Disposable Syringe with permanently attached needle, Safety Syringe with permanently attached needle
K221308 · Jiangxi Maidikang Medical Devices Co., Ltd. · FMF · Oct 17, 2022 · General Hospital
Device Facts
Record ID
K221308
Device Name
Disposable Syringe with permanently attached needle, Safety Syringe with permanently attached needle
Applicant
Jiangxi Maidikang Medical Devices Co., Ltd.
Product Code
FMF · General Hospital
Decision Date
Oct 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Safety Syringe with permanently attached needle is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the skin. The Safety sheath of Syringe is designed to aid in the prevention of needle stick injuries and reduce the potential of syringe reuse. The Disposable Syringe with permanently attached needle is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.
Device Story
Manual piston syringes with permanently attached needles; available in disposable and safety-engineered configurations. Safety version includes sliding sheath to prevent needlestick injuries and discourage reuse. Used by healthcare professionals in clinical settings for fluid aspiration and injection. Device operates via manual plunger depression; no electronic or automated components. Safety sheath provides physical barrier post-injection. Benefits include reduced risk of accidental needlestick and prevention of syringe reuse. Output is the delivery of medication or aspiration of fluids.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via ISO 7886-1, ISO 7886-4, ISO 7864, ISO 9626, and ISO 23908. Biocompatibility confirmed per ISO 10993 series. Sterilization validated per ISO 11135.
Technological Characteristics
Materials: PP (barrel, plunger, cap, sleeve), Polyisoprene (piston), Stainless Steel 304 (needle), PC (safety connector). Manual operation. EO sterilization (SAL 10^-6). Complies with ISO 7886-1, ISO 7864, ISO 9626, ISO 23908, ASTM F1980-16, ASTM F88/F88M-15, ASTM F1929-2015, ASTM F1886/F1886M-16.
Indications for Use
Indicated for health care professionals for general purpose aspiration of fluid from vials/ampoules and liquid injection below the skin surface. Applicable to adult and pediatric populations requiring subcutaneous or similar injections.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
Syringe with permanently attached needle (K192551)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 17, 2022
Jiangxi Maidikang Medical Devices Co., Ltd Wu Huansheng Management Room 308, floor 3, building 1, Jiangxi pharmaceutical co., Ltd, No 999, kelun Avenue, pharmaceutical industrial park Yichun, Jiangxi China
Re: K221308
Trade/Device Name: Safety Syringe with permanently attached needle; Disposable Syringe with permanently attached needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: FMF, MEG Dated: September 8, 2022 Received: September 14, 2022
Dear Wu Huansheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal. ObGyn. General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221308
#### Device Name
Safety Syringe with permanently attached needle Disposable Syringe with permanently attached needle
#### Indications for Use (Describe)
The Safety Syringe with permanently attached needle is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the skin. The Safety sheath of Syringe is designed to aid in the prevention of needle stick injuries and reduce the potential of syringe reuse.
The Disposable Syringe with permanently attached needle is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) summary
### l Submitter
Device submitter: Jiangxi Maidikang Medical Devices Co., Ltd.
Room 308, floor 3, building 1, jiangxi pharmaceutical co., Ltd, No 999, kelun Avenue, pharmaceutical industrial park, Yuanzhou District, Yichun City, Jiangxi province
Contract manufacturer: Anhui Tiankang Medical Technology Co., Ltd.
No.228, Weiyi Road, Economic Development Zone, Tianchang City, Anhui, China.
510(k) Number: K221308
Contact person: Name: Wu Huansheng Title: Management Phone: +86-13617915949 E-mail: 3487486481 @qq.com
Preparation Date: May 6th, 2022
### II Device
Trade Name of Device: Safety Syringe with permanently attached needle Disposable Syringe with permanently attached needle Common Name: Piston syringe Regulation Number: 21CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: II Product code: FMF, MEG Review Panel: General Hospital III Predicate Devices
| Trade name: | 1ml Luer Slip or Luer Lock Syringe |
|-----------------|----------------------------------------------------------------------------|
| | Syringe with permanently attached needle (used as Predicate Device) |
| | Safety Syringe with permanently attached needle (used as Predicate Device) |
| Common name: | Piston Syringe and antistick syringe |
| Classification: | Class II, 21 CFR 880.5860 |
| Product Code: | FMF, MEG |
{4}------------------------------------------------
Premarket Notification: K192551 Manufacturer: Jiangsu Caina Medical Co., Ltd.
# IV Device description
The proposed Syringes include Disposable Syringe with permanently attached needle and a Safety Syringe with permanently attached needle. The Disposable Syringe with permanently attached needle have one kind of product configuration, and the Safety Syringe with permanently attached needle has two kinds of product configurations. The proposed syringes are available in different combination of syringe volumes and/or needle sizes, and the syringe is sterilized by Ethylene Oxide Gas to achieve a SAL of 10-6 and supplied sterile in packaging to maintain the sterility of the device during the shelf life of three years.
| Model | Syringe<br>Volume | Needle Length | Needle<br>Gauge | Wall<br>type | Needle<br>Bevel |
|-----------------------------------------------------------------------|-------------------|----------------------------------------------|-----------------|--------------|-----------------|
| Disposable Syringe<br>with permanently<br>attached needle<br>-TKSPN01 | 0.5ml; 1ml | 3/8" (10mm);<br>1/2" (13mm);<br>5/8" (16mm); | 23G | TW | LB |
| | | 3/4" (20mm);<br>7/8" (22mm);<br>1" (25mm); | 24G | RW | LB |
| | | | 25G | RW | LB |
| | | 3/8" (10mm);<br>1/2" (13mm);<br>5/8" (16mm); | 26G | RW | LB |
| | | 3/4" (20mm); | 27G | RW | LB |
| | | 3/8" (10mm);<br>1/2" (13mm);<br>5/8" (16mm); | 28G | RW | LB |
| | | 3/10" (8mm); | 29G | RW | LB |
| | | 3/8" (10mm);<br>1/2" (13mm); | 30G | RW | LB |
| | | 5/8" (16mm); | 31G | RW | LB |
| Safety Syringe with<br>permanently attached<br>needle -TKSSPN01 | 0.3ml | 3/8" (10mm);<br>1/2" (13mm);<br>5/8" (16mm); | 27G | RW | LB |
| | | | 28G | RW | LB |
| | | 3/10" (8mm); | 29G | RW | LB |
| | | 3/8" (10mm);<br>1/2" (13mm); | 30G | RW | LB |
| | | 5/8" (16mm); | 31G | RW | LB |
Table 1 specification of proposed device
{5}------------------------------------------------
| 0.5ml; 1ml | | 3/8" (10mm);<br>1/2" (13mm);<br>5/8" (16mm); | 25G | RW | LB |
|----------------------------------------------------------------|--------------|----------------------------------------------|-----|----|----|
| | | | 26G | RW | LB |
| | | | 27G | RW | LB |
| | | | 28G | RW | LB |
| | | 3/10" (8mm);<br>3/8" (10mm);<br>1/2" (13mm); | 29G | RW | LB |
| | | | 30G | RW | LB |
| | 5/8" (16mm); | 31G | RW | LB | |
| Safety Syringe with<br>permanently attached<br>needle-TKSSPN02 | 1ml | 3/4" (20mm); | 21G | TW | LB |
| | | 7/8" (22mm); | 22G | TW | LB |
| | | 1" (25mm); | 23G | TW | LB |
| | | 1 1/4" (32mm); | 24G | RW | LB |
| | | 1 1/2" (38mm); | 25G | RW | LB |
| | | 3/4" (20mm) | 26G | RW | LB |
| | | 3/4" (20mm) | 27G | RW | LB |
# V Indications for use
# Disposable Syringe with permanently attached needle
The Disposable Syringe with permanently attached needle is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin.
# Safety Syringe with permanently attached needle
The Safety Syringe with permanently attached needle is intended for use by health care professionals for general purpose aspiration of fluid from vials, ampoules and liquid injection below the surface of the skin. The Safety sheath of Syringe is designed to aid in the prevention of needle stick injuries and reduce the potential of syringe reuse.
### VI Comparison of technological characteristics with the predicate devices
The Disposable Syringe with permanently attached needle and Safety Syringe with permanently attached needle have the same intended use, technology, design and performance specifications to the legally marketed predicate device. The main differences between subject devices and predicate devices are the specification of syringe volume, needle gauge and needle length. This difference does not affect indication for use. The difference between the subject device and the predicate does not affect the safety and effectiveness of the subject device T
{6}------------------------------------------------
| Device<br>feature | Subject Device | Predicate Device K192551 | Comment |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Product | Disposable Syringe with<br>permanently attached<br>needle | Syringe with permanently<br>attached needle | / |
| Syringe type | Piston syringe | Piston syringe | Same |
| Indications<br>for use | The Disposable Syringe<br>with permanently attached<br>needle is intended for use<br>by health care professionals<br>for general purpose<br>aspiration of fluid from vials,<br>ampoules and liquid<br>injection below the surface<br>of the skin. | The Syringe with<br>permanently attached<br>needle is intended for use<br>by health care professionals<br>for general purpose<br>aspiration of fluid from vials,<br>ampoules and liquid<br>injection below the surface<br>of the skin. | Same |
| Product code | FMF | FMF | Same |
| Regulation<br>number | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | II | II | Same |
| Principle of<br>operation | For Manual Use Only, For<br>Single Use Only | For Manual Use Only, For<br>Single Use Only | Same |
| Connector<br>Type | Attached needle | Attached needle | Same |
| Needle<br>gauge | 23G, 24G, 25G, 26G, 27G,<br>28G, 29G, 30G, 31G | 21G, 23G, 25G, 26G, 27G,<br>28G, 29G, 30G, 31G | Difference<br>1 |
| Needle<br>Length | 8mm, 10mm, 13mm, 16mm,<br>20mm, 22mm, 25mm | 8mm, 10mm, 13mm, 16mm,<br>20mm, 25mm | |
| Needle wall<br>type | RW, TW | RW, TW | |
| Needle bevel | 11°±2° | 11°±2°, 15°±2° | |
| Syringe<br>Volume | 0.5ml, 1ml | 0.3ml, 0.5ml, 1ml | |
| Device<br>feature | Subject Device | Predicate Device K192551 | Comment |
| Main<br>structure and<br>materials | (1) Needle cap (PP)<br>(2) Needle (Stainless Steel<br>304)<br>(3) Piston (Polyisoprene)<br>(4) Plunger (PP)<br>(5) Barrel (PP) | Syringe with permanently<br>attached needle-type A<br>(1) needle cap (PP or PE)<br>(2) needle (Stainless Steel<br>304)<br>(3) piston (Polyisoprene)<br>(4) plunger (PP)<br>(5) barrel (PP)<br>(6) end cap (PP or PE)<br>Syringe with permanently<br>attached needle-type B<br>(1) needle cap (PP or PE)<br>(2) needle (Stainless Steel<br>304)<br>(3) piston (Polyisoprene)<br>(4) plunger (PP or ABS)<br>(5) barrel (PP) | Difference<br>2 |
| Performance<br>specifications | Complies with ISO 7886-1;<br>ISO 9626 and ISO 7864 | Complies with ISO 7886-1;<br>ISO 9626 and ISO 7864 | Same |
| Single Use | Yes | Yes | Same |
| Sterilization | EO Sterilization | EO Sterilization | Same |
| SAL | 10-6 | 10-6 | Same |
| Biocompatibility | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Same |
| Labeling | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | Same |
Table 6-1 Substantial equivalence discussion for Disposable Syringe with permanently attached needle
{7}------------------------------------------------
Difference 1
The needle bevel of subject device (11°±2°) is smaller than the predicate device's bevel range of 11°±2ând 15° £°. The subject device's 0.5ml and 1ml syringe volume is smaller than the predicated device's syringe volume range of 0.3ml, 0.5ml and 1ml. And the needle gauge and length of subject devices are similar to the predicate device. This difference does not affect intended use. In addition, differences were addressed through ISO 7886-1, ISO 7864 and ISO 9626. The differences
{8}------------------------------------------------
on syringe volume, needle bevel, needle gauge and length do not raise different question of safety and effectiveness.
### Difference 2
The configuration of subject device is same as type B of predicate device, and compared with type A, the difference is that the subject device does not have an end cap. but the syringe with attached needle is widely used in the clinical. Whether there is an end cap or not will not affect the indication for use of the equipment itself. This difference does not raise new questions about safety and effectiveness.
The materials of subject device are similar to the predicated device. The proposed syringes' biocompatibility can be demonstrated by the biocompatibility tests. The differences on configuration and materials do not raise new questions about safety and effectiveness.
| Device<br>feature | Subject Device | Predicate Device K192551 | Comment |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Product | Safety Syringe with<br>permanently attached<br>needle | Safety Syringe with<br>permanently attached<br>needle | / |
| Syringe type | Piston syringe | Piston syringe | Same |
| Indications<br>for use | The Safety Syringe with<br>permanently attached<br>needle is intended for use<br>by health care professionals<br>for general purpose<br>aspiration of fluid from vials,<br>ampoules and liquid<br>injection below the surface<br>of the skin. The Safety<br>sheath of Syringe is<br>designed to aid in the<br>prevention of needle stick<br>injuries and reduce the<br>potential or syringe reuse. | The Safety Syringe with<br>permanently attached<br>needle is intended for use<br>by health care professionals<br>for general purpose<br>aspiration of fluid from vials,<br>ampoules and liquid<br>injection below the surface<br>of the skin. The Safety<br>sheath of Syringe is<br>designed to aid in the<br>prevention of needle stick<br>injuries and reduce the<br>potential or syringe reuse. | Same |
| Product code | MEG | MEG | Same |
| Regulation<br>number | 21 CFR 880.5860 | 21 CFR 880.5860 | Same |
| Class | II | II | Same |
| Device<br>feature | Subject Device | Predicate Device K192551 | Comment |
| Principle of operation | For Manual Use Only, For<br>Single Use Only | For Manual Use Only, For<br>Single Use Only | Same |
| Connector<br>Type | Attached needle | Attached needle | Same |
| Needle<br>gauge | 21G, 22G, 23G, 24G, 25G,<br>26G, 27G, 28G, 29G, 30G,<br>31G | 21G, 23G, 25G, 26G, 27G,<br>28G, 29G, 30G, 31G | Difference<br>3 |
| Needle<br>Length | 8mm, 10mm, 13mm, 16mm,<br>20mm, 22mm, 25mm,<br>32mm, 38mm | 8mm, 10mm, 13mm, 16mm,<br>20mm, 25mm | |
| Needle wall<br>type | RW, TW | RW, TW | |
| Needle bevel | 11°±2° | 11°±2°, 15°±2° | |
| Syringe<br>Volume | 0.3ml, 0.5ml, 1ml | 0.3ml, 0.5ml, 1ml | Same |
| Main<br>structure and<br>materials | (1) Protecitve cap (PP)<br>(2) Needle (Stainless Steel<br>304)<br>(3) Safety mechanism-<br>Connector base (PC)<br>(4) Piston (Polyisoprene)<br>(5) Safety mechanism-<br>Sliding sleeve (PP)<br>(6) Plunger (PP)<br>(7) Barrel (PP) | (1) needle cap (PP or PE)<br>(2) needle (Stainless Steel<br>304)<br>(3) safety mechanism (PC)<br>(4) piston (Polyisoprene)<br>(5) safety mechanism (PP)<br>(6) plunger (PP)<br>(7) barrel (PP) | Same |
| Performance<br>specifications | Complies with ISO 7886-1;<br>ISO 9626 and ISO 7864 | Complies with ISO 7886-1;<br>ISO 9626 and ISO 7864 | Same |
| Single Use | Yes | Yes | Same |
| Sterilization | EO Sterilization | EO Sterilization | Same |
| SAL | 10-6 | 10-6 | Same |
| Biocompatibility | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Contact level: blood path,<br>indirect, limited contact (<24<br>hours).<br>Conforms to the<br>requirement of ISO 10993<br>series Standards | Same |
| Labeling | Complied with 21 CFR part<br>801 | Complied with 21 CFR part<br>801 | Same |
Table 6-2 Substantial equivalence discussion for Safety Syringe with permanently attached needle
{9}------------------------------------------------
{10}------------------------------------------------
### Difference 3
The needle bevel of subject device (11°±2°) is smaller than the predicate device's bevel range of 11°±2'and 15°±2°. And the needle gauge and length of subject devices are similar to the predicate device, the difference is just in dimension. This difference does not affect intended use. In addition, differences were addressed through ISO 7864 and ISO 9626. The differences on syringe volume, needle bevel, needle gauge and length do not raise different question of safety and effectiveness.
# VII Performance data
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing
Biocompatibility of the Disposable Syringe with permanently attached needle and Safety Syringe with permanently attached needle were evaluated in accordance with ISO 10993-1:2018 for the body contact category of "External communication device - Blood path indirect" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended:
| Cytotoxicity | ISO 10993-5: 2009 |
|---------------------------|--------------------|
| Skin sensitization | ISO 10993-10: 2010 |
| Hemolysis | ISO 10993-4: 2017 |
| Intracutaneous reactivity | ISO 10993-10: 2010 |
| Acute systemic toxicity | ISO 10993-11: 2017 |
| Pyrogenicity | ISO 10993-11: 2017 |
| Particulate matter | USP <788> |
# Sterilization and shelf life testing
The sterilization method has been validated to ISO11135, which has thereby determined the routine control and monitoring parameters. Bacteria Endotoxin Limit is carried out according to USP42-NF37 <85> Bacterial Endotoxins Test.
The testing is performed according to the following standards:
| EO residue | ISO 10993-7:2008 |
|--------------------|------------------|
| ECH residue | ISO 10993-7:2008 |
| Bacteria Endotoxin | USP42-NF37 <85> |
The shelf life of 3 year is determined based on stability studies which include ageing test according to FDA recognized standard ASTM F1980-16.
Package integrity testing was conducted on the final, packaged, and sterile devices after environmental conditioning and simulated transportation. All packaging deemed acceptable for protection of product and sterility maintenance.
The testing is performed according to the following standards:
Seal strength
ASTM F88/F88M-15
{11}------------------------------------------------
Blue Dye Penetration Seal Integrity (Visual Inspection)
ASTM F 1929-2015 ASTM F 1886/ F 1886M-16
# Performance testing
Performance testing is performed according to the following standards:
- ISO 7886-1: 2017, Sterile hypodermic syringes for single use- Part 1: Syringes for A manual use.
- > ISO 7886-4: 2018, Sterile hypodermic syringes for single use- Part 4: Syringes with re-use prevention feature
- > ISO 7864: 2016, ISO 9626:2016, Stainless Steel Needle Tubing For The Manufacture of Medical Devices.
- > ISO 9626:2016, Stainless Steel Needle Tubing for The Manufacture of Medical Devices.
- > ISO 23908: 2011 Sharps injury protection - Requirements and test methods -Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling.
# VIII Clinical Test Conclusion
No clinical study is included in this submission.
# IX Conclusion
The Disposable Syringe with permanently attached needle and Safety Syringe with permanently attached needle are substantially equivalent to its predicate devices. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.