The EleGARD™ Patient Positioning System is cardiopulmonary board which may elevate a patient's head and thorax including during airway management; during manual CPR, manual CPR adjuncts, CPR with the LUCAS Chest Compression Systems; and patient transport.
Device Story
EleGARD is an electrically powered cardiopulmonary backboard used in hospital and pre-hospital settings by trained medical personnel. The device is placed under a patient to provide a stable platform for airway management and CPR. It features an electro-mechanical system to elevate the patient's head and thorax from a supine position. It is compatible with manual CPR, manual CPR adjuncts, and the LUCAS Chest Compression System. The device assists healthcare providers in positioning patients during emergency procedures and transport, potentially improving clinical outcomes by facilitating head and thorax elevation during resuscitation efforts.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility (ISO 10993-1: cytotoxicity, sensitization, irritation), electrical safety (ANSI/AAMI/ES 60601-1), EMC (IEC 60601-1-2), and EMS environment safety (IEC 60601-1-12). Functional performance testing verified the raising/lowering mechanism, software verification, and compatibility with the LUCAS backplate.
Technological Characteristics
Electrically powered cardiopulmonary backboard. Materials evaluated per ISO 10993-1. Standards met: ANSI/AAMI/ES 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-1-12 (EMS environment). Features electro-mechanical lifting mechanism for head/thorax elevation. Compatible with LUCAS Chest Compression System.
Indications for Use
Indicated for patients requiring elevation of the head and neck, including those needing airway management or undergoing CPR. Contraindicated if the patient cannot be positioned safely or weighs over 350 pounds.
Regulatory Classification
Identification
A cardiopulmonary resuscitation board is a device consisting of a rigid board which is placed under a patient to act as a support during cardiopulmonary resuscitation.
Predicate Devices
Minnesota Resuscitation (now AdvancedCPR Solutions) Heads Up Platform
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 13, 2020
Minnesota Resuscitation Solutions, d.b.d AdvancedCPR Solutions c/o Paul Dryden Manager ProMedic LLC 131 Bay Point Dr. NE St. Petersburg, Florida 33704
Re: K191689
Trade/Device Name: EleGARD Patient Positioning System Regulation Number: 21 CFR 880.6080 Regulation Name: Cardiopulmonary Resuscitation Board Regulatory Class: Class I. reserved Product Code: FOA, Dated: January 14, 2020 Received: January 15, 2020
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Geeta Pamidimukkala Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K191689
Device Name
EleGARD™ Patient Positioning System
Indications for Use (Describe)
The EleGARD™ Patient Positioning System is cardiopulmonary board which may elevate a patient's head and thorax including during airway management; during manual CPR, manual CPR adjuncts, CPR with the LUCAS Chest Compression Systems; and patient transport.
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17) 443-6740 EF
Page 1 of 1
SC Publishing Services (301)
{3}------------------------------------------------
### 510(k) Summary K191689
#### Date Prepared: 13-Feb-2020
#### I Submitter
Minnesota Resuscitation Solutions, d.b.d AdvancedCPR Solutions
5201 Eden Ave., Suite 300
Edina, MN 55436
Tel – 763.259.3722
| Submitter Contact: | Philip Tetzlaff, CTO |
|---------------------------|----------------------------|
| Submission Correspondent: | Paul Dryden, ProMedic, LLC |
II Device
| Proprietary or Trade Name: | EleGARD™ Patient Positioning System |
|----------------------------|------------------------------------------------------|
| Common/Usual Name: | Cardiopulmonary resuscitation board |
| Classification Name: | Cardiopulmonary resuscitation board (21CFR 880.6080) |
| Regulatory Class: | I |
| Product Code: | FOA |
III Predicate Device
| | Minnesota Resuscitation (now AdvancedCPR Solutions)<br>Heads Up Platform - Class I exempt, FOA, 21CFR 880.6080 |
|--|----------------------------------------------------------------------------------------------------------------|
|--|----------------------------------------------------------------------------------------------------------------|
#### IV Device Description
The EleGARD™ patient positioning system is an electrically powered device on which a patient is placed. It can be used to elevate a patient's head and thorax during a number of different procedures, e.g., airway management and different CPR techniques.
The EleGARD™ acts as a back board for airway management and for performing CPR. It allows the user to position the patient in supine or with the head and thorax elevated. The EleGARD™ is a cardiopulmonary back board which is placed under the patient.
It may be used with the patient lying flat or may elevate the patient's thorax and head. The device would be used in emergency conditions when healthcare professionals need to: provide airway management, perform CPR with manual / hand compressions, adjunctive manual CPR devices, and a LUCAS mechanical compression device, perform manual bag / mask ventilation, and transport a patient.
#### V Indications for Use
The EleGARD™ Patient Positioning System is a cardiopulmonary board which may elevate a patient's head and thorax: including during airway management; during manual CPR, manual CPR adjuncts, CPR with the LUCAS Chest Compression Systems; and patient transport.
{4}------------------------------------------------
## Patient Population:
Patients who may benefit from elevation of the head and neck, including those patients in need of airway management, elevation of the head, and those undergoing CPR.
### Contraindications:
It is recommended that EleGARD™ not be used under the following conditions:
- When it is not possible to position the patient safely or correctly on the EleGARD™ .
- If the patient weighs more than 350 pounds
#### Environments of Use:
Hospital and pre-hospital
#### VII Comparison of Technological Characteristics and Performance with the Predicate
### Table 1 - Table of the Similarities and Differences of Predicate vs. Proposed Device
| | Proposed<br>EleGARD™ Patient<br>Positioning | Predicate<br>Head Up Platform |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Procode | FOA - cardiopulmonary resuscitation<br>board<br>CFR 880.6080 | FOA - cardiopulmonary resuscitation<br>board<br>CFR 880.6080 |
| Indications for Use | The EleGARD™ Patient Positioning<br>System is cardiopulmonary board<br>which may elevate a patient's head<br>and thorax including during airway<br>management; during manual CPR,<br>manual CPR adjuncts, CPR with the<br>LUCAS Chest Compression Systems;<br>and patient transport | Intended to assist in elevating the head<br>and shoulders of a patient from a<br>supine position to approximately 30<br>degrees and may be used during<br>various procedures, i.e., airway<br>management procedures, CPR, etc. |
| Prescriptive | Trained medical personnel | Trained medical personnel |
| Environments of use | Hospital and pre-hospital | Hospital and pre-hospital |
| Technology | Method to elevate the head and torso<br>via manual and electro-mechanical<br>lifting and hold in place | Manual mechanical means to elevate<br>the head and torso and hold in place |
| Performance | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-1-12<br>Cleaning of patient contacting surfaces<br>Compatibility of Lucas backplate<br>connection | Device was not electrical |
| Contraindications | It is recommended that EleGARD™<br>not be used under the following<br>conditions:<br>• When it is not possible to position<br>the patient safely or correctly on the<br>EleGARD<br>• If the patient weighs more than 350<br>pounds | It is recommended that EleGARD™<br>not be used under the following<br>conditions:<br>• When it is not possible to position<br>the patient safely or correctly on the<br>EleGARD<br>• If the patient weighs more than 350<br>pounds |
| Patient contacting<br>materials | Evaluated via ISO 10993-1 | Evaluated via ISO 10993-1 |
{5}------------------------------------------------
# Substantial Equivalence Rationale
The EleGARD™ is viewed as substantially equivalent to the predicate device because:
### Indications -
- . Both the EleGARD™ and predicate device are designed to elevate a patient's head and thorax during airway management, various modalities of CPR and patient transport. The differences in the indications is that the subject device can be used with the LUCAS chest compression system. Performance testing was completed to support this change in indication.
### Environment of Use -
- All devices have the same environments of use: hospital and pre-hospital to be used by trained medical personnel
### Technology -
- The technology is to stabilize the body position and to elevate the patient's head and . thorax during CPR. The positions of the head and thorax are similar but the means to elevate is different between the proposed EleGARD™ and the predicate Head Up Platform.
- The difference is electro-mechanical elevations vs. manual elevation. The practical nature of having electro-mechanical assistance in place of manual mechanical spring strut to elevate the patient when needed has been evaluated and does not raise any new concerns for safety or effectiveness that are not typical of other electro-mechanical devices.
#### VIII Performance Data
### Biocompatibility of Materials -
The patient contacting materials of the EleGARD™ have been evaluated according to ISO 10993-1. Testing included by Cytotoxicity, Sensitization and Irritation.
# Electrical, EMC, EMI testing -
We have evaluated the proposed device per ANSI/AAMI/ES 60601-1. IEC 60601-1-2, and IEC 60601-1-12.
The proposed device met the requirements of the standards.
# Bench testing -
The proposed device was tested for functional performance of raising and lowering the device to the specified positions; Electrical safety ( ES160601-1); EMC (IEC60601-1-2); and Safety for use in EMS environments (IEC 60601-1-12), software verification and system verification (Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices) and compatibility of the backplate connection.
#### IX Conclusion
The sponsor has demonstrated through performance testing, design and features to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.