The Moving Life ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
Device Story
The ATTO Mobility Scooter is a battery-powered, three-wheeled, collapsible electric scooter designed for elderly or disabled individuals. It is operated by the user via a thumb throttle and tiller. The device features a front-wheel brushless DC hub motor, spring-applied electronic-release brakes, and a regenerative braking system. Models ATTO SPORT and ATTO SPORT MAX include additional manual front-wheel disc brakes and software-based speed control. The device is used in indoor and outdoor settings. The user controls speed and direction; the microprocessor-based controller monitors system performance. The device provides mobility, allowing users to navigate environments independently. It is collapsible for transport and storage.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 7176 series (sections 2-6, 9-11, 14, 16, 21, 25) and EN 12184:2014 standards confirmed performance and safety.
Technological Characteristics
Three-wheeled electric scooter; 48V DC brushless hub motor; Li-ion battery (48V, 5.2Ah); microprocessor-based controller; spring-applied electronic-release brakes; regenerative braking; manual front disc brake (SPORT models); airless NPT tires; collapsible frame. Designed per ISO 7176 and EN 12184:2014.
Indications for Use
Indicated for disabled or elderly persons requiring indoor/outdoor transportation assistance.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The logo is simple and professional, and it is easily recognizable.
May 19, 2023
Moving Life Ltd Igor Naroditsky RA Adviser Havat Alenby Netzer Sereni, 7039500 Israel
Re: K222703
Trade/Device Name: ATTO Mobility Scooter (ATTO):ATTO Mobility Scooter (ATTO SPORT);ATTO Mobility Scooter (ATTO SPORT MAX) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: April 17, 2023 Received: April 17, 2023
Dear Igor Naroditsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Tushar Bansal -S
Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222703
Device Name
ATTO Mobility Scooter (ATTO) ATTO Mobility Scooter (ATTO SPORT) ATTO Mobility Scooter (ATTO SPORT MAX)
Indications for Use (Describe)
ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY (K222703)
## ATTO Mobility Scooter
#### A. Contact Details
Applicant Name: Moving Life Ltd Havat Allenby, Netzer Sereni, 7039500 Israel Contact Name: Igor Naroditsky, Regulatory Adviser Phone: + 972 54 4384386 Date Prepared: Aug 15, 2022
#### B. Device Name:
Trade Name: ATTO Mobility Scooter
Models: ATTO, ATTO SPORT, ATTO SPORT MAX
Common Name: Vehicle, Motorized 3-Wheeled
Classification Name: Motorized three-wheeled vehicle; INI; 890.3800
#### C. Legally Marketed Predicate Device(s)
| 510(k) Number | Product Code | Trade Name | Manufacturer |
|---------------|--------------|-----------------------|-----------------|
| K160909 | INI | ATTO Mobility Scooter | Moving Life Ltd |
#### D. Device Description
ATTO Mobility Scooter is an electrical battery powered, three-wheeled scooter intended to provide mobility for elderly or disabled individuals in a variety of indoor and outdoor settings. The ATTO is a collapsible electric scooter which has a front-wheel brushless DC hub motor. The scooter can be collapsed in 12 seconds and can also be disassembled with ease.
The scooter controlled through a thumb throttle and protected by an electronic-release brake system, has a driving range up to 20 km between charges. It is capable of carrying a driver weighing up to 136 (for ATTO MAX) kg. It moves both directions and handles a 6° slope. It can come to a full stop within 1.1m when on a horizontal plane and within 2.2 m when on a 6° slope.
The differences between the ATTO Mobility Scooter and its predicates are:
- । Introduction of the two additional models, ATTO SPORT, and ATTO SPORT MAX with
- । Additional Manual Front Wheal brake for extra safety
{4}------------------------------------------------
- Software Speed control feature (in ATTO SPORT, and ATTO SPORT MAX only) ।
- -Enhanced LCD display
- -Extra wide rear wheel
- -Rear light handle
- Powerful LED Headlights -
- -Shock-absorbing airless NPT tires
- -Regenerative braking control system
- Modified controller -
- -Modified charger and battery
#### ய் Intended Use/indications for use
The Moving Life ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
#### F. Substantial Equivalence Comparison
The proposed ATTO Mobility Scooter is as safe and effective as the ATTO Mobility Scooter (K160909) by Moving Life. The Scooter has the same intended uses and same indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Moving Life ATTO Mobility Scooter and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Moving Life ATTO Mobility Scooter is as safe and effective as the ATTO Mobility Scooter (K160909) by Moving Life. Thus, the proposed ATTO Mobility Scooter is substantially equivalent.
| Model | Predicate ATTO Mobility<br>Scooter (K160909) | Proposed ATTO Mobility Scooter<br>(K222703) |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Moving Life ATTO Mobility<br>Scooter is an indoor/outdoor<br>scooter that provides<br>transportation for a disabled or<br>elderly person. | The Moving Life ATTO Mobility<br>Scooter is an indoor/outdoor<br>scooter that provides<br>transportation for a disabled or<br>elderly person. |
| Front wheel size | 203 X 27 mm | 8 in |
| Rear wheel size | 228 X 57 mm | 9 in |
| Ground clearance | 100 mm | 100 mm |
| Number of wheels | 3 | 3 |
| Length | 1200 mm | 1200 mm |
| Width | 560 mm | 560 mm |
| Max load | 100 kg | 120 kg for ATTO and ATTO Sport.<br>136 kg for ATTO SPORT MAX |
| Required width of<br>angled corridor | 1000 mm | 1000 mm |
#### G. Substantial Equivalence Chart
{5}------------------------------------------------
| Model | Predicate ATTO Mobility<br>Scooter (K160909) | Proposed ATTO Mobility Scooter<br>(K222703) |
|----------------------------|------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Turn radius | 1350 mm | 1350 mm |
| Battery | Li-Ion battery 48V 5.2 ah | Li-ion Battery 5.2 Ah; DC 48 V;<br>249.6 Wh |
| Charger | AC input.100-240 volt,<br>50/60Hz, DC output 54.6 volt.<br>2 Amp for lithium-ion battery | AC input. 100-240 volt,<br>50/60Hz, DC output 54.6 volt. 2<br>Amp for lithium-ion battery |
| Type of controller | Microprocessor based,<br>proportional and integral<br>controller which continually<br>monitors the ATTO systems. | Microprocessor based,<br>proportional and integral controller<br>which continually monitors the<br>ATTO systems. |
| Maximum speed<br>(forward) | 6.4 km/h | 6.4 km/h for ATTO and ATTO<br>Max. 9.6 km/h for Max SPORT |
| Maximum speed<br>(reverse) | 4 km/h | 4 km/h |
| Travel distance | 15 km | 20/17 km |
| Electrical System: | 48-volt DC | 48-volt DC |
| Drive system | Front wheel, direct drive. | Front wheel, direct drive. |
| Motor | 48-volt DC, Brushless electric<br>motor | 48-volt DC, Brushless electric motor |
| Brake | Spring Applied Electrical<br>Released Brakes with<br>freewheel mode | Spring Applied Electrical<br>Released Brakes with freewheel<br>mode with regenerative braking<br>control system. Additional front<br>disk brake for extra safety |
| Tiller | Adjustable locking for driving<br>comfort | Adjustable locking for driving<br>comfort |
| Braking Time | Not publicly available | .5s |
| Braking Distance | Not publicly available | 1.1m |
| Speed Control | Velocity and current control | Velocity and current control |
| Seat | Folding seat | Folding seat |
#### H. Non-clinical Testing
Moving Life ATTO Mobility Scooter is designed in accordance with the ISO7176 series including all relevant sections i.e., Section 2, Section 3, Section 4, Section 5, Section 6, Section 9, Section 10, Section 11, Section 14, Section 14, Section 16, Section 21 and Section 25, etc. ATTO performance and safety testing is conducted according to EN 12184:2014. All Performance testing met the predetermined acceptance values. The Moving Life ATTO Mobility Scooter functioned as intended and its functionality observed was as expected.
#### l. Clinical Testing
No clinical testing is included in this submission.
#### J. Technological Characteristics
ATTO Mobility Scooter is a same as the commercially available ATTO Mobility Scooter (K160909) by Moving Life and based on the same fundamental technology as the predicate devices. Both are
{6}------------------------------------------------
electric scooters that are battery operated and have automatic braking systems. Batteries and battery chargers are provided with each scooter. Performance parameters of the devices are very similar. There are some minor technology differences between the proposed ATTO Mobility Scooter and the commercially available ATTO Mobility Scooter (K160909) by Moving Life. None of these differences raises new issues of safety and effectiveness.
#### K. Conclusion
The safety and effectiveness of the ATTO Mobility Scooter were demonstrated by the testing in compliance with national and international standards. The intended use, basic technology, and the features of the ATTO Mobility Scooter are similar to the predicate device. No new issues of safety and effectiveness is raised by the differences between the proposed ATTO Mobility Scooter and its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.