NP1 Instrument, Creo Electrosurgical System

K200003 · Creo Medical, Ltd. · NEY · Oct 28, 2020 · General, Plastic Surgery

Device Facts

Record IDK200003
Device NameNP1 Instrument, Creo Electrosurgical System
ApplicantCreo Medical, Ltd.
Product CodeNEY · General, Plastic Surgery
Decision DateOct 28, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Creo Electrosurgical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories. The System is not intended for use in cardiac procedures. The NP1 instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue. The NP1 Instrument is not intended for use in cardiac procedures.

Device Story

System delivers microwave (MW) energy for soft tissue coagulation/ablation; comprises electrosurgical generator, footswitch, interface cable, and NP1 instrument. Generator produces 5.8 GHz MW energy; controlled by proprietary software and hardware circuits to ensure output within specifications. NP1 instrument features 1.4m shaft, nitinol antenna, ceramic tip, and stainless-steel bipolar electrodes; couples MW energy to tissue via dielectric heating. Used in clinical settings by physicians; generator operated via footswitch or membrane switches. MW energy causes dipolar water molecule oscillation, frictional heating, and thermal conduction for tissue ablation. System provides controlled thermal effects; benefits include precise soft tissue coagulation. Generator and footswitch are reusable; NP1 and interface cable are single-use. Healthcare providers monitor system output via generator interface; output directly affects tissue ablation volume and thermal spread.

Clinical Evidence

Bench testing only. Included simulated use studies in ex vivo muscle, liver, and kidney tissue to characterize ablation volumes; biocompatibility testing per ISO 10993-1; electrical safety/EMC testing per IEC 60601-2-2, 60601-2-6, and 60601-1-2; and mechanical performance testing (tensile strength, temperature evaluation, standing wave ratio).

Technological Characteristics

Microwave energy source (5.8 GHz); stainless-steel bipolar electrodes; nitinol antenna; ceramic tip. Software-controlled generator output. Dimensions: 1.4 m shaft length, 2.7 mm diameter. Sterilization: NP1 (ISO 11135, SAL 10^-6). Connectivity: Proprietary interface cable. Standards: IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-6, ISO 10993-1.

Indications for Use

Indicated for coagulation (ablation) of soft tissue in patients requiring such procedures. Not indicated for use in cardiac procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 28, 2020 Creo Medical Ltd. % Phil Triolo President and Senior Consultant Phil Triolo and Associates LC 86 Skycrest Lane Salt Lake City, Utah 84108 Re: K200003 Trade/Device Name: Creo Electrosurgical System with NP1 Accessory Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: NEY Dated: December 29, 2019 Received: September 21, 2020 Dear Phil Triolo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K200003 Device Name Creo Electrosurgical System with NP1 Accessory Indications for Use (Describe) The Creo Electrosurgical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories. The System is not intended for use in cardiac procedures. The NP1 instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue. The NP1 Instrument is not intended for use in cardiac procedures. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------|--| | <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | | | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K200003 # 510(k) Summary Date Prepared: October 26, 2020 ### 1. SUBMITTER - 510(k) Sponsor: Creo Medical Ltd Creo House, Unit 2, Beaufort Park Way Chepstow NP16 5UH United Kingdom - Contact Persons: Phil Triolo, PhD, RAC President and Senior Consultant Phil Triolo and Associates LC Phone: 801-699-9846 / Email: philt@philt.com Shaun Preston Senior Design and Development Engineer Creo Medical Ltd. +44 1291 606005 (office) / +44 7539 732362 (mobile) Email: Shaun.Preston@creomedical.com - Prepared By: Phil Triolo and Associates, LC 86 Skycrest Lane SLC UT 84108 ### 2. DEVICE | Trade/Device Name: | Creo Electrosurgical System with NP1 Accessory | | |----------------------|----------------------------------------------------------------|--| | Common Name: | Electrosurgical System | | | Classification Name: | Electrosurgical cutting and coagulation device and accessories | | | Regulation number: | CFR 878.4400 | | | Product Code: | NEY (Class II) | | ## 3. PREDICATE DEVICES K163105 Emprint Ablation System {4}------------------------------------------------ #### SYSTEM DESCRIPTION 4. The Creo Medical Electrosurgical System with NP1 Accessory includes the following: - . NP1 Instrument - . Electrosurgical Generator - . Footswitch - Interface Cable with Sterile Sheaths The Electrosurgical Generator is designed to deliver bipolar radiofrequency (RF) energy and microwave (MW) energy for the purpose of cutting and coagulating tissue. Only the microwave output is used to power the NP1 Instrument. The Electrosurgical Generator output is actuated via a two-pedal Footswitch. One pedal activates the bipolar RF energy output for cutting; the other pedal activates the MW energy output for coagulation. The Electrosurgical Generator incorporates proprietary software developed by Creo Medical for generating and controlling the two energies delivered. The Electrosurgical Generator and Footswitch are non-sterile and reusable. The single-use Interface Cable connects compatible Creo electrosurgical instruments to the Electrosurgical Generator and is supplied with a sterile sheath that is fitted over its distal end during use to minimize the potential for contamination of the sterile field. The sterile, single-use NP1 soft tissue ablation accessory (Instrument) consists of a handle, 1.4 m long x 2.7 mm diameter shaft, active nitinol antenna and pointed ceramic tip. Stainless-steel bipolar electrodes separated by dielectric materials couple microwave energy to target tissue when NP1 is in use. NP1 is only compatible for use with the Creo Medical Electrosurgical System, and the System is only intended for use with compatible Creo instruments and accessories. ### 5. INDICATIONS FOR USE The Creo Electrosurqical System is intended to provide microwave energy for coagulation (ablation) of soft tissue and is for use only with compatible Creo Medical instruments and accessories. The System is not intended for use in cardiac procedures. The NP1 instrument, when used with the Creo Medical Electrosurgical System, is intended for use in coagulation (ablation) of soft tissue. The NP1 Instrument is not intended for use in cardiac procedures. ### 6. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES The Electrosurgical Generator (when used to power the NP1 Instrument) is designed to provide microwave energy at 5.8 GHz. The generator output voltage, power, current, duty cycle, and duration are under the control of software and hardware circuits to assure the energy delivered to the NP1 instrument via the Interface Cable is within specifications. When NP1 is connected to the Electrosurgical Generator and selected for use using either the Footswitch or membrane switches on the Generator, only microwave energy is delivered. The NP1 Instrument is a flexible electrosuraical accessory that employs stainless-steel bipolar electrodes separated by dielectric materials to couple microwave energy delivered by the generator to target tissue as directed by the user. NP1 is only compatible for use with the Creo Medical Electrosurgical System, and the System is only intended for use with compatible Creo instruments and accessories. {5}------------------------------------------------ #### Creo Electrosurgical System with NP1 Accessory ### K200003 The subject and predicate devices are based on the following same technological elements, as both systems use microwave energy to effect ablation of soft tissue. Both devices utilize the fundamental mechanism for electromagnetic energy coupling to tissue employed by all microwave-powered ablation devices. The interaction between the microwave electromagnetic fields and soft tissue primarily results in rapid oscillation of dipolar water molecules, frictional resistance to such movement, and resultant active localized tissue heating. Thermal conduction, due to the temperature gradient created by the active tissue heating region, results in adjacent tissue heating (thermal spread). The mechanism of microwave energy absorption is known as dielectric heating. For a specified tissue type and microwave frequency, the rate of microwave energy absorption is predictable, which allows for control of the energy absorption and subsequent thermal effects. The following technological differences between the subject and predicate devices were compared: | Difference | Evidence supporting substantial equivalence | |----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | The NP1 Instrument is supplied with pulsed<br>62 W microwave power at 5.8 GHz<br>whereas the predicate antenna is supplied<br>with microwave power at 2.45 GHz | Results of simulated use studies conducted in ex<br>vivo muscle, liver, and kidney tissue were used to<br>characterize the ablation volumes created by the<br>new and predicate systems, which have been<br>determined to be equivalent. | | The predicate antenna is actively cooled by<br>water circulated through the antenna; NP1<br>is not cooled. | Evaluations in ex vivo bovine tissue demonstrate<br>that the temperature of applied parts does not<br>exceed permissible limits and that no non-target<br>tissue is thermally injured as a result of NP1<br>treatments at its maximum power setting. | | Different patient-contacting materials | Results of biocompatibility and performance<br>evaluations demonstrate an acceptable biological<br>risk and that acceptance criteria are met. | | NP1 Interface Cable is single use,<br>predicate cable is reusable | Both cables conduct microwave energy for ablation<br>and neither poses a significant risk for infection. | | NP1 System is controlled by software and<br>hardware, predicate only by hardware | EMC, electrical safety, and verification test results<br>demonstrate acceptable safety and performance. | ### 7. PERFORMANCE DATA The system meets all design specifications, risk mitigation requirements, and applicable medical device standard requirements. Sterilization/Shelf-life Per ISO 11135 and AAMI TIR28 standard evaluation, the NP1 Sterility Assurance Level is 10° The shelf-life for the NP1 Instrument is 6 months. The Interface Sheath is 3 years. {6}------------------------------------------------ ### Creo Electrosurgical System with NP1 Accessory #### Biocompatibility The biocompatibility evaluation for the NP1 Instrument was conducted in accordance with ISO 10993-1:2009 and the FDA guidance document, Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process: Guidance for Industry and Food and Drug Administration Staff," issued by the FDA on June 16, 2016. According to Table A.1: Biocompatibility Evaluation Endpoints in the FDA Biocompatibility Guidance, the antenna assembly of NP1 and the distal portion of its shaft are categorized, from a biological risk perspective, as externally communicating with limited (< 24h) contact with tissue. The battery of tests whose results support a finding of a low and acceptable biological and toxicological risk included the following: - Cytotoxicity - . Sensitization - . Irritation - . Systemic toxicity - . Pyrogen Testing ### Electrical Safety, Microwave Safety and Electromagnetic Compatibility Electrical safety and EMC tests were conducted on the Creo Medical Electrosurgical System with NP1 Accessory. The system complies with FDA-recognized versions of IEC 60601-2-2 and IEC 60601-2-6 standards for safety and the IEC 60601-1-2 standard for EMC. #### Software Verification and Validation Testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury or death to the patient or operator. #### Mechanical and Ablation Performance Testing - . Handle and shaft maximum temperature evaluation - . Post-activation blanch testing - Tensile strength testing of shaft/antenna components - Strength testing for handle connection, antenna locking knob and shaftadjustment locking knob - Standing wave ratio / Reflected power threshold testing - . Simulated use testing in ex-vivo tissue (ablation zone repeatability and predicate comparison studies) Results of mechanical and ablation performance tests, including bench tests and simulated use studies conducted in ex-vivo tissue, demonstrate that the NP1 Electrosurgical System meets design specifications and performs as intended, and support Creo's assertion of substantial equivalence to the predicate device. {7}------------------------------------------------ K200003 #### CONCLUSION 8. The nonclinical bench testing, simulated use testing, hardware and software verification and validation and standards testing demonstrate there are no new or different questions of safety or effectiveness raised by any differences between the subject device and predicate device. The Creo Medical Electrosurgical System with NP1 Accessory as determined by the indications for use, technological characteristics and the performance testing demonstrate it is as safe and effective as the legally marketed device identified and is substantially equivalent to the identified predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...