K183134 · Urocure, LLC · OTN · Feb 7, 2019 · General, Plastic Surgery
Device Facts
Record ID
K183134
Device Name
ArcTV Transvaginal Sling System
Applicant
Urocure, LLC
Product Code
OTN · General, Plastic Surgery
Decision Date
Feb 7, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ArcTV Transvaginal Sling System is a retropubic sling indicated for a transvaginal (TV) placement of a midurethral sling for the treatment of adult female stress urinary inconting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
Device Story
Sterile, single-use retropubic sling system; includes detachable handle, two delivery needles, and sling assembly. Sling assembly consists of loosely knitted polymeric mesh with integrated bioresorbable suture, two removable plastic insertion sheaths, and two sling connectors. Physician uses handle to direct needles through tissue for transvaginal placement of mid-urethral sling. Integrated suture allows for sling adjustment post-placement without additional tools. Mesh and suture remain permanently implanted; needles, handle, sheaths, and connectors are not implanted. Used in clinical setting to treat female stress urinary incontinence; provides suburethral support to improve patient continence.
Clinical Evidence
Bench testing only. Testing included dimensional, fatigue, tensile, packaging, usability, and distribution testing. Biocompatibility testing performed per ISO 10993-1:2009 for limited duration contact (<24 hours) for needles/handle and permanent implant (>30 days) for mesh/suture.
Technological Characteristics
System components: handle, delivery needles, sling assembly (polymeric mesh, bioresorbable suture, plastic sheaths, sling connectors). Biocompatibility per ISO 10993-1:2009. Sterilization: Ethylene Oxide (EO), SAL 10^-6. Single-use. Mechanical delivery system.
Indications for Use
Indicated for adult females with stress urinary incontinence (SUI) caused by urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the words "FDA U.S. FOOD & DRUG" on the first line and "ADMINISTRATION" on the second line.
February 7, 2019
UroCure LLC % Ming-Cheng Chew Regulatory Consultant Libra Medical, Inc. 8401 73rd Ave No., Suite 63 Brooklyn Park, MN 55428
Re: K183134
Trade/Device Name: ArcTV Transvaginal Sling System Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: OTN Dated: January 9, 2019 Received: January 11, 2019
Dear Ming-Cheng Chew:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
for
Benjamin R. Fisher. Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K183134
Device Name ArcTV Transvaginal Sling System
Indications for Use (Describe)
The ArcTV Transvaginal Sling System is a retropubic sling indicated for a transvaginal (TV) placement of a midurethral sling for the treatment of adult female stress urinary inconting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD).
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY 1
### 1.1 ADMINISTRATIVE INFORMATION
Date of Summary Preparation: 1/9/2018
#### CONTACT INFORMATION 1.2
| Primary Submission Contact | Ming-Cheng Chew |
|------------------------------|----------------------------|
| | Regulatory Consultant, |
| | Libra Medical Inc. |
| | Phone: 763-232-3701 |
| | Email: mcchew@libramed.com |
| Secondary Submission Contact | Sew-Wah Tay, PhD |
| | Regulatory Consultant, |
### DEVICE INFORMATION 1.3
| Trade Name | ArcTV Transvaginal Sling System |
|---------------------------|-------------------------------------------------------------------------------------------------------------|
| Common Name | Urinary sling |
| Classification Name | Surgical Mesh |
| Classification Regulation | 878.3300 |
| Class | II |
| Panel | Gastroenterology/Urology |
| Product Code | OTN |
| Product Code Name | Mesh, Surgical, Synthetic, Urogynecologic, For Stress<br>Urinary Incontinence, Retropubic Or Transobturator |
Libra Medical Inc. Phone: 612-801-6782
Email: swtay@libramed.com
### 1.4 510(K) TYPE AND REASON FOR SUBMISSION
This 510(k) is a traditional 510(k) and is submitted to obtain marketing clearance for a new device - the ArcTV Transvaginal Sling System.
#### PREDICATE DEVICE 1.5
RetroArc Repubic Sling System (K132655)
#### DEVICE DESCRIPTION 1.6
The ArcTV Transvaginal Sling System is a sterile, single-use system, consisting of a Handle, two delivery Needles and a Sling assembly. The tip portion of each delivery Needle is configured to allow for passage through tissue. The opposite end of the Needle is configured to connect with
{4}------------------------------------------------
the Handle. The Handle is detachable and is used to direct both delivery Needles through tissue. The Sling assembly includes one piece of loosely knitted polymeric Mesh with an integrated bioresorbable Suture, two removable plastic insertion Sheaths, and two Sling Connectors. The Suture is an integral feature of the Mesh. The Suture and Mesh allow for adjustment of the Sling after initial placement in the patient without the use of additional tools. The two plastic Sheaths facilitate the placement of the Sling.
The Sling Connectors, Sheaths, Needles, and Handle are used to facilitate placement of the Sling assembly, and are not implanted. The Mesh and the bioresorbable Suture are permanently implanted.
### 1.7 INDICATIONS FOR USE
The ArcTV Transvaginal Sling System is a retropubic Sling indicated for a transvaginal (TV) placement of a mid-urethral Sling for the treatment of adult female stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency (ISD)
### 1.8 PERFORMANCE DATA
The ArcTV Transvaginal Sling System has been tested to meet the device's intended use and to ensure conformance to the product specifications.
The ArcTV meets all its physical and performance specifications including:
- Dimensional ●
- Fatigue
- Tensile
- Packaging Testing
- Usability Testing ●
- Distribution Testing
In addition, the ArcTV was tested for biocompatibility per ISO 10993-1:2009 for external communicating device with tissue for limited duration contact with (<24 hours) for the Needle and Handle, and for permanent implant in tissue (>30 days) for the Mesh and Suture.
### 1.9 SUBSTANTIAL EQUIVALENCE
| Device<br>Characteristics | ArcTV Transvaginal Sling System | AMS RetroArc Retropubic Sling<br>System (K132655) |
|---------------------------|----------------------------------------------------------------------------------------------------|---------------------------------------------------|
| Intended Use | The ArcTV Transvaginal Sling System<br>is intended to treat female stress urinary<br>incontinence. | Same |
{5}------------------------------------------------
| Device<br>Characteristics | ArcTV Transvaginal Sling System | AMS RetroArc Retropubic Sling<br>System (K132655) |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The ArcTV Transvaginal Sling System<br>is a retropubic Sling indicated for a<br>transvaginal (TV) placement of a mid-<br>urethral Sling for the treatment of adult<br>female stress urinary incontinence<br>(SUI) resulting from urethral<br>hypermobility and/or intrinsic sphincter<br>deficiency (ISD) | The RetroArc Retropubic Sling<br>System is intended for the placement<br>of a suburethral Sling for the<br>treatment of female stress urinary<br>incontinence (SUI) resulting from<br>urethral hypermobility and/or<br>intrinsic sphincter deficiency (ISD). |
| Entry Method | Transvaginal Retropubic | Same |
| Delivery System | Handle with detachable curved Needle | Same |
| Sling | Sheathed polypropylene mesh with<br>adjustment bioresorbable suture | Same |
| Sterilization | EO, SAL 10⁻⁶, single-use | Same |
| Packaging | Molded tray with Tyvek lid and<br>separate pouch for Sling | Same |
# 1.10 CONCLUSION
Based on the test data and the same intended use, the ArcTV is substantially equivalent to its predicate. The minor differences do not raise any new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.