K223049 · Icentia, Inc. · DSH · Mar 22, 2023 · Cardiovascular
Device Facts
Record ID
K223049
Device Name
CardioSTAT® ECG Test Solution
Applicant
Icentia, Inc.
Product Code
DSH · Cardiovascular
Decision Date
Mar 22, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K223049 · Mar 22, 2023
CardioSTAT® ECG Test Solution
Icentia, Inc.
Peer-reviewed clinical literature; Post-market surveillance data (Canada and UK); Retrospective clinical cohort data
RWE is used to support the device's arrhythmia detection performance, the reliability of home-based self-application (VSL position), and patient wear time equivalence compared to predicate devices.
Arrhythmia detection; Home-based monitoring; Post-market surveillance; Wear time; Literature review
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Validation of a novel single lead ambulatory ECG monitor - Cardiostat - Compared to a standard ECG Holter monitoring; Comparative study; Follow-up/Duration: 24 hours
212 patients; Sample Size: 212
Standard 24h Holter ECG monitoring
Arrhythmia detection accuracy (AF detection)
A pandemic response to home delivery for ambulatory ECG monitoring: Development and validation; Retrospective cohort comparison; Follow-up/Duration: 2 weeks; Study Period: Pre-pandemic vs. pandemic response
Patients receiving standard in-clinic application vs. patients receiving mail-out home-installation; Sample Size: 94 (47 per group)
In-clinic application model
Data quality and patient compliance
Icentia Post Market data; Post-market surveillance analysis; Follow-up/Duration: 14 days
Patients in Canada and UK; Sample Size: 33,486 ECG tests
Not applicable for this study
Patient wear time
Indications for Use
The CardioSTAT® ECG Test Solution is intended to capture, analyze, and report symptomatic and/or continuous ECG information for long-term ambulatory monitoring (up to 14 days). It is indicated for use on adult patients who may be asymptomatic or who may suffer from transient symptoms, such as palpitations, shortness of breath, dizziness, lightheadedness, presyncope, syncope, fatigue, or anxiety. The reported ECG metrics include single-lead analysis on a beat by beat basis, heart rate measurement and rhythm annotation. The report does not contain diagnostic interpretation; the report is provided for review to the intended user to render a diagnosis based on clinical judgment and experience.
Device Story
CardioSTAT is a prescription-only, non-invasive, single-patient-use ambulatory ECG monitoring system. It consists of a torso-worn recorder and the ECG Analysis and Reporting Tool (EART). The recorder captures continuous single-channel ECG data for up to 14 days; it snaps onto two off-the-shelf electrodes secured with adhesive tape. After the monitoring period, the patient mails the device to a data analysis center. Certified Cardiographic Technicians (CCT) use EART to analyze the data and generate a clinical report for the physician. The system provides automated beat and noise detection; all other rhythm annotations and classifications are performed manually by the CCT. The device is not for real-time monitoring, ST-segment shift detection, or automated diagnostic interpretation. Physicians use the final report to render diagnoses based on clinical judgment. The system supports both in-clinic and home-application models.
Clinical Evidence
Clinical evidence includes literature-based studies. A study of 212 patients comparing CardioSTAT to standard 24-hour Holter monitoring showed 99% agreement for AF detection. A second study (n=94) validated the home-application model and vertical sternal lead (VSL) placement, demonstrating data quality and patient compliance equivalent to in-clinic models. Post-market data from 33,486 tests showed a mean wear time of 13.1 days, exceeding the 10.4-day mean reported for the Zio patch in a systematic literature review.
Indicated for adult patients, asymptomatic or symptomatic (palpitations, shortness of breath, dizziness, lightheadedness, presyncope, syncope, fatigue, anxiety), requiring long-term ambulatory ECG monitoring up to 14 days.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
Zio® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service (K143513)
Reference Devices
Carnation™ Ambulatory Monitor Model 3000 (K143067)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a solid blue square and the agency name in a lighter blue. The text reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
March 22, 2023
Icentia Inc. % Christina Henza Regulatory Affairs Ultra Lifescience Solutions Inc 2811 Milton Ave #409 Janesville, Wisconsin 53545
Re: K223049
Trade/Device Name: CardioSTAT® ECG Test Solution Regulation Number: 21 CFR 870.2800 Regulation Name: Medical magnetic tape recorder Regulatory Class: Class II Product Code: DSH, DQK Dated: February 21, 2023 Received: February 21, 2023
Dear Christina Henza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shruti N. Mistry -S
for
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223049
Device Name CardioSTAT ECG Test Solution
### Indications for Use (Describe)
The CardioSTAT ECG Test Solution is intended to capture, and report symptomatic and/or continuous ECG information for long-term ambulatory monitoring (up to 14 days). It is indicated for use on adult patients who may be asymptomatic or who may suffer from transient symptoms, such as palpitations, shortness, lightheadedness, presyncope, syncope, fatigue, or anxiety. The reported ECG metrics include single-lead analysis on a beat by beat basis, heart rate measurement and thythm annotation. The report does not contain diagnostic interpretation; the report is provided for review to the intended user to render a diagnosis based on clinical judgment and experience.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "icentia" in a bold, sans-serif font. The letters are black, except for a small blue square above the "i". To the right of the word, there is a small "TM" symbol, indicating a trademark.
# 510(k) Summary
(21 CFR 807.92)
Date Prepared: March 20, 2023
### 5.1. GENERAL INFORMATION
### Submitter:
Icentia Inc. 2750 Einstein Street Quebec. QC Canada G1P 4R1 Phone: +1 800-431-9148 Contact Person: Danièle Grondines, Regulatory Affairs Manager Phone: +1 800-431-9148 ext.187 Email: daniele.grondines@icentia.com
#### DEVICE INFORMATION 5.2.
Trade Name: CardioSTAT® ECG Test Solution Common Name: Single-Use Ambulatory Electrocardiogram (ECG) Recorder Regulatory Class: Class II Classification Name: Medical Magnetic Tape Recorder [21CFR§870.2800], Programmable Diagnostic Computer [21CFR§870.1425] Product Code: DSH, DQK Special Controls: At this time, no special controls have been issued for devices with product codes DSH, DQK.
### 5.3. PREDICATE DEVICE(S)
Zio® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service, [K143513] Reference device: Carnation™ Ambulatory Monitor Model 3000, [K143067] The predicates have not been subject to a design-related recall.
#### 5.4. INDICATIONS FOR USE
# System Indication for USE
The CardioSTAT® ECG Test Solution is intended to capture, analyze, and report symptomatic and/or continuous ECG information for long-term ambulatory monitoring (up to 14 days). It is indicated for use on adult patients who may be asymptomatic or who may suffer from transient symptoms, such as palpitations, shortness of breath. dizziness, lightheadedness, presyncope, syncope, fatigue, or anxiety. The reported ECG metrics include single-lead analysis on a beat by beat basis, heart rate measurement and rhythm annotation. The report does not contain diagnostic interpretation; the report
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Icentia. The logo is black with a blue square above the "i" and a trademark symbol to the right of the "a". The font is sans-serif and bold.
is provided for review to the intended user to render a diagnosis based on clinical judgment and experience.
#### 5.5. DEVICE DESCRIPTION
The CardioSTAT® ECG Test Solution is a complete by prescription-only system for continuous ambulatory collection of ECG single-channel data for up to 14 days, followed by analysis and reporting of collected data. The CardioSTAT® ECG Test Solution is composed of two major components: (1) CardioSTAT® ECG Recorder and (2) ECG Analysis and Reporting Tool (EART) .
The CardioSTAT® ECG Recorder (1) is a nonsterile, non-invasive, and single-patientuse recorder worn on the patient's torso that provides a continuous, single-channel, ECG recording for up to 14 days. The Recorder snaps onto two off-the-shelf FDAcleared long-term monitoring electrodes on which OFIX® adhesive tape is applied to ensure long-term adherence to the patient. After the prescribed recording duration, the device is taken off by the patient and mailed to the data analysis center for analysis where data is analyzed by Certified Cardiographic Technicians (CCT) using (2) ECG Analysis and Reporting Tool (EART). The EART is intended for use only by trained technicians. A clinical report is then sent to the physician. The report does not contain diagnostic interpretation; the reported analysis is provided for review by the physician to render a diagnosis based on clinical judgment and experience.
A home use kit (CardioSTAT® Home Application) is also offered which includes all required devices to allow for remote prescription and home application of the CardioSTAT® recorder. The included devices (surgical razor, alcohol prep pad, electrodes, OFIX® tape) outside of this submission are currently available on the market.
The CardioSTAT® device is not designed to record ECG signals suitable to identify ST segment shifts and the EART software is not designed to detect or measure ST segment shifts.
The ECG signal collected by the CardioSTAT® ECG Test Solution is not intended for automated analysis other than beat and noise detection.
The ECG signal collected by the CardioSTAT® ECG Test Solution is not intended for third party ECG analysis systems.
The CardioSTAT® recorder is not a real-time device and all clinical data is reviewed after the device has been returned to Icentia.
The CardioSTAT® ECG Test Solution software provides automated detection of beats and noise only. All other classifications, rhythms and anomalies must be annotated by hand by the user in the EART software. Detections of beats and noise must be verified and confirmed by the user, or manually corrected with their respective annotation tools.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE 5.6. DEVICE
{5}------------------------------------------------
The proposed (subject device) CardioSTAT® Electrocardiogram (ECG) Test Solution is substantially equivalent to the primary predicate device 'Zio® SR ECG Monitoring Service (K143513 cleared on 06/19/2015)' with the reference device Carnation™ Ambulatory Monitor Model 3000 (CAM) (K143067 cleared on 12/22/2014) to support the difference in placement and the allowance for the option of user upload or manufacturer service, based on availability and user preference.
The predicate and reference devices are class II legally marketed devices as shown by the associated 510(k) numbers and clearance dates. All of the devices use the same main product code, DSH; Medical magnetic tape recorder. The subject device and the predicate use a secondary product code of DQK (Programmable diagnostic computer).
The regulation intended use covers the proposed and predicate devices. Comparison of the predicate device indications shows 6 main elements: (1) functionality, (2) duration, (3) demographics, (4) recording type, (5) reported metrics, and (6) report type. For the reference device, some of this information was not included in the indication statement so the relevant information was extracted from the IFU. brochure, and sample report to make the comparison.
All of the devices have the same primary technological characteristics. The CardioSTAT® ECG Test Solution is composed of two major components: (A) CardioSTAT® ECG Recorder and (B) ECG Analysis and Reporting Tool (EART) . Both the predicate (Zio Recorder & Zio ECG Utilization Service (ZEUS) and reference device (CAM & BDxCONNECT) have similar main components.
The main technological differences include:
- Separation of device and electrodes raise a question concerning electrode i. performance. Electrode performance is standardized (ANSI/AAMI EC12:2000 (R2015) Disposable ECG Electrodes) and all of these electrodes are compliant to the standard. The primary predicate (Zio) has integrated the two accessories of the subject device (OFIX® & 3M electrodes) and the reference device has taken integration of the devices further incorporating the battery into the electrode to create a "battrode", however the function and performance of the components remains equivalent based on compliance with standard requirements. This question of electrode performance applies to the new device and the predicate / reference devices equally and is thus not considered a different or new question of safety and efficacy.
- Electrode application in vertical and horizontal positions raises a question of ii. signal acquisition quality. The reference device is presented to support technological differences related to vertical placement. Clinical data in the form of literature and International Post Market Surveillance data is included in support of the horizontal application. This question of signal acquisition quality applies to the new device and the predicate / reference devices equally and is thus not considered a different or new question of safety and efficacy.
- iii. Report contents raise a question of whether the report includes all information required by the clinician for diagnosis. The full ECG signal data is made available to the Certified Cardiographic Technicians in all three devices. The clinical report of all three devices contain the specifically identified content by
{6}------------------------------------------------
the CCT. This question of report content applies to the new device and the predicate / reference devices equally and is thus not considered a different or new question of safety and efficacy.
The identified questions of safety and efficacy listed above apply to both the new device and the predicate/reference devices and so the new device does not raise different or new questions of safety and efficacy. Therefore, the proposed (subject device) CardioSTAT®Electrocardiogram (ECG) Test Solution meets substantial equivalence requirements with regards to the legally marketed predicate device Zio® SRECG Monitoring Service (K143513 cleared on 06/19/2015) and reference device Carnation™ Ambulatory Monitor Model 3000 (CAM) (K143067 cleared on 12/22/2014)
### 5.7. PERFORMANCE DATA
The CardioSTAT® ECG Test Solution has been designed and tested to comply with all applicable FDA Recognized Standards. There are no device-specific quidance documents, special controls, and/or requirements in the device-specific classification regulation regarding performance data that are applicable to the subject device. All necessary performance testing has been conducted to support determination of substantial equivalence. Results confirm by examination and provision of objective evidence that the design output met the design input requirements in conformance with the following recognized consensus standards:
- ISO 14971 Third Edition 2019-12, Medical Devices - Application of Risk Management to Medical Devices.
- ANSI AAMI IEC 62304:2006, Medical device software Software life cycle ● processes.
- ISO 15223-1 Third Edition 2016-11-01, Medical Devices Symbols to be Used ● with Medical Device Labels, Labelling And Information to be Supplied -Part 1: General Requirements.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the CardioSTAT® ECG Test Solution and conformance has been shown to the following standards:
- . ANSI/AAMI ES60601-1:2005 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text), Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD).
- IEC 60601-1-2 Edition 4.0 2014-02. Medical Electrical Equipment Part 1-2: ● General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Disturbances – Requirements and Tests.
- IEC 60601-1-11 Edition 2.0 2015-01. Medical Electrical Equipment Part 1-11: ● General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment.
- ANSI AAMI IEC 60601-2-47:2012/(R)2016, Medical electrical equipment -- Part ● 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "icentia" in a bold, sans-serif font. The letters are black, except for a small blue square above the "i". To the right of the word, there is a small "TM" symbol, indicating a trademark. The logo is simple and modern.
# Biocompatibility
The biocompatibility evaluation for the CardioSTAT® ECG Test Solution was conducted in accordance with the FDA guidance Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process, Document issued on: September 4, 2020 and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity ●
- . Sensitization
- Irritation ●
The CardioSTAT® ECG Recorder is considered a surface device for contact with intact skin for a duration likely to exceed 24 hours but not 30 days.
## Animal
No Animal or clinical studies have been conducted in support of the CardioSTAT® ECG Test Solution.
# Clinical
Clinical data in the form of literature is included as follows:
In support of CardioSTAT® ECG Test Solution arrythmia detection performance when applied horizontally as compared to a standard Holter recorder.
Nault I. André P. Plourde B. Leclerc F. Sarrazin JF. Philippon F. O'Hara G. Molin F, Steinberg C, Roy K, Blier L, Champagne J.
Validation of a novel single lead ambulatory ECG monitor - Cardiostat™ -Compared to a standard ECG Holter monitoring
J Electrocardiol. 2019 Mar-Apr:53:57-63. doi: 10.1016/j.jelectrocard.2018.12.011. Epub 2018 Dec 19.
In this paper, arrhythmia detection accuracy was compared between CardioSTAT® ECG Test Solution and a standard 24 h Holter ECG monitoring. Two hundred twelve patient recordings were compared. AF was diagnosed in 73 patients. Agreement between CardioSTAT® ECG Test Solution ECG and standard Holter monitoring was 99% for AF detection. Authors conclude that CardioSTAT® ECG Test Solution showed excellent correlation with the standard Holter ECG monitoring for AF detection. Added value of the CardioSTAT® ECG Test Solution includes longer monitoring duration, less cumbersome installation and water resistance.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "icentia" in a bold, sans-serif font. The word is mostly black, but there is a small blue square above the "i". To the right of the word, there is a small "TM" symbol, indicating a trademark.
In support of CardioSTAT® ECG Test Solution performance in the VSL (Vertical Sternal Lead) position and in support of its performance for a home application modality.
Benjamin H, Bischof M, Goldshtein D, Fecteau P, Newman D.
A pandemic response to home delivery for ambulatory ECG monitoring: Development and validation
J Electrocardiol. 2021 Jan-Feb;64:72-75. doi: 10.1016/j.jelectrocard.2020.11.003. Epub 2020 Nov 18.
In response to the COVID-19 pandemic, a protocol was designed for mail-out devices and educational materials created to teach patients how to install the CardioSTAT® ECG recorder for 2 weeks of continuous ambulatory ECG monitoring. We compared data collection from two sequential patient populations; one who received standard device application in the same clinic in the months before the pandemic response, and another, who received their device by mail for self-installation at home in the VSL position. Patients received a single phone call when the device was mailed and were able to contact the manufacturer as needed for support. A total of 47 devices were assessed from each group. Each group was similar in age (70 vs 65 years), and clinical indication for monitoring. Overall, it was found that a mail-delivered home-based recording platform can be reliably used to acquire clinical data with similar data quality and patient compliance as a conventional in-clinic model for long term ambulatory ECG monitoring.
In support of CardioSTAT® ECG Test Solution performance with regards to patient wear time in long-term (up to 14 days) ambulatory environments.
Yennikomshian M. Jarvis J. Patton C. Yee C. Mortimer R. Birnbaum H. Topash M. Cardiac arrhythmia detection outcomes among patients monitored with the Zio patch system: a systematic literature review
ISSN: 0300-7995 (Print) 1473-4877 (Online) Journal homepage: https://www.tandfonline.com/loi/icmo20
This meta-analysis reports on twenty-three publications (22 unique studies). The unweighted mean wear time of the Zio patch was 10.4 days (median ranging from 5 to 14 days). Findings from the review suggest that long-term, continuous, uninterrupted monitoring with Zio results in longer patient wear times and higher cardiac arrhythmia detection rates compared with outcomes reported in previous reviews of short-duration (24-48 h) cardiac rhythm recording studies.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the word "icentia" in a stylized font. The word is primarily black, with a small blue square above the "i". To the right of the word, there is a small "TM" symbol, indicating a trademark.
lcentia Post Market data from Canada and UK is compared to the study above. When prescribed for 14 days, CardioSTAT® ECG Test Solution wear time in post-market data from Canada and UK (33486 ECG tests) was 13.1 days (standard deviation of 1day and median of 13.9 days). Based on the Zio patch literature review and CardioSTAT® postmarket data, the CardioSTAT® ECG Test Solution wear time is shown to be equivalent to or longer than the Zio patch device.
#### CONCLUSION 5.8.
The CardioSTAT® ECG Test Solution is substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.