K200304 · Philips Ultrasound, Inc. · IYN · Mar 6, 2020 · Radiology
Device Facts
Record ID
K200304
Device Name
EPIQ Diagnostic Ultrasound System
Applicant
Philips Ultrasound, Inc.
Product Code
IYN · Radiology
Decision Date
Mar 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The intended use of the EPIQ Diagnostic Ultrasound System is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal Echo), Cardiac Pediatric, Cerebral Vascular, Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal. The clinical environments where the Philips EPIO Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients. When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance. The systems are intended to be installed and used only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces user responsibility for sound clinical judgement and best clinical procedure.
Device Story
EPIQ Diagnostic Ultrasound System; software-controlled console with transducers; acquires/displays ultrasound data via various modes. Input: ultrasound signals from transducers; Output: diagnostic images/flow analysis on console display. Used in clinics, hospitals, point-of-care by trained healthcare professionals. Integrates with Philips EchoNavigator for live X-ray/Echo guidance during cardiovascular procedures. Assists clinicians in diagnosis and interventional guidance; supports clinical decision-making through real-time visualization of anatomy and fluid flow.
Clinical Evidence
No clinical data required for substantial equivalence determination. Evidence based on non-clinical verification testing, including electrical safety (ANSI/AAMI 60601-1), electromagnetic compatibility (IEC 60601-1-2), ultrasonic field characterization (IEC 62359), and biocompatibility (ISO 10993 series).
Technological Characteristics
General-purpose diagnostic ultrasound system; console-based with plug-in transducers. mC12-3 transducer: curved array, 12-3 MHz frequency range, 160 elements. Connectivity: optional peripherals (ECG, printers). Software-controlled. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, and ISO 10993 standards.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring abdominal, cardiac (adult/pediatric/fetal), vascular (cerebral/peripheral), cephalic (neonatal), fetal/obstetric, gynecological, intraoperative, musculoskeletal, urological, and small organ (breast/thyroid/testicle) examinations. Used in clinics, hospitals, and point-of-care settings by trained healthcare professionals.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Philips EPIQ Series Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound System (K182857)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
March 6, 2020
Philips Ultrasound, Inc. % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K200304
Trade/Device Name: EPIO Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: February 5, 2020 Received: February 6, 2020
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K200304
Device Name
## EPIQ Diagnostic Ultrasound System
Indications for Use (Describe)
The intended use of the EPIQ Diagnostic Ultrasound System is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use:
Abdominal, Cardiac Adult, Cardiac other (Fetal Echo), Cardiac Pediatric, Cerebral Vascular, Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.
The clinical environments where the Philips EPIO Diagnostic Ultrasound Systems and Philips Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.
The systems are intended to be installed and used only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces user responsibility for sound clinical judgement and best clinical procedure.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Philips EPIQ Series Diagnostic Ultrasound Systems
This summary of safety and effectiveness information is submitted in accordance with 21CFR \$807.92
- 1. Submitter's name, address, telephone number, contact person. Philips Ultrasound, Inc. 22100 Bothell Everett Hwy Bothell, WA 98021-8431 Linda Schulz, Sr. Regulatory Affairs Specialist Contact person: Email: linda.schulz@philips.com Tel: 425-487-7127 Fax: 425-487-8666 Hebe Sun, Regulatory Affairs Manager Secondary Contact: Email: hebe.sun@philips.com Date prepared: January 15, 2020
- 2. Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
- Common name: Diagnostic ultrasound system and transducers
- Proprietary name: EPIQ Series Diagnostic Ultrasound System
These devices are classified as follows:
| Classification Description | 21 CFR Section | Product Code |
|------------------------------------------|----------------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | ITX |
As stated in 21 CFR, parts 892.1550, 892.1560, and 892.1570, each of these generic types of devices has been classified as Class II.
{4}------------------------------------------------
#### 3. Indications for Use
The intended use of the EPIQ Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use:
Abdominal, Cardiac Adult, Cardiac other (Fetal Echo), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal.
The clinical environments where the Philips EPIQ Series Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live Echo Guidance.
The systems are intended to be installed and used only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces user responsibility for sound clinical judgement and best clinical procedure.
#### 4. Device Description
The proposed Philips EPIQ Series Diagnostic Ultrasound Systems are general purpose, software controlled, diagnostic ultrasound systems. Their function is to acquire ultrasound data and to display the data in various modes of operation. The devices consist of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics and optional peripherals (ECG, printers).
Transducers are connected to the system using a plug in electrical connection. Other than the introductions of the new mC12-3 Transducer, the device description, accessories and components are unchanged. See Table 1 below for technical comparison of the subject device and the predicate device.
#### 5. Substantially Equivalent Devices
Philips Ultrasound believes the proposed Philips EPIQ Series Diagnostic Ultrasound Systems with mC12-3 Transducer is substantially equivalent to the following currently marketed devices:
Primary Predicate:
#### K1812857
Philips EPIQ Series Diagnostic Ultrasound System and Philips Affiniti Diagnostic Ultrasound System
A comparison of the subject device Philips EPIQ Series Diagnostic Ultrasound System with mC12-3 Transducer to the currently marketed and predicate device, Philips EPIQ Series Diagnostic Ultrasound Systems is provided in Table 1 below:
{5}------------------------------------------------
### TABLE 1
Comparison of the proposed Philips EPIQ Series Diagnostic Ultrasound System to the currently marketed and predicate device, Philips EPIQ Series Diagnostic Ultrasound System
| Feature | Subject Device<br>mC12-3<br>Transducer | Primary Predicate C8-5<br>Transducer<br>(K182857) | Reference Device<br>C9-2 Transducer<br>(K182857) | Comparison | |
|-----------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|--------------------------------------|----------------------------------------------------------|
| Indications<br>for Use | Same as EPIQ<br>K182857 | Fetal/Obstetric, Abdominal, Intraoperative<br>(Vascular), Intraoperative (Cardiac), Pediatric,<br>Small Organ (Breast, Thyroid, Testicle),<br>Cephalic (Neonatal), Cephalic (Adult),<br>Transrectal, Transvaginal, Musculoskeletal<br>(Conventional), Musculoskeletal (Superficial),<br>Gynecological, Other: Urology, Cardiac Adult,<br>Cardiac Pediatric, Transesophageal (Cardiac),<br>Cardiac other (Fetal), Peripheral Vessel,<br>Cerebral Vascular, | | Same | |
| Clinical application | Abdominal,<br>Pediatric, Neonatal<br>Cephalic,<br>Peripheral Vessel | Abdominal,<br>Pediatric, Neonatal<br>Cephalic,<br>Peripheral Vessel | Fetal / OB,<br>abdominal,<br>pediatric, musculo-<br>skeletal, GYN,<br>Urology, peripheral<br>vessel, and Fetal<br>Echo | Indications<br>identical to C8-<br>5 | |
| Acoustic Output | $ISPTA.3 \le 720$<br>(mW/cm2)<br>$TI \le 6.0$<br>$MI \le 1.9$ | Same | Same | Same | |
| Physical<br>Characteristics | Broadband<br>Frequency<br>Range | 12-3 MHz | 8-5 MHz | 9-2 MHz | Within range of<br>all EPIQ<br>Transducers<br>(18-2 MHz) |
| | Geometrical<br>Configuration | Curved Array | Curved Array | Curved Linear | Same as C8-5 |
| | Total number<br>of elements | 160 | 128 | 192 | Within the same<br>range |
| | Field of View | 94.9° | 122° | 102° | Narrower field<br>of view |
{6}------------------------------------------------
#### 6. Safety Considerations
The subject Philips EPIQ Series Diagnostic Ultrasound System with mC1203 Transducer is a Track 3 Device and complies with the referenced standard as well as the FDA ultrasound guidance document, Guidance for Industry and FDA Staff - Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers, issued on June 27, 2019.
#### 7. Nonclinical Performance Data
Philips Ultrasound performed the following testing to ensure the safety and effectiveness of the proposed mC12-3 Transducer with the Philips EPIO Series Diagnostic Ultrasound System:
- ANSI/AAMI 60601-1: Medical electrical equipment. General requirements for basic safety and essential performance, 2005, Amendment 1, 2012
- 트 IEC 60601-1-2 Medical Electrical Equipment - Part 1-2, General Requirements for Basic Safety and Essential Performance - Collateral Standard Electromagnetic Compatibility, 2014
- . IEC 60601-2-37: Medical electrical equipment. Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment, 2015
- . IEC 62359, Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields. 2017
- . ISO 10993-1: Biological evaluation of medical devices – Part 1:Evaluation and testing within a risk management process; 2018
- . 10993-5 Biological evaluation of medical devices - Part 5:Tests for in vitro cytotoxicity. 2009
- . ISO 10993-10: Biological evaluation of medical devices - Part 10:Tests for irritation and skin sensitization, 2010
- . ISO 10993-12: Biological evaluation of medical devices - part 12: sample preparation and reference materials, 2012
There are no device specific performance standards the proposed Philips EPIQ Series Diagnostic Ultrasound System with mC12-3 Transducer.
Non-Clinical verification testing has been performed addressing system level requirements according to system and design specifications, and risk control measures.
#### 8. Clinical Data
The proposed Philips EPIQ Series Diagnostic Ultrasound System did not require clinical data for determination of substantial equivalence.
#### 9. Sterilization
No components are supplied sterile.
#### 10. Conclusion
For testing, all pre-determined acceptance criteria were met. Results of these tests show that the proposed Philips EPIQ Series Diagnostic Ultrasound Systems with the new mC12-3 transducer meet their intended use. The subject device and predicate device:
{7}------------------------------------------------
- indicated for the diagnostic ultrasonic imaging and fluid flow analysis. .
- have the same gray-scale and Doppler capabilities.
- 트 use essentially the same technologies for imaging, Doppler functions and signal processing.
- . have acoustic output levels within the Track 3 FDA limits.
- manufactured under equivalent quality systems.
- I manufactured of materials with equivalent bio safety.
- . designed and manufactured to the same electrical and physical safety standards.
The differences between the subject device and predicate device do not raise new questions of safety and/or effectiveness. Therefore, the proposed Philips EPIQ Series Diagnostic Ultrasound System with mC12-3 Transducer is substantially equivalent to the predicate Philips EPIQ Series Diagnostic Ultrasound Systems in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.