SMARTRELEASE Endoscopic Soft Tissue Release System
Applicant
Microaire Surgical Instruments, LLC
Product Code
HRX · Orthopedic
Decision Date
Nov 6, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MicroAire® SMARTRELEASE® Endoscopic Soft Tissue Release System is indicated for use in minimally invasive ligament or fascia release: • Carpal tunnel release in the wrist • Cubital tunnel release in the elbow
Device Story
System comprises endoscope, pistol-grip handpiece with trigger, blade assemblies (standard/low profile), and sterilization tray. Endoscope provides visualization via connection to standard C-mount video camera systems and external light sources. Handpiece facilitates mechanical raising and retracting of surgical blade. Blade assembly performs ligament/fascia release via single retracting cut. Used in surgical settings by physicians for carpal/cubital tunnel release. Reusable components (handpiece, endoscope, manual instruments) undergo cleaning and heat sterilization; blade assemblies are single-use, gamma-sterilized. System enables minimally invasive access, potentially reducing surgical trauma compared to open procedures.
Clinical Evidence
Bench testing performed: blade actuation, horizontal/vertical deflection, sharpness, and piercing. Cadaveric study conducted with multiple surgeons to evaluate system performance per instructions for use. Package, distribution, and biocompatibility testing completed. No clinical human trial data reported.
Technological Characteristics
Components: 300/400 series stainless steel blades, 300 series stainless steel tray with silicone mat. Endoscope: standard arthroscope, C-mount compatible, no electrical source. Handpiece: mechanical pistol-grip/trigger. Sterilization: heat sterilization for reusable components (meets ANSI AAMI ST 77:2003), gamma sterilization for single-use blades. Connectivity: standard video/light source interfaces.
Indications for Use
Indicated for patients requiring minimally invasive ligament or fascia release for carpal tunnel release in the wrist or cubital tunnel release in the elbow.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 6, 2018
MicroAire Surgical Instruments LLC % Ms. Vikki O'Connor Regulatory Affairs Consultant Ambriel Associates, Inc. 411 Walnut St. Unit 9236 Green Cove Springs, Florida 32043
Re: K181819
Trade/Device Name: SMARTRELEASE Endoscopic Soft Tissue Release System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX, KCT, EMF Dated: October 4, 2018 Received: October 5, 2018
Dear Ms. O'Connor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long H. Chen - Digitally signed by Long H.
Chen - Chen - Chen -s Date: 2018.11.06 08:16:55 S -05'00'
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K181819
#### Device Name
MicroAire SMARTRELEASE® Endoscopic Soft Tissue Release System
Indications for Use (Describe)
The MicroAire SMARTRELEASE® Endoscopic Soft Tissue Release System is indicated for use in minimally invasive ligament or fascia release:
- · Carpal tunnel release in the wrist
- Cubital tunnel release in the elbow
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: MicroAire SMARTRELEASE®
# Endoscopic Soft Tissue Release System
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular, 21 CFR Part 807.92, the following summary of information is provided:
| Submitter: | MicroAire® Surgical Instruments Inc. |
|----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 3590 Grand Forks Blvd. |
| | Charlottesville, VA 22911 USA |
| Contact Person | Vikki M. O'Connor |
| | Regulatory Consultant |
| | Phone : 207-214-8535 |
| | Email : vikki0730@yahoo.com |
| Correspondance | Mr. Glenn Gerstenfeld |
| | Vice President, QA/RA and Compliance Officer |
| | Phone: 434-975-8344 |
| | FAX: 434-975-4144 |
| | Email: glenn.gerstenfeld@microaire.com |
| Date Prepared | November 5, 2018 |
| Trade Name | MicroAire SMARTRELEASE® Endoscopic Soft Tissue |
| | Release System |
| Proposed Class | Class II |
| Classification Name | Diagnostic Devices - Arthroscope, 21 CFR 888.1100, |
| and Number | Class II |
| | Sterilization Container, 21 CFR 880.6580, Class II |
| | Manual Surgical Instrument, 21 CFR 878.4800, Class I |
| Product Code | HRX, KCT, EMF |
| Predicate Devices | Deployment Handle and Blade Assembly Primary |
| | Predicate – A.M. Surgical Mountable Endoscopic Blade - |
| | K080133 (Class I), A.M. Surgical STRATOS® Endoscopic |
| | Release System – Carpal and Cubital Tunnel Release |
| | (Class I, no 510k). Reference: 3M Carpal Tunnel System - |
| | K881703 (Class II) Carpal Tunnel Release. |
| | Arthroscope Primary Predicate - Henke Sass Wolfe |
| | Arthroscope: K080560 (Class II); |
| | Sterilization Tray Primary Predicate - Paragon Medical |
| | Inc. Surgical Instrument Delivery System: K032119 (Class |
| | II); |
| Device Description | MicroAire's SMARTRELEASE® Endoscopic Soft Tissue<br>Release System is comprised of four (4) components: The<br>SMARTRELEASE® Endoscope, the SMARTRELEASE®<br>Handpiece, the SMARTRELEASE® Standard and Low<br>Profile Blade Assemblies and the SMARTRELEASE®<br>Sterilization Tray. In addition, the system includes a set of<br>FDA Class 1 Manual Surgical Instruments. The<br>SMARTRELEASE® Handpiece, the SMARTRELEASE®<br>Endoscope, and the Manual Surgical Instruments are re-<br>usable after cleaning and subsequent sterilization in the<br>SMARTRELEASE® Sterilization Tray. The<br>SMARTRELEASE® Blade Assemblies are gamma<br>sterilized, single-use devices. |
| Intended Use | The MicroAire® SMARTRELEASE® Endoscopic Soft<br>Tissue Release System is indicated for use in minimally<br>invasive ligament or fascia release:<br>• Carpal tunnel release in the wrist<br>• Cubital tunnel release in the elbow |
| Subject Device<br>Summary of the<br>Technological<br>Characteristics | The SMARTRELEASE® Endoscope is a standard<br>endoscope used in endoscopic / arthroscopic surgery. It<br>connects to any standard C-mount video camera system<br>for visualization and has a light post and two Light Post |
| | Adapters that attach to six of the most common light<br>sources. There is no electrical source and the endoscope<br>is re-usable after cleaning and heat sterilization.<br><br>The SMARTRELEASE® Handpiece has a pistol grip and<br>trigger to facilitate raising and retracting the surgical blade.<br>The SMARTRELEASE® Blade Assembly is designed to<br>cut the carpal tunnel ligament or fascia release in the<br>cubital tunnel with one retracting cut.<br><br>The SMARTRELEASE® Sterilization Tray is large enough,<br>and segmented, to hold and heat sterilize all devices and<br>instruments required to perform a carpal or cubital tunnel<br>surgery. The sterilization tray meets the requirements set<br>forth in ANSI AAMI ST 77:2003. |
| Comparison to<br>Predicate(s) | The SMARTRELEASE® Endoscope is made by contract<br>manufacturer Henke Sass Wolf and is identical to the<br>Henke Sass Wolf Arthroscope (K080560). This is the same<br>arthroscope used in the 3M Carpal Tunnel System.<br><br>MicroAire's SMARTRELEASE® Handpiece and Blade<br>Assembly is similar in technology, materials and design to<br>the predicates. The subject and predicate devices all<br>provide technology to access the wrist or elbow to release<br>the ligament or fascia during carpal or cubital tunnel<br>surgery. The Handpieces are made of similar materials,<br>hold the Blade Assembly and can deploy the blade to<br>make a cut and retract the blade for removal of the device<br>from the surgical area(s). The blades are made of 300/400<br>series stainless steel and are sharp enough to cut fascia / |
| | ligament and are biocompatible. |
| | The subject and predicate tray are made from 300 Series<br>Stainless Steel and hold all devices that require re-<br>sterilization. The subject sterilization tray includes a silicon<br>mat. |
| Summary of<br>Supporting Data | The following functional testing was performed to verify the<br>following basic functions of the Handpiece and Blade<br>Assembly: Blade Assembly Actuation, Horizontal and<br>Vertical Deflection, Blade Sharpness and Blade Piercing.<br>The system was also evaluated by multiple surgeons per<br>the instructions for use during a cadaver study. All test<br>results demonstrated the MicroAire SMARTRELEASE Soft<br>Tissue Release System can perform it's intended use<br>safely and effectively.<br>In addition, the SMARTRELEASE® Endoscopic Soft<br>Tissue Release System passed package and distribution<br>testing and applicable biocompatibility testing. |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
### Summary / Conclusion
MicroAire's SMARTRELEASE® Endoscopic Release System is comprised of the SMARTRELEASE® Handpiece and SMARTRELEASE® Blade Assembly(s) which are FDA Class I Manual Surgical Instruments. The addition of the dedicated SMARTRELEASE® Endoscope to the system and a Sterilization Tray for re-sterilizing the Endoscope, Handpiece and Class I instruments make the overall system a Class II device.
Basic device functional tests and a cadaveric study by multiple surgeons were performed to evaluate the system for the intended use. All test results demonstrated the MicroAire SMARTRELEASE Soft Tissue Release System can perform the intended use safely and effectively.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.