K223086 · Ceribell, Inc. · GXY · Mar 22, 2023 · Neurology
Device Facts
Record ID
K223086
Device Name
Ceribell Instant EEG Headcap
Applicant
Ceribell, Inc.
Product Code
GXY · Neurology
Decision Date
Mar 22, 2023
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 882.1320
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Ceribell Instant EEG Headcap is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired.
Device Story
The Ceribell Instant EEG Headcap is a single-use, non-sterile, disposable electrode array designed for rapid EEG signal acquisition in clinical settings. It consists of an elastic spandex-blend fabric cap, an elastic chin strap, and 9 to 19 silver/silver chloride (Ag/AgCl) electrodes mounted in soft thermoplastic elastomer (TPE) holders. The device is placed on the patient's scalp; electrodes are positioned according to the International 10-20 or 10-10 systems. An integrated cable connects the headcap to an external EEG recording or monitoring device via a standard DIN 42-802 touch-proof safety socket. The device facilitates rapid setup for clinicians, enabling EEG signal collection for diagnostic or monitoring purposes. It does not perform signal processing or analysis itself; it serves as a passive interface between the patient's scalp and the recording equipment.
Clinical Evidence
Bench testing only. No clinical data provided. Performance was validated against FDA guidance 'Cutaneous Electrodes for Recording Purposes - Performance Criteria for Safety and Performance Based Pathway'. Testing included electrical performance (AC impedance, DC offset voltage, offset instability, internal noise, bias current tolerance), shelf-life testing, and biocompatibility (cytotoxicity, sensitization, irritation/intracutaneous reactivity per ISO 10993).
Technological Characteristics
Disposable EEG electrode array. Materials: Spandex/nylon fabric cap, TPE electrode mounts, Ag/AgCl-coated photopolymer electrodes. Connectivity: Wired via DIN 42-802 touch-proof safety socket (1.5mm). Standards: ANSI/AAMI EC12, IEC 60601-1, ISO 10993-1/5/10. Dimensions: Head size range 26cm-66cm. Cable length: 0.1m-3.0m. Passive device; no software or energy source.
Indications for Use
Indicated for use in routine clinical settings for adults and children requiring rapid placement of EEG electrodes for signal acquisition.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 21, 2023
Ceribell, Inc. Raymond Woo, PhD Chief Technical Officer 360 N Pastoria Ave Sunnyville, California 94085
Re: K223086
Trade/Device Name: Ceribell Instant EEG Headcap Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: January 24, 2023 Received: January 25, 2023
Dear Dr. Woo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
ivalence determination does not mean that FDA
Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Heather L. Dean -S
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223086
Device Name Ceribell Instant EEG Headband
Indications for Use (Describe)
The Ceribell Instant EEG Headcap is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@.fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary is being submitted in accordance with the requirements of 21 CFR 807.92.
#### Applicant Information:
Ceribell, Inc. 360 North Pastoria Avenue Sunnyvale, California 94085
#### Contact Person:
Raymond Woo, PhD CTO Telephone: (650) 556-4349 E-mail: ray@ceribell.com
#### Device Information:
| Trade Name: | Ceribell Instant EEG Headcap |
|----------------------|--------------------------------------|
| Common Name: | Cutaneous electrode |
| Classification Name: | Cutaneous electrode (21CFR 882.1320) |
| Device Class: | II |
| Product Code: | GXY |
#### Predicate Device:
K200162, Wuhan Greentek Disposable EEG Electrodes (MODEL: DL, E-CAP, FLEX-CAP)
#### Date Prepared:
January 24, 2023
#### Device Description:
The Ceribell Instant EEG Headcap is a single-use, non-sterile, disposable EEG electrode device that includes a minimum of 9 EEG electrodes that are placed on the subject's scalp. The Headcap is intended to collect and provide EEG signals to an EEG recording or monitoring device.
The Ceribell Instant EEG Headcap is comprised of the following components:
- An elastic fabric headcap
- An elastic chin strap
- A minimum of 9 silver/silver chloride (Ag/AgCl) electrodes
- A cable attached to the headcap to allow connection to an EEG acquisition/recording device ●
#### Indications for Use:
{4}------------------------------------------------
The Ceribell Instant EEG Headcap is intended for use in routine clinical settings where rapid placement of a number of EEG electrodes is desired.
### Comparison of Intended Use and Technological Characteristics with the Predicate Device:
Compared to the predicate device, the subject device has the same intended use, similar product design and the same product effectiveness as the predicate device as summarized in the following table.
| Attribute | Subject Device<br>Ceribell Instant EEG<br>Headcap (Ceribell, Inc.) | Predicate Device<br>Disposable EEG<br>Electrodes, K200162<br>(Wuhan Greentek Pty<br>Ltd.) | Comparison |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Regulation | Class II per 21 CFR<br>882.1320, E Cutaneous<br>electrode | Class II per 21 CFR<br>882.1320, E<br>Cutaneous electrode | Same |
| Product Code | GXY, Electrode, cutaneous | GXY, Electrode,<br>cutaneous | Same |
| Indications for<br>Use | The Ceribell Instant EEG<br>Headcap is intended for<br>use in routine clinical<br>settings where rapid<br>placement of a number of<br>EEG electrodes is desired. | Disposable EEG<br>Electrodes (MODEL:<br>DL, E-CAP, FLEX-CAP)<br>is intended for use in<br>routine clinical<br>settings where rapid<br>placement of a<br>number of EEG<br>electrodes is desired. | Same |
| Intended Patient<br>population | Adults and children | Adults and children | Same |
| Environment of<br>Use | Electrophysiological | Electrophysiological | Same |
| Where Used | On the head | On the head | Same |
| Device<br>Description | Disposable EEG electrode<br>array with three sizes<br>consisting of:<br>• Between 9 and 19<br>silver/silver-chloride<br>(Ag/AgCl) electrodes<br>• Cable to connect to an<br>EEG recording device<br>• A spandex fabric<br>headcap to secure the<br>electrodes to the<br>patient | Disposable EEG<br>electrode array with<br>three models (DL, E-<br>CAP, FLEX-CAP)<br>consisting of:<br>• Between 2 to 128<br>silver/silver-<br>chloride (Ag/AgCl)<br>electrodes<br>• Cable to connect to<br>an EEG recording<br>device<br>• A spandex fabric<br>headcap to secure<br>the electrodes to<br>the patient | Same |
| Attribute | Subject Device<br>Ceribell Instant EEG<br>Headcap (Ceribell, Inc.) | Predicate Device<br>Disposable EEG<br>Electrodes, K200162<br>(Wuhan Greentek Pty<br>Ltd.) | Comparison |
| Number of<br>Electrodes | 9 -19 | 2 - 128 electrodes | Yes, the number of electrodes of the<br>subject device falls within the<br>possible customizable number of<br>electrodes of the predicate device<br>(between 2 to 128 electrodes). |
| Electrode<br>Locations | The placement of the<br>electrodes is according to<br>the International 10-20<br>system of electrode<br>placement or the<br>American<br>Electroencephalographic<br>Society positioning system<br>(10-10). The number of the<br>electrodes in use is<br>according to the needs of<br>clinical practice. | The placement of the<br>electrodes is<br>according to the<br>International 10-20<br>system of electrode<br>placement or the<br>American<br>Electroencephalograp<br>hic Society positioning<br>system (10-10). The<br>number of the<br>electrodes in use is<br>according to the<br>needs of clinical<br>practice. | Same |
| Available Sizes | Various sizes (overall head<br>size range 26cm - 66cm) | Various sizes (babies<br>to large: overall head<br>size range 26cm -<br>66cm) | Same |
| Cap Material | Spandex blend: Black<br>Nylon Knitted Fabric, 82%<br>Nylon and 18% Spandex | Spandex | The biocompatibility testing and<br>electrical performance testing<br>completed are the same and follow<br>the FDA guidance document<br>"Cutaneous Electrodes for Recording<br>Purposes- Performance Criteria<br>for Safety and Performance Based<br>Pathway," August 2020. |
| Electrode<br>Mounts | Soft thermoplastic<br>elastomer (TPE) | Silicone | The electrical performance and<br>biocompatibility testing completed<br>are the same and follow the FDA<br>guidance document "Cutaneous<br>Electrodes for Recording Purposes-<br>Performance Criteria<br>for Safety and Performance Based<br>Pathway," August 2020. |
| Attribute | Subject Device<br>Ceribell Instant EEG<br>Headcap (Ceribell, Inc.) | Predicate Device<br>Disposable EEG<br>Electrodes, K200162<br>(Wuhan Greentek Pty<br>Ltd.) | Comparison |
| Electrode<br>Material | Silver/silver-chloride-<br>coated photopolymer base | Silver/silver chloride<br>ink printed on PET<br>(polyethylene<br>terephthalate) or<br>silver/silver chloride-<br>plated ABS base | The electrical performance and<br>biocompatibility testing completed<br>are the same and follow the FDA<br>guidance document "Cutaneous<br>Electrodes for Recording Purposes-<br>Performance Criteria<br>for Safety and Performance Based<br>Pathway," August 2020. |
| Type of<br>Connector | Touch-proof safety socket<br>DIN 42-802 (ø1.5mm) | Touch-proof safety<br>socket DIN42-802<br>(ø1.5mm) | The electrical performance testing<br>completed met the same<br>requirements per FDA guidance<br>document "Cutaneous Electrodes for<br>Recording Purposes- Performance<br>Criteria<br>for Safety and Performance Based<br>Pathway," August 2020. |
| Biocompatibility<br>Requirements | ISO 10993-1<br>ISO 10993-5 (Cytotoxicity)<br>ISO 10993-10<br>(Sensitization, Irritation or<br>Intracutaneous Reactivity) | ISO 10993-1<br>ISO 10993-5<br>(Cytotoxicity)<br>ISO 10993-10<br>(Sensitization,<br>Irritation or<br>Intracutaneous<br>Reactivity) | The biocompatibility testing met the<br>same requirements per FDA<br>guidance document "Use of<br>International Standard ISO 10993-1,"<br>issued Sep 2020, for surface device<br>with intact skin contact. |
| Cable | 0.1m-3.0m integrated<br>single cable | 0.1m-3.0m standard<br>ribbon cable and lead<br>wires | Both designed in conformance with<br>AAMI/ ANSI ES60601-1:2005(R)2012<br>and A1:2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012 (consolidated text)<br>Medical electrical equipment- Part 1:<br>General requirements for basic<br>safety and essential performance<br>(IEC 60601-1:2005, mod). The<br>electrical connection compliance<br>met the same requirements per FDA<br>guidance document "Cutaneous<br>Electrodes for Recording Purposes-<br>Performance Criteria<br>for Safety and Performance Based<br>Pathway," August 2020. |
| Attribute | Subject Device<br>Ceribell Instant EEG<br>Headcap (Ceribell, Inc.) | Predicate Device<br>Disposable EEG<br>Electrodes, K200162<br>(Wuhan Greentek Pty<br>Ltd.) | Comparison |
| Electrical<br>Performance<br>Requirements | ANSI/AAMI EC12<br>• Average AC Impedance:<br>≤ 2 kΩ (individual pairs ≤<br>3kΩ).<br>• DC Offset Voltage: ≤ 100<br>mV<br>• Combined Offset<br>Instability and Internal<br>Noise: ≤ 150 μV<br>• Bias Current Tolerance: ≤<br>100 mV | ANSI EC12<br>• AC Impedance: < 2<br>kΩ (at 10 Hz)<br>• DC Offset Voltage: < 100 mV<br>• Combined Offset<br>Instability and<br>Internal Noise: < 150 μV<br>• Bias Current<br>Tolerance: < 100 mV | The electrical performance testing<br>met the same requirements per FDA<br>guidance document "Cutaneous<br>Electrodes for Recording Purposes-<br>Performance Criteria<br>for Safety and Performance Based<br>Pathway," August 2020. |
| Electrical<br>Connection<br>Compliance | • Conductive Connection<br>Compliance (Patient<br>Leads or Patient Cables)<br>per ES 60601-1<br>consensus standard<br>• IEC 60601-1 clause<br>8.5.2.3<br>• 21 CFR 898.12 | AAMI/ANSI ES60601-<br>1 clause 8.5.2.3 | The electrical connection compliance<br>met the same requirements per FDA<br>guidance document "Cutaneous<br>Electrodes for Recording Purposes-<br>Performance Criteria<br>for Safety and Performance Based<br>Pathway," August 2020. |
{5}------------------------------------------------
==============================================================================================================================================================================
{6}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{7}------------------------------------------------
The results of completed testing demonstrate that any differences in technology do not raise different questions of safety and effectiveness for the subject device compared to the predicate device.
## Performance Data:
In accordance with FDA guidance document, "Cutaneous Electrodes for Recording Purposes- Performance Criteria for Safety and Performance Based Pathway", issued on August, 14, 202, the following performance data were provided to demonstrate safety and efficacy in substantial equivalence determination:
- . Electrical Performance
- o AC Impedance per FDA-recognized consensus standard, ANSI/AAMI EC12 Disposable ECG Electrodes
- o Offset Voltage per FDA-recognized consensus standard, ANSI/AAMI EC12 Disposable ECG Electrodes
- Combined offset instability and internal noise per FDA-recognized consensus standard, ANSI/AAMI EC12 Disposable ECG Electrodes
- o Bias Current Voltage (DC Voltage Offset) per FDA-recognized consensus standard, ANSI/AAMI EC12 Disposable ECG Electrodes
- Shelf life testing per FDA-recognized consensus standards, ANSI/AAMI EC12 Disposable ECG Electrodes and IEC 60601-2-2 Medical electrical equipment- Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
{8}------------------------------------------------
- . Biocompatibility testing per ISO 10993-1, ISO 10993-5, and ISO 10993-10
#### Summary:
The Ceribell Instant EEG Headcap has the same intended use as the predicate device. In addition, it has similar technological characteristics; performance data demonstrates that any differences in technological characteristics do not raise different questions of safety or effectiveness. Therefore, the Ceribell Instant EEG Headcap is substantially equivalent to the cleared predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.