K183671 · Lg Electronics.Inc · OHS · Jan 22, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183671
Device Name
LG PRA.L DERMA LED MASK
Applicant
Lg Electronics.Inc
Product Code
OHS · General, Plastic Surgery
Decision Date
Jan 22, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LG PRA.L DERMA LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles.
Device Story
Device is an over-the-counter LED mask for full-face wrinkle treatment; emits red (637 nm) and infrared (854 nm) light. System comprises mask with internal LEDs, controller for power/operation, and power supply unit/charging cradle. User places mask over face; controller initiates treatment; device automatically shuts off after 9-minute session. Includes protective eye-shield to prevent ocular/ophthalmic irradiation. Used by consumers in home setting. Benefits include non-invasive wrinkle reduction.
Clinical Evidence
Bench testing only. Performance data included measurement of wavelength, average output power, and total irradiance. Safety and performance evaluated against IEC 60601-1 (basic safety), IEC 60601-1-2 (electromagnetic compatibility), IEC 62471-1 (optical hazard), ISO 14971 (risk management), and IEC 62366-1 (usability). Biocompatibility of patient-contacting materials (polypropylene, polycarbonate, silicone) supported by prior use in medical devices.
Technological Characteristics
Materials: Polypropylene, polycarbonate, silicone. Energy: LED-based light therapy (637 nm red, 854 nm IR). Power: 25 mW/cm² total irradiance; AC 100-240V input, 5V 2A output; Lithium-ion battery. Form factor: Wearable mask with controller. Connectivity: None. Sterilization: Not specified. Software: Simple timer-based control logic.
Indications for Use
Indicated for the treatment of full face wrinkles in the general population via over-the-counter use.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font below.
LG Electronics, Inc. % Do-Hyun Kim CEO BT Solutions. Inc. 904. Eonju-ro 86-gil 5. Seoul, 06210 Kr
January 22, 2019
Re: K183671
Trade/Device Name: LG PRA.L Derma LED Mask Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OHS Dated: December 21, 2018 Received: December 27, 2018
Dear Do-Hyun Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
| Neil R | Digitally signed by<br>Neil R Ogden -S |
|----------|----------------------------------------|
| Ogden -S | Date: 2019.01.22<br>15:41:21 -05'00' |
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### LG PRA.L DERMA LED MASK
#### Indications for Use Statement
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2020<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Indications for Use | |
| 510(k) Number (if known) | N/A K183671 |
Device Name LG PRAL DERMA LED MASK
-
#### Indications for Use (Describe)
The LG PRA.L DERMA LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
# 5. 510(k) Summary
Date: December 21, 2018
| 1. Applicant / Submitter: | LG Electronics, Inc.<br>222 LG-ro, Jinwi-myeon, Pyeongtaek-si,<br>Gyeonggi-do, 17709, Republic of Korea |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submission Correspondent: | Do-Hyun Kim<br>BT Solutions, Inc.<br>904, Eonju-ro 86-gil 5,<br>Gangnam-gu, Seoul, 06210, Republic of Korea<br>Tel: +82-2-538-9140 Fax: +82-2-539-9140<br>Email: smanager@btsolutions.co.kr |
## 3. Device Name and Code:
| Device Name: | LG PRA.L DERMA LED MASK |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | LG PRA.L DERMA LED MASK |
| Classification Name: | Light Based Over The Counter Wrinkle Reduction |
| Product Code: | OHS |
| Regulation Number: | 878.4810 |
| Classification: | Class II |
| Review Panel: | General & Plastic Surgery (ODE)<br>Division of Surgical Devices (DSD)<br>General Surgery Devices Branch One - Light Based/Laser (GSDB1) |
# 4. Predicate Device:
- Primary Predicate Device: K170984 LG BEAUTY LED MASK by LG Electronics, Inc. ●
# 5. Device Description:
The LG PRA.L DERMA LED MASK is a device which allows emission of LED light in the RED (637 nm) and IR (854 nm) spectrum on to the face. The LG PRA.L DERMA LED MASK includes the mask that contains LEDs on the inner surface, the controller which turns on and off the device, and the power supply unit that delivers electrical power to the controller for operation or for charging the battery contained in the controller. Users place the lightweight mask over the face and use the controller to operate the LG PRA.L DERMA LED MASK. The device will automatically turn off after each treatment. Emitted light from LEDs are not intended for ocular or ophthalmic treatment. To prevent irradiation of LED lights to eyes during the treatment, LG PRA.L DERMA LED MASK has protective eye-shield which blocks light energy from LEDs.
# 6. Indication for use:
The LG PRA.L DERMA LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles.
{4}------------------------------------------------
## LG PRA.L DERMA LED MASK
#### 510(k) Summary
## 7. Performance Data:
Non-clinical tests: Measurement of wavelength, average output power, and total irradiance (power density, in units of mW/cm2) of treatment LEDs were performed. Other performance, such as optical hazard, electromagnetic compliance, etc, were tested using following consensus standards:
- Basic safety and essential performance of the LG PRA.L DERMA LED MASK is tested and evaluated according to the FDA-recognized consensus standard, IEC 60601-1.
- . Effect to the device by electromagnetic disturbances were tested and evaluated using IEC 60601-1-2.
- . Optical hazard was evaluated using IEC 62471-1.
- Risk management was recorded by referring to ISO 14971. ●
- Usability was documented using IEC 62366-1.
The portion of the device that touches patient body is made of polypropylene, polycarbonate plastic, and silicone, which have been used for other medical devices without any biocompatibility risk.
## 8. Substantial Equivalence:
The LG PRA.L DERMA LED MASK is an over the counter device that is intended for the use in the treatment of full face wrinkles. The subject device has the same device characteristics as the predicate (unmodified) device. They have the same intended use, material, and use concept and employ the same manufacturing processes. The differences are in shape, dimensions and LED total number. The performance test shown that there is no technical characteristic difference in the modified device. The external design changes do not raise any concerns regarding the device performance and safety. Therefore, LG PRA.L DERMA LED MASK is substantially equivalent to the legally marketed predicate/unmodified device.
| | Predicate/Unmodified Device | Proposed/Modified Device | SE |
|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Manufacturer | LG Electronics, Inc. | LG Electronics, Inc. | - |
| Device Name | LG BEAUTY LED MASK | LG PRA.L DERMA LED MASK | - |
| 510(K) # | K170984 | - | - |
| Class | 2 | 2 | YES |
| Product Code | OHS | OHS | YES |
| Intended Use | The LG BEAUTY LED MASK is<br>an over the counter device that is<br>intended for the use in the<br>treatment of full face wrinkles. | The LG PRA.L DERMA LED<br>MASK is an over the counter<br>device that is intended for the use<br>in the treatment of full face<br>wrinkles. | YES |
| Materials | Mask frame: PC<br>Eye frame: PP<br>Eye shield: Silicone<br>Viewing shield: ABS plastic<br>Controller frame: PP<br>Adaptor frame: PP<br>Cradle frame: PP | Mask frame: PC<br>Eye frame: PP<br>Eye shield: Silicone<br>Viewing shield: ABS plastic<br>Controller frame: PP<br>Adaptor frame: PP<br>Cradle frame: PP | YES |
{5}------------------------------------------------
## LG PRA.L DERMA LED MASK
| Dimension<br>(mm) | LED Mask: 180 x 214.5 x 92.4<br>Controller: 41 x 31 x 103<br>Adaptor: 87 x 44 x 28<br>Cradle: 190 x 222 x 156 | LED Mask: 178.8 x 225 x 89<br>Controller: 40 x 30 x 110<br>Adaptor: 87x 44 x 28<br>Cradle: 180 x 229 x 167.8 | YES |
|---------------------|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----|
| Net weight<br>(g) | LED Mask: 207 g<br>Controller: 82 g<br>Adaptor: 75 g<br>Cradle: 375 g | LED Mask: 225 g<br>Controller: 84 g<br>Adaptor: 75 g<br>Cradle: 380 g | YES |
| LED<br>Emission | RED (637 nm) and IR (854 nm) | RED (637 nm) and IR (854 nm) | YES |
| Number of<br>LEDs | Total 120 LEDs (60 of them are<br>for 637 nm and the rest 60 of<br>them are for 854 nm) | Total 160 LEDs (80 of them are<br>for 637 nm and the rest 80 of<br>them are for 854 nm) | NO |
| LED power | 25 mW/cm² total | 25 mW/cm² total | YES |
| Electronic<br>power | Input: AC 100V ~ 240V<br>Frequency: 50Hz/60Hz<br>Output: 5V 2A | Input: AC 100V ~ 240V<br>Frequency: 50Hz/60Hz<br>Output: 5V 2A | YES |
| Battery | Lithium-ion | Lithium-ion | YES |
| Treatment<br>time | 9 minutes daily<br>5 days per week for 8 weeks | 9 minutes daily<br>5 days per week for 8 weeks | YES |
#### 510(k) Summary
# 9. Conclusion
On the basis of the information provided in this Summary, LG Electronics believes LG PRA.L DERMA LED MASK is substantially equivalent to the legally marketed predicate/unmodified device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.