CT CoPilot™ is intended for automatic labeling, visualization and volumetric quantification of segmentable structures from sets of CT images of the brain. This software is intended to automate the current manual process of identifying, labeling and quantifying structures identified on CT images of the brain and to provide automated registration and reformatting of data.
Device Story
CT CoPilot™ is a software-based medical device for post-acquisition quantitative analysis of brain CT images. It takes CT image sets as input and performs automatic registration, reformatting, labeling, and volumetric quantification of segmentable structures. The device coregisters current data to the most recent prior exam for longitudinal comparison, generating voxel-by-voxel subtraction maps of pixel density changes in Hounsfield Units (HU). Output includes numerical volumes and annotated images with color-coded graphical overlays, provided in DICOM format for display on third-party PACS workstations. Used by trained personnel in Neuro CT imaging, the software includes a 'measurement index' safety feature that evaluates alignment accuracy against a neuroanatomic atlas; if the index exceeds specified limits, processing terminates and an error report is generated. The device assists clinicians by automating manual identification and quantification tasks, though it is not intended for sole patient management decisions.
Clinical Evidence
Bench testing only. Accuracy testing on 100 CT head scans demonstrated registration variability <15 degrees for yaw, roll, and pitch. Segmentation accuracy tested on 179 scans from 34 patients with ventriculostomy catheters showed high correlation (95-99%) with manual expert segmentation for Lateral Ventricle Volume, Total CSF Volume, Intra-Cranial Volume, and Midline Shift Index. Test-retest reliability was equivalent to manual expert segmentation.
Technological Characteristics
Software-only device installed on off-the-shelf PC. Operates on Linux. Uses registration, alignment, and segmentation methods based on a neuroanatomic atlas. Connectivity via DICOM standard. Complies with ISO 14971 (risk management), ISO 62304 (software development), and NEMA PS 3.1-3.20 (DICOM).
Indications for Use
Indicated for patients aged 18 or older undergoing CT imaging of the brain. Used for automatic labeling, visualization, and volumetric quantification of segmentable brain structures.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 7, 2016
ZepMed, LLC. % Mr. James Monroe CEO Monroe Medical Device Consulting, LLC 319 Shilling Drive SOMERSET NJ 08873
Re: K161322 Trade/Device Name: CT CoPilot™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: November 11, 2016 Received: November 14, 2016
Dear Mr. Monroe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K161322
Device Name CT CoPilot™
Indications for Use (Describe)
CT CoPilot™ is intended for automatic labeling, visualization and volumetric quantification of segmentable structures from sets of CT images of the brain. This software is intended to automate the current manual process of identifying, labeling and quantifying structures identified on CT images of the brain and to provide automated registration and reformatting of data.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | <span> <span style="text-decoration: overline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> | <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
|--|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete.
time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber."
FORM FDA 3881 (8/14)
Page 1 1
PSC Publishing Services (301)443-6740
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "ZEPMED" in a bold, sans-serif font. The letters "ZEP" are in black, while the letters "MED" are in a dark gray color. The letters are all capitalized and evenly spaced.
### 05 - 510k Summary
CT Co-Pilot
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared:
Dr. Douglas J. Bates, MD ZepMed, LLC. 2465 Avenida de La Playa La Jolla, CA 92037
Phone: (858) 692-0795, Facsimile: (858) 412 - 3711
Contact Person: James W. Monroe; jwmonroe@monroemdc.com, (908) 809-0081
Date Prepared: November 11, 2016
#### Name of Device and Name/Address of Sponsor
# CT CoPilot™
ZepMed, LLC. 2465 Avenida de La Playa La Jolla, CA 92037
#### Common or Usual Name
Picture Archiving and Communication System
#### Classification Name
System, Image Processing, Radiological (892.2050)
Classification Panel Radiology
Product Codes LLZ
Device Class II
Predicate Devices NeuroQuant (K061855)
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "ZEPMED" in a bold, sans-serif font. The letters are a dark gray color, with the "Z", "E", and "P" appearing slightly darker than the "M", "E", and "D". The letters are all capitalized and evenly spaced, creating a clean and modern look. The background is plain white.
#### Intended Use / Indications for Use
CT CoPilot™ is intended for automatic labeling, visualization and volumetric quantification of segmentable structures from sets of CT images of the brain. This software is intended to automate the current manual process of identifying, labeling and quantifying structures identified on CT images of the brain and to provide automated registration and reformatting of data.
## Device Description
CT CoPilot™ is intended for use in automating post-acquisition quantitative analysis of CT images of the brain for patients aged 18 or older. CT CoPilot™ performs automatic reformatting, labeling and quantification of segmentable structures from a set of CT images. Output of the software provides these values as numerical volumes and images which have been annotated with graphical color overlays, with each color representing a specific segmental structure. When CT imaging is performed more than once on a patient, the current data is coregistered to the most recent processed prior exam of the same patient, facilitating comparison between the studies using CT CoPilot™. Voxel-by-voxel subtraction maps of the pixel density change in Hounsfield Units (HU) are generated in up to 3 dimensions between the current and most recent processed prior exam of the patient.
CT CoPilot™ incorporates registration, alignment, and segmentation methods similar to a previous 510(k) cleared device, known as NeuroQuant (K061855), to automatically label and quantify the volume of segmentable structures in MRI images of the head. CT CoPilot™ output is provided in standard DICOM format as additional series of images (with appropriate descriptors) and reports that can be displayed on most third-party commercial DICOM workstations. CT CoPilot™ is intended to provide visualization and quantification data for CT scans of the brain. CT CoPilot™ includes safety procedures and error reporting similar to those adopted by NeuroQuant to identify cases that may not be processed for any reason. CT CoPilot™ is intended to be used by trained personnel in Neuro CT imaging. Patient management decisions should not be made based solely on the results of CT CoPilot™ quantitative data.
In laboratory testing, CT CoPilot™ demonstrates the following registration accuracy based on 100 randomly acquired CT head scans on both normal patients and those with abnormal
{5}------------------------------------------------
pathologies: 1) The inter-subject variability of the angle formed by the inter-hemispheric plane and the vertical line, as measured on axial views (yaw), is less than 15 degrees. 2) The intersubject variability of the positioning of the inter-hemispheric plane and the vertical line, as measured on coronal views (roll), is less than 15 degrees. 3) The inter-subject variability of the positioning of the AC/PC plane and the horizontal line, as measured on sagittal views, is less than 15 degrees. Laboratory testing of CT CoPilot™ software on 179 scans from 34 patients with ventriculostomy catheters demonstrates the following correlation coefficients between the automatic software segmentation accuracy of relevant anatomical structures when compared against the same group of medical expert manually segmented subjects: (Lateral Ventricle Volume = 98%, Total CSF Volume = 98%, Intra-Cranial Volume = 99%, Midline Shift Index = 95%). Laboratory testing of CT CoPilot™ segmentation reliability demonstrates equivalent test-retest performance as expert manually segmented subjects. The accuracy testing of CT CoPilot™ registration and segmentation performance was similar to the predicate device (NeuroQuant) without additional safety risk.
CT CoPilot™ consists of proprietary software developed by ZepMed, Inc. installed on an offthe-shelf personal computer. The output of CT CoPilot™ is intended for Picture Archive and Communications System (PACS) display systems. PACS display systems must contain sufficient functionality to display color images (either MR or Secondary Capture), and to display report text and graphics output either in DICOM Structured Report or display of Adobe PDF or JPEG files.
The accuracy of the automatic cross-sectional registration and segmentation in CT CoPilot™ is affected by the subject's deviation from the features embedded in the reference neuroanatomic Atlas. In general, the accuracy of the CT CoPilot™ system software may decrease when the subject's head includes pathologic features not present in the Atlas, lacks features present in the Atlas, or is structurally different than that defined in the preexisting neuroanatomic Atlas. The accuracy of the CT CoPilot™ single image and serial image analysis program may also be degraded by patient motion or by artifacts which are introduced into the patient scanning process, or if patient positioning severely deviates from expectations. Alignment of subject to Atlas space is an important software step in the overall process of registration and segmentation. CT CoPilot™ uses the mechanism for
{6}------------------------------------------------
determination of alignment accuracy instantiated in the predicate device (NeuroQuant) for establishing the relationship between image quality and overall alignment accuracy. Specifically, a "measurement index" is determined for each study based on the deviation of each image from the normalized anatomic index embedded in the program. The measurement index is determined by the deviation of the image volumes from normal atlas space. The automatically determined "measurement index" is used to define the limit of anatomic alignment variance which has been determined through laboratory testing of CT CoPilot™ to be accepted by the program. As a safety feature, the CT CoPilot™ software calculates and reports a measurement index number which reflects the adequacy of overall single image and serial image alignment. If the measurement index exceeds specified limits, an error report is generated to inform the user and further processing is terminated.
Factors that may degrade the technical quality and accuracy of CT CoPilot™ registration and segmentation results include:
- a) Patient's motion during scan.
- b) Artifacts affecting overall image quality.
- c) Reconstruction artifacts.
- d) Pathological and or anatomical deviations from the Atlas.
- e) Large initial alignment deviations between the patient and the Atlas.
## Non-Clinical Testing
- ISO 14971 Second Edition 2007, Medical Devices Application Of Risk Management o To Medical Devices.
- ISO 62304: The Harmonized Standard for Medical Device Software Development .
- . NEMA PS 3.1 - 3.20 (2011), Digital Imaging And Communications In Medicine (Dicom) Set.
## Comparison to Predicate Device
ZepMed's CT CoPilot™ , is substantially equivalent to the predicate devices listed below with respect to intended use/indications for use, principles of operation and technological characteristics.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "ZEPMED" in a bold, sans-serif font. The letters "ZEP" are in a darker shade of gray, while the letters "MED" are in a lighter shade of gray. The letters are all capitalized and evenly spaced.
# Substantial Equivalence Table
| Attribute | CT CoPilot™ | NeuroQuant | Equivalence |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| | | | |
| 510(k) | (Subject Device) | K061855 | |
| Product Code | LLZ | LLZ | Yes |
| Intended Use | CT CoPilot™ is<br>intended for automatic<br>labeling, visualization<br>and volumetric<br>quantification of<br>segmentable structures<br>from sets of CT images<br>of the brain. This<br>software is intended to<br>automate the current<br>manual process of<br>identifying, labeling<br>and quantifying<br>structures identified on<br>CT images of the brain<br>and to provide<br>automated registration<br>and reformatting of<br>data. | NeuroQuant™ is intended<br>for automatic labeling,<br>visualization and<br>volumetric quantification<br>of segmentable brain<br>structures from sets of MR<br>images. This software is<br>intended to automate the<br>current manual process of<br>identifying, labeling and<br>quantifying the volume of<br>segmental brain structures<br>identified on MR images. | Yes |
| Data Source | CT Scanner | MRI Scanner | Different |
| Display images | Reformatted, realigned<br>axial, coronal, and sagittal<br>images. | Reformatted, realigned<br>axial, coronal and sagittal<br>images. | Yes |
| Quantitative<br>Metrics | CSF volumes, Intracranial<br>volume, Midline shift. | CSF volumes, Intracranial<br>volume, Brain structure<br>volumes. | Similar |
| Physical<br>Characteristics | Software package<br>• Operates off-the-shelf<br>software (multiple<br>vendors) | Software package<br>• Operates off-the-shelf<br>software (multiple<br>vendors) | Yes |
| Operating<br>System | OS:Linux | OS: Linux,Mac,Windows | Yes |
| DICOM<br>compatible | Yes | Yes | Yes |
| Performance<br>measurement<br>Testing | Reproducibility and<br>Accuracy testing | Reproducibility and<br>Accuracy testing | Yes |
| Safety | Measurement data can be<br>viewed, accepted or<br>rejected by a physician | Measurement data can be<br>viewed, accepted or<br>rejected by a physician | Yes |
| In plane voxel<br>Resolution<br>(Input Data) | 0.1-1mm | 1mm | Similar |
| Slice<br>Thickness<br>(Input Data) | 0.2-1mm | 1.2mm | Similar |
| Automatic<br>Alignment | Yes | Yes | Yes |
| Registration<br>Target Data | Atlas, Prior | Atlas | Similar |
| Skull<br>Stripping | Yes | Yes | Yes |
| Automatic<br>Segmentation | Yes | Yes | Yes |
| Error<br>Detection | Yes | Yes | Yes |
| Output Image<br>Data Format | 3D Volumetric, 2D MPR<br>-1mm isotropic volume<br>-1-5mm thick MPR | 3D Volumetric<br>-1mm isotropic volume | Similar |
| Color-coded<br>Segmentation<br>Series<br>(Output Data) | Yes | Yes | Yes |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "ZEPMED" in a bold, sans-serif font. The letters "ZEP" are in a darker shade, while "MED" is in a lighter shade of gray. The letters are closely spaced, creating a solid block of text.
# Substantial Equivalence Discussion:
Both CT CoPilot™ and NeuroQuant (K061855) allow for the visualization and volumetric quantification of brain images. Similarly, both systems incorporate post-acquisition quantitative analysis of acquired images and transform and display the results in DICOM standard formatted images and reports. Both systems perform reproducibility and accuracy testing. The difference between the post processing systems is that CT CoPilot™ is analyzing CT images of the brain
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "ZEPMED" in a bold, sans-serif font. The letters "ZEP" are in a darker shade of gray, while the letters "MED" are in a lighter shade of gray. The letters are all capitalized and have a modern, geometric design. The overall impression is one of a strong, professional brand.
whereas NeuroQuant analyzes MRI images of the brain. Although CT CoPilot™ utilizes CT scans and NeuroQuant utilizes MRI scans, they both perform automated segmentation and quantification of segmental structures within the brain. Post-processing of the images with the Subject device (CT CoPilot™) and predicate device (NeuroQuant) use similar algorithms for segmentation and registration of scans to an Atlas as well as similar safety procedures to assess the quality of segmentation and registration accuracy.
CT CoPilot™ is substantially equivalent in performance, technology, and characteristics. The differences between the two devices do not raise any new questions of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.