The Electronic Pulse Stimulator MDTS100 is to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household and work activities.
Device Story
MDTS100 is an OTC, battery-powered (3V, 2 AAA) transcutaneous electrical nerve stimulator (TENS) for at-home pain relief. Device consists of a host unit, output cables, and reusable skin electrodes. User selects from seven stimulation modes via front-panel buttons and adjusts intensity; device delivers pulsed monophasic rectangular waveforms to muscles. Features include dual independent output channels, automatic timing (default 20 minutes), and output load detection. Healthcare providers do not operate the device; users self-administer therapy. Output affects muscle tissue to provide temporary pain relief. Device is not life-supporting, not implantable, and non-sterile.
Clinical Evidence
Bench testing only. No clinical data provided. Device validated through electrical safety and EMC testing per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-10. Biocompatibility of electrode pads confirmed via reference to K090198 and ISO 10993 standards. Software verification performed per FDA guidance.
Indicated for temporary relief of pain associated with sore and aching muscles in shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to exercise or normal household/work activities. For OTC use.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is a common symbol associated with healthcare and medicine. The logo is simple and recognizable, and it is used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 21, 2016
Beijing Choice Electronic Technololgy Co., Ltd. Lei Chen Quality Director No.9 Shuangyuan Road, Badach Hi-tech Zone Shijingshan District Beijing, 100041 CN
Re: K160508
Trade/Device Name: Electronic Pulse Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator For Pain Relief Regulatory Class: Class II Product Code: NUH Dated: February 22, 2016 Received: February 24, 2016
Dear Lei Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
| William J. | Digitally signed by William J. H |
|--------------|----------------------------------|
| Heetderks -A | |
| DN: | c=US, o=U.S. Government |
|-------|------------------------------|
| ou= | People, |
| | 0.9.2342.19200300.100.1.1=00 |
| cn= | William J. Heetderks -A |
| Date: | 2016.11.21 21:22:10 -05" |
| for | Carlos L. Peña, PhD, MS |
|-----|--------------------------------------------|
| | Director |
| | Division of Neurological |
| | and Physical Medicine Devices |
| | Office of Device Evaluation |
| | Center for Devices and Radiological Health |
Enclosure
{2}------------------------------------------------
# Section II Indications for Use Statement
## Indications for Use
510(k) Number (if known): Device Name: Electronic Pulse Stimulator Model:
Indications for Use:
The Electronic Pulse Stimulator MDTS100 is to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household and work activities.
| Prescription Use | |
|-----------------------------|--|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | V |
|------------------------|---|
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
### 3.1 Submitter Information
### Manufacturer Name:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 4104,No.A12 Yuquan Road Haidian District 100143 Beijing, P.R.China
### Contact Person:
Mr.Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88798300 Ext 6020 Fax:215-4052545 Email: cc@choicemmed.com Date prepared: October 31, 2016
## 3.2 Proposed Device Information
Device Common Name: Transcutaneous Electrical Nerve Stimulator, OTC
Device Trade/Proprietary Name: Electronic Pulse Stimulator
Model: MDTS100
Classification Name: Transcutaneous Electrical Nerve Stimulator for pain relief
Regulation Number: 21 CFR 882.5890
Product Code: NUH
Class: II
Use: Over -The-Counter Use
{4}------------------------------------------------
Panel: Neurology
## 3.3 Predicate Device
510(k) Number: K122744
Common Name: Transcutaneous electrical nerve stimulator OTC
Device Trade/Proprietary Name: Prospera OTC TENSE Electronic Pulse Massager
Model: PL029
Classification Name: Transcutaneous Electrical Nerve Stimulator for pain relief
Product Code: NUH, NGX
Regulation Number: 21 CFR 882.5890
Device Class: II
Panel: Neurology
Use: Over-The-Counter Use
Manufacturer: Prospera Corporation
### Intended Use:
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities(arm), and lower extremities(leg) due to strain from exercise or normal household work activities.
## 3.4 Device Description
The MDTS100 Electronic Pulse Stimulator is an electrically powered device intended for over the counter use and used to apply an electrical current to electrodes on a user's skin to relieve pain. The proposed device is intended for at home use in delivering electric pulses to tired and sore muscles.
The proposed device consists of controlling host device, output connecting cable and skin electrodes. The proposed device has many safety measures such as output load detection, timing indication and so on. It is adopted battery as the power supply. The output waveform is adopted "square waveform" as the basic wave. It simulates the various output modes by various frequency combination. The waveform is mainly for single rectangular. The output waveform intensity is continuous adjustable. The working time is controlled automatic.
The MDTS100 is adopted the method that inputting the specific low frequency pulse
{5}------------------------------------------------
### Premarket Notification 510(k) Submission-510(k) Summary
current into the human body through the human skin to treat the pain.
The MDTS100 Electronic Pulse Stimulator has seven stimulation modes. These stimulation modes can be selected by pressing the MODE button on the front of the unit.
The MDTS100 comes with 2 independent output channels by pressing the CHA button.
Channel A&B are selected by default. Pressing the & buttons will decrease & increase the intensity of stimulation. LCD screen display the left remaining stimulation time in the lower left corner. The default time is 20 minutes.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for over the counter use.
The device does not contain drug or biological products.
The device is software-driven and software validation is provided in software.
## 3.5 Indication for Use
The Electronic Pulse Stimulator MDTS100 is to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household and work activities.
{6}------------------------------------------------
## 3.6 Comparison list of technical characteristics
| Parameter | Proposed Device | Predicate Device |
|---------------------------------------------------------------------------------------------------|------------------------------------------------|--------------------|
| 510(k) number | (to be assigned) | K122744 |
| Device Name and Model | MDTS100 | PL029 |
| Manufacturer | Beijing Choice Electronic Technology Co., Ltd. | Prospera |
| Power Source | DC 3V, 2 AAA batteries | 3V Battery |
| -Method of Line Current Isolation | N/A for DC current | Battery Supply N/A |
| -patient Leakage Current | -- | -- |
| -Normal Condition( $\mu$ A) | ≤10 $\mu$ A | 2.0 $\mu$ A |
| -Single Fault Condition( $\mu$ A) | ≤50 $\mu$ A | 3.0 $\mu$ A |
| Average DC current through electrodes when device is on but no pulses are being applied( $\mu$ A) | ≤10 $\mu$ A | 0 $\mu$ A |
| Number of Output Modes | 7 | 8 |
| Number of Synchronous or Output Alternating? | Alternating | Alternating |
| Channels Method of Channel Isolation | No | By software |
| Regulated Current or Regulated Voltage? | Regulated Voltage | Voltage control |
| Software/Firmware/Microprocessor Control? | Yes | Yes |
Table 3-1 Basic Unit Characteristics Comparison Table between the Proposed Device (MDTS100) and Predicate Device
{7}------------------------------------------------
## Premarket Notification 510(k) Submission—510(k) Summary
Maximum Output Current (specify units) (+/- 20
%)
| Automatic Overload Trip? | No | No | |
|------------------------------------------------------------------------------------------------------------|--------------------------------------|-----------------------------------------|--------------|
| Automatic No-Load Trip? | Yes | No | |
| Automatic Shut Off? | Yes | Yes | |
| User Override Control? | Yes | Yes | |
| Display: | On/OFF Status? | Yes | Yes |
| | Low Battery? | Yes | No |
| | Voltage/Current Level? | Yes | No |
| Timer Range (minutes) | 20 minutes | 5minutes, 10 minutes | |
| Compliance with Voluntary Standards? | Yes | Yes | |
| Compliance with 21 CFR 898? | Yes | Yes | |
| Weight (lbs., oz.) | 0 lb., 2.2 oz. ( Battery Excluded ) | 0 lb, 8.21oz. | |
| Dimensions (in.) [W x H x D] | 2.17x 5.12x0.87 | 2.24x7.87x0.83 | |
| Housing Materials and Construction | Enclosure: ABS | Enclosure: ABS | |
| Table 3-2 Output Specification Comparison Table between the Proposed Device (MDTS100) and Predicate Device | | | |
| Parameter | Proposed Device | Predicate Device | |
| Device Name and Model | MDTS100 | PL029 | |
| Waveform | pulsed monophasic | Monophasic | |
| Shape | Rectangular | Rectangular | |
| Maximum Output Voltage (volts) (+/- 20 %) | 150V@500 Ω<br>160V @2kΩ<br>165V@10kΩ | 49.6V@500 Ω<br>99.2V@2kΩ<br>114V @10k Ω | |
| | | | |
| | | 16.5mA @10k Ω | 0.6mA @10k Ω |
| Duration of primary (depolarizing) phase (µsec) | | 0µsec | 40msec |
| Pulse Duration (µsec) | | 50~140µsec | 120~6800µsec |
| Frequency (Hz) [or Rate (pps)] | | 0.9Hz~82Hz | 0.5~86Hz |
| Net Charge (microcoulombs (µC) per pulse) @500Ω | | 42µC | 18000µC |
| Maximum Phase Charge, (µC) @500Ω | | 42µC | 23µC |
| Maximum Current Density, (mA/cm2, r.m.s.) | | 3.3mA/cm2@1.57"×1.57"Electrode<br>Pad | 1.4mA/cm2 |
| Maximum Average Current (average absolute value), mA | | 2.8mA | --- |
| Maximum Average Power Density, (W/cm2),<br>(using smallest electrode conductive surface rate) | | 0.02W/cm2@1.57"×1.57"Electrode<br>Pad | 0.23 W/cm2 |
| Burst Mode | (a) Pulses per burst | 197 | 1 |
| | (b) Burst per second | 0.185 | 0-7 |
| | (c) Burst duration (seconds) | 2.4s | 1 |
| | (d) Duty Cycle: Line(b) x Line(c) | 0.44 | 7 |
| ON TIME (seconds) | | <1second | 40ms |
| OFF TIME (seconds) | | <1second | 18ms |
| Additional Features | | --- | --- |
18mA @500 Ω
3.2mA@2k Ω
300mA @500 Ω
80mA@2k Ω
{8}------------------------------------------------
## Premarket Notification 510(k) Submission—510(k) Summary
{9}------------------------------------------------
### Premarket Notification 510(k) Submission-510(k) Summary
Conclusion: The data in Table 3-1 and Table 3-2 indicates that the technical characteristics, features, specifications and intended use of the two devices are very similar. Both systems are designed to relief pain in the muscle areas. Outputs are almost the identical, each can produce square electronic pulses and positive-going waveform output. Both devices have selection modes to allow users to select preset programs and adjust intensity level by pressing the keypad on the main unit. Both devices are transiting pulses by the use of Electrode Pads. Technological characteristics, features, specifications, materials and intended uses of the new device are substantially equivalent to the quoted predicated device. The parameters such as waveform, frequency, pulse width, output current and voltage, although not exactly similar to this particular predicate, they do not raise any new questions of safety or effectiveness. The differences that exist between the devices are insignificant in terms of safety and effectiveness.
{10}------------------------------------------------
## 3.7 Functional and Safety Testing:
Non-clinical tests were performed on the subject matter device in order to validate the design and to assure conformance with the following voluntary design standards in connection with medial device electrical safety, electromagnetic compatibility, and particular requirements for the safety of nerve and muscle stimulator:
### EMC and electrical Safety
IEC 60601-1:2005 Medical electrical equipment-Part 1: General requirements for basic safety and essential performance
IEC 60601-1-2:2007 Medical electrical equipment-Part1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic compatibility-Requirements and tests.
IEC 60601-1-11: 2010 Medical electrical equipment-Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
IEC 60601-2-10: 2012Medical electrical equipment-Part 2-10: Particular requirements for the safety of nerve and muscle stimulator.
### Biocompatibility
The electrode pads are the mainly body contacting parts. Electrode pads, as accessories of the subject matter device, are adopted the Self-adhesive Electrode manufactured by Wuxi Jiajian Medical Instrument Co., Ltd. and have been cleared by FDA in June 23, 2009 as K090198. The patient contacting electrodes comply with ISO 10993 and are considered Class A ≤24 hours contact.
### Software
In addition to the compliance of voluntary standards, the software verification has been carries out according to the FDA Guidance for the Content of Premarket Submission for Software Contained in Medical Devices.
### Cleaning
The cleaning instructions as described in the Instruction Manual have been tested to be sufficient. The test result was presented in the system test report in Performance Testing.
## 3.8 Conclusion
The Electronic Pulse Stimulator MDTS100 has the similar intended use and technological characteristics as the predicate device. Moreover, bench testing and safety report documentation demonstrate that the submitted device could maintain the same safety and
{11}------------------------------------------------
### Premarket Notification 510(k) Submission—510(k) Summary
effectiveness as that of the predicate device. In other words, the differences do not affect the intended use and do not raise any new questions of safety or effectiveness or alter the fundamental scientific technology of the device. Thus, the proposed device MDTS100 is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.