K223336 · Canon Medical Systems Corporation · LLZ · Jan 9, 2023 · Radiology
Device Facts
Record ID
K223336
Device Name
Vitrea Software Package, VSTP-002A
Applicant
Canon Medical Systems Corporation
Product Code
LLZ · Radiology
Decision Date
Jan 9, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
Vitrea Software Package is an application package developed for use on Vitrea®, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea Software Package has the following additional indications: Auto MPR application is a post processing software of CT brain images that is intended to align images into a standard anatomical position for review. It provides tools to reformat images parallel to a standard anatomical position.
Device Story
Vitrea Software Package (VSTP-002A) is a software application suite for the Canon Medical Informatics Vitrea workstation; used by physicians to process, review, and analyze multi-dimensional digital images. The Auto MPR application takes 3D volumetric CT brain images as input; automatically aligns images into a standard anatomical position; and reformats images for review. The device operates within the Vitrea workstation environment; no external network or internet connectivity. Output is intended for PACS display systems. The device automates manual image alignment processes; assisting physicians in specialized measurements and analysis; potentially improving workflow efficiency and diagnostic review consistency.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed Auto MPR output image alignment into standard anatomical positions and assessed performance under various conditions. Results demonstrated the device met all established specifications.
Technological Characteristics
Software-based medical image management and processing system. Operates on Canon Medical Informatics Vitrea workstations. Conforms to ISO 13485, IEC 62304, IEC 60601-1-6, IEC 60601-1-9, and IEC 62366-1. Standalone software application within the Vitrea workstation environment; no external network connectivity.
Indications for Use
Indicated for use by clinicians to process, review, and analyze multi-dimensional digital images from various imaging devices. The Auto MPR application is indicated for post-processing CT brain images to align them into a standard anatomical position for review.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
Canon Medical Systems Corporation % Mr. Orlando Tadeo Sr. Manager, Regulatory Affairs Canon Medical Systems USA, Inc. 2441 Michelle Drive TUSTIN CA 92780
#### Re: K223336
January 9, 2023
Trade/Device Name: Vitrea Software Package, VSTP-002A Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: November 17, 2022 Received: November 17, 2022
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K223336
Device Name Vitrea Software Package, VSTP-002A
#### Indications for Use (Describe)
Vitrea Software Package is an application package developed for use on Vitrea®, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea Software Package has the following additional indications:
Auto MPR application is a post processing software of CT brain images that is intended to align images into a standard anatomical position for review. It provides tools to reformat images parallel to a standard anatomical position.
Type of Use (Select one or both, as applicable)
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
1. SUBMITTER'S NAME Fumiaki Teshima Sr. Manager, Quality Assurance Department Canon Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
## 2. ESTABLISHMENT REGISTRATION 9614698
### 3. OFFICIAL CORRESPONDENT/CONTACT PERSON
Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Department Canon Medical Systems USA, Inc. 2441 Michelle Dr. Tustin, CA 92780 (714) 669-7459
- 4. DATE PREPARED October 31, 2022
- 5. DEVICE NAME(S) Vitrea Software Package, VSTP-002A
- 6. TRADE NAME(S) Vitrea Software Package, VSTP-002A
#### 7. COMMON NAME
Radiological Image Processing Software
#### 8. DEVICE CLASSIFICATION
Class II (per 21 CFR 892.2050, Medical Image Management and Processing System) Medical Image Management and Processing System- Product Code: 90LLZ [per 21 CFR 892.2050]
Image /page/3/Picture/15 description: The image shows the text "Made For life" in a bold, sans-serif font. The words are arranged horizontally, with "Made" and "For" in a slightly larger font size than "life". The text is black against a white background, creating a high contrast and making it easily readable.
K223336
{4}------------------------------------------------
#### 9. PREDICATE DEVICE
| Product | Marketed by | Regulation Number | Regulation Name | Product Code | 510(k) Number | Clearance Date |
|-----------------------------------------------------------------------|------------------------------------|---------------------|---------------------------------------------------------------|--------------|---------------|----------------|
| Vitrea<br>Software<br>Package,<br>VSTP-001A<br>(Primary<br>Predicate) | Canon<br>Medical<br>Systems<br>USA | 21 CFR<br>§892.2050 | Medical<br>Image<br>Management<br>and<br>Processing<br>System | LLZ | K213504 | 2/16/2021 |
| CT CoPilot<br>(Reference<br>Predicate) | ZepMed,<br>LLC. | 21 CFR<br>§892.2050 | Medical<br>Image<br>Management<br>and<br>Processing<br>System | LLZ | K161322 | 12/07/2016 |
#### 10. REASON FOR SUBMISSION
New software application
#### 11. DEVICE DESCRIPTION
The Vitrea Software Package, VSTP-002A, is a portfolio of applications software designed to be used in the Canon Medical Informatics Vitrea workstation. VSTP-002A currently includes a post processing application, Auto MPR, which use CT brain image data, obtained from Canon CT Systems, to assist physicians in performing specialized measurements and analysis.
Auto MPR is a software application that aligns CT brain images into a standard anatomical position for review.
#### 12. INDICATIONS FOR USE
Vitrea Software Package is an application package developed for use on Vitrea, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. Vitrea Software Package has the following additional indications:
Auto MPR application is a post processing software of CT brain images that is intended to align images into a standard anatomical position for review. It provides tools to reformat images parallel to a standard anatomical position.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Canon Medical Systems USA, INC. The word "Canon" is in red, and the rest of the text is in black. The logo is simple and clean, and it is easy to read. The text is left-aligned.
#### 13. SUBSTANTIAL EQUIVALANCE
The Vitrea Software Package, VSTP-002A, perform in a manner similar to and are intended for the same use as VSTP-001A, 510(k) cleared per K213504. Both are an application package developed for use on Vitrea, a medical diagnostic system that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices. The software application, Auto MPR, included in Vitrea Software Package, VSTP-002A, performs in a manner similar to and is intended for the same use as CT CoPilot, 510(k) cleared per K161322. Both are post processing software used to provide auto reformatting for standard viewing of CT brain images and are intended to automate the current manual process of image alignment. See below for a brief comparison of the technological characteristics between the subject and the predicate device:
{6}------------------------------------------------
| Device Name | Subject | Primary Predicate | Reference Predicate | |
|------------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|--------------------------------------------------------------------------------------|
| | Vitrea Software Package<br>(VSTP-002A) V1.0 | Vitrea Software Package<br>(VSTP-001A) V8.10 | CT CoPilot | |
| Available Applications | Auto MPR | Cerebral Aneurysm<br>Analysis, MR Wall<br>Motion Tracking, MR<br>Coronary Tracking,<br>SUREVolume Synthesis,<br>Angio Viewer, US<br>Cardiac Fusion,<br>Ultrasound Clinical<br>Applications, Spectral<br>Stone Analysis, Spectral<br>Composition Analysis,<br>Embolization Plan,<br>Spectral Analysis | N/A | |
| Auto MPR | | | | |
| - | Anatomical Region | Brain | N/A | Brain |
| - | Type of Input Data | 3D Volumetric CT image | N/A | CT image |
| - | Type of Output Data | The output is intended<br>for PACS display system. | N/A | The output is<br>intended for PACS<br>display system.<br>- 3D Volumetric<br>- 2D MPR |
| - | Auto reformat for<br>standard viewing of<br>brain images. | Available<br>(OM plane base) | N/A | Available |
| - | Auto Labeling of<br>segmentable brain<br>structures. | N/A | N/A | Available |
| - | Volumetric<br>quantification of<br>brain images (CSF<br>volumes,<br>Intracranial volume,<br>Midline shift) | N/A | N/A | Available |
{7}------------------------------------------------
## 14. SAFETY
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC62304, IEC60601-1-6, IEC60601-1-9, and IEC62366-1.
### 15. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrates that the requirements for the application have been met. Bench studies were conducted to test Auto MPR output image alignment into a standard anatomical position and assessed the impact of various conditions on Auto MPR image alignment. Results of both these studies demonstrated that Auto MPR met established specifications and performed as intended.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also referenced as for this submission.
### 16. CYBERSECURITY
The VSTP-002A is a portfolio of software applications developed by Canon that are integrated into the Canon Medical Informatics Vitrea workstations. The Vitrea workstation includes all cybersecurity controls and is responsible for importing images into an internal database. The software applications included in VSTP-002A pulls applicable images from the Vitrea directory to be processed and then places the processed images back into the directory within Vitrea. This software has no connections to the internet, wired or wireless networks, etc. All import and export of data is within the domain of the Canon Medical Informatics product.
#### 17. CONCLUSION
The Vitrea Software Package, VSTP-002A, performs in a manner similar to and is intended for the same use as the primary predicate device. The software application, Auto MPR, included in the Vitrea Software Package, VSTP-002A, performs in a manner similar to and is intended for the same use as the reference predicate device. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.