Alexandrite and Nd:YAG Laser models CANLASE and TORLASE
K222673 · Canadian Pioneer Medical Technology Corporation (Cpmt Laser) · GEX · Nov 17, 2022 · General, Plastic Surgery
Device Facts
Record ID
K222673
Device Name
Alexandrite and Nd:YAG Laser models CANLASE and TORLASE
Applicant
Canadian Pioneer Medical Technology Corporation (Cpmt Laser)
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
755nm: Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). 1064nm: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles. Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)
Device Story
Dual-head laser system (Alexandrite 755nm and Nd:YAG 1064nm); optically combined beam paths; single delivery system. High-intensity light penetrates skin; delivers therapeutic heat for photocoagulation, hemostasis, and selective melanin targeting. Used in clinical settings by trained professionals. Output affects tissue via thermal energy; benefits include hair reduction, lesion clearance, scar improvement, and nail clarity. System includes laser rail and delivery handpiece.
Clinical Evidence
Bench testing only. No clinical data provided. Performance supported by biocompatibility testing (cytotoxicity, irritation, sensitization per ISO 10993) and electrical/EMC safety testing (IEC 60601-1, IEC 60601-2-22, IEC 60825-1, IEC 60601-1-2).
Technological Characteristics
Class IV laser system; dual-wavelength (755nm/1064nm); ambient air cooling; spot sizes 1.5-26mm; pulse duration 0.25-100ms; frequency 0.5-10Hz. Complies with IEC 60601-1, IEC 60601-2-22, IEC 60825-1, and IEC 60601-1-2 standards.
Indications for Use
Indicated for hair reduction, treatment of benign pigmented/vascular lesions, wrinkles, soft tissue coagulation/hemostasis, and onychomycosis in patients of all Fitzpatrick skin types (I-VI) including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
November 17, 2022
Canadian Pioneer Medical Technology Corporation (CPMT LASER) Rashid Sayah Managing Director 210 Drumlin Circle. Concord Vaughan, Ontario L4K 3E3 Canada
Re: K222673
Trade/Device Name: Alexandrite and Nd:YAG Laser models CANLASE and TORLASE Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: September 6, 2022 Received: September 6, 2022
Dear Rashid Sayah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Jianting Wang -S" in a large, sans-serif font. The text is black against a white background. The letters are clearly legible and evenly spaced.
Jianting Wang Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222673
#### Device Name
Alexandrite and Nd:YAG laser models CANLASE , TORLASE
#### Indications for Use (Describe)
755mm:
Temporary hair reduction. Stable long-term or permanent reduction targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I-VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
1064nm:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesion size. for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Temporary increase of clear nail in patients with onychomycosis ( e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K222673
# Section 5 510(k) summary
#### l Submitter
Canadian Pioneer Medical Technology Corporation (CPMT LASER) 210 Drumlin Circle #2, Vauqhan, Ontario, L4K 3E3, Canada
Contact person: Dr. Rashid Reza Mir Sayah Managing Director Phone: 4377727788 Email: Canadianpioneer@yahoo.com Canadianpioneermedical@gmail.com Date of preparation: Sep 06, 2022
#### II Subject Device
Trade/Device Name: Alexandrite Nd:YAG laser , Models: CANLASE , TORLASE Regulation Number: 21 CFR 878.4810 Requlation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: II Product code: GEX Manufacturer: Canadian Pioneer Medical Technology Corporation
### III Predicate Devices
Trade/Device Name: GentleMax Pro Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: II Product code: GEX 510(k) number: k201111
Manufacturer: Candela Corporation
#### IV Device description
Alexandrite Nd:YAG laser Models CPMT CANLASE , CPMT TORLASE
contain two separate laser heads (Alexandrite and Nd:YAG), which produce laser light outputs of 755 nm and 1064 nm, respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system.
{5}------------------------------------------------
This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat.
## V Indications for use
#### 755nm:
Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.
Treatment of benign pigmented lesions. Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
### 1064nm:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)
# VI Comparison of technological characteristics with the predicate devices
The indication of proposed device is covered by the predicated devices.
| Device feature | | |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Alexandrite and Nd:YAG Laser | Laser treatment system |
| | (subject device) | GentleMax Family of Laser Systems<br>(K201111) |
| Product code | GEX | GEX |
| Manufacturer | Canadian Pioneer Medical<br>Technology corporation (CPMT<br>LASER) | Candela Corp. |
| Regulation<br>number | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Indications for<br>use | 755nm:<br>Temporary hair reduction. Stable<br>long-term or permanent reduction<br>through selective targeting of<br>melanin in hair follicles. Permanent<br>hair reduction is defined as<br>long-term stable reduction in the<br>number of hairs regrowing after a<br>treatment regime. Permanent hair<br>reduction is defined as the<br>long-term, stable reduction in the<br>number of hairs regrowing when<br>measured at 6, 9, and 12 months<br>after the completion of a treatment<br>regime. On all skin types<br>(Fitzpatrick I- VI) including tanned<br>skin.<br>Treatment of benign pigmented<br>lesions. Treatment of wrinkles.<br>The photocoagulation of<br>dermatological vascular lesions<br>(such as port-wine stains,<br>hemangiomas, telangiectasias).<br>1064nm:<br>Removal of unwanted hair, for<br>stable long term or permanent hair<br>reduction and for treatment of PFB.<br>Permanent hair reduction is<br>defined as the long-term, stable<br>reduction in the number of hairs<br>regrowing when measured at 6, 9 | 755nm:<br>Temporary hair reduction.<br>Stable long-term or permanent<br>reduction through selective<br>targeting of melanin in hair<br>follicles. Permanent hair<br>reduction is defined as<br>long-term stable reduction in<br>the number of hairs regrowing<br>after a treatment regime.<br>Permanent hair reduction is<br>defined as the long-term,<br>stable reduction in the number<br>of hairs regrowing when<br>measured at 6, 9, and 12<br>months after the completion of<br>a treatment regime. On all skin<br>types (Fitzpatrick I- VI)<br>including tanned skin.<br>Treatment of benign<br>pigmented lesions. Treatment<br>of wrinkles.<br>The photocoagulation of<br>dermatological vascular<br>lesions (such as port-wine<br>stains, hemangiomas,<br>telangiectasias).<br>1064nm:<br>Removal of unwanted hair, for<br>stable long term or permanent<br>hair reduction and for |
| and 12 months after the | treatment of PFB. Permanent | |
| completion of a treatment regime. | hair reduction is defined as the | |
| The lasers are indicated on all skin | long-term, stable reduction in | |
| types Fitzpatrick I-VI including | the number of hairs regrowing | |
| tanned skin. Photocoagulation and | when measured at 6, 9, and | |
| hemostasis of pigmented and | 12 months after the | |
| vascular lesions such as but not | completion of a treatment | |
| limited to port wine stains, | regime. The lasers are | |
| hemangioma, warts, | indicated on all skin types | |
| telangiectasia, rosacea, venus | Fitzpatrick I-VI including | |
| lake, leg veins and spider veins. | tanned skin. Photocoagulation | |
| Coagulation and hemostasis of soft | and hemostasis of pigmented | |
| tissue. Benign pigmented lesions | and vascular lesions such as | |
| such as, but not limited to, lentigos | but not limited to port wine | |
| (age spots), solar lentigos (sun | stains, hemangioma, warts, | |
| spots), cafe au lait macules, | telangiectasia, rosacea, venus | |
| seborrheic keratosis, nevi, | lake, leg veins and spider | |
| chloasma, verrucae, skin tags, | veins. Coagulation and | |
| keratosis, tattoos (significant | hemostasis of soft tissue. | |
| reduction in the intensity of black | Benign pigmented lesions | |
| and/or blue-black tattoos) and | such as, but not limited to, | |
| plaques. | lentigos (age spots), solar | |
| The laser is indicated for | lentigos (sun spots), cafe au | |
| pigmented lesions to reduce lesion | lait macules, seborrheic | |
| size, for patients with lesions that | keratosis, nevi, chloasma, | |
| would potentially benefit from | verrucae, skin tags, keratosis, | |
| aggressive treatment, and for | tattoos (significant reduction in | |
| patients with lesions that have not | the intensity of black and/or…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.