K223147 · Myblend · OHS · May 14, 2023 · General, Plastic Surgery
Device Facts
Record ID
K223147
Device Name
myLEDmask
Applicant
Myblend
Product Code
OHS · General, Plastic Surgery
Decision Date
May 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
myLEDmask is an over-the-counter device that is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full-face wrinkles.
Device Story
myLEDmask is an over-the-counter photobiomodulation device for face and neck wrinkle reduction. It uses 288 LEDs (144 red at 630 nm; 144 near-infrared at 850 nm) to deliver light energy to the skin. The device features a protective eye-shield to prevent ocular exposure. Users select one of three care programs based on skin phototype (fair, moderately dark, or dark), which determines the treatment duration and energy dose. The device is powered by a rechargeable NiMH battery and controlled by an internal microcontroller that manages the timer and automatically shuts off after the session. It is intended for home use by the patient. The healthcare provider does not interact with the device output; the patient uses the device to potentially improve skin appearance by stimulating cell functions via low-power light radiation.
Clinical Evidence
No clinical performance data included. Usability was evaluated with a panel of representative users per FDA guidance. Bench testing only.
Indicated for the treatment of full-face wrinkles in adults using an over-the-counter device.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 14, 2023
myBlend % José Perez Lead Project Manager Ceiso 69 Rue de Paris Hall B Orsay, 91400 France
Re: K223147
Trade/Device Name: myLEDmask Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: October 5, 2022 Received: October 5, 2022
Dear José Perez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jianting Wang -S
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K223147
Device Name myLEDmask
Indications for Use (Describe)
myLEDmask is an over-the-counter device that is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full-face wrinkles.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K223147
#### PAPIER à ENTETE
#### 510(k) SUMMARY
#### Prepared on 2022-06-06 2023-04-25
## 1. Applicant / Submitter:
| 510(k) Owner | My Blend (SASU) |
|--------------|---------------------------------------------------------------------------------------|
| | 9 Rue du Commandant Pilot, 92200 Neuilly-sur-Seine, France<br>Tel : +33 1 56 60 65 35 |
## 2. Submission correspondent
| Contact Person | Cnnsultant | José PEREZ/ Laetitia ROBIN |
|----------------|----------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | CEISO 69 Rue de PARIS, 91400 ORSAY<br>Phone: +33 6 27 30 01 94 / +33 6 42 19 30 64<br>Email: jose.perez@ceiso.fr / laetitia.robin@ceiso.fr |
| | Manufacturer | Clédia HETTINGER<br>Assistant & Coordinator<br>Phone: +33 6 89 08 11 79<br>Address: 12 Avenue de la Porte des Ternes 7017 Paris<br>Mail: cledia.hettinger@clarins.com |
| | Specification Developper/<br>Contract Manufacturer | Hanan BERNARDELLI<br>CEO MOSTLEDS<br>Phone: +33 6 85 14 26 08<br>Address: 1432 Quai Marcel Dassault 92210 Saint-Cloud France<br>Email Address: hanan@mostleds.com |
## 3. Device Name and Code
| Device /trade Name | MyLEDMask |
|--------------------|------------------------------------------------------|
| Classification | 878.4810 |
| Product Code | OHS : light based over the counter wrinkle reduction |
| Panel | General and Plastic Surgery |
| Class | 2 |
#### 4. Predicate devices
| Applicant/Submitter : | LED Technologies Inc | LA LUMIERE, LLC | BioPhotas, Inc.<br>Anaheim, CA 92805<br>USA | LG Electronics, Inc<br>Gyeonggi-do, 17709,<br>Rep. of Korea |
|-----------------------|----------------------|-------------------------------------|---------------------------------------------|-------------------------------------------------------------|
| Device Name : | DPL II PANEL | PRO X OTC 5 Light<br>Therapy device | Biophotas Celluma3 | LG PRA.L DERMA LED<br>MASK |
| Product Code : | OHS | OHS | OHS | OHS |
| 510(k) | K171390 | K140471 | K171323 | K183671 |
{4}------------------------------------------------
#### 5. Device description
Image /page/4/Picture/2 description: The image shows a white LED face mask. The mask is shaped like a human face and has eye and mouth openings. The mask is attached to a white base with a control panel that has several buttons and the words "myBlend" printed on it. The mask appears to be a beauty device.
MyLEDmask is a device used to perform photobiomodulation (based on lightemitting diode (LED) light) on the face and neck.
Light radiates from the inner surface of the device onto the face and neck. This light is generated by LEDs with two different visible spectrum wavelengths: red (630 nm) and near-infrared (850 nm).
3 Care programs (LEDs activation duration) based on 3 skin phototypes (skin tone) are available.
- Phototype 1 Fair Skin -
-
- Phototype 2 Moderately Dark Skin
- -Phototype 3 Dark Skin
At the end of care duration, device will automatically switch off
It is operated by a microcontroller and powered by rechargeable batteries with an external independent Charger
To prevent irradiation of LED lights to eyes during the treatment, MyLEDmask has protective eye-shield which blocks light energy from LEDs.
### 6. Indications for Use
MyLedMask is an over-the-counter device that is intended to emit energy in the visible and infrared region of the spectrum for use in the treatment of full-face wrinkles.
### 7. Summary of the technological characteristics
As described in the Product Code "OHS - Light based OTC for wrinkle reduction", MyLEDmask principle of operation is based on generation of Red and Near infrared Light, to interact with skin, in order to reduce wrinkle, as it is long time recognized that application of low-power lights radiation in these wavelength range stimulates and enhances some cell functions.
## 8. Substantial Equivalence/ non-Clinical performance data:
| Property | MyLedMask | DPL II PANEL | PRO X OTC 5 Light<br>Therapy device | Celluma | LG PRA.L DERMA LED<br>MASK |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>Number | Pending | K171390 | K140471 | K171323 | K183671 |
| Manufacturer | MyBlend | LED Technologies Inc. | LA LUMIERE, LLC | BIOPHOTAS Inc. | LG Electronics, Inc. |
| Device Trade<br>name | MyLedMAsk | DPL II PANEL | PRO X OTC 5 Light Therapy<br>device | Celluma | LG PRA.L DERMA LED<br>MASK |
| Class | 2 | | | | |
| Product Code | OHS - Light based OTC for wrinkle reduction | | | | |
| Indications<br>for Use | MyLedMask is an over-the-<br>counter device that is intended<br>to emit energy in the visible<br>and infrared region of the<br>spectrum for use in the<br>treatment of full-face wrinkles | The dp®II Panel System is<br>an Over The Counter<br>(OTC) device intended for<br>use in treating wrinkles. | The Pro X OTC 5 is an Over-<br>The-Counter device<br>intended for use in the<br>treatment of facial wrinkles.<br>It is for people with wrinkles<br>on their face and who have | The BioPhotas Celluma3<br>is intended to emit<br>energy in the visible and<br>infrared region of the<br>spectrum for use in the<br>treatment of full-face<br>wrinkles. | The LG PRA.L DERMA LED<br>MASK is an over-the-<br>counter device that is<br>intended for the use in<br>the treatment of full-face<br>wrinkles. |
| Property | MyLedMask | DPL II PANEL | PRO X OTC 5 Light<br>Therapy device | Celluma | LG PRA.L DERMA LED<br>MASK |
| | | | Fitzpatrick skin types I, II | | |
| | | | and/or III.<br>Calculated at 325 cm² | | |
| Treatment<br>Area (cm2) | About 72 cm² estimated from<br>the LED Mask dimension | 415 cm² | (Overall total Energy 332 J<br>and Total Dose Energy 1.02<br>J/cm²) | Active treatment area of<br>15″ x 8″ (381 mm x 203<br>mm): about 77 cm² | 178.8 x 225 x 89: about<br>50 cm² estimated from<br>the LED Mask dimension |
| Materials | Mask inner face: MAKROLON<br>2805<br>Protective eye cup: Silicon<br>TN750<br>Elastic band holder: Neopren<br>and Polycotton | Not known | Not known | Not known | Mask frame: PC<br>Eye frame: PP<br>Eye shield: Silicone<br>Viewing shield: ABS<br>plastic<br>Controller frame: PP<br>Adaptor frame: PP<br>Cradle frame: PP |
| Dimension<br>(mm) | LED mask: 176 x 342 x 122<br>Storage case: 365 x 225 x 150 | Not known | Not known | Not known | LED Mask: 178.8 x 225 x<br>89<br>Controller: 40 x 30 x 110<br>Adaptor: 87x 44 x 28<br>Cradle: 180 x 229 x 167.8 |
| Net weight<br>(g) | LED mask: 650 g<br>Storage case: 623 g | Not known | Not known | Not known | LED Mask: 225 g /<br>Controller: 84 g<br>Adaptor: 75 g / Cradle:<br>380 g |
| LED Emission | Red: 630 nm (OSRAM LS T67F)<br>NIR: 850 nm (OSRAM SFH 4253) | 605 nm, 630 nm, 660 nm,<br>880 nm | red light (620-630nm)<br>infrared light (855nm) | Red: 640nm+/-25nm<br>NIR: 880nm+/-50nm | RED (637 nm)<br>IR (854 nm) |
| Number of<br>LEDs | Total 288 LEDs<br>- 144 for 630 nm and 144 for<br>850 nm<br>- 244 for the face and 44 for the<br>neck | Not known | Not known | Not known | Total 160 LEDs (80 of<br>them are<br>for 637 nm and the rest<br>80 of<br>them are for 854 nm) |
| LED Power | 18.7 mW/cm²<br>(average of 6 measurement<br>location (including one on the<br>neck) between 10 to 27<br>mW/cm2) | 70.16 mW/cm² | 16 mW/cm² (red)<br>2.9 mW/cm2 (infrared)<br>Total : 18.9 mW/cm² | 6.5 mW/cm² | 25 mW/cm² total |
| Irradiance | According to skin phototype:<br>- 1 fair skin: 6.3 J/cm²<br>- 2 moderately dark skin: 12.5 J<br>/cm²<br>- 3 dark skin: 15.6 J /cm² | 12.6 J/cm²<br>Calculated from Power<br>(70.16 mW/cm2) and<br>Treatment duration 3 min<br>(180 s) | 0.86 J/cm² (red)<br>0.16 J/cm² (infrared)<br>Total : 1.02 J/cm² | 11.7 J/cm² | 13.5 J/cm²<br>Calculated from Power<br>(25 mW/cm²) and<br>Treatment duration 9 min<br>(540 s) |
| Treatment<br>Time | According to skin phototype,<br>daily<br>- 1 fair skin: 5 min 35 (335 s)<br>- 2 moderately dark skin: 11<br>min 10 (670 s)<br>- 3 dark skin: 13 min 55 (835 s)<br>during 6 to 8 weeks | 3 minutes daily<br>5 days per week for 8<br>weeks | 900 seconds per treatment<br>60 times | 3 treatments per week<br>(1800 seconds)- 4 weeks | 9 minutes daily<br>5 days per week for 8<br>weeks |
| Total Energy<br>Dose (J/cm2) | For 8 weeks, 3.5 Treatments<br>(48 h spaced) per week:<br>According to skin phototype:<br>- 1 fair skin: 176 J<br>- 2 moderately dark skin: 350 J<br>- 3 dark skin: 437 J | For 8 weeks, 5<br>Treatments per week :<br>504 J/cm² | For 60 treatments of 900 s :<br>61.2 J/cm² | For 4 weeks, 3<br>Treatments per week :<br>140 J | For 8 weeks, 5<br>Treatments per week :<br>540 J |
| Treatment<br>Control | Device uses a timer and<br>software to control<br>treatment duration.<br>Software Safety Class: A (Low<br>level of concern) | Not known | Not known | Device uses a timer and<br>software to control<br>treatment duration.<br>Software Safety Class: A<br>(Low level of concern) | Not known |
| Battery | NiMH 1.7 Ah | Not known | Not known | Not known | Lithium-ion |
| Power Supply | Voltage: 100 to 240 volts, AC<br>Frequency: 50-60Hz<br>Intensity: 0.35 A | 120-240 V AC | Not known | 110-120V | Input: AC 100V ~ 240V<br>Frequency: 50Hz/60Hz<br>Output: 5V 2A |
| Property | MyLedMask | DPL II PANEL | PRO X OTC 5 Light<br>Therapy device | Celluma | LG PRA.L DERMA LED<br>MASK |
| Electrical<br>Safety | 60601-2-57:2011<br>60601-1:2012<br>60601-1-2:2014 | 60601-1:2006<br>60601-1-2:2007 | 60601-1:2006<br>60601-1-2:2007<br>62471 | 60601-1:2012<br>60601-1-2:2007 | Not known |
{5}------------------------------------------------
{6}------------------------------------------------
### 9. Summary of the Clinical performance data
No clinical Performance data are included in this submission. However, based on FDA guidance, usability has been evaluated with panel of representative users.
#### Overall Conclusions 10.
Comparison between device described in this 510(k) and predicate devices shows a substantial equivalence based on:
- Same Indications for Use, -
- -Same technological and technical characteristics (Principle of operation)
- -Results of non-clinical tests
Slight differences do not raise any questions regarding safety and effectiveness.
Therefore, it can be concluded that device described in this 510(k) is « as Safe and effective »as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.