K222293 · Urocure, LLC · OTN · Sep 21, 2022 · General, Plastic Surgery
Device Facts
Record ID
K222293
Device Name
ArcSP Suprapubic Sling System
Applicant
Urocure, LLC
Product Code
OTN · General, Plastic Surgery
Decision Date
Sep 21, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Implantable Sling intended to treat female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
Device Story
ArcSP Suprapubic Sling System is a surgical mesh device for treating female stress urinary incontinence. System comprises a knitted polypropylene mesh with integrated bioresorbable suture, plastic insertion sheaths, and two stainless-steel needle delivery assemblies. Device is used in a clinical setting by a physician. During surgery, needles facilitate placement of the sling mid-urethra via a suprapubic approach. The mesh and suture are permanently implanted to provide support to the urethra, addressing hypermobility or intrinsic sphincter deficiency. The delivery needles are single-use tools and are not implanted. The device functions by providing mechanical support to the urethra, which helps restore continence and improve patient quality of life.
Clinical Evidence
No clinical data. Bench testing only. Performance specifications including dimensional, needle functional, sling functional, fatigue, tensile, packaging, usability, distribution, and biocompatibility testing were conducted or leveraged from the predicate.
Technological Characteristics
Materials: knitted polypropylene mesh, bioresorbable suture, stainless-steel needles, plastic sheaths. Energy: None (mechanical). Connectivity: None. Sterilization: Ethylene Oxide (EO), SAL 10^-6. Form factor: Single-use sterile kit with delivery needles and sheathed mesh. Biocompatibility: ISO 10993-1:2009 compliant.
Indications for Use
Indicated for placement mid-urethra in females for treatment of stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
September 21, 2022
UroCure LLC % Denise Lenz Regulatory Consultant Libra Medial, Inc. 8401 73rd Ave N. Suite 63 Brooklyn Park, MN 55428
K222293 Re:
> Trade/Device Name: ArcSP Suprapubic Sling System Regulation Number: 21 CFR§ 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTN Dated: August 31, 2022 Received: September 6, 2022
Dear Denise Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica K. Nguyen, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K22293
Device Name ArcSP Suprapubic Sling System
Indications for Use (Describe)
The ArcSP Suprapubic Sling System is indicated to be placed at the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 7 510(K) SUMMARY
### 7.1 ADMINISTRATIVE INFORMATION
Date of Summary Preparation: July 29, 2022
#### 7.2 SUBMITTER INFORMATION
John Nealon Submitter UroCure, LLC 701 North 3rd Street, Suite 110, Minneapolis, MN 55401 Phone: 612-850-6814 Email: john.nealon(@urocure.com
#### 7.3 CONTACT INFORMATION
| Primary Submission Contact | Denise Lenz |
|------------------------------|---------------------------|
| | Regulatory Consultant, |
| | Libra Medical Inc. |
| | Phone: 763-232-3701 |
| | Email: dlenz@libramed.com |
| Secondary Submission Contact | Sew-Wah Tay, PhD |
| Secondary Submission Contact | Sew-Wah Tay, PhD<br>Regulatory Consultant,<br>Libra Medical Inc.<br>Phone: 612-801-6782<br>Email: swtay@libramed.com |
|------------------------------|----------------------------------------------------------------------------------------------------------------------|
|------------------------------|----------------------------------------------------------------------------------------------------------------------|
#### DEVICE INFORMATION 7.4
| Trade Name | ArcSP Suprapubic Sling System |
|--------------------------------------------|-------------------------------|
| Common Name | Urinary sling |
| Classification Name | Surgical Mesh |
| Classification Regulation | 878.3300 |
| Class | II |
| Panel | Gastroenterology/Urology |
| Product Code | OTN |
| Prior Regulatory Submissions<br>for Device | None |
### 7.5 510(K) TYPE AND REASON FOR SUBMISSION
This 510(k) is a Special 510(k) and is submitted to obtain marketing clearance for the ArcSP Suprapubic Sling System, a modification to the delivery device for the predicate ArcTV
{4}------------------------------------------------
Transyaginal Sling. The implantable portion of the device is identical to the predicate (ArcTV), with the delivery tool being the only difference from the predicate. ArcTV.
### 7.6 PREDICATE DEVICE
The predicate is ArcTV Transvaginal Sling System (K183134).
#### DEVICE DESCRIPTION 7.7
The ArcSP Suprapubic Sling System consists of the following:
- 1 ArcSP Sling with bio-resorbable Suture with plastic sheath ●
- 2 ArcSP Handle and Needle Delivery Assemblies .
The Handle and Needle delivery assembly is a sterile, single-use system, consisting of a two curved, stainless-steel Needles with attached Handles. The conical tip of each delivery Needle is designed to allow for passage through tissue. The Needles are a tool and together with the Sling Connectors and Sheaths, facilitate placement of the Sling and are not implanted.
The ArcSP Sling assembly is identical to the predicate. ArcTV Sling assembly and includes one knitted polypropylene mesh with an integrated bioresorbable Suture, two removable plastic insertion Sheaths, and two Sling Connectors. The Suture is an integral feature of the Mesh. The integrated Suture helps to minimize deformation of the mesh during sling insertion and placement and allows for adjustment of the Sling after removal of the Sheaths. The two plastic Sheaths cover and facilitate placement of the Sling. The Mesh and the bioresorbable Suture are permanently implanted.
### 7.8 INTENDED USE
Implantable Sling intended to treat female stress urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
#### 7.9 INDICATIONS FOR USE
The ArcSP Suprapubic Sling System is indicated to be placed mid-urethra for the treatment of female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
# 7.10 PERFORMANCE DATA
The ArcSP Suprapubic Sling System has been tested to meet the device's intended use and to ensure conformance to the product specifications.
The ArcSP meets all its physical and performance specifications including:
- Dimensional .
- Needle Functional Testing ●
- *Sling Functional Testing ●
- *Fatigue
- *Tensile
- *Packaging Testing
{5}------------------------------------------------
- Usability Testing ●
- *Distribution Testing ●
- *Biocompatibility
* No testing was repeated for this submission since the Sling implant assembly and packaging are identical to the predicate, ArcTV.
# 7.11 DEVICE COMPARISON
The ArcSP Suprapubic Sling System consists of the following:
- 1 ArcSP Sling with bio-resorbable Suture and plastic sheath
- 2 ArcSP Handle and Needle Delivery Assemblies .
The ArcSP Suprapubic Sling is identical to the predicate sling, ArcTV. The delivery approach of the sling is suprapubic for the subject device (ArcSP) versus transvaginal for the predicate device (ArcTV); therefore, the shape of the needle has changed to facilitate a suprapubic delivery. The final placement of the sling is the same for both. The two different Needles are to accommodate physician's preference of approach to implant.
The following table compares the subject device, ArcSP to the predicate device, ArcTV and the reference device, SPARC.
| Device<br>Characteristics | Subject Device<br>ArcSP Suprapubic Sling<br>System | Predicate Device<br>ArcTV Transvaginal<br>Sling System (K183134) | Reference Device<br>SPARC Sling System<br>(K081613) |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same as predicate<br>Implantable Sling<br>intended to treat female<br>stress urinary<br>incontinence resulting<br>from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency (ISD). | Implantable Sling<br>intended to treat female<br>stress urinary<br>incontinence resulting<br>from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency (ISD). | Implantable Sling<br>intended to treat female<br>stress urinary<br>incontinence resulting<br>from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency (ISD). |
| Device<br>Characteristics | Subject Device<br>ArcSP Suprapubic Sling<br>System | Predicate Device<br>ArcTV Transvaginal<br>Sling System (K183134) | Reference Device<br>SPARC Sling System<br>(K081613) |
| Indication for<br>Use | Same as predicate<br>The polypropylene sling<br>is indicated to be placed<br>mid-urethra for the<br>treatment of female stress<br>urinary incontinence<br>(SUI) resulting from<br>urethral hypermobility<br>and/or intrinsic sphincter<br>deficiency (ISD). | The polypropylene sling<br>is indicated to be placed<br>mid-urethra for the<br>treatment of female stress<br>urinary incontinence<br>(SUI) resulting from<br>urethral hypermobility<br>and/or intrinsic sphincter<br>deficiency (ISD). | Intended for the<br>placement of pubourethral<br>sling for the treatment of<br>female stress urinary<br>incontinence (SUI)<br>resulting from urethral<br>hypermobility and/or<br>intrinsic sphincter<br>deficiency. |
| Regulation<br>Number and<br>Product Code | Same as predicate | 878.3300<br>OTN | 878.3300<br>OTN |
| Regulatory<br>Classification<br>Name | Same as predicate | Surgical mesh | Surgical mesh |
| Device<br>Configuration | Same as predicate | Single use, sterile device | Single use, sterile device |
| Entry Method | Same as technological<br>reference | Transvaginal Retropubic | Suprapubic Retropubic |
| Delivery System | Same as technological<br>reference | Handle with detachable<br>curved Needle | Curved needle with<br>attached handle |
| Sling | Same as predicate | Sheathed polypropylene<br>mesh with adjustment<br>bioresorbable suture | Sheathed polypropylene<br>mesh with adjustment<br>bioresorbable suture |
| Sterilization | Same as predicate | EO, SAL 10-6-, single use | EO, SAL 10-6-, single use |
| Biocompatibility | Same as predicate | Meets ISO 10993-1:2009,<br>Biological evaluation of<br>medical devices<br>requirements | Meets ISO 10993-1:2009,<br>Biological evaluation of<br>medical devices<br>requirements |
| Packaging | Same as predicate | Molded tray with Tyvek<br>lid and separate pouch for<br>Sling | Molded tray with Tyvek<br>lid and separate pouch for<br>Sling |
Table 7-1: Comparison of Device Characteristics to Marketed Predicate Device
{6}------------------------------------------------
{7}------------------------------------------------
### 7.12 SUBSTANTIAL EQUIVALENCE
The ArcSP Suprapubic Sling System covered by this substantially equivalent to the predicate ArcTV Transvaginal Sling System device (K183134). The minor difference in the shape and size of the needle to optimize the ArcSP's top-down needle passage and technique do not raise new safety and efficacy as the shape, size and stiffness of the needle is the same as the technological reference.
## 7.13 CONCLUSION
Since the implantable sling assembly system is identical to the predicate, ArcTV and the intended use is the same, the ArcSP is substantially equivalent to its predicate. The minor difference in the shape, size and stiffness of the needle does not raise any new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.