The Electronic Pulse Stimulator MDTS100 is to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (leg) due to strain from exercise or normal household and work activities.
Device Story
Electronic Pulse Stimulator (MDTS100) is a battery-powered transcutaneous electrical nerve stimulator (TENS) for OTC home use. Device delivers pulsed monophasic electrical current to muscles via adhesive electrode pads placed on skin. User selects from 7 stimulation modes via front-panel button. Device features include automatic shut-off, no-load trip, and patient override control. Although device includes Bluetooth hardware, functionality is disabled in this model. Healthcare providers do not operate the device; patients self-administer therapy to relieve muscle soreness. Output parameters (voltage, current, frequency) are fixed within safety limits to ensure therapeutic benefit without requiring clinical supervision.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-2-10), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993-1, -5, -10), and software verification/validation.
Indicated for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities, and lower extremities due to strain from exercise or normal household and work activities in adults.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 21, 2022
Beijing Choice Electronic Technology Co., Ltd. c/o Haiying Zhao Quality Director North Building 3F, No.9 Shuangyuan Road Badachu Hi-tech Zone, Shijingshan District Beijing, China 100041
Re: K221992
Trade/Device Name: Electronic Pulse Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: September 15, 2022 Received: September 19, 2022
Dear Haiying Zhao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela D. Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K221992
Device Name Electronic Pulse Stimulator
Indications for Use (Describe)
The Electronic Pulse Stimulator MDTS100 is to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (leg) due to strain from exercise or normal household and work activities.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section II 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the device.
## 3.1 Submitter Information
#### ● Manufacturer Name:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. 2nd Floor, 3rd Floor and Room 410-412 4th Floor, No. 2 Building, No. 9 Shuangyuan Road , Shijingshan District BEIJING, 100041, P.R.China
#### ● Contact Person:
Haiying Zhao Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88204631 Fax: 861088204632 Email: cc@choicemmed.com
- Date prepared :June 24, 2022 ●
### 3.2 Proposed Device Information
Device Common Name: Transcutaneous Electrical Nerve Stimulator, OTC
Device Trade/Proprietary Name: Electronic Pulse Stimulator
Model: MDTS100
Classification Name: Stimulator, Nerve, Transcutaneous,Over-The-Counter
Regulation Number: 882.5890
Product Code: NUH
Class: II
Panel: Neurology
{4}------------------------------------------------
## 3.3 Predicate Device
#### 510(k) Number: K160508
Common Name: Transcutaneous Electrical Nerve Stimulator, OTC
Device Trade/Proprietary Name: Electronic Pulse Stimulator
Model: MDTS100
Classification Name: Stimulator, Nerve, Transcutaneous, Over-The-Counter
Product Code: NUH
Regulation Number: 882.5890
Device Class: II
Panel: Neurology
Manufacturer: Beijing Choice Electronic Technology Co., Ltd.
Intended Use: The Electronic Pulse Stimulator MDTS100 is to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household and work activities.
## 3.4 Device Description
The Electronic Pulse Stimulator is powered by internal battery. It is intended for at home use in delivering electric pulses to tired and sore muscles. These pulses are generated by the device & delivered through electrode pads placed on the skin. The device has 7 stimulation modes: Model-Mode 7. These stimulation modes can be selected by pressing the MODE button on the front of the unit.
The device generates small pulses of electrical current. It transmits these pulses to the user's skin through adhesive electrode pads, which activates the nerves below and temporarily relieves pain related to soreness.
Its accessories include connecting wires, electrode pads, 2 AAA batteries.
The device is for over the counter use.
The device does not contain drug or biological products.
The device includes Bluetooth functionality, however, in this model of the device, the Bluetooth functionality is not used.
{5}------------------------------------------------
### 3.5 Comparison list of the technological characteristics
| Comparison Elements | | | | Proposed Device | Predicate Device | Remark |
|---------------------------------------------------------------------------------------------------|----------------------|----------------------------|------------------------------------------------|---------------------------------------------------------|----------------------------------------------|--------|
| 510(k) number | | | | NA | K160508 | |
| Device Name/Model | | | | Electronic Pulse Stimulator MDTS100 | Electronic<br>Pulse<br>Stimulator<br>MDTS100 | |
| Manufacturer | | | Beijing Choice Electronic Technology Co., Ltd. | Beijing<br>Choice<br>Electronic<br>Technology Co., Ltd. | Same | |
| | -Method<br>Isolation | of<br>Line | Current | N/A for DC current | N/A for DC current | Same |
| Power | -Patient | -Normal condition | | < 10μΑ | <10μA | Same |
| Source(s) | Leakage<br>Current | -Single<br>condiction | fault | <50uA | <50μA | Same |
| Average DC current through<br>electrodes when device is on but no<br>pulses are being applied(uA) | | | < 10μΑ | < 10μΑ | Same | |
| Number of Output Modes | | | 7 | 7 | Same | |
| Number of Output<br>-Synchronous<br>or<br>Channels<br>Alternating? | | | Alternating | Alternating | Same | |
| | | of<br>-Method<br>Isolation | Channel | No | No | Same |
Table 2-1 Basic Unit Characteristics Comparison Table between the Proposed Device and Predicate Device
{6}------------------------------------------------
| Regulated Current or Regulated Voltage? | Regulated Voltage | Regulated Voltage | Same | |
|-------------------------------------------------------------------------------|--------------------------|-----------------------------|---------|------|
| Software/Firmware/Microprocessor Control? | Yes | Yes | Same | |
| Automatic Overload Trip | No | No | Same | |
| Automatic No-Load Trip? | Yes | Yes | Same | |
| Automatic Shut Off? | Yes | Yes | Same | |
| Patient Override Control? | Yes | Yes | Same | |
| Indicator<br>Display: | - On/Off Status? | Yes | Yes | Same |
| | - Low Battery? | Yes | Yes | Same |
| | -Voltage/Current Level? | Yes | Yes | Same |
| Timer Range (minutes) | 20 minutes | 20 minutes | Same | |
| Compliance with Voluntary Standards? | Yes | Yes | Same | |
| Compliance* with 21 CFR 898?<br>(*Becomes mandatory beginning May 9,<br>2000) | Yes | Yes | Same | |
| Weight | 2.3 oz(Battery Excluded) | 2.2 oz. (Battery Excluded ) | Similar | |
| Dimensions (in.) [W x H x D] | 2.17x 5.12x0.79 | 2.17x 5.12x0.87 | Similar | |
| Housing Materials and Construction | Enclosure: ABS | Enclosure: ABS | Same | |
### Premarket Notification 510(k) Submission—Section III 510(k) Summary
{7}------------------------------------------------
| Comparison Elements | Proposed Device | Predicate Device | Remark | |
|------------------------------------------------------------------------------------|------------------------------------------------|----------------------------------------------------------------------------------------------------|----------|------------|
| Waveform (e.g., pulsed monophasic, biphasic) | pulsed monophasic | pulsed monophasic | Same | |
| Shape (e.g., rectangular, spike, rectified sinusoidal) | Rectangular | Rectangular | Same | |
| Maximum Output Voltage (specify units)<br>(+/- 20 %) | 150V @ 500 Ω<br>160V @ 2 kΩ<br>165V @ 10 kΩ | 150V @ 500 Ω<br>160V @ 2 kΩ<br>165V @ 10 kΩ | Same | |
| Maximum Output Current (specify units)<br>(+/- %) | 300mA @ 500 Ω<br>80mA @ 2 kΩ<br>16.5mA @ 10 kΩ | 300mA @ 500 Ω<br>80mA @ 2 kΩ<br>16.5mA @ 10 kΩ | Same | |
| Duration of primary (depolarizing) phase (µsec) | 0 µsec | 0 µsec | Same | |
| Pulse Duration((µsec)) | 35 µsec | 35 µsec | Same | |
| Frequency(Hz) | 0.9Hz~82Hz | 0.9Hz~82Hz | Same | |
| Net Charge (µC per pulse)<br>(If zero, state method of achieving zero net charge.) | 42µC @ 500 Ω | 42µC @ 500 Ω | Same | |
| Maximum Phase Charge, (µC)@500Ω | 42µC | 42µC | Same | |
| Maximum Current Density, (mA/cm | 3.3mA/cm2@1.57"×1.57"Electrode Pad | 3.3mA/cm2@1.57"×1.57"Electrode Pad | Same | |
| Maximum Average Current (average absolute value), mA | 2.8mA | 2.8mA | Same | |
| Burst Mode (i.e., pulse trains) | a. Pulses per burst | 197 | 197 | Same |
| | b. Bursts per second | 0.185 | 0.185 | Same |
| | c. Burst duration (seconds) | 2.4s | 2.4s | Same |
| | d. Duty Cycle [Line (b) x Line (c)] | 0.44 | 0.44 | Same |
| ON Time (seconds) | | ≤1second | ≤1second | Same |
| OFF Time (seconds) | | ≤1second | ≤1second | Same |
| Cybersecurity | | FDA Guidance for Content of Premarket Submission for Management of Cybersecurity in Medical Device | / | Difference |
| Wireless Testing | | Conform to FCC and wireless coexistence test | / | Difference |
{8}------------------------------------------------
#### Premarket Notification 510(k) Submission-Section III 510(k) Summary
#### Conclusion:
The proposed device and the predicate device have same performance specification. There are very small differences in the weight and dimensions of the two devices, which have no impact on the safety and effectiveness.
Compared with the predicate device the proposed device has bluetooth function. We have conducted the wireless test and cybersecurity analysis to verify the basic safety and the essential performance. The Bluetooth function in this model of the device is not used.
{9}------------------------------------------------
### 3.6 Intended use
The Electronic Pulse Stimulator MDTS100 is to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household and work activities.
## 3.7 Testing
### Biocompatibility Testing
The material in contact with the patient of the proposed device is self adhesive electrode, which has been conducted the biocompatibility test according to the requirement of the ISO 10993-1, ISO 10993-5, ISO 10993-10.
### Electrical safety and electromagnetic compatibility (EMC)
The proposed device Electronic Pulse Stimulator MDTS100 is tested in accordance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10 to evaluate the electrical safety and EMC.
#### Performance Test-Bench
We have also conducted IEC 60601-2-10 and other performance tests including maximum output amplitudes, pulse durations, pulse repetition frequencies, maximum output current, supply voltage fluctuations, supply voltage fluctuations, Cleaning Test , High and Low Temperature & Humidity Test, Shelf-life Test as so on .
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
{10}------------------------------------------------
#### Cybersecurity testing
Cybersecurity testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Postmarket Management of Cybersecurity in Medical Devices." Cybersecurity testing was performed because of the Bluetooth function. Please note that the Bluetooth function in this model of the device is not used.
#### Wireless Testing
We have conducted the FCC test and wireless coexistence for the proposed device MDTS 100. The test reports are presented in Electromagnetic Compatibility and Electrical Safety. Wireless testing was performed because of the Bluetooth function. Please note that the Bluetooth function in this model of the device is not used.
#### Clinical Testing
The clinical testing was not needed in this submission.
### 3.8 Determination of substantial equivalence
The proposed device of the Electronic Pulse Stimulator MDTS100 has the same classification information, same intended use, same design principle, same product design and specifications as the predicate device. So, the proposed device are Substantially Equivalent (SE) to the predicate device which is US legally market device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.