The ZyMot ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures. The ZyMot Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.
Device Story
Devices (ZyMōt ICSI and ZyMōt Multi) prepare motile sperm from semen for ART procedures; operate via sperm motility (swim-up) principle; used in clinical laboratory settings by embryologists/technicians. ZyMōt ICSI: 5 micro-channels; semen loaded into inlet; motile sperm collected from outlet after 30-minute incubation at 37°C. ZyMōt Multi: 850µl or 3ml volumes; semen added to lower chamber; separation medium added to upper chamber; microporous filter separates chambers; motile sperm swim through filter into upper chamber during 30-minute incubation at 37°C. Output is a concentrated sample of motile sperm for clinical use in fertilization procedures; enhances sperm quality for ART; single-use, radiation-sterilized.
Clinical Evidence
Bench testing only. Sterilization validated per ISO 11137-2:2012. Shelf-life supported by accelerated aging (ASTM F1980-07), dye penetration (ASTM F1929-98), and seal strength (ASTM F88/F88M-15). Endotoxin levels ≤2.15 EU/device (USP<85). Human sperm survival assay (HSSA) confirmed ≥80% motility compared to control after 24 hours. Performance testing on donor semen showed significant increases in motile and progressively motile sperm percentages post-separation across all device versions.
Indicated for infertile couples undergoing assisted reproductive technology (ART) procedures, specifically ICSI, IVF, or IUI, requiring the preparation of motile sperm from semen samples.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 8, 2018
DxNow, Inc. Kevin Sly Senior Advisor to DxNow, Inc. 401 Professional Drive, Suite 130 Gaithersburg, Maryland 20879-3429
Re: K173075
Trade/Device Name: ZyMot ICSI Sperm Separation Device, ZyMot Multi Sperm Separation Device (850µl, 3ml) Regulation Number: 21 CFR 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: Class II Product Code: MQK Dated: February 6, 2018 Received: February 6, 2018
Dear Kevin Sly:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K173075
Device Name ZyMōt ICSI Sperm Separation Device ZyMot Multi Sperm Separation Device (850ul, 3ml)
### Indications for Use (Describe)
The ZyMot ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures.
The ZyMot Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (IVF) and intrauterine insemination (IUI) procedures.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary K173075
#### 1. Submitter
DxNow, Inc. 401 Professional Drive, Suite 130 Gaithersburg, MD 20879-3429 Contact: Bill Sharp, CEO Email: bsharp@dxnowinc.com Phone: 240-454-9893
#### 2. Correspondent
Kevin Sly Senior Advisor, DxNow, Inc. Email: ksly@dxnowinc.com Phone: 240-751-6521
#### 3. Date Prepared: March 6, 2018
#### 4. Device Identification:
| Trade name: | ZyMōt ICSI Sperm Separation Device<br>ZyMōt Multi Sperm Separation Device (850µl, 3ml) |
|----------------------|----------------------------------------------------------------------------------------|
| Common name: | Sperm Separation Device |
| Classification name: | Assisted Reproduction Labware (21 CFR 884.6160) |
| Product code: | MQK (Labware, Assisted Reproduction) |
| Regulatory class: | II |
#### ട. Predicate Device
Qualis (K133295). This predicate device has not been subject to any design related recalls.
#### 6. Device Description
The ZyMōt ICSI Sperm Separation Device and ZyMōt Multi Sperm Separation Device (850μl, 3ml) are used to prepare motile sperm for assisted reproductive technology (ART) procedures. They separate motile sperm from the semen based on the mobility (i.e., swimup nature) of motile sperm.
{4}------------------------------------------------
The ZyMōt ICSI Sperm Separation Device has five micro-channels, each accommodating 2ul of semen and processes up to five separations from one semen sample. Each channel has an inlet port for applying the semen sample and an outlet port for collecting the motile sperm. The ports are connected by a fluid-filled micro-channel in which separation occurs.
The ZyMōt Multi Sperm Separation Device has two processing volumes, 850µl and 3ml. Each version has an inlet port that communicates with the lower sample chamber. The sample chamber is separated from the upper collection chamber by a microporous filter. Untreated semen is added through the inlet port. After incubation, the separated motile sperm are collected from the upper chamber through the outlet port.
The subject devices are radiation-sterilized devices with a sterility assurance level (SAL) of 10-6. They are individually packaged and for single-use only.
#### 7. Indications for Use
The ZyMōt ICSI Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI) procedures.
The ZyMōt Multi Sperm Separation Device is intended for preparing motile sperm from semen for use in the treatment of infertile couples by intracytoplasmic sperm injection (ICSI), in vitro fertilization (IVF) and intrauterine insemination (IUI) procedures.
| Device/Predicate<br>Device(s): | K173075 (subject devices) | K133295 (predicate device) |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The ZyMōt ICSI is intended for preparing<br>motile sperm from semen for use in the<br>treatment of infertile couples by<br>intracytoplasmic sperm injection (ICSI)<br>procedures.<br>The ZyMōt Multi is intended for preparing<br>motile sperm from semen for use in the<br>treatment of infertile couples by<br>intracytoplasmic sperm injection (ICSI), in<br>vitro fertilization (IVF) and intrauterine<br>insemination (IUI) procedures. | The Qualis is intended for preparing<br>motile human sperm for use in the<br>treatment of infertile couples by<br>intracytoplasmic sperm insertion<br>(ICSI) fertilization. |
| Design | ZyMōt ICSI – A disposable culture dish<br>containing five micro-channels, each<br>connected to an inlet port and an outlet<br>port<br>ZyMōt Multi – A disposable culture dish<br>containing a separation chamber and an<br>inlet port. The separation chamber has a<br>lower sample chamber and an upper | A disposable culture dish with four<br>chambers connected by micro-<br>channels |
#### 8. Comparison of Intended Use and Technological Characteristics with the Predicate Device
{5}------------------------------------------------
| | collection chamber. The two chambers<br>are separated by a microporous filter. | |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanism of<br>action | ZyMōt ICSI — The inlet port is preloaded<br>with sperm separation medium and then<br>loaded with semen. After incubation at<br>37°C for 30 min, the motile sperm are<br>collected from the outlet port.<br>ZyMōt Multi – The semen is added to the<br>inlet port to fill the lower sample<br>chamber; then, the separation medium is<br>added to the upper collection chamber.<br>The loaded device is incubated at 37°C for<br>30 min to allow motile sperm to swim up<br>and cross the filter to migrate into the<br>over-laying separation medium in the<br>upper collection chamber. | The semen sample is placed in<br>Chamber A and separation medium<br>is placed in Chamber B. Fluids from<br>both chambers flow via the micro-<br>channels into the central micro-<br>channel where the two fluids pass<br>side-by-side in laminar flow. Motile<br>sperm can swim across the interface<br>of the laminar flow streams and<br>pass into the separation medium<br>stream but non-motile sperm and<br>debris cannot. Motile sperm that<br>cross into the separation medium<br>flow are carried into Chamber C<br>where they are collected. Non-<br>motile sperm and debris remain in<br>the semen sample flow from<br>Chamber A into Chamber D. |
| Material | ZyMōt ICSI – Polymethylmethacrylate,<br>borosilicate glass, flash-spun high-density<br>polyethylene fibers<br>ZyMōt Multi - Polymethylmethacrylate,<br>polycarbonate, flash-spun high-density<br>polyethylene fibers | Cyclo-olefin polymer |
The subject and predicate devices have different Indications for Use statements; however, they have the same intended use – preparing motile sperm from semen for use in ART procedures.
The subject and predicate devices have the same fundamental design incorporating a culture dish with an inlet port/chamber for loading of semen and an outlet port/chamber for collection of motile sperm. Both devices rely on sperm motility to separate motile sperm from non-motile sperm or non-sperm cells. The predicate device incorporates microfluidic technology, whereas the subject devices do not. However, this difference does not raise different questions of safety and effectiveness as compared to the predicate device, as both devices (and other cleared devices) have the same fundamental mode of action (i.e., using motility of sperm to prepare pools of motile sperm for use in ART procedures). In addition, the subject and predicate devices have different materials. However, differences in materials do not raise different questions of safety and effectiveness as compared to the predicate device.
#### g. Summary of Non-Clinical Performance Testing:
The following studies have been performed to support substantial equivalence to the predicate devices:
{6}------------------------------------------------
- . Sterilization validation study per ISO 11137-2:2012
- . Transportation simulation study per ASTM D5276-98
- . Package integrity testing following accelerated aging per ASTM F1980-07 to support a 12-month shelf-life:
- * Dye penetration testing per ASTM F1929-98
- * Seal strength testing per ASTM F88/F88M-15
- Endotoxin testing per USP<85>: ≤2.15 EU/device .
- . Human sperm survival assay (HSSA) before and after accelerated aging to support a 12-month shelf-life:
Donor sperm were exposed to the subject devices for 30 minutes and then, incubated at room temperature for 24 hours. The rate of motile sperm after incubation was compared to that of the control (no exposure to the subject device). The acceptance criterion is ≥80% of the control motility at 24 hours after exposure for 30 minutes.
- Performance testing: .
Each version of the subject device was used to separate motile sperm from donor semen samples. The separation procedures followed the Instructions for Use, and the percentage of motile sperm and progressively motile sperm in the output samples were compared to those of the input samples. A summary of the results is provided in the table below:
| Device | Percentage of motile sperm<br>(before/after separation) | Percentage of progressively motile<br>sperm (before/after separation) |
|-------------------|---------------------------------------------------------|-----------------------------------------------------------------------|
| ZyMōt ICSI | 29.7% / 87% | 26.3% / 74.2% |
| ZyMōt Multi 850µl | 67% / 83.6% | 44.7% / 68.6% |
| ZyMōt Multi 3ml | 67% / 86.5% | 44.7% / 75.4% |
#### 10. Conclusion
The subject and predicate devices have the same intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.