K160909 · Moving Life, Ltd. · INI · Oct 4, 2016 · Physical Medicine
Device Facts
Record ID
K160909
Device Name
ATTO Mobility Scooter
Applicant
Moving Life, Ltd.
Product Code
INI · Physical Medicine
Decision Date
Oct 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Moving Life ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
Device Story
ATTO Mobility Scooter is a collapsible, battery-powered, three-wheeled electric scooter; provides mobility for elderly or disabled individuals in indoor/outdoor settings. Input: user operation via thumb throttle on a two-handle tiller. Transformation: microprocessor-based controller with PID velocity and current control regulates a 48V brushless DC hub motor; electronic-release brake system provides stopping. Output: motorized transport; max speed 6.4 km/h forward, 4 km/h reverse; 15 km range. Operated by the user; device can be collapsed or disassembled into two parts for storage/transport. Benefits: provides independent mobility for users weighing up to 100 kg; handles 6° slopes.
Clinical Evidence
No clinical testing is included in this submission. Bench testing performed in accordance with ISO 7176 series (sections 2-7, 9-11, 13-16, 21) and IEC 62133 for battery/charger safety.
Technological Characteristics
Three-wheeled electric scooter; 48V 5.2Ah Li-Ion battery; 48V brushless DC hub motor; microprocessor-based PID velocity/current controller; spring-applied electrical-release brakes. Dimensions: 1200mm length, 560mm width. Standards: ISO 7176 series, IEC 62133.
Indications for Use
Indicated for disabled or elderly persons requiring indoor/outdoor transportation.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 4, 2016
Moving Life Ltd. % Mr. Igor Naroditsky RA adviser Igor Naroditsky, Medical Device Certification LLC Emeq Hashalom 26/2 Yogneam Illit, 20692 Israel
Re: K160909
Trade/Device Name: ATTO Mobility Scooter Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: July 30, 2016 Received: August 8, 2016
Dear Mr. Naroditsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160909
Device Name ATTO Mobility Scooter
Indications for Use (Describe)
ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
# K160909
#### Contact Details A.
Applicant Name: Nino A. Ransenberg, Chairman Moving Life Ltd 2 Ha'ofe St. Kadima, lsrael 6092000
Contact Name: Igor Naroditsky, RA Adviser Phone: + 972 54 4384386 Fax: + 972 73 3913 998
#### B. Device Name:
Trade Name: ATTO Mobility Scooter
Model: Type I
Common Name: Vehicle, Motorized 3-Wheeled
Classification Name: Motorized three-wheeled vehicle; INI; 890.3800
#### C. Legally Marketed Predicate Device(s)
| 510(k) Number | Product Code | Trade Name | Manufacturer |
|---------------|--------------|------------|-----------------------|
| K110165 | INI | Luggie | Freerider Corporation |
#### D. Device Description
ATTO Mobility Scooter is an electrical battery powered, three-wheeled scooter intended to provide mobility for elderly or disabled individuals in a variety of indoor and outdoor settings. The ATTO is a collapsible electric scooter which has a front-wheel brushless DC hub motor. The scooter can be collapsed in 12 seconds, and can also be disassembled with ease.
The scooter controlled through a thumb throttle and protected by an electronic-release brake system, has a driving range of 15 km between charges. It is capable of carrying a driver
{4}------------------------------------------------
weighing up to 100 kg. It moves in both directions, and can handle a 6° slope. It can come to a full stop within 1.1m when on a horizontal plane and within 2.2 m when on a 6° slope.
#### ய் Intended Use/indications for use
The Moving Life ATTO Mobility Scooter is an indoor/outdoor scooter that provides transportation for a disabled or elderly person.
#### Substantial Equivalence Comparison ட்
The Moving Life ATTO Mobility Scooter is as safe and effective as the Luggie (K110165). The Moving Life ATTO Mobility Scooter has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Moving Life ATTO Mobility Scooter and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Moving Life ATTO Mobility Scooter is as safe and effective as the Luggie (K110165). Thus, the ATTO Mobility Scooter is substantially equivalent.
| Model | ATTO Mobility Scooter | Luggie |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| K number | K160909 | K110165 |
| Indication for Use | The Moving Life ATTO Mobility<br>Scooter is an indoor/outdoor<br>scooter that provides<br>transportation for a disabled or<br>elderly person. | The device provides<br>transportation for an elderly or<br>disabled person. It can be<br>used in variety of indoor and<br>outdoor settings. |
| Front wheel size | 203 x 27 mm | 200 x 50 mm |
| Rear wheel size | 228 x 57 mm | 200 x 50 mm |
| Ground clearance | 100 mm | 64 mm / 2.5 inches |
| Number of wheels | 3 | 3 (double front wheel) |
| Length | 1200 mm | 1000 mm |
| Width | 560 mm | 455 mm / 18 inches |
| Max load | 100 kg | 114 kg / 250 pounds |
| Required width of angled<br>corridor | 1000 mm | 762 mm |
| Turn radius | 135.0 cm | 103.75 cm / 40.85 inches |
| Battery | Li-Ion battery 48V 5.2 ah | Li-Ion battery 24V 8.5 ah |
| Charger | AC input.100-240 volt, 50/60Hz,<br>DC output 54.6 volt. 2 Amp for<br>lithium-ion battery | AC input.100-240 volt,<br>50/60Hz, DC output 24 volt. 5<br>Amp for li-ion battery |
| Type of controller | Microprocessor based,<br>proportional and integral<br>controller which continually<br>monitors the ATTO systems. | Microprocessor based which<br>continually monitors the<br>Luggie scooter's systems |
| Maximum speed (forward) | 6.4 km/h | 6 km/h |
| Maximum speed (reverse) | 4 km/h | 3.52 km/h |
| Travel distance | 15 km | 18 km |
| Electrical System: | 48 volt DC | 24 volt DC |
| Drive system | Front wheel, direct drive. | Rear wheel, direct drive via<br>sealed drive axle |
| Motor | 48 volt DC, Brushless electric<br>motor with 3 phases & sensors | 24 volt DC. Permanent<br>magnet totally enclosed for<br>outdoor use. Internal brushes.<br>3.0 Amp (no load) |
| Brake | Spring Applied Electrical<br>Released Brakes with freewheel<br>mode | Automatic dynamic<br>regenerating braking system<br>with spring activated<br>magnetic solenoid parking<br>brake and freewheel facility. |
| Tiller | Adjustable locking for driving<br>comfort | Same |
| Speed Control | PID velocity and current control | Proportional state of the art<br>design for safety and smooth<br>operation incorporated with<br>'Fault' diagnosis. |
| Seat | Folding seat | Contoured adjustable seat<br>with height adjustment.<br>Armrest are optional with<br>multiple width adjustments |
{5}------------------------------------------------
#### G. Non-clinical Testing
Moving Life ATTO Mobility Scooter is designed and tested in accordance with the ISO 7176 standard series, including all relevant sections i.e. Section 2, Section 3, Section 4, Section 5, Section 6, Section 7, Section 9, Section 10, Section 10, Section 11, Section 13,
{6}------------------------------------------------
Section 14, Section 15, Section 16, and Section 21. The battery and charger conform to the IEC 62133 standard. All Performance tests met the predetermined acceptance values. The Movinq Life ATTO Mobility Scooter functioned as intended and its functionality observed was as expected.
#### H. Clinical Testing
No clinical testing is included in this submission.
#### . Technological Characteristics
An electrical battery powered, three-wheeled scooter intended to provide mobility for elderly or disabled individuals in a variety of indoor and outdoor settings is the technological principle for both the subject and predicate devices. It is based on the use of electrical motor powered by rechargeable battery through electrical controller mounted on same three wheel platform. Thus, based on the following similar technological elements:
- . DC motor, is used to rotate the wheel forward or backward.
- Li lon rechargeable battery, to power the system.
- . The controller provides a movement and speed control.
- Three wheel based platform provides a stable and flexible platform for indoor and outdoor use.
- Brake system utilizes a same dynamic principles and technology.
- Both systems are operated via two handles tiller.
The following technological differences exist between the subject and predicate devices:
- The subject device reach maximum 15 kilometer of travel distance in comparing to 18 in the predicate.
- The subject device utilizes 48 V 5.2 ah Li-Ion battery in comparing to 24 V 8.5ah at the predicate.
- The subject device has only off board charger in comparing to the On and Off o board charger attached to the predicate device.
- The subject device has 48 V electrical systems in comparison to the 24V in the predicate.
- The subject device use a front wheel driving system in comparing to rear wheel driving in the predicate.
- . A speed control archived through PID velocity and current control in the subject device in compare to the proportional state of the art design for safety and smooth operation incorporated with 'Fault' diagnosis in the predicate.
{7}------------------------------------------------
At a high level, the subject and predicate devices are based on the following same technological elements: Both are electric scooters that are battery operated and have automatic braking systems. Batteries and battery chargers are provided with each scooter. Performance parameters of the devices are very similar. There are some minor technological differences between the proposed ATTO Mobility Scooter and the commercially available Luggie (K110165). None of these differences raise new issues of safety and effectiveness.
The additional and minor technology differences between the proposed ATTO Mobility Scooter Type I and its predicate, the Luggie (K110165) is folding mechanism of the proposed device and the face that it can be disassembled into two parts. The proposed device has a mechanism retracts the rear wheels while folding the seat. In addition, the proposed device has a separation mechanism that allows the disassembly of the vehicle to two main parts.
#### J. Conclusion
The safety and effectiveness of the ATTO Mobility Scooter were demonstrated by the testing in compliance with national and international standards. The intended use, basic technology, and the features of the ATTO Mobility Scooter are similar to the predicate device. Therefore, the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.