The LensHooke CA0 Sperm Separation Device and the LensHooke CA1 Sperm Separation Device are intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.
Device Story
LensHooke CA0 and CA1 are single-use, radiation-sterilized sperm separation devices; used in clinical laboratory settings to prepare motile sperm from liquefied semen samples. Device consists of a base and filter layer(s) (CA0: one layer; CA1: two layers). Process involves adding 1 mL liquefied semen to base, placing filter(s), and adding sperm washing medium via filter port. Device incubated at 37°C for 30 minutes; motile sperm swim up and across filter(s) into separation medium. Separated sperm collected from filter port for clinical use. Output provides improved motile and progressively motile sperm concentrations for assisted reproduction procedures; benefits patient by isolating high-quality sperm for fertilization.
Clinical Evidence
Bench testing only. Performance evaluated via Human Sperm Survival Assay (HSSA) and sperm separation efficiency. HSSA confirmed ≥80% control motility at 24 hours post-exposure. Sperm separation performance demonstrated average total motility improvement of 24.4% (CA0) and 25.9% (CA1), and progressive motility improvement of 26.6% (CA0) and 27.1% (CA1) compared to input samples. Sterilization validated per ISO 11137-1; package integrity per ASTM F1980-21, F1929-15, and F88/F88M-15; endotoxin levels <20 EU/device per USP<85> and ANSI/AAMI ST72.
Technological Characteristics
Polycarbonate construction; single-use; radiation-sterilized (SAL 10^-6). CA0 features one filter layer; CA1 features two filter layers. Operates via passive migration (swim-up) of motile sperm through microporous filter(s) at 37°C. Shelf-life 24 months. Standards: ISO 11137-1 (sterilization), ISTA 3A (transport), ASTM F1980-21 (aging), ASTM F1929-15 (dye penetration), ASTM F88/F88M-15 (seal strength), USP<85>/ANSI/AAMI ST72 (endotoxin).
Indications for Use
Indicated for preparing motile sperm from semen for use in the treatment of infertile couples undergoing ICSI, IVF, or IUI procedures.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Predicate Devices
ZyMōt Multi Sperm Separation Device (850µl, 3 mL) (K173075)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
October 31, 2022
Bonraybio Co., LTD. % Feng-Yu Lee Principal Regulatory Consultant IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212 San Juan Capistrano, CA 92675
Re: K221810
> Trade/Device Name: LensHooke CA0 Sperm Separation Device LensHooke CA1 Sperm Separation Device Regulation Number: 21 CFR§ 884.6160 Regulation Name: Assisted reproduction labware Regulatory Class: II Product Code: MOK Dated: September 28, 2022 Received: September 29, 2022
Dear Feng-Yu Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Monica D. Garcia, Ph.D. Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221810
### Device Name
LensHooke CA0 Sperm Separation Device LensHooke CA1 Sperm Separation Device
### Indications for Use (Describe)
The LensHooke CA0 Sperm Separation Device and the LensHooke CA1 Sperm Separation Device are intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K221810 Page 1 of 4
Image /page/3/Picture/1 description: The image contains the logo for Bonraybio. The logo consists of a teal-colored circle with a white plant-like design inside, followed by the text "bonraybio" in a sans-serif font. The text is in a dark gray color and is positioned to the right of the circular icon.
# 510(K) Summary K221810
#### Submitter/Correspondent: 1
- 1.1 Submitter:
BonrayBio Co., Ltd. 4F., No.118, Gongye 9th Rd., Dali Dist., Taichung City 41280, Taiwan Contact Person: Clare Huang TEL: +886-4-24912385 FAX: +886-4-24912885
- 1.2 Submission Correspondent:
IVDD Regulatory Consultant 29122 Rancho Viejo Road, Suite 212, San Juan Capistrano, CA 92675 Contact Person: Feng-Yu Lee TEL: 1-949-218-0929 FAX: 1-949-218-0928
- 1.3 Date Summary Prepared: October 27, 2022
## 2 Subject Device:
| Proprietary Name: | LensHooke CA0 Sperm Separation Device<br>LensHooke CA1 Sperm Separation Device |
|--------------------|--------------------------------------------------------------------------------|
| Common Name: | Sperm Separation Device |
| Regulation Name: | Assisted Reproduction Labware |
| Regulation Number: | 884.6160 |
| Product Code: | MQK |
| Regulatory Class: | II |
#### 3 Predicate Device:
ZyMōt Multi Sperm Separation Device (850µl, 3 mL) Device Company: DxNow, Inc. 510(k) Number: K173075
The predicate device has not been subject to a design-related recall.
#### 4 Device Description
The LensHooke CA0 Sperm Separation Device and LensHooke CA1 Sperm Separation Device are used to prepare motile sperm from semen based on the mobility of motile sperm.
The LensHooke Sperm Separation Device has a base and filter layers to separate and process
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for "bonraybio". The logo consists of a circular icon on the left and the company name on the right. The icon features a stylized plant or leaf design in a light blue color, enclosed within a circle of the same color. The company name, "bonraybio", is written in a sans-serif font in a dark gray color.
the semen sample. The LensHooke Sperm Separation Device has two models, CA0 with one filter layer and CA1 with two filter layers. Liquefied semen (1 mL) is added to the base and sperm washing medium is added to the filter via a filter port. After incubation, the separated sperm are collected from the filter layer via the filter port.
The subject devices are radiation-sterilized with a sterility assurance level (SAL) of 10 °. The LensHooke Sperm Separation Devices are individually packaged and for single-use only.
# 5 Indications for Use
The LensHooke CA0 Sperm Separation Device and the LensHooke CA1 Sperm Separation Device are intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.
# 6 Comparison of Intended Use and Technological Characteristics with the Predicate Device:
| | LensHooke® CA0 and CA1<br>Sperm Separation Device<br>(Subject Device) | ZyMōt Multi Sperm Separation Device<br>(850 µL, 3 mL)<br>(Predicate Device - K173075) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The LensHooke CA0 Sperm Separation Device<br>and the LensHooke CA1 Sperm Separation<br>Device are intended for preparing motile<br>sperm from semen for use in the treatment<br>of infertile couples by intracytoplasmic sperm<br>injection (ICSI), in vitro fertilization (IVF) and<br>intrauterine insemination (IUI) procedures. | The ZyMōt Multi Sperm Separation<br>Device is intended for preparing motile<br>sperm from semen for use in the<br>treatment of infertile couples by<br>intracytoplasmic sperm injection (ICSI),<br>in vitro fertilization (IVF) and<br>intrauterine insemination (IUI)<br>procedures. |
| Design | A disposable device containing a base, filter<br>layer(s), and a port on the filter for media<br>loading and sperm collection after<br>processing.<br>- CA0 with one filer layer<br>- CA1 with two filter layers | ZyMōt Multi – A disposable culture<br>dish containing a separation chamber<br>and an inlet port. The separation<br>chamber has a lower sample chamber<br>and an upper collection chamber. The<br>two chambers are separated by a<br>microporous filter. |
| Mechanism of<br>action | Liquefied semen is added to the base. The<br>filter(s) are placed on the base and<br>separation medium added through the layer<br>port.<br>The loaded device is incubated at 37°C for<br>30 minutes to allow motile sperm to swim up<br>and across the filter(s) to migrate into the<br>separation medium in the upper filter layer. | ZyMōt Multi - The semen is added to<br>the inlet port to fill the lower sample<br>chamber; then, the separation<br>medium is added to the upper<br>collection chamber.<br>The loaded device is incubated at 37°C<br>for 30 minutes to allow motile sperm<br>to swim up and cross the filter to<br>migrate into the over-laying<br>separation medium in the upper<br>collection chamber |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Bonraybio. The logo consists of a circular icon on the left and the company name "bonraybio" on the right. The icon features a stylized wheat stalk inside a light blue circle, while the company name is written in a simple, sans-serif font.
| Material | Polycarbonate | Polymethylmethacrylate, polycarbonate, flash-spun high-density polyethylene fibers |
|---------------------------------|------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Human Sperm<br>Survival Testing | ≥80% of the control motility at 24 hours after exposure for 30 minutes | ≥80% of the control motility at 24 hours after exposure for 30 minutes |
| Endotoxin | ≤20 EU/device | ≤2.5 EU/device |
| Sterile | Yes | Yes |
| Shelf-Life | 24 months | 12 months |
The subject and predicate devices have the same indications for use and intended use. The subject device and predicate device have the same fundamental design incorporating a chamber for loading semen, a filter for motile sperm to migrate through, and a port for collection of motile sperm. However, there are technological differences between the subject and predicate devices, including the number of filters, materials, endotoxin specifications, and shelf-life duration). These differences do not raise different questions of safety and effectiveness.
#### 7 Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows :
The following testing has been performed on the LensHooke CAO and CA1 Sperm Separation Devices:
- Sterilization Validation Study per ISO 11137-1:2006
- Transportation Simulation Study per ISTA 3A (2018)
- Package integrity testing following accelerated aging per ASTM F1980-21 to support a 2-year shelf-life:
- o Dye penetration testing per ASTM F1929-15
- Seal strength testing Per ASTM F88/F88M-15 O
- Endotoxin testing per USP<85> and ANSI/AAMI ST72:2019: <20 EU/device ●
- Human Sperm Survival Assay (HSSA) before and after accelerated aging to support a 2-year shelf life:
- Donor sperm were exposed to the subject devices for 30 minutes and then, O incubated at room temperature for 24 hours. The rate of motile sperm after incubation was compared to that of the control (no exposure to the subject device). The acceptance criterion is ≥80% of the control motility at 24 hours after exposure for 30 minutes
- Sperm Separation Performance: ●
- Each version of the subject device was used to separate motile sperm from O donor semen samples. The separation procedures followed the Instructions for Use, and the percentage of motile sperm and progressively motile sperm in the output samples were compared to those of the input samples. The results showed an average improvement in total motility of 24.4% and 25.9% for the CA0 and CA1 devices, respectively. The results showed an average improvement in progressive motility of 26.6% and 27.1% for the CA0 and CA1 devices, respectively.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for "bonraybio". The logo consists of a teal-colored circular icon with a stylized plant or wheat stalk inside, followed by the company name "bonraybio" in a simple, sans-serif font. The text is in a dark gray or black color, providing a clear contrast against a white background.
#### 8 Conclusions
The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.