K143513 · iRhythm Technologies, Inc. · DSH · Jun 19, 2015 · Cardiovascular
Device Facts
Record ID
K143513
Device Name
ZIO SR ECG Monitoring Service
Applicant
iRhythm Technologies, Inc.
Product Code
DSH · Cardiovascular
Decision Date
Jun 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2800
Device Class
Class 2
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
ECG rhythm analysis
—
—
—
—
—
—
—
Indications for Use
The ZIO® SR ECG Monitoring Service is intended to capture, and report symptomatic and/or continuous electrocardiogram (ECG) information for long-term monitoring (up to 14 days). It is indicated for use on adult patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reported ECG metrics include single-lead analysis on a beat by beat basis, heart rate measurement and rhythm analysis. The report does not contain diagnostic interpretation; the reported analysis is provided for review by the intended user to render a diagnosis based on clinical judgment and experience.
Device Story
ZIO SR ECG Monitoring Service captures continuous single-lead ECG data for up to 14 days. System includes a wearable patch recorder with Bluetooth, a wireless gateway, and a central processing service. Patient wears patch; patch records ECG continuously and allows patient-triggered 90-second ECG strip transmission via gateway. Data uploaded to iRhythm Clinical Center (iCC) post-wear or via wireless transfer. Algorithms process ECG data to generate metrics (beat-by-beat analysis, heart rate, rhythm). Certified Cardiographic Technicians (CCTs) review and edit algorithm outputs using a QA tool. Final report provided to clinicians for diagnostic decision-making. Benefits include improved patient compliance and symptom-rhythm correlation for cardiac event detection.
Clinical Evidence
Bench testing only. Performance testing confirmed design outputs met requirements per consensus standards (ISO 14971, IEC 60601 series, ISO 10993-1, EC12). Automated ECG analysis performance quantified; sensitivity and positive predictivity levels met requirements.
Technological Characteristics
Single-lead ambulatory ECG monitor; Bluetooth connectivity; wireless gateway; disposable electrodes (EC12:2000). Complies with IEC 60601-1, 60601-1-2, 60601-1-11, 60601-2-47. Biocompatible materials (ISO 10993-1). Software-based analysis with manual technician review.
Indications for Use
Indicated for adult patients 18+ years, asymptomatic or experiencing transient symptoms (palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, anxiety) requiring long-term ECG monitoring up to 14 days.
Regulatory Classification
Identification
A medical magnetic tape recorder is a device used to record and play back signals from, for example, physiological amplifiers, signal conditioners, or computers.
Predicate Devices
ZEUS (ZIO ECG Utilization Services) System (K142681)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 19, 2015
iRhythm Technologies, Inc. Rich Laguna Director QA/RA 650 Townsend Street, Ste 380 San Francisco, California 94103
Re: K143513
Trade/Device Name: ZIO® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service Regulation Number: 21 CFR 870.2800 Regulation Name: Medical Magnetic Tape Recorder Regulatory Class: Class II Product Code: DSH, DOK, DXH Dated: May 15, 2015 Received: May 18, 2015
Dear Rich Laguna:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Mitchell Stein
forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# SECTION 4 INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K143513
Device Name: ZIO® SkyRunner (SR) Electrocardiogram (ECG) Monitoring Service
#### Indications for Use:
The ZIO® SR ECG Monitoring Service is intended to capture, and report symptomatic and/or continuous electrocardiogram (ECG) information for long-term monitoring (up to 14 days). It is indicated for use on adult patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reported ECG metrics include single-lead analysis on a beat by beat basis, heart rate measurement and rhythm analysis. The report does not contain diagnostic interpretation; the reported analysis is provided for review by the intended user to render a diagnosis based on clinical judgment and experience.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use (21 CFR Part 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# SECTION 5 510(k) SUMMARY
This summary of the 510(k) premarket notification for the ZEUS System is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
{4}------------------------------------------------
IRHYTHM TECHNOLOGIES, INC.
K143513
## SECTION 5 510(k) SUMMARY
510(k) Notification K
## GENERAL INFORMATION
## Applicant:
iRhythm Technologies, Inc. 650 Townsend Street, Suite 380 San Francisco, CA 94103 U.S.A Phone: 415-632-5700 Fax: 415-632-5701
# Contact Person:
Rich Laguna Director Quality & Regulatory Affairs eMail: rlaguna@irhythmtech.com Phone: 415-632-5749 Fax: 415-632-5701
Date Prepared: December 10, 2014
## DEVICE INFORMATION
## Classification:
Recorder, Magnetic Tape, Medical, 21 CFR§870.2800 Computer, diagnostic, programmable, 21CFR§870.1425 Transmitters and receivers, electrocardiograph, telephone, 21CFR§870.2920
#### Product Codes: DSH, DQK, DXH
Trade Name: ZIO®®
Generic/Common Name: Medical magnetic tape recorder
# PREDICATE DEVICE(S)
ZEUS (ZIO ECG Utilization Services) System [K142681] ZIO PATCH [K121319] EPI MINI [K121628]
{5}------------------------------------------------
## SECTION 5 510(k) SUMMARY
## INDICATIONS FOR USE
The ZIO® SR ECG Monitoring Service is intended to capture, and report symptomatic and/or continuous electrocardiogram (ECG) information for long-term monitoring (up to 14 days). It is indicated for use on adult patients 18 years or older who may be asymptomatic or who may suffer from transient symptoms such as palpitations, shortness of breath, dizziness, light-headedness, pre-syncope, syncope, fatigue, or anxiety. The reported ECG metrics include single-lead analysis on a beat by beat basis, heart rate measurement and rhythm analysis. The report does not contain diagnostic interpretation; the reported analysis is provided for review by the intended user to render a diagnosis based on clinical judgment and experience.
## PRODUCT DESCRIPTION
The ZIO® SR ECG Monitoring Service consists of three components: (1) ZIO SR Patch Recorder with Bluetooth technology, (2) ZIO SR Wireless Gateway, and (3) ZIO ECG Utilization Service System.
The ZIO® SR Patch is a single-patient-use ECG monitor that provides a continuous, singlechannel recording in addition to symptomatic data transmission for up to 14 days. The ZIO® SR Patch records ECG without patient interaction, with the goal of improving patient compliance via simplicity of operation. Patients have the option of pressing a convenient trigger button which marks the continuous record and initiates a wireless transfer of an ECG strip of 90-second duration. The wireless transfer of data is enabled by the ZIO® SR Gateway, which requires proximity and reception but no patient interaction. The patient is encouraged to fill out a log to document symptomatic events, which will support symptom-rhythm correlation in the ZIO SR Report. Alternatively, the patient can enter symptom logs and view received transmissions via an online patient portal.
At the conclusion of the wear period (up to 14 days), the patient removes the ZIO® SR Patch and returns it by mail to an iRhythm data processing center.
Upon receipt of symptomatic or continuous ECG data at iRhythm's Clinical Center (iCC) the ECG data is downloaded, the data is processed through the algorithm and delivered to the QA Tool module where the results are reviewed and/or adjusted by iRhythm Certified Cardiographic Technicians (CCT's) for accuracy. iRhythm employed and trained Patch in-take and CCT personnel follow internal procedures for processing and are of algorithm performance anomalies. Any software anomalies are visible to and manually corrected by iRhythm Technologies CCT's during the QA review and/or Patch Report edits. The CCT generates a final report of the ECG findings contained within the data; thereby enabling the provision of a complete ECG processing and analysis service.
Upon explicit request from a clinician responsible for the patient's healthcare, longer segments of ECG data from the continuous recording on the Patch can also be wirelessly retrieved during the wear period.
Automated ECG analysis performance was quantified for any claimed analysis metrics. The resulting statistics demonstrate sensitivity and positive predictivity levels which satisfy requirements and minimize safety or efficacy concerns.
{6}------------------------------------------------
## SECTION 5 510(k) SUMMARY
# SUBSTANTIAL EQUIVALENCE
The indications for use for the ZIO® SR ECG Monitoring Service are substantially equivalent to the indications for use for the predicate devices. The performance testing results demonstrate that any differences in the technological characteristics between the devices do not raise any new issues of safety or efficacy. Therefore, the ZIO® SR ECG Monitoring Service is substantially equivalent to the predicate device.
## NONCLINICAL TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
There are no required FDA performance standards for the ZIO® SR ECG Monitoring Service. All necessary performance testing was conducted on the ZIO® SR ECG Monitoring Service to support determination of substantial equivalence to the predicate devices. The results confirm by examination and provision of objective evidence that the design output met the design input requirements in conformance with the following list of recognized consensus standards:
- . ISO 14971:2007/(R)2010 (Corrected 4 October 2007), medical devices - applications of risk management to medical devices
- IEC 60601-1:2005/(R)2012 and A1:2012, c1:2009/(r)2012 and a2:2010/(r)2012 (consolidated text) medical electrical equipment - part 1: general requirements for basic safety and essential performance
- IEC 60601-1-2:2007, Medical electrical equipment part 1-2: general requirements for . basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests
- IEC 60601-1-11 Edition 1.0 2010-04, medical electrical equipment - part 1-11: general requirements for basic safety and essential performance - collateral standard: requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- . IEC 60601-2-47:2012, Medical electrical equipment -- part 2-47: particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems. (Cardiovascular)
- o EC12:2000/(R)2010, disposable ecg electrodes. (Cardiovascular)
- ISO 10993-1:2009/(R) 2013, biological evaluation of medical devices -- part 1: evaluation and testing within a risk management process. (Biocompatibility)
## CONCLUSION
The ZIO® SR ECG Monitoring Service is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.