The EleGARD™ Patient Positioning System (EleGARD) is a cardiopulmonary board which may elevate a patient's head and thorax: including during airway management; during manual CPR, manual CPR, adjuncts, CPR with the LUCAS® Chest Compression System or ARM XR Automated Chest Compressor, and patient transport. EleGARD is indicated for adults only when used with the LUCAS Chest Compression System or the ARM XR Automated Chest Compressor. When not used with an automated compression system, EleGARD can be used for adults and children, not including infants and neonates.
Device Story
EleGARD is an electrically powered cardiopulmonary backboard used in hospital and pre-hospital settings by trained medical personnel. It functions as a platform to stabilize and elevate a patient's head and thorax during airway management, manual CPR, or while using automated chest compression systems (LUCAS or ARM XR). The device uses electro-mechanical lifting to adjust and hold the patient's position. By elevating the head and thorax, the device aims to assist in airway management and CPR procedures. It is designed to integrate with specific automated compression backplates, allowing for seamless use during resuscitation efforts. The device is operated by clinicians to improve patient positioning, potentially facilitating better airway access and CPR efficacy.
Clinical Evidence
Bench testing only. Testing verified the mechanical connection, patient positioning, and integration of the Defibtech ARM XR backplate with the EleGARD system. Biocompatibility testing was provided for device coverings per ISO 10993-1. Electrical and EMC performance was established via previous testing (K191689) and deemed unaffected by the modification.
Technological Characteristics
Electrically powered cardiopulmonary backboard; electro-mechanical lifting mechanism; supports manual and automated CPR backplates (LUCAS, ARM XR). Materials include patient-contacting coverings tested per ISO 10993-1. Complies with ANSI/AAMI/ES 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for adults (when used with LUCAS or ARM XR automated chest compressors) or adults and children (excluding infants/neonates) for head and thorax elevation during airway management, manual CPR, CPR adjuncts, or patient transport. Contraindicated for patients >350 lbs or when safe positioning is not possible.
Regulatory Classification
Identification
A cardiopulmonary resuscitation board is a device consisting of a rigid board which is placed under a patient to act as a support during cardiopulmonary resuscitation.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 29, 2022
AdvancedCPR Solutions LLC % Paul Dryden Consultant ProMedic Consulting LLC 131 Bay Point Dr NE Saint Petersburg, Florida 33704
Re: K221901
Trade/Device Name: EleGARD Regulation Number: 21 CFR 880.6080 Regulation Name: Cardiopulmonary Resuscitation Board Regulatory Class: Class I Product Code: FOA Dated: June 28, 2022 Received: June 30, 2022
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Wolloscheck, Ph.D. For Payal Patel Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221901
Device Name
EleGARD™ Patient Positioning System
#### Indications for Use (Describe)
The EleGARD™ Patient Positioning System (EleGARD) is a cardiopulmonary board which may elevate a patient's head and thorax: including during airway management; during manual CPR, manual CPR, adjuncts, CPR with the LUCAS® Chest Compression System or ARM XR Automated Chest Compressor, and patient transport. EleGARD is indicated for adults only when used with the LUCAS Chest Compression System or the ARM XR Automated Chest Compressor. When not used with an automated compression system, EleGARD can be used for adults and children, not including infants and neonates.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared: | 28-Jul-22 |
|----------------------------|----------------------------------------------------------------|
| AdvancedCPR Solutions LLC | |
| 5201 Eden Ave., Suite 300 | |
| Edina, MN 55436 | |
| Tel – 763.259.3722 | |
| Official Contact: | Philip Faris - CEO |
| Proprietary or Trade Name: | EleGARD™ Patient Positioning System |
| Common/Usual Name: | Cardiopulmonary board |
| Classification Name: | FOA - Cardiopulmonary resuscitation board (21<br>CFR 880.6080) |
| Predicate Device: | K191689 - AdvancedCPR Solutions LLC |
| Reference Device | K211289 - Defibtech RMU 2000 |
## Device Description:
The EleGARD™ patient positioning system is an electrically powered device on which a patient is placed. It can be used to elevate a patient's head and thorax during a number of different procedures, e.g., airway management and different CPR techniques.
The EleGARD™ acts as a back board for airway management and for performing CPR. It allows the user to position the patient in supine or with the head and thorax elevated. The EleGARD™ is a cardiopulmonary back board which is placed under the patient.
There are several different CPR methods which use a back board, stretcher or bed, e.g., manual CPR, manual CPR Adjuncts, and with the Defibtech Chest Compression System and ARM XR backplate or with the LUCAS Chest Compression System as previously cleared under K191689.
### Indications for Use:
The EleGARD™ Patient Positioning System (EleGARD) is a cardiopulmonary board which may elevate a patient's head and thorax: including during airway management; during manual CPR, manual CPR adjuncts, CPR with the LUCAS® Chest Compression System or ARM XR Automated Chest Compressor; and patient transport. EleGARD is indicated for adults only when used with the LUCAS Chest Compression System or the ARM XR Automated Chest Compressor. When not used with an automated compression system, EleGARD can be used for adults and children, not including infants and neonates.
### Patient Population:
Patients who may benefit from elevation of the head and neck, including those patients in need of airway management, elevation of the head, and those undergoing CPR. Specifically, when used with the automated compression systems the population is limited to adults and when not used with an automated compression system, EleGARD can be used for adults and children, not including infants and neonates.
{4}------------------------------------------------
# 510(k) Summary
# Contraindications:
It is recommended that EleGARD™ no be used under the following conditions:
- It is recommended that EleGARD™ no be used under the following conditions:
•
- If the patient weighs more than 350 pounds .
# Environments of Use:
Hospital and pre-hospital
{5}------------------------------------------------
# 510(k) Summary
# Table 1 - Table of the Similarities and Differences of Predicate vs. Proposed Device
| | Proposed | Predicate | Comments |
|---------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| | EleGARD™ Patient Positioning System with Defibtech | EleGARD™ Patient Positioning Systems with | |
| | ARM XR backplate | LUCAS backplate<br>K191689 | |
| Procode / Classification /<br>CFR | FOA - cardiopulmonary board<br>CFR 880.6080 | FOA - cardiopulmonary board<br>CFR 880.6080 | Similar |
| Indications for Use | The EleGARD™ Patient Positioning System (EleGARD) is a<br>cardiopulmonary board which may elevate a patient's head and<br>thorax: including during airway management; during manual<br>CPR, manual CPR adjuncts, CPR with the LUCAS® Chest<br>Compression System or ARM XR Automated Chest<br>Compressor; and patient transport. EleGARD is indicated for<br>adults only when used with the LUCAS Chest Compression<br>System or the ARM XR Automated Chest Compressor. When<br>not used with an automated compression system, EleGARD<br>can be used for adults and children, not including infants and<br>neonates. | The EleGARD™ Patient Positioning System is<br>cardiopulmonary board which may elevate a<br>patient's head and thorax including during airway<br>management; during manual CPR, manual CPR<br>adjuncts, CPR with the LUCAS Chest<br>Compression Systems; and patient transport | Similar |
| Prescriptive | Trained medical personnel | Trained medical personnel | Similar |
| Environments of use | Hospital and pre-hospital | Hospital and pre-hospital | Similar |
| Technology | Method to elevate the head and torso via manual and electro-<br>mechanical lifting and hold in place | Method to elevate the head and torso via manual<br>and electro-mechanical lifting and hold in place | Similar |
| Used with Automatic<br>Compression /<br>Decompression Systems | Yes, adding Dibetech ARM XR® | Yes with LUCAS | Similar, but different<br>Automatic Compression /<br>Decompression Systems |
| Performance | From K191689<br>IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-12<br>Cleaning | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-12<br>Cleaning | Similar |
| Contraindications | It is recommended that EleGARD™ not be used under the<br>following conditions:<br>• When it is not possible to position the patient safely or<br>correctly on the EleGARD<br>• If the patient weighs more than 350 pounds | It is recommended that EleGARD™ not be used<br>under the following conditions:<br>• When it is not possible to position the patient<br>safely or correctly on the EleGARD<br>• If the patient weighs more than 350 pounds | Similar |
{6}------------------------------------------------
# Substantial Equivalence Rationale and Discussion of Differences
The EleGARD™ is viewed as substantially equivalent to the predicate device because:
## Indications -
The EleGARD™ is designed to elevate a patient's head and thorax during airway management, various modalities of CPR and patient transport.
Discussion - The indications for use are similar to the predicate K191689 and within the intended use of the accessory backplate of the reference K211289. The patient population was clarified in the indications for to be consistent with and without use of automatic compression systems. There were no differences between the predicate and subject device.
## Environment of Use -
All devices have the same environments of use: hospital to be used by trained medical personnel
Discussion - The environments of use and personnel are similar to the predicate device.
## Technology -
The technology is to stabilize the body position and to elevate the patient's head and thorax during CPR. The positioning of the head and thorax is similar to the predicate. The only modification is the option of using the Defibtech ARM XR backplate, K211289, for the LUCAS backplate to connect to the applicable device.
Discussion – There is no change in technology between the proposed device and the predicate and the option of the Defibtech ARM XR backplate is similar to that of the LUCAS backplate. There are no technological differences in the backplates and their use with the EleGARD™ that would raise different risk concerns.
# Non-clinical Testing Summary
# Biocompatibility of Materials -
There are no direct patient contacting materials of the EleGARD™
Discussion - The sponsor offers 2 different coverings, one is identical to K191689 and the other has supportive ISO 10993-1 testing.
# Electrical, EMC, EMI testing -
Testing was performed for K191689. The modification has not effect on this performance. K191689 included testing per ANSI/AAMI/ES 60601-1 and IEC 60601-1-2, and the device performed as intended and met the requirements.
### Bench testing -
Testing and verification was performed to evaluate the ability to connect the Defibtech ARM XR backplate with the EleGARD™. This testing included Patient Position, connection to Optional Associate Equipment, Carrying, a review of any different Risk Mitigation, there were none, and the ability to fit within the EleGARD™ and that the placement of the Defibtech ARM XR backplate can be attached to the main ARM XR unit.
Discussion - Upon completion of the tests, it was found to meet its performance requirements.
# Substantial Equivalence Conclusion -
The proposed addition of the Defibtech ARM XR backplate does not raise different questions of safety compared to the predicate and thus can be found substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.