanshan Shenzhen, Guangdong 518000 China
Re: K193273
Trade/Device Name: Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: Class II Product Code: FMF
Dear Christy Young:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 5, 2021. Specifically, FDA is updating this SE Letter (E.G., typo in manufacturer name) as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Rumi Young, OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, 301-796-6005, Rumi. Young@fda.hhs.gov.
> Sincerely. Digitally signed by Rumi Rumi Young -S Date: 2021.02.26 Young -S Date: 2021.02.2 Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logos of the Department of Health and Human Services (HHS) and the Food and Drug Administration (FDA). The HHS logo is a circular seal with an emblem in the center. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 5, 2021
Prominsemed Hangzhou Meditech Co., Ltd. % Christy Young Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town. Nanshan Shenzhen, 518000 Cn
Re: K193273
Trade/Device Name: Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF Dated: November 14, 2020 Received: December 8, 2020
Dear Christy Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rumi Young - Digitally signed by Rumi Young -S Date: 2021.01.05 09:58:10 -05'00'
Rumi Young Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K193273
Device Name Insulin Syringe
Indications for Use (Describe)
Insulin Syringe is intended for subcutaneous injection of U-40 insulin in the treatment of diabetes.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
### 510(k) Number: K193273
#### 1. Contact Details
### 1.1 Applicant information
| Applicant Name | Promisemed Hangzhou Meditech Co., Ltd. | | |
|----------------|--------------------------------------------------------------------------------------------------|--|--|
| Address | No. 12, Longtan Road, Cangqian Street, Yuhang District, Hangzhou City<br>311121 Zhejiang, China. | | |
| Phone No. | + 86(0571)88772985 | | |
| Fax No. | + 86(0571)88772985 | | |
| Contact person | Liqing Yang | | |
| Date Prepared | October 12, 2020 | | |
#### 1.2 Submission Correspondent
| Image: logo<br>卓远天成 | Shenzhen Joyantech Consulting Co., Ltd<br>1713A, Block A, Zhongguan Times Square, Liuxian Avenue, Xili Town,<br>Nanshan District, Shenzhen, Guangdong Province, China. |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone No. | +86-755-86069197 |
| Contact person | Joyce Yang; |
| Contact person's e-mail | joyce@cefda.com; cefda@foxmail.com |
| Website | http://www.cefda.com |
#### 2. Device information
| Trade name | Insulin Syringe |
|---------------------|-----------------|
| Common name | Insulin Syringe |
| Model/Type | Type7, Type8 |
| Classification | II |
| Classification name | Piston syringe |
| Product code | FMF |
| Regulation No. | 21 CFR 880.5860 |
#### 3. Legally Marketed Predicate Device
| | Primary Predicate Device |
|---------------|-----------------------------------|
| Trade Name | Disposable Insulin Syringes |
| 510(k) Number | K162180 |
| Product Code | FMF |
| Manufacturer | Berpu Medical Technology Co.,Ltd. |
{5}------------------------------------------------
#### 4. Device Description
The proposed device Insulin Syringe, a sterile device consisting of a calibrated barrel with plunger, is intended to be used to administer an injection of insulin to a patient subcutaneously. A non-retractable integrated needle is included. The syringe is made of plastic and silicone materials and allowing smooth plunger movement. This is a single-use device.
#### 5. Intended Use/Indications for Use
Insulin Syringe is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.
| Element of<br>comparison | Proposed Device:<br>Insulin Syringe | Predicate Device:<br>Disposable Insulin Syringe<br>(K162180) | Comment |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Product Code | FMF | FMF | Same |
| Regulation number | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Syringe type | Piston syringe | Piston syringe | Same |
| Intended use | Insulin Syringe is intended for<br>subcutaneous injection of<br>U-40 and U-100 insulin in the<br>treatment of diabetes. | The disposable insulin<br>syringe is intended for<br>medical purposes for the<br>manual aspiration of U-100<br>insulin, and for the injection of<br>insulin into parts of the body<br>below the surface skin. | Similar<br>(Note1) |
| Principle of operation | The insulin is injected to<br>subcutaneous tissue by<br>pushing force generated<br>through pushing plunger rod<br>of the insulin syringe. | The insulin is injected to<br>subcutaneous tissue by<br>pushing force generated<br>through pushing plunger rod<br>of the insulin syringe. | Same |
| Specific drug use | Insulin | Insulin | Same |
| Length | 120mm | 120mm | Same |
| Volume | 0.3ml, 0.5ml, 1.0ml | 0.3ml, 0.5ml, 1.0ml | Same |
| Needle length | 6mm, 8mm,12mm | 5mm, 6mm, 8mm, 12mm | Similar<br>(Note2) |
| Needle gauge | 32G, 31G, 30G, 29G, 28G | 31G, 30G, 29G, 28G, 27G | Similar<br>(Note2) |
| Needle dimensions | 0.23mm,0.25mm, 0.30mm,<br>0.33mm, 0.36mm, | 0.25mm, 0.30mm, 0.33mm,<br>0.36mm, 0.40mm | Similar<br>(Note2) |
| Needle tip<br>configuration | 3 bevels<br>Primary bevel length:<br>(0.84±0.15) mm<br>Primary bevel angle: 10° ± 2° | 3 bevels | Same |
| Element of<br>comparison | Proposed Device:<br>Insulin Syringe | Predicate Device:<br>Disposable Insulin Syringe<br>(K162180) | Comment |
| | | | |
| | Secondary bevel length:<br>$(0.64±0.15)$ mm<br>Secondary bevel angle: 17°<br>$±2°$<br>(The secondary bevel is<br>rotated 35° $±2°$ to form 2<br>bevels) | | |
| Nozzle type | Not applicable | Not applicable | Same |
| Numbering of scale | At every five units for the<br>0.3mL and 0.5mL syringes,<br>and at every 10units for<br>1.0mL | At every five units for the<br>0.3mL and 0.5mL syringes,<br>and at every 10units for<br>1.0mL | Same |
| Gradations legibility | Legible | Legible | Same |
| Needle cover<br>dimensions | Length:25mm, Diameter:<br>6mm | Length:25mm, Diameter:<br>6mm | Same |
| Needle cover color | Red (U-40) and orange<br>(U-100) | Orange (U-100) | Similar<br>(Note 1) |
| Lubricant<br>composition | Aminofuntional siloxane | Aminofuntional siloxane | Same |
| Lubricant<br>amount/cm² | The lubricant is not form pools<br>of fluid on the interior surface<br>of the syringe or outside<br>surfaces of the needle tube | The lubricant is not form pools<br>of fluid on the interior surface<br>of the syringe or outside<br>surfaces of the needle tube | Same |
| Barrel transparency | Transparent | Transparent | Same |
| Reuse durability | Single Use | Single Use | Same |
| Needle cover<br>strength | <15N | <15N | Same |
| Hub/needle bond<br>strength | >22N | >22N | Same |
| Biocompatibility | No cytotoxicity<br>No irritation reactivity<br>No significant evidence of<br>skin sensitization<br>No significant evidence of<br>systemic toxicity<br>No evidence of Hemolysis<br>No evidence of pyrogens | No cytotoxicity<br>No irritation reactivity<br>No significant evidence of<br>skin sensitization<br>No significant evidence of<br>systemic toxicity<br>No evidence of Hemolysis<br>No evidence of pyrogens | Same |
| Configuration<br>and<br>Materials | Needle: Stainless Steel<br>(SUS304)<br>Barrel: Polypropylene | Needle: Stainless Steel<br>(SUS304)<br>Barrel: Polypropylene (PP) | Similar<br>(Note3) |
| Element of<br>comparison | Proposed Device:<br>Insulin Syringe | Predicate Device:<br>Disposable Insulin Syringe<br>(K162180) | Comment |
| | Plunger: Polypropylene<br>Piston: Polyisoprene rubber<br>Needle cap: Polyethylene<br>Protective end cap(only type<br>8): Polyethylene | Plunger: Polypropylene (PP)<br>Piston: Polyisoprene Rubber<br>Needle cover: Polypropylene (PP)<br>Protective end cap:<br>Polypropylene (PP) | Same |
| Label | Device name, indication,<br>instruction for use,<br>precaution, warning, shelf life,<br>manufacturer | Device name, indication,<br>instruction for use,<br>precaution, warning, shelf life,<br>manufacturer | |
| Sterilization method<br>and SAL | Sterilized by ethylene oxide<br>gas SAL = 10-6 | Sterilized by ethylene oxide<br>gas SAL = 10-6 | |
| Sterilization<br>validation standard | ISO 11135 | ISO 11135 | |
#### 6. Substantial Equivalence Comparison
{6}------------------------------------------------
{7}------------------------------------------------
lssue:
Note1: The subject device and predicate device have the same indication, that is, for insulin injection. The subject device has another type of syringe for injection of U-40 insulin. The sizes, graduated scale and tolerance on graduated capacity of U-40 insulin syringe are same as the U-100 insulin syringe. The difference between U-40 and U-100 insulin syringe is delivery different insulin concentration. For U-40 insulin syringe, a red needle cap is used for color marking in accordance with ISO 8573. In addition, the tolerance on graduated capacity has been validated per ISO 8573 Annex H. Therefore, this difference does not affect the Substantially Equivalent between the subject device and predicate device.
Note 2: The needle length of proposed device is covered by the predicate device.
The proposed device has extra 32G than the predicate device, and the 32G needle is similar the 31G in dimension. This difference does not affect intended use. In addition, the legally marketed similar device such as Safety Insulin Needle (K181447) has the same needle gauge.
The gauge (32G) for proposed device is different from the predicate device. However, this difference is just in dimension. Different gauge will be selected by physician per patient's condition. This difference does not affect intended use. Differences in needle length and gauge between the predicate and subject device were addressed through ISO 8537:2016 performance testing. Therefore, this difference does not affect substantially equivalence on safety and effectiveness.
Note 3: The materials of needle cap and protective end cap are different between the subject device and predicate device. The biocompatibility test of the subject device was conducted to demonstrate that the subject device met the biocompatibility requirements. So this difference does not raise any safety and effectiveness problems.
#### 7. Non-clinical studies and tests performed
The Insulin Syringes have been designed and tested to meet the requirements of voluntary standards and FDA guidance documents applicable to the subject and predicate device. The results
{8}------------------------------------------------
of the non-clinical testing supported the conclusion of substantial equivalence.
#### Performance Testing
The Insulin Syringes have been designed and successfully tested to meet the applicable requirements outlined in ISO 8537:2016 Sterile single-use syringes, with or without needle, for insulin.
#### Biocompatibility Testing
The material of the Insulin Syringes have successfully passed testing as outlined in ISO 10993-1 for devices categorized as External communicating devices, Limited exposure.
ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Test for Irritation and Sensitization
ISO 10993-11: 2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity
ISO 10993-4:2017 Biological evaluation of medical devices-Part 4: Selection of tests for interactions with blood.
ASTM F 756-17 Standard Practice for Assessment of Hemolytic Properties of Materials
The United States Pharmacopeia <151> (Pyrogen test)
The United States Pharmacopeia <788> Particulate matters
#### Sterilization and Shelf-life Testing
Sterilization of the Insulin Syringes has been validated according to ISO 11135.Testing demonstrated maximum levels of residues of ethylene oxide and ethylene chlorohydrins do not exceed the limits presented in ISO 10993-7. Shelf-life testing supports a shelf life of 5-years after sterilization.
#### 8. Clinical study
No prospective clinical trials were conducted in support of this 510(K).
#### 9. Conclusion
Based on the comparison and analysis above, the proposed devices Insulin Syringes are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.