K201111 · Candela Corporation · GEX · May 26, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201111
Device Name
GentleMax Pro Plus
Applicant
Candela Corporation
Product Code
GEX · General, Plastic Surgery
Decision Date
May 26, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
755mm: Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). 1064nm: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques. The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles. 1064nm: Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)
Device Story
Dual-wavelength laser system (755nm Alexandrite, 1064nm Nd:YAG) for dermatological procedures; delivers high-intensity light to target melanin or vascular structures; includes Dynamic Cooling Device (DCD) using cryogen bursts to protect epidermis; operated by clinicians in professional settings; system uses optical fiber delivery; output parameters (energy, spot size, pulse duration) controlled via user interface; therapeutic heat induces photocoagulation/hemostasis; benefits include hair reduction, lesion clearance, and scar improvement.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-2-22), electromagnetic compatibility (IEC 60601-1-2), and laser safety (IEC 60825-1). Biocompatibility established via ISO 10993-5 and ISO 10993-10 testing.
Technological Characteristics
Class IV laser; flashlamp-excited solid-state Alexandrite (755nm) and Nd:YAG (1064nm); lens-coupled optical fiber delivery; cryogen-based DCD skin cooling; green diode aiming beam (520-550nm); repetition rate 1-10 Hz; pulse duration 0.25-100ms; max fluence 400-520 J/cm2; firmware-controlled user interface.
Indications for Use
Indicated for patients of all skin types (Fitzpatrick I-VI), including tanned skin, requiring hair reduction, treatment of benign pigmented/vascular lesions, wrinkle treatment, soft tissue coagulation/hemostasis, scar reduction, or temporary increase of clear nail in onychomycosis patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Candela GentleMAX Family of Lasers System (K140122)
Submission Summary (Full Text)
{0}------------------------------------------------
May 26, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Candela Corporation Yverre Bobay Director, Regulatory Affairs 251 Locke Dr Marlborough, Massachusetts 01752
Re: K201111
Trade/Device Name: GentleMax Pro Plus Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: April 24, 2020 Received: April 27, 2020
Dear Yverre Bobay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K201111
Device Name GentleMAX Family of Laser Systems
# Indications for Use (Describe)
755mm:
Temporary hair reduction. Stable long-term or permanent reduction targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas).
1064nm:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment hair reduction is defined as the long-term. stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesion size. for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary GentleMax Pro Plus System
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92
### 1. DATE PREPARED
April 24, 2020
### 2. APPLICANT NAME
Yverre Bobay Global Director, Regulatory Candela Corporation 251 Locke Drive Marlborough MA 01752 M: (617) 669-6181 yverreb@candelamedical.com
### 3. Official Correspondent
Greg Wallender VP, Global Quality & Regulatory Affairs Candela Corporation 530 Boston Post Road Wayland, MA 01778 M: (925) 206-5439 gregw@candelamedical.com
#### 4. DEVICE INFORMATION
| Proprietary Name: | GentleMax Family of Laser Systems |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Dermatology Laser System |
| Classification Name: | Laser surgical instrument for use in General and Plastic<br>surgery and in dermatology (21 CFR Section 878.4810,<br>Product Code GEX) |
| Product Code: | GEX |
| Device Classification: | Class II |
| Laser Classification: | Class IV |
### 5. PREDICATE DEVICE
Candela GentleMAX Family of Lasers System (K140122).
{5}------------------------------------------------
#### 6. INTENDED USE/INDICATION FOR USE
There are no new Intended Uses for the GentleMax Pro Plus. The Indications for Use Statement is unchanged for the GentleMax Pro Plus is as follows:
## 755mm:
Temporary hair reduction. Stable long-term or permanent reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as long-term stable reduction in the number of hairs regrowing after a treatment regime. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
## 1064nm:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.
The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.)
{6}------------------------------------------------
#### 7. DEVICE DESCRIPTION
The GentleMax Pro Plus contains two separate laser heads (Alexandrite and Nd:YAG), which produce laser light outputs of 755 nm and 1064 nm, respectively. The output of each laser head is optically combined on the laser rail, so that their beam paths are identical as they exit the laser system. This allows the use of a single delivery system which can output either 755 nm or 1064 nm wavelengths. The laser system creates a beam of high intensity light that penetrates deep into the skin tissue where it delivers a controlled amount of therapeutic heat. The Dynamic Cooling Device (DCD) protects the upper layers of the skin with a cooling burst of cryogen.
### 8. TECHNOLOGICAL CHARACTERISTICS
The new GentleMax Pro Plus system has the same design, technological characteristics, operating principles, and intended use as the Candela GentleMax Family of Laser System (K140122). The devices share the same technical features, such as calibration port, wavelengths, laser medium, delivery systems, power supply, cooling system, electronics, firmware and user display screen. The devices share the same operating principles, such as energy but the Gentle Max Pro Plus offers a faster repetition rate, and offers larger spot sizes. Any minor differences do not raise any new types of safety or effectiveness questions because the GentleMax Pro Plus parameters are similar to the predicates.
| Specifications | Modified Candela<br>GentleMax Pro Plus | Current Candela<br>GentleMax Pro |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Candela Corp. | Candela Corp. |
| K Number | Not Assigned | K140122 |
| Product Code | GEX | GEX |
| Regulation<br>Number | 878.4810 | 878.4810 |
| Device Class | II | II |
| Indications<br>for Use | The GentleMax Family of Laser Systems is<br>indicated for the following at the specified<br>wavelength:<br>755 nm<br>Temporary hair reduction. Stable long-term<br>or permanent reduction through selective<br>targeting of melanin in hair follicles.<br>Permanent hair reduction is defined as long- | The GentleMax Family of Laser Systems is<br>indicated for the following at the specified<br>wavelength:<br>755 nm<br>Temporary hair reduction. Stable long-term<br>or permanent reduction through selective<br>targeting of melanin in hair follicles.<br>Permanent hair reduction is defined as long- |
| | | |
| | term stable reduction in the number of hairs regrowing after a treatment regime.<br>Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.<br>Treatment of benign pigmented lesions.<br>Treatment of wrinkles.<br>The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). | term stable reduction in the number of hairs regrowing after a treatment regime.<br>Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. On all skin types (Fitzpatrick I- VI) including tanned skin.<br>Treatment of benign pigmented lesions.<br>Treatment of wrinkles.<br>The photocoagulation of dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). |
| | 1064nm<br>Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick 1-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.<br>The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. | 1064nm<br>Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime. The lasers are indicated on all skin types Fitzpatrick 1-VI including tanned skin. Photocoagulation and hemostasis of pigmented and vascular lesions such as but not limited to port wine stains, hemangioma, warts, telangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, chloasma, verrucae, skin tags, keratosis, tattoos (significant reduction in the intensity of black and/or blue-black tattoos) and plaques.<br>The laser is indicated for pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. |
| | Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.<br>Treatment of wrinkles.<br>1064nm<br>Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.) | Reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.<br>Treatment of wrinkles.<br>1064nm<br>Temporary increase of clear nail in patients with onychomycosis (e.g., dermatophytes, Trichophyton rubrum and T. mentagrophytes, and/or yeast Candida Albicans, etc.) |
| Laser Type | Flashlamp-excited, Solid state Alexandrite and Nd:YAG laser | Flashlamp-excited, Solid state Alexandrite and Nd:YAG laser |
| Aiming Beam | Green diode laser<br>(520-550 nm) | Green diode laser<br>(520-550 nm) |
| Wavelengths<br>(nm) | 755 & 1064 | 755 & 1064 |
| Maximum | 68 Joules (J) ALEX; 90 J | 53 joules (J) ALEX; 80 J |
| Energy (J) | Nd:YAG | Nd: YAG |
| Accuracy of<br>Output | ± 20% | ± 20% |
| Maximum<br>Fluence<br>(J/cm2) | 400 J/cm2 (ALEX)<br>520 J/cm2 (YAG) | 400 J/cm2 (ALEX)<br>520 J/cm2 (YAG) |
| Spot Size<br>(mm) | 1.5 mm, 3 mm, 3x10 mm, 5 mm,<br>6 mm, 8 mm, 10 mm, 12 mm, 15<br>mm, 18 mm, 20 mm, 22 mm, 24<br>mm, 26 mm | 1.5 mm, 3 mm, 3x10 mm, 5 mm,<br>6 mm, 8 mm, 10 mm, 12 mm, 15<br>mm, 18 mm |
| Repetition<br>Rate (Hz) | 1 – 10 Hz | 1 - 10 Hz |
| Pulse<br>Duration (ms) | 0.25 - 100 ms (new 2ms) | 0.25 - 100 ms |
| Operating<br>Modes | Pulse | Pulse |
| Energy<br>Delivery | Footswitch & finger switch | Footswitch & Handpiece switch |
| Beam<br>Delivery | Lens-coupled optical fiber | Lens-coupled optical fiber |
| Voltage and<br>Power (30A<br>configuration) | 200 V to 240 V~, 50/60 Hz,<br>single phase, 4,600 VA or 20 A at<br>230 V~ | 200 V to 240 V~, 50/60 Hz,<br>single phase, 4,600 VA or 20 A at<br>230 V~ |
| Voltage and<br>Power (20A<br>configuration) | 200 V to 240 V~, 50/60 Hz,<br>single phase, 3,600 VA or 16 A at<br>230 V~ | 200 V to 240 V~, 50/60 Hz,<br>single phase, 3,600 VA or 16 A at<br>230 V~ |
| System<br>Cooling | Ambient Air | Ambient Air |
| Skin Cooling | Yes<br>Cryogen (DCD)…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.