SmartRelease Endoscopic Soft Tissue Release System
Applicant
Microaire Surgical Instruments
Product Code
HRX · Orthopedic
Decision Date
Feb 2, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MicroAire® SmartRelease® Endoscopic Soft Tissue Release System is intended for use in minimally invasive ligament or fascia release.
Device Story
Endoscopic surgical system for minimally invasive ligament/fascia release. Components: reusable handpiece, endoscope, accessory instruments; single-use disposable blade assemblies. Surgeon inserts blade assembly into surgical site (wrist, elbow, or finger) via small incision; attaches standard video camera/light source to endoscope for visualization. Trigger mechanism on handpiece advances shuttle to raise cutting blade along track in blade case; surgeon cuts target tissue (transverse carpal ligament, cubital tunnel fascia, or A1 pulley) while withdrawing instrument. Released trigger returns blade to retracted position via internal spring. Used in clinical/surgical settings by surgeons. Provides direct visualization of deep tissue structures; enables precise, minimally invasive release; reduces surgical trauma compared to open procedures.
Clinical Evidence
Bench testing and user validation. User validation involved 20 surgeons performing trigger finger release on 80 fingers and 40 thumbs in cadaver labs. Results confirmed no usability risks, safe/effective use, and fulfillment of user needs. Shelf life supported by real-time aging testing.
Technological Characteristics
Reusable handpiece (aluminum, stainless steel, Radel), endoscope (stainless steel, titanium, PEEK, Sapphire glass), and accessory instruments (stainless steel). Single-use blade assemblies (polyetherimide, polyacrylamide, stainless steel). Mechanical actuation via trigger-driven shuttle and actuator. No electrical energy or battery. Sterilization: Gamma radiation (25-40kGy) for blades; steam autoclave for reusable components. Complies with ISO 10993-1, ISO 14971, ISO 11137, ISO 11607, and ANSI AAMI ST79.
Indications for Use
Indicated for carpal tunnel release in the wrist, cubital tunnel release in the elbow, and trigger finger release in the hand.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
SmartRelease® Endoscopic Soft Tissue Release System (K181819)
Submission Summary (Full Text)
{0}------------------------------------------------
February, 2, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below that.
MicroAire Surgical Instruments Glenn Gerstenfeld Vice President, QA/RA & Compliance Officer 3590 Grand Forks Blvd Charlottesville, Virginia 22911
#### Re: K223406
Trade/Device Name: SmartRelease® Endoscopic Soft Tissue Release System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, KCT, EMF Dated: November 4, 2022 Received: November 9, 2022
#### Dear Glenn Gerstenfeld:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sara S. Thompson -S
For
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K223406
#### Device Name
SmartRelease® Endoscopic Soft Tissue Release System
Indications for Use (Describe)
The MicroAire® SmartRelease® Endoscopic Soft Tissue Release System is indicated for:
- · Carpal tunnel release in the wrist
- Cubital tunnel release in the elbow
- · Trigger finger release in the hand
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Section 5: 510(k) Summary
## SmartRelease® Endoscopic Soft Tissue Release System
#### 1. Submission Sponsor
MicroAire® Surgical Instruments, LLC
3590 Grand Forks Blvd.
Charlottesville, VA 22911
USA
Contact: Glenn Gerstenfeld
Title: VP, QA/RA and Compliance Officer
Phone: (434) 975-8344
Fax: (434) 975-4144
#### 2. Date Prepared
November 4, 2022
#### 3. Device Identification
Trade/Proprietary Name: SmartRelease Endoscopic Soft Tissue Release System Common/Usual Name: SmartRelease Classification Name: Arthroscope Regulation Number: 21 CFR 888.1100 Product Code: HRX, KCT, EMF Class: Class II Classification Panel: Orthopedic
#### 4. Predicate Device(s)
The predicate device for this submission is MicroAire's SmartRelease® Endoscopic Soft Tissue Release System (K181819). This predicate has not been the subject of a design-related recall. No reference devices were used in this submission.
{4}------------------------------------------------
#### 5. Intended Use / Indication for Use Statement
Intended Use: The MicroAire® SmartRelease® Endoscopic Soft Tissue Release System is intended for use in minimally invasive ligament or fascia release.
Indications for Use: The MicroAire® SmartRelease® Endoscopic Soft Tissue Release System is indicated for:
- Carpal tunnel release in the wrist
- . Cubital tunnel release in the elbow
- Trigger finger release in the hand
#### 6. Device Description
MicroAire's SmartRelease Endoscopic Soft Tissue Release System is comprised of the following components: the SmartRelease Endoscope, the SmartRelease Handpiece, and the SmartRelease Standard and Onyx Blade Assemblies. The system includes a set of accessory Manual Surgical Instruments. The SmartRelease Handpiece, the SmartRelease Endoscope, and the Manual Surgical Instruments are reusable after cleaning and subsequent sterilization. The SmartRelease Blade Assemblies are gamma sterilized and intended for single-use only.
#### Carpal Tunnel:
The MicroAire SmartRelease Endoscopic Soft Tissue Release System is comprised of an endoscope and a handpiece that holds a Disposable Blade assembly. The carpal tunnel device attaches to any standard video camera and light source used in endoscopic/arthroscopy procedures. The surgeon introduces the Disposable Blade assembly into the carpal tunnel through an incision in the proximal wrist flexion crease. Viewing the deep side of the transverse carpal ligament (flexor retinaculum) through a window at the tip of the instrument, the surgeon elevates the blade to cut the ligament as the instrument is withdrawn.
#### Cubital Tunnel:
The MicroAire SmartRelease Endoscopic Soft Tissue Release System is comprised of an endoscope and a handpiece that holds a Disposable Blade assembly. The cubital tunnel device attaches to any standard video camera and light source used in endoscopic procedures. The surgeon introduces the disposable blade assembly into the cubital tunnel through a 3-cm longitudinal incision between the medial epicondyle and olecranon. Viewing the roof of the cubital tunnel fascia through a window at the tip of the instrument, the surgeon elevates the blade to cut the instrument is withdrawn.
#### Trigger Finger:
The MicroAire SmartRelease Endoscopic Soft Tissue Release System is comprised of an endoscope and a handpiece that holds a Disposable Blade assembly. The trigger finger device attaches to any standard video camera and light source used in endoscopic procedures. The surgeon introduces the
{5}------------------------------------------------
disposable blade assembly into the finger through a 1-cm incision crease of the affected finger. Viewing the roof of the A1 Pulley through a window at the tip of the instruments, the surgeon elevates the blade to cut the pulley as the instrument is withdrawn.
#### 7. Substantial Equivalence Discussion
#### Technological Characteristics
This submission is to add an Indication for Use to the SmartRelease Endoscopic Soft Tissue Release System and does not include any design changes to the product family. There have been no design changes between the originally cleared SmartRelease Endoscopic Soft Tissue Release System and the proposed product family. Thus, the two SmartRelease devices have technological characteristics as there have been no changes to the design, material, chemical composition, energy source, manufacturing, packaging, or sterilization of the device, as summarized in the table below.
| Technological<br>Characteristic | SUBJECT<br>SmartRelease Endoscopic Soft<br>Tissue Release System | PREDICATE<br>510(K) K181819<br>SmartRelease Endoscopic Soft<br>Tissue Release System | Comparison |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Design | | | |
| Mechanism of<br>Action | The blade assembly locks onto the front of the handpiece. The endoscope is inserted through the handpiece into the blade assembly. A light source and camera are attached to the endoscope. When the blade assembly is inserted into the surgical site, the endoscope provides a view of the distal end of the blade assembly and tissue to be resected. When the user pulls the trigger on the handpiece, it advances the shuttle into the proximal end of the blade assembly, pushing the actuator forward. This forward movement of the actuator raises the cutting blade along a track in the blade case. With the cutting blade fully raised, the user can slowly pull the assembly back out of the surgical site, allowing the cutting blade to | The blade assembly locks onto the front of the handpiece. The endoscope is inserted through the handpiece into the blade assembly. A light source and camera are attached to the endoscope. When the blade assembly is inserted into the surgical site, the endoscope provides a view of the distal end of the blade assembly and tissue to be resected. When the user pulls the trigger on the handpiece, it advances the shuttle into the proximal end of the blade assembly, pushing the actuator forward. This forward movement of the actuator raises the cutting blade along a track in the blade case. With the cutting blade fully raised, the user can slowly pull the assembly back out of the surgical site, allowing the cutting blade to | Identical to<br>predicate device |
| Technological<br>Characteristic | SUBJECT<br>SmartRelease Endoscopic Soft<br>Tissue Release System | PREDICATE<br>510(K) K181819<br>SmartRelease Endoscopic Soft<br>Tissue Release System | Comparison |
| | resect the tissue of interest.<br>When the trigger is released, the<br>force of the shuttle on the<br>actuator is removed, and a spring<br>in the blade assembly returns the<br>actuator and cutting blade to their<br>original, retracted position. | resect the tissue of interest.<br>When the trigger is released, the<br>force of the shuttle on the<br>actuator is removed, and a spring<br>in the blade assembly returns the<br>actuator and cutting blade to their<br>original, retracted position. | |
| Materials | • Handpiece – aluminum,<br>stainless steel, Radel | • Handpiece – aluminum,<br>stainless steel, Radel | Identical to<br>predicate device |
| | • Blade Assemblies –<br>polyetherimide,<br>polyacrylamide, stainless steel | • Blade Assemblies –<br>polyetherimide,<br>polyacrylamide, stainless steel | |
| | • Endoscope – stainless steel,<br>titanium, PEEK, Sapphire glass,<br>CASTIN lead-free solder | • Endoscope - stainless steel,<br>titanium, PEEK, Sapphire glass,<br>CASTIN lead-free solder | |
| | • Sterilization Tray – stainless<br>steel, silicone | • Sterilization Tray – stainless<br>steel, silicone | |
| | • Accessory Instruments -<br>stainless steel | • Accessory Instruments -<br>stainless steel | |
| Single-Use /<br>Reusable | • Blade Assemblies are single<br>use, disposable devices | • Blade Assemblies are single<br>use, disposable devices | Identical to<br>predicate device |
| | • Handpiece, endoscope,<br>sterilization tray, and accessory<br>instruments are reusable<br>devices | • Handpiece, endoscope,<br>sterilization tray, and accessory<br>instruments are reusable<br>devices | |
| Shelf Life | Blade Assemblies have 4 year shelf<br>life | Blade Assemblies have 4 year shelf<br>life | Identical to<br>predicate device |
| Energy Output | None | None | Identical to<br>predicate device |
| Battery Operated | No | No | Identical to<br>predicate device |
| AC Powered | No | No | Identical to<br>predicate device |
| Technological<br>Characteristic | SUBJECT<br>SmartRelease Endoscopic Soft<br>Tissue Release System | PREDICATE<br>510(K) K181819<br>SmartRelease Endoscopic Soft<br>Tissue Release System | Comparison |
| Complies with ISO<br>10993-1 | Yes, complies with ISO 10993-<br>1:2018 | Yes, complies with ISO 10993-<br>1:2018 | Identical to<br>predicate device |
| Electrical Safety<br>Testing | N/A | N/A | Identical to<br>predicate device |
| Manufacturing | | | |
| Handpiece | Components machined both<br>internally and by external<br>suppliers. Anodized by external<br>supplier. Assembled, labeled, and<br>packaged internally. | Components machined both<br>internally and by external<br>suppliers. Anodized by external<br>supplier. Assembled, labeled, and<br>packaged internally. | Identical to<br>predicate device |
| Endoscope | Manufactured, labeled, and<br>packaged complete by external<br>supplier. | Manufactured, labeled, and<br>packaged complete by external<br>supplier. | Identical to<br>predicate device |
| Blade Assemblies | Components molded/<br>manufactured by external<br>suppliers. Assembled, packaged,<br>and labeled internally. Sterilized<br>by external supplier. | Components molded/<br>manufactured by external<br>suppliers. Assembled, packaged,<br>and labeled internally. Sterilized<br>by external supplier. | Identical to<br>predicate device |
| Sterilization Tray | Manufactured, labeled, and<br>packaged complete by external<br>supplier. | Manufactured, labeled, and<br>packaged complete by external<br>supplier. | Identical to<br>predicate device |
| Accessory<br>Instruments | Manufactured and labeled<br>complete by external supplier.<br>Packaged internally. | Manufactured and labeled<br>complete by external supplier.<br>Packaged internally. | Identical to<br>predicate device |
| Technological<br>Characteristic | SUBJECT<br>SmartRelease Endoscopic Soft<br>Tissue Release System | PREDICATE<br>510(K) K181819<br>SmartRelease Endoscopic Soft<br>Tissue Release System | Comparison |
| Packaging | | | |
| Reusable Devices | • Handpiece – placed in a Korrvu<br>insert (plastic film and<br>corrugated board), then placed<br>in corrugated shipper box.<br>• Endoscope – placed in a<br>custom-fit foam insert, then<br>placed in corrugated shipper<br>box.<br>• Sterilization Tray – shrink-<br>wrapped with plastic film, then<br>placed in corrugated shipper<br>box.<br>• Accessory Instruments – placed<br>in poly bags, heat-sealed on<br>both ends. | • Handpiece – placed in a Korrvu<br>insert (plastic film and<br>corrugated board), then placed<br>in corrugated shipper box.<br>• Endoscope – placed in a<br>custom-fit foam insert, then<br>placed in corrugated shipper<br>box.<br>• Sterilization Tray – shrink-<br>wrapped with plastic film, then<br>placed in corrugated shipper<br>box.<br>• Accessory Instruments – placed<br>in poly bags, heat-sealed on<br>both ends. | Identical to<br>predicate device |
| Sterile Devices<br>(blade assemblies) | Device placed in PETG Tray with<br>Tyvek lid, heat sealed on all four<br>sides. Tray placed in SBS box<br>(single- or 6-pack configuration). | Device placed in PETG Tray with<br>Tyvek lid, heat sealed on all four<br>sides. Tray placed in SBS box<br>(single- or 6-pack configuration). | Identical to<br>predicate device |
| Sterilization | | | |
| Sterilization | • Blade Assemblies are provided<br>sterile via gamma radiation, 25-<br>40kGy<br>• Handpiece, endoscope, and all<br>accessories are provided non-<br>sterile, to be cleaned and<br>sterilized via steam autoclave<br>by user prior to each use | • Blade Assemblies are provided<br>sterile via gamma radiation, 25-<br>40kGy<br>• Handpiece, endoscope, and all<br>accessories are provided non-<br>sterile, to be cleaned and<br>sterilized via steam autoclave<br>by user prior to each use | Identical to<br>predicate device |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
#### Performance Data
As the subject device is identical in construction, geometry, functionality, and Intended Use to its cleared predicate, previous testing provided in the cleared K181819 demonstrates that the SmartRelease
{9}------------------------------------------------
Endoscopic Soft Tissue Release System is safe and effective for its Intended Use. Nothing has changed with the SmartRelease System to alter that conclusion.
Shelf life testing originally submitted in K181819 was based on product that had undergone accelerated aging conditioning. Since that submission was cleared, performance testing has been completed on product that has undergone real time aging to support the labeling shelf life. All products under consideration passed this testing.
A new User Validation was performed with appropriate users for the subject device's new Indication for Use to evaluate the user interfaces and to confirm that the intended user groups can appropriately use the device and demonstrate that the device is safe and effective for the new Indication for Use. Twenty users participated in cadaver labs with the subject device, walking through the procedure for trigger finger release on both fingers and thumbs on multiple hands each, for a total evaluation of 80 fingers and 40 thumbs. The results confirm that the SmartRelease Endoscopic Soft Tissue Release System 1) presents no usability risks, 2) can be safely and effectively used as intended, and 3) meets the User Needs.
#### Intended Use
The Intended Use is identical between the subject device and predicate. The addition of the "Trigger finger release in the hand" Indication for Use does not change the Intended Use nor raise any concerns regarding safety or effectiveness, as discussed above.
#### 8. Statement of Substantial Equivalence
The subject device, SmartRelease Endoscopic Soft Tissue Release System, is the identical product as its predicate, cleared under K181819. It has the same technological characteristics, same performance testing and results, and the same Intended Use as its predicate. The added Indication for Use does not raise new or different questions of safety or effectiveness, as demonstrated in the User Validation. Therefore, the subject SmartRelease Endoscopic Soft Tissue Release System is substantially equivalent to the predicate device.
#### 9. Referenced Standards
The SmartRelease Endoscopic Soft Tissue Release System also complies in whole or in part to the following consensus standards.
- ISO 14971:2019, Medical devices – Application of risk management to medical devices
- ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . ISO 15223-1:2021, Medical devices – Symbol to be used with information to be supplied by the manufacturer - Part 1: General requirements
- ISO 20417:2021, Medical devices – Information to be supplied by the manufacturer
{10}------------------------------------------------
- IEC 62366-1:2020, Medical devices – Part 1: Application of usability engineering to medical devices
- . ISO 11137-1:2006, Sterilization of health care products – Radiation – Part 1: Requirements for development validation and routine control of a sterilization process for medical devices
- . ISO 11137-2:2013, Sterilization of health care products – Radiation – Part 2: Establishing the sterilization dose
- ISO 11607-1:2019, Packaging for terminally sterilized medical devices – Part 1: Requirements for materials sterile barrier systems and packaging systems
- . ISO 11607-2:2019, Packaging for terminally sterilized medical devices – Part 2: Validation requirements for forming sealing and assembly processes
- . ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- ASTM F2096-11, Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
- ANSI AAMI ST77:2013/(R)2018, Containment devices for reusable medical device sterilization
- ANSI AAMI ST79:2017, Comprehensive guide to steam sterilization and sterility assurance in ● health care facilities
#### 10. Conclusion
SmartRelease Endoscopic Soft Tissue Release System is substantially equivalent to the predicate device in technical characteristics, design features, operating principles, functional and performance characteristics, and for the intended uses in the stated medical specialties. The SmartRelease Endoscopic Soft Tissue Release System is designed to comply with applicable federal safety and performance standards. The User Validation data support the new indication claim. There are no new safety and effectiveness issues.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.