5000 Compact Series Ultrasound Systems (Ultrasound System 5500 G, Ultrasound System 5500 P, Ultrasound System 5500 W, Ultrasound System 5500 CV, Ultrasound System 5300 G, Ultrasound System 5300 P, Ultrasound System 5300 W)
5000 Compact Series Ultrasound Systems (Ultrasound System 5500 G, Ultrasound System 5500 P, Ultrasound System 5500 W, Ultrasound System 5500 CV, Ultrasound System 5300 G, Ultrasound System 5300 P, Ultrasound System 5300 W)
Applicant
Philips Ultrasound, LLC
Product Code
IYN · Radiology
Decision Date
Sep 27, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The 5000 Compact Series Ultrasound Systems (Ultrasound System 5500G, Ultrasound System 5500P, Ultrasound System 5500W, Ultrasound System 5500CV, Ultrasound System 5300G, Ultrasound System 5300P, Ultrasound System 5300W) are intended for diagnostic Ultrasound imaging in B (or 2D), 3D/4D, Color Doppler, Continuous Wave Doppler, Pulse Wave Doppler, Tissue Doppler, M-mode (including anatomical M-mode), Harmonics (Tissue and Contrast), Color Power Angio (CPA), and Combined modes. 5000 Compact series is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Abdominal, Adult Cephalic, Cardiac Adult, Cardiac Pediatric, Carotid, Cerebral vascular, Fetal Echo, Fetal/Obstetric, Gynecological, Intra-operative (Vascular), Lung, Musculoskeletal (conventional), Musculoskeletal (superficial), Neonatal Cephalic, Ophthalmic, Pediatric, Peripheral Vessel, Small Organs (breasts, testicles, thyroid), Transesophageal (cardiac), Trans-rectal, Trans-vaginal, Urology. The clinical environments where the 5000 Compact Series Ultrasound Systems can be used are: hospitals, surgery centers, clinics, physicians' office, diagnostic centers, critical care and emergency room environments, point-of-care clinical settings for diagnosis of patients. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. However, nothing stated in the user information reduces the user's responsibility for sound clinical judgment and best clinical procedure. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed.
Device Story
Portable, laptop-style diagnostic ultrasound system; battery or AC powered. Acquires ultrasound data via transducers; displays in various modes (B, 3D/4D, Doppler, M-mode, etc.). Operated by trained healthcare professionals in clinical environments (hospitals, clinics, point-of-care). System includes monitor, control panel, transducer/ECG receptacles. Outputs digital images and measurements for clinical review. Healthcare providers use output for diagnosis and fluid flow analysis. Benefits include portability for point-of-care use and efficient diagnostic imaging.
Clinical Evidence
No clinical data required. Substantial equivalence demonstrated through design features, indications for use, fundamental scientific technology, and non-clinical performance testing (verification/validation).
Indicated for diagnostic ultrasound imaging and fluid flow analysis in abdominal, cephalic (adult/neonatal), cardiac (adult/pediatric/fetal), vascular (carotid/cerebral/peripheral/intra-operative), fetal/obstetric, gynecological, lung, musculoskeletal, ophthalmic, small organ (breast/testicle/thyroid), transesophageal, trans-rectal, trans-vaginal, and urological applications. For use by trained healthcare professionals in hospitals, clinics, surgery centers, physician offices, diagnostic centers, and critical/emergency care settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
September 27, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
Philips Ultrasound LLC % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114
# Re: K222648
Trade/Device Name: 5000 Compact Series Ultrasound Systems (Ultrasound System 5500 G, Ultrasound System 5500 P, Ultrasound System 5500 W, Ultrasound System 5500 CV, Ultrasound System 5300 G, Ultrasound System 5300 P, Ultrasound System 5300 W) Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 31, 2022 Received: September 1, 2022
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Yanna Kang, Ph.D.
Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222648
#### Device Name
5000 Compact Series Ultrasound Systems (Ultrasound System 5500G, Ultrasound System 5500W, Ultrasound System 5500CV, Ultrasound System 5300G, Ultrasound System 5300P, Ultrasound System 5300W)
#### Indications for Use (Describe)
The 5000 Compact Series Ultrasound Systems (Ultrasound System 5500G, Ultrasound System 5500P, Ultrasound System 5500W, Ultrasound System 5500CV, Ultrasound System 5300C, Ultrasound System 5300P, Ultrasound System 5300W) are intended for diagnostic Ultrasound imaging in B (or 2D), 3D/4D, Color Doppler, Continuous Wave Doppler, Pulse Wave Doppler, Tissue Doppler, M-mode (including anatomical M-mode), Harmonics (Tissue and Contrast), Color Power Angio (CPA), and Combined modes.
5000 Compact series is indicated for diasound imaging and fluid flow analysis in the following applications: Abdominal, Adult Cephalic, Cardiac Adult, Cardiac Pediatric, Carotid, Cerebral vascular, Fetal Echo, Fetal/Obstetric, Gynecological, Intra-operative (Vascular), Lung, Musculoskeletal (conventional), Musculoskeletal (superficial), Neonatal Cephalic, Ophthalmic, Peripheral Vessel, Small Organs (breasts, testicles, thyroid), Transesophageal (cardiac), Trans-rectal, Trans-vaginal, Urology.
The clinical environments where the 5000 Compact Series Ultrasound Systems can be used are: hospitals, surgery centers, clinics, physicians' office, diagnostic centers, critical care and emergency room environments, point-of-care clinical settings for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. However, nothing stated in the user information reduces the user's responsibility for sound clinical judgment and best clinical procedure. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;"> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial;"> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K22648
# 5000 Compact Series Ultrasound Systems (Ultrasound System 5500G, Ultrasound System 5500P, Ultrasound System 5500W, Ultrasound System 5500CV, Ultrasound System 5300G, Ultrasound System 5300P, Ultrasound System 5300W)
This summary of safety and effectiveness information is submitted in accordance with 21 CFR 8807.92
### 1. Submitter's name, address, telephone number, contact person
- Manufacturer: Philips Ultrasound LLC 22100 Bothell Everett Hwy Bothell, WA 98021-8431
- Contact Person: Shilpa Rapaka Senior Regulatory Affairs Specialist 22100 Bothell Everett Hwy Bothell, WA 98021-8431 Phone: 203 213 6862
- Secondary Contact: Tamara Daniels Senior Regulatory Affairs Manager 22100 Bothell Everett Hwy Bothell, WA 98021-8431 Phone: 203 213 6862
- Date Prepared: September 27, 2022
# 2. Name of the device, including the trade of proprietary name if applicable, the common or usual name, and the classification name, if known:
| Proprietary Name: | 5000 Compact Series Ultrasound Systems<br>Marketed under the following models:<br>Ultrasound System 5500G<br>Ultrasound System 5500P<br>Ultrasound System 5500W<br>Ultrasound System 5500CV<br>Ultrasound System 5300G<br>Ultrasound System 5300P<br>Ultrasound System 5300W |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Names: | Ultrasonic pulsed doppler imaging system<br>Ultrasonic pulsed echo imaging system<br>Diagnostic ultrasonic transducer |
{4}------------------------------------------------
Regulation Description:
| Classification Name | 21 CFR Section | Product Code |
|---------------------------------------------|----------------|--------------|
| System, imaging, pulsed doppler, ultrasonic | 892.1550 | IYN |
| System, imaging, pulsed echo, ultrasonic | 892.1560 | IYO |
| Transducer, ultrasonic, diagnostic | 892.1570 | ITX |
As stated in 21 CFR parts 892.1550, 892.1560, and 892.1570, each of these generic types of devices have been classified as Class II.
Device Class: Class II
# 3. Indications for Use
The 5000 Compact Series Ultrasound Systems (Ultrasound System 5500G, Ultrasound System 5500P, Ultrasound System 5500W, Ultrasound System 5500CV, Ultrasound System 5300G, Ultrasound System 5300P, Ultrasound System 5300W) are intended for diagnostic Ultrasound imaging in B (or 2D), 3D/4D, Color Doppler, Continuous Wave Doppler, Pulse Wave Doppler, Tissue Doppler, M-mode (including anatomical M-mode), Harmonics (Tissue and Contrast), Color Power Angio (CPA), and Combined modes.
5000 Compact series is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Abdominal, Adult Cephalic, Cardiac Adult, Cardiac Pediatric, Carotid, Cerebral vascular, Fetal Echo, Fetal/Obstetric, Gynecological, Intra-operative (Vascular), Lung, Musculoskeletal (conventional), Musculoskeletal (superficial), Neonatal Cephalic, Ophthalmic, Pediatric, Peripheral Vessel, Small Organs (breasts, testicles, thyroid), Transesophageal (cardiac), Trans-rectal, Trans-vaginal, Urology.
The clinical environments where the 5000 Compact Series Ultrasound Systems can be used are: hospitals, surgery centers, clinics, physicians' office, diagnostic centers, critical care and emergency room environments, point-of-care clinical settings for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. However, nothing stated in the user information reduces the user's responsibility for sound clinical judgment and best clinical procedure. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed.
# 4. Device Description Summary
The proposed Philips 5000 Compact Series Ultrasound Systems (Ultrasound System 5500G, Ultrasound System 5500P, Ultrasound System 5500W, Ultrasound System 5500CV, Ultrasound System 5300G, Ultrasound System 5300P, Ultrasound System 5300W) are general purpose, compact, battery or alternating current (AC) powered, software controlled, diagnostic ultrasound
{5}------------------------------------------------
systems. The ultrasound system is housed in a portable, laptop-style chassis. The ultrasound system function is to acquire ultrasound data and to display the data in various modes of operation. The ultrasound system components include the monitor, control panel, transducer receptacles, ECG/physio receptacles, and AC adapter/battery charger. The ultrasound system can be attached to an optional cart. The cart height is adjustable to accommodate a range of operator heights and operating positions. The cart has wheels for maneuverability and may provide isolated power for system and peripheral components (such as black and white image printer or a report printer).
The 5000 Compact Series Ultrasound Systems comprises of different configurations which differ in terms of features, transducer options, and license options (features available as purchasable options in addition to the standard features available in the system. The types of options available include clinical options, QLAB Advanced Quantification Software, protocols, imaging capabilities, connectivity capabilities, etc.). The 5500G is the primary configuration with the most available capabilities. Identical software architecture is used in all configurations and all configurations are derived from a common codebase. Similarly, the hardware is same for all configurations. All the 5000 Compact Series Ultrasound System configurations are compatible with existing Philips transducers cleared in Philips Affiniti Ultrasound System (K200304). 5500 Compact Series POC (point-of-care) configurations leverage POC features from the CX50 platform (K162329) to enable efficient POC ultrasound capability.
# 5. Substantially Equivalent Devices
Primary Predicate Device: Philips Affiniti Ultrasound System (K200304)
Reference Device: Philips CX 50 Ultrasound Systems (K162329)
# 6. Technological Comparison to Predicate Devices
The proposed 5000 Compact Series Ultrasound Systems are substantially equivalent to the currently marketed predicate devices. The following is an overview of the similarities between the proposed 5000 Compact Series Ultrasound System and the currently marketed predicate and reference devices.
- The systems emplov the same fundamental scientific technology for imaging. doppler ● functions, and signal processing and have the same grayscale and doppler capabilities
- The systems are all are indicated for diagnostic ultrasonic imaging and fluid flow analysis ● for use by the same type of users in the same type of use environments
- The systems have the same clinical intended uses and indications for use
- The systems have the same imaging modes ●
- The transducers are identical ●
- The systems are manufactured with materials which have been evaluated and found to be . safe for the intended use of the device
{6}------------------------------------------------
- . The systems have acoustic power levels which are within the Track 3 FDA limits
- . The systems have similar capability in terms of performing measurements, capturing digital images, reviewing, and reporting studies
- . The systems have been designed and manufactured to the same electrical and physical safety standards manufactured under equivalent quality systems
A comparison of the subject 5000 Compact Series Ultrasound Systems and the predicate and reference devices is shown in Table 8-1 below.
Table 8-1. Comparison of the proposed 5000 Compact Series Ultrasound Systems to the currently marketed predicate and reference devices
| Standard Feature/<br>Technology<br>Characteristics | Proposed Philips<br>5000 Compact Series<br>Ultrasound Systems | Predicate<br>EPIQ/Affiniti<br>Ultrasound System<br>(K200304) | Reference CX50<br>Ultrasound Systems<br>(K162329) |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | Class II | Class II | Class II |
| Product Codes | IYN<br>IYO<br>ITX | IYN<br>IYO<br>ITX | IYN<br>IYO<br>ITX |
| Classification<br>Regulation Numbers | 21 CFR 892.1550<br>21 CFR 892.1560<br>21 CFR 892.1570 | 21 CFR 892.1550<br>21 CFR 892.1560<br>21 CFR 892.1570 | 21 CFR 892.1550<br>21 CFR 892.1560<br>21 CFR 892.1570 |
| Indications for Use<br>and Modes of<br>Operation | Abdominal,<br>Adult Cephalic,<br>Cardiac Adult,<br>Cardiac Pediatric,<br>Carotid,<br>Cerebral vascular,<br>Fetal Echo,<br>Fetal/Obstetric,<br>Gynecological,<br>Intra-operative<br>(Vascular),<br>Lung,<br>Musculoskeletal<br>(conventional),<br>Musculoskeletal<br>(superficial),<br>Neonatal Cephalic,<br>Ophthalmic,<br>Pediatric,<br>Peripheral Vessel,<br>Small Organs<br>(breasts, testicles,<br>thyroid), | Fetal/Obstetric,<br>Abdominal,<br>Intraoperative<br>(Vascular),<br>Intraoperative<br>(Cardiac),<br>Pediatric, Small<br>Organ (Breast,<br>Thyroid, Testicle),<br>Cephalic (Neonatal),<br>Cephalic (Adult),<br>Transrectal,<br>Transvaginal<br>Musculoskeletal<br>(Conventional),<br>Musculoskeletal<br>(Superficial)<br>Gynecological, Other:<br>Urology,<br>Cardiac Adult,<br>Cardiac Pediatric,<br>Transesophageal<br>(Cardiac),<br>Cardiac other (Fetal). | Ophthalmic,<br>Intracardiac echo,<br>Intraoperative,<br>Laparoscopic, Fetal,<br>Abdominal, Pediatric,<br>Small Organ,<br>Adult Cephalic,<br>Neonatal Cephalic,<br>Trans-vaginal,<br>Musculoskeletal,<br>Gynecological,<br>Cardiac Adult,<br>Cardiac pediatric,<br>Trans-Esophageal.<br>(Cardiac), Peripheral<br>Vessel, Other<br>(Carotid), Trans-rectal<br><b>Modes of Operation:</b><br>Sparq<br>B (or 2-D), M-mode<br>(including<br>Anatomical M-mode),<br>Pulse Wave Doppler,<br>Continuous Wave |
| | Transesophageal<br>(cardiac),<br>Trans-rectal,<br>Trans-vaginal,<br>Urology | Peripheral Vessel,<br>Cerebral Vascular,<br>Modes of Operation:<br>B (2D), M, PWD,<br>CWD, Color Doppler,<br>Combined (B+PWD,<br>B+Color,<br>B+Amplitude, B+M,<br>B+M+Color,<br>B+Color+PWD,<br>B+Amplitude+PWD,<br>B+CWD;<br>B+Color+CWD;<br>B+Amplitude+CWD),<br>SonoCT, Imaging for<br>guidance of biopsy,<br>Panoramic Imaging,<br>Color Power Angio<br>(CPA), Harmonic<br>Imaging, Contrast,<br>3D/4D Imaging,<br>XRES | Doppler, Color<br>Doppler, Tissue<br>Doppler Imaging and<br>Harmonics (Tissue<br>and Contrast)<br>CX50<br>B (or 2-D), M-mode<br>(including<br>Anatomical M-mode),<br>Pulse Wave Doppler,<br>Continuous Wave<br>Doppler, Color<br>Doppler, Tissue<br>Doppler Imaging and<br>Harmonics (Tissue<br>and Contrast) modes |
| Transducers | Modes of Operation:<br>B (or 2D),<br>3D/4D,<br>Color Doppler,<br>Continuous Wave<br>Doppler,<br>Pulse Wave Doppler,<br>Tissue Doppler,<br>M-mode (including<br>anatomical M-mode),<br>Harmonics (Tissue<br>and Contrast),<br>Color Power Angio<br>(CPA), Combined<br>modes | Affiniti<br>3D9-3v, C5-1, C6-2,<br>C8-5, C9-2, C9-4V,<br>C10-4ec, D2cwc,<br>D2tcd, D5cwc, BP10-<br>5ec, eL18-4, L12-3,<br>L12-3ERGO, L12- 4,<br>L12-5 50, L15-7io,<br>L18-5, S4-2, S5-1<br>S7-3t, S8-3, S12-4,<br>V6-2, VL13-5, X7-<br>2t, V9-2, mC7-2<br>EPIQ<br>C5-1, C8-5, C9-2,<br>C10-3v, C10-4ec,<br>D2cwc, D5cwc,<br>eL18-4, L12-3, L12-5<br>50, L15-7io, L18-5,<br>S5-1, S7-3t, S8-3, S8-<br>3t, S12-4, V6-2, X7-<br>2t, X8-2t, D2tcd, X5-<br>1, X6-1, VL13-5, X7-<br>2, X8-2ti, XL14-3,<br>3D9-3v, V9-2, mC7-<br>2, mC12-3 | C5-1<br>C8-5<br>C9-3io<br>C9-3v<br>C10-3v<br>D2cwc<br>D5cwc<br>L10-4 lap<br>L12-3<br>L12-5 50<br>L15-7io<br>S5-1<br>S8-3<br>S12-4<br>S7-3t<br>X7-2t<br>C6-2<br>C9-4v<br>L12-4<br>S4-2 |
| | No new transducers<br>introduced as part of<br>5000 Compact Series.<br>All transducers that<br>are part of the 5000<br>Compact Series are<br>already cleared in<br>K200304 or<br>K162329. | | |
| Power Supply | Same | Rechargeable lithium<br>ion batteries, AC<br>Adapter | Rechargeable lithium<br>ion batteries, AC<br>Adapter |
| Optional<br>Accessories/Peripherals | Various hardcopy, recording and printing products, black-and-white image printer, USB storage devices, Biopsy guides, Footswitch, Barcode scanner | Various hardcopy, recording and printing devices: digital video recorder (DVR), digital video disc (DVD), report printer, thermal printer (B/W, color), USB, PercuNav Instruments, Biopsy guides, Footswitch | Various hardcopy, recording and printing devices: report printer, DVD, USB, Biopsy guides, Footswitch |
| Form factor | Portable Laptop that can be placed on a mobile cart | Cart based, with an articulating monitor arm | CX50= Portable Laptop that can be placed on a mobile cart<br>Sparq= Cart based, with an articulating monitor arm |
| Battery Life | System operates via internal battery or AC power.<br>System battery supports up to 30 min scanning. Extended battery up to 2 hours with optional batteries | System operates on AC power. A battery is provided to put the system in Sleep mode for up to 30 minutes allowing the system to be transported to a new location without powering down the system. | The system operates via an internal battery or AC Power. System battery supports up to 30 minute scanning. |
| Wireless Network<br>Functions | Integrated on PC module<br>Wired and wireless | Integrated on PC module<br>Wired and wireless | Wired Ethernet<br>Wireless USB<br>Adapter |
| Software Features | Imaging Modes<br>Patient Data Entry<br>Patient Database and Storage<br>Generic Measurement Tools<br>Analysis and Calculation Packages<br>Annotations | Imaging Modes<br>Patient Data Entry<br>Patient Database and Storage<br>Generic Measurement Tools<br>Analysis and Calculation Packages<br>Annotations | Imaging Modes<br>Patient Data Entry<br>Patient Database and Storage<br>Generic Measurement Tools<br>Analysis and Calculation Packages<br>Annotations |
| Clinical Options | Abdomen Clinical Option<br>Adult Cardiology<br>Clinical Option | Abdomen Clinical Option<br>Adult Cardiology<br>Clinical Option | Abdomen Clinical Option<br>Adult Cardiology<br>Clinical Option |
| | Advanced | Advanced | Fetal Echo Clinical |
| | Capabilities Limited | Capabilities Limited | Option |
| | Clinical Option | Clinical Option | GYN Clinical Option |
| | Breast Clinical Option | Breast Clinical Option | Musculoskeletal |
| | Fetal Echo Clinical | Fetal Echo Clinical | Clinical Option |
| | Option | Option | OB Clinical Option |
| | GYN Clinical Option | GYN Clinical Option | Pediatric Cardiology |
| | Musculoskeletal | Musculoskeletal | Clinical Option |
| | Clinical Option | Clinical Option | Pediatric GI Clinical |
| | OB Clinical Option | OB Clinical Option | Option |
| | Pediatric Cardiology | Pediatric Cardiology | Small Parts including |
| | Clinical Option | Clinical Option | Breast Clinical Option |
| | Pediatric GI Clinical | Pediatric GI Clinical | Cerebrovascular |
| | Option | Option | including TCD |
| | Small Parts Clinical | Small Parts Clinical | Clinical Option |
| | Option | Option | Vascular Clinical |
| | TCD Clinical Option | TCD Clinical Option | Option |
| | Urology Clinical | Urology Clinical | Peripheral Vascular |
| | Option | Option | Clinical Option |
| | Vascular Clinical | Vascular Clinical | Regional Anesthesia |
| | Option | Option | Clinical Option |
| | POC Clinical Option | | Acute Care Clinical |
| | | | Option |
| Performance Options | Advanced Cardiac UI | Advanced Cardiac UI | Anatomic M mode |
| | Anatomic Mmode | Anatomic Mmode | Auto Doppler |
| | Auto Doppler | Auto Doppler | Auto Scan |
| | Auto Scan | Auto Scan | iScan |
|…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.