K221672 · Creo Medical, Ltd. · NEY · Feb 14, 2023 · General, Plastic Surgery
Device Facts
Record ID
K221672
Device Name
Creo Electrosurgical System with NP1 Instrument
Applicant
Creo Medical, Ltd.
Product Code
NEY · General, Plastic Surgery
Decision Date
Feb 14, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NP1 MicroBlate Fine Instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue. The NP1 MicroBlate Fine Instrument is not intended for use in cardiac procedures. The Electrosurgical Generator provides microwave (MW) energy to the compatible Creo Medical AB1 MicroBlate Flex and NP1 MicroBlate Fine instruments, intended for coagulation (ablation) of soft tissue. The Electrosurgical Generator provides microwave (MW) energy to the compatible Creo Medical HS1 SlypSeal Instrument, intended for coagulation (hemostasis and cauterization) of soft tissue. The Electrosurgical Generator provides microwave (MW) and radiofrequency (RF) energy to the compatible Creo Medical RS2 Speedboat Instrument, intended for coagulation (hemostasis and cauterization) and cutting soft tissue. The Electrosurgical System is not intended for use in cardiac procedures.
Device Story
System comprises electrosurgical generator, footswitch, interface cable, and NP1 instrument; used for soft tissue coagulation/ablation. Generator delivers 5.8 GHz microwave energy to NP1 instrument via bipolar electrodes; dielectric heating mechanism causes rapid oscillation of water molecules, frictional heating, and localized tissue ablation. Operated by physician in clinical setting; footswitch controls energy delivery. System provides controlled thermal effects for soft tissue management; enables transluminal access to target tissues (e.g., liver, kidney, pancreas).
Clinical Evidence
No clinical trials performed. Evidence consists of a GLP-compliant preclinical study in a porcine model demonstrating successful transluminal navigation and ablation of soft tissues (liver, kidney, pancreas). Supporting evidence provided by review of clinical literature and FDA TPLC database.
Technological Characteristics
Microwave energy at 5.8 GHz; bipolar electrodes with dielectric materials. NP1 instrument: 1.6m flexible shaft, 2.7mm diameter, nitinol antenna, ceramic tip. System includes generator, footswitch, and interface cable. Sterilization: EO (ISO 11135). Standards: IEC 60601-1, IEC 60601-2-2, IEC 60601-2-6, ISO 10993 series, ISO 14971, IEC 62304.
Indications for Use
Indicated for coagulation (ablation) of soft tissue in patients requiring electrosurgical intervention. Not indicated for cardiac procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Creo Electrosurgical System with NP1 Accessory (K200003)
Submission Summary (Full Text)
{0}------------------------------------------------
February 14, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Creo Medical Ltd. % Phil Triolo President and Senior Consultant Phil Triolo and Associates LC 86 Skvcrest Ln Salt Lake City, Utah 84108
Re: K221672
Trade/Device Name: Creo Electrosurgical System with NP1 Instrument Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: NEY Dated: June 7, 2022 Received: June 9, 2022
Dear Phil Triolo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Colin K. C
Colin Keijing Chen, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K221672
Device Name
Creo Electrosurgical System with NP1 Instrument
Indications for Use (Describe)
The NP1 MicroBlate Fine Instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue.
The NP1 MicroBlate Fine Instrument is not intended for use in cardiac procedures.
The Electrosurgical Generator provides microwave (MW) energy to the compatible Creo Medical AB1 MicroBlate Flex and NP1 MicroBlate Fine instruments, intended for coagulation (ablation) of soft tissue.
The Electrosurgical Generator provides microwave (MW) energy to the compatible Creo Medical HS1 SlypSeal Instrument, intended for coagulation (hemostasis and cauterization) of soft tissue.
The Electrosurgical Generator provides microwave (MW) and radiofrequency (RF) energy to the compatible Creo Medical RS2 Speedboat Instrument, intended for coagulation (hemostasis and cauterization) and cutting soft tissue.
The Electrosurgical System is not intended for use in cardiac procedures.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K221672
Date Prepared: February 12, 2023
## 1. SUBMITTER
- 510(k) Sponsor: Creo Medical Ltd Creo House, Unit 2, Beaufort Park Way Chepstow NP16 5UH United Kingdom
- Contact Persons: Tiffany Powell Global Regulatory Specialist, Creo Medical Ltd. Phone +44 (0) 7760 150775 / Fax +44 (0) 1291 606015 Email Tiffany.Powell@creomedical.com
Phil Triolo, PhD, RAC President and Senior Consultant Phil Triolo and Associates LC Phone: 801-699-9846 / Fax 801-328-2399 Email: philt@philt.com
- Prepared By: Phil Triolo and Associates, LC 86 Skvcrest Lane SLC UT 84108 www.philt.com
#### 2. DEVICE
| Trade/Device Name: | Creo Electrosurgical System with NP1 Instrument |
|----------------------|----------------------------------------------------------------|
| Common Name: | Electrosurgical System |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulation number: | CFR 878.4400 |
| Product Code: | NEY (Class II) |
#### 3. PREDICATE DEVICE
K200003 Creo Electrosurgical System with NP1 Accessory
{4}------------------------------------------------
#### 4. SYSTEM DESCRIPTION
The Creo Medical Electrosurgical System with NP1 Instrument includes the following:
- . NP1 Instrument
- . Electrosurgical Generator
- . Footswitch
- . Interface Cable with Sterile Sheaths
The Electrosurgical Generator is designed to deliver bipolar radiofrequency (RF) energy and microwave (MW) energy for the purpose of cutting and coagulating tissue. Only the microwave output is used to power the NP1 Instrument. The Electrosurgical Generator output is actuated via a two-pedal Footswitch. One pedal activates the bipolar RF energy output for cutting (not active for use of NP1); the other pedal activates the MW enerqy output for coaqulation. The Electrosurgical Generator incorporates proprietary software developed by Creo Medical for generating and controlling the two energies delivered. The Electrosurgical Generator and Footswitch are non-sterile and reusable.
The single-use Interface Cable provides a conductive pathway between compatible Creo electrosurgical instruments and the output of the Electrosurgical Generator and is supplied with a sterile sheath that is fitted over the cable's distal end during use to minimize the potential for contamination of the sterile field.
The sterile, single-use NP1 soft tissue ablation instrument consists of a handle, 1.6 m long x 2.7 mm diameter flexible shaft, active nitinol antenna and pointed ceramic tip. Bipolar electrodes separated by dielectric materials couple microwave energy to target tissue when NP1 is activated. NP1 is only compatible for use with the Creo Medical Electrosurgical System, and the System is only intended for use with compatible Creo instruments and accessories.
#### INDICATIONS FOR USE 5.
The NP1 MicroBlate Fine instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue.
The NP1 MicroBlate Fine Instrument is not intended for use in cardiac procedures.
The Electrosurgical Generator provides microwave (MW) energy to the compatible Creo Medical AB1 MicroBlate Flex and NP1 MicroBlate Fine instruments, intended for coagulation (ablation) of soft tissue.
The Electrosurgical Generator provides microwave (MW) energy to the compatible Creo Medical HS1 SlypSeal Instrument, intended for coagulation (hemostasis and cauterization) of soft tissue.
The Electrosurgical Generator provides microwave (MW) and radiofrequency (RF) energy to the compatible Creo Medical RS2 Speedboat Instrument, intended for coagulation (hemostasis and cauterization) and cutting soft tissue.
The Electrosurgical System is not intended for use in cardiac procedures.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE 6. DEVICES
{5}------------------------------------------------
The Electrosurgical Generator (when used to power the NP1 Instrument) is designed to provide microwave energy at 5.8 GHz. The generator output voltage, power, current, duty cycle, and duration are under the control of software circuits to assure the energy delivered to the NP1 instrument via the Interface Cable is within specifications. When NP1 is connected to the Electrosurgical Generator and selected for use using either the Footswitch or membrane switches on the Generator, only microwave energy is delivered.
The NP1 Instrument is a flexible electrosurgical accessory that employs bipolar electrodes separated by dielectric materials to couple microwave energy delivered by the generator to target tissue as directed by the user. NP1 is only compatible for use with the Creo Medical Electrosurgical System, and the System is only intended for use with compatible Creo instruments and accessories.
This 510k is being submitted to remove a contraindication for transluminal use of the instrument that was required to be included in the instructions for use of the instrument by the FDA. The technological characteristics and specifications of the predicate and modified devices are identical. Microwave energy at 5.8 GHz is delivered to targeted tissue by the NP1 antenna when it is supplied with power by the Generator. The interaction between the microwave electromagnetic fields and soft tissue primarily results in rapid oscillation of dipolar water molecules, frictional resistance to such movement, and resultant active localized tissue heating. Thermal conduction, due to the temperature gradient created by the active tissue heated volume with surrounding tissue, results in adjacent tissue heating (thermal spread). The mechanism of microwave energy absorption is known as dielectric heating. For a specified tissue type and microwave frequency, the rate of microwave energy absorption is predictable, which allows for control of the energy absorption and subsequent thermal effects.
#### PERFORMANCE DATA 7.
No bench or clinical data were required to demonstrate substantial equivalence with the predicate device or to justify the elimination of the contraindication for transluminal use from the NP1 instrument IFU.
A preclinical study conducted in a porcine model under GLP requirements demonstrates that the NP1 instrument can be transluminally navigated to targeted soft (liver, kidney, pancreas) tissue, deliver microwave energy, and ablate the accessed tissue. Review of the relevant clinical literature and FDA's TPLC database supports the preclinical study results obtained in the preclinical study and the elimination of the transluminal contraindication from instrument labeling.
#### SOFTWARE 8.
The "new" NP1 .instrument that is the subject of the 510k incorporates software updates and fixes that did not require submission of a new 510k as they did not affect the substantial equivalence to the predicate device.
#### BIOCOMPATIBILITY 9.
The "new" NP1 instrument that is the subject of this 510k did not incorporate any changes in formulation, processing, sterilization, or geometry, and no chemicals were added (e.g., plasticizers, fillers, color additives, cleaning agents, mold release agents) that required additional biocompatibility studies to establish an acceptable biological risk.
{6}------------------------------------------------
## 10. STERILIZATION
The "new"NP1 instrument was adopted, in accordance with the recommendations of AAMI TIR 28, into an EO product family shown to achieve a Sterility Assurance Level of at least 10°. Sterilization processing conditions were validated in accordance with ISO 11135. The 6-month shelf-life has not changed.
## 11. STANDARDS
The following standards were referenced in this 510k to support substantial equivalence:
- ES60601- 1:2005/(R)2012\\ and A1:2012, C1:2009/(R)2012 . ANSI AAMI and A2:2010/(R)2012 (Consolidated Text)
- IEC 60601-1-2 Edition 4.0 2014-02 ●
- IEC 60601-2-2 Edition 6.0 2017-03 ●
- IEC 60601-2-6 Edition 2.1 2016-04 ●
- IEC 60601-1-6 Edition 3.1 2013-10 ●
- IEC 60601-2-18 Edition 3.0 2009-08
- ISO 10993-1 Fifth Edition 2018-11
- ISO 10993-4 Third Edition 2017-04 .
- ISO 10993-5 Third Edition 2009-06-01 ●
- ISO 10993-7 Second Edition 2008-10-15/[Amd 1:2019(en)] ●
- ISO 10993-10:2010 Third Edition 2010-08-01 ●
- ISO 10993-11 Third Edition 2017-09 ●
- ISO 11135 Second Edition 2014-07-15 ●
- IEC 62366-1 Edition 1.0 2015-02 [Including CORRIGENDUM 1 (2016)] ●
- ISO 14971 Second Edition 2007-03-01 ●
- ANSI/AAMI/IEC 62304:2006/A1:2016 ●
# 12. CONCLUSION
The information and data provided in K221672 demonstrate that the "new" NP1 Instrument with the contraindication for transluminal use removed is substantially equivalent to and as safe and effective as the predicate NP1 instrument that included a contraindication for transluminal use in its labeling.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.