K201166 · Merit Medical Systems, Inc. · KNW · Aug 28, 2020 · Gastroenterology, Urology
Device Facts
Record ID
K201166
Device Name
Temno Elite Biopsy System
Applicant
Merit Medical Systems, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Aug 28, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Temno Elite Biopsy System is intended for use in obtaining biopsies from soft tissues such as liver, kidney,breast,prostate, spleen,lung,lymph nodes,thyroid,and various soft tissue masses. It is NOT intended for use in bone. The Valved Coaxial Introducer Needle and Standard Coaxial Introducer Needles are intended for use as guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney,prostate,spleen, lung,lymph nodes,thyroid, and various soft tissue masses. It is NOT intended for use in bone.
Device Story
Single-use, full-core biopsy device; available in multiple gauge sizes and lengths. Includes biopsy device with optional valved or standard coaxial introducer needles. Features color-coded gauge indicators and depth stops. Operates via manual trigger mechanism to obtain soft tissue samples. Used by clinicians in clinical settings to acquire tissue for diagnostic evaluation. Output is a core tissue sample; aids in clinical decision-making by providing histological material for pathology. Benefits patient through minimally invasive tissue acquisition for disease diagnosis.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional verification, tensile strength, adjustable throw accuracy, multiple sample collection, device visibility, simulated use, and a full battery of biocompatibility tests (cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity) per ISO 10993 standards.
Technological Characteristics
Single-use, sterile, manual core biopsy instrument. Materials include stainless steel needle tubing (ISO 9626). Features color-coded gauge identification (ISO 6009). Includes small-bore connectors (ISO 80369-7). Sterilized via ethylene oxide (ISO 11135). No electronic components or software.
Indications for Use
Indicated for obtaining core biopsy samples from soft tissues (liver, kidney, breast, prostate, spleen, lung, lymph nodes, thyroid, and soft tissue masses) in patients requiring tissue sampling. Not indicated for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Achieve Programmable Automatic Biopsy Systems (K141552)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 28, 2020
Merit Medical Systems, Inc. Niloufar Samimi Senior Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095
Re: K201166
Trade/Device Name: Temno Elite Biopsy System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology-Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: July 30, 2020 Received: August 3, 2020
Dear Niloufar Samimi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel G. Walter, Jr. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201166
Device Name Temno Elite Biopsy System
#### Indications for Use (Describe)
The Temno Elite Biopsy System is intended for use in obtaining biopsies from soft tissues such as liver, kidney,breast,prostate, spleen,lung,lymph nodes,thyroid,and various soft tissue masses. It is NOT intended for use in bone.
The Valved Coaxial Introducer Needle and Standard Coaxial Introducer Needles are intended for use as guiding needle in obtaining core biopsy samples from soft tissue such as liver, kidney,prostate,spleen, lung,lymph nodes,thyroid, and various soft tissue masses. It is NOT intended for use in bone.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size: 1em;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 1em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Temno Elite Biopsy System Premarket Notification 510(k)
# 5.0 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4583<br>Niloufar Samimi<br>26 April 2020<br>1721504 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Temno Elite Biopsy System<br>Biopsy System<br>Instrument, Biopsy<br>II<br>KNW<br>876.1075<br>Gastroenterology/ Urology |
| Predicate<br>Device | Premarket Notification Predicate #1:<br>Trade Name:<br>Classification Name:<br>Premarket Notification: | Bard® Mission® Disposable Core Biopsy<br>Instrument, Biopsy<br>K171953 |
| | Manufacturer: | Bard Peripheral Vascular, Inc |
| Predicate<br>Device | Premarket Notification Predicate #2:<br>Trade Name: | Achieve Programmable Automatic Biopsy<br>Systems |
| | Classification Name:<br>Premarket Notification:<br>Manufacturer: | Instrument, Biopsy<br>K141552<br>CareFusion |
| | These predicates have not been subject to a design-related recall. | |
| Device<br>Description | The subject device is a single use full core biopsy device. It is available<br>in several gauge sizes and lengths. The device has printed gauge size<br>indicator that is color coded according to the various gauge sizes (<br>yellow= 20G, pink= 18G, purple= 16 G, green= 14G, and blue=12G.<br>The needles have a protective sheath. | |
| | The Temno Elite Biopsy system includes Biopsy device and has the<br>option of including a Valved Coaxial Introducer Needle, or a Standard<br>Coaxial Introducer Needle. The depth stop on both the Valved Coaxial | |
{4}------------------------------------------------
Introducer Needle and the Standard Coaxial Introducer Needle is color coded to match the gauge size of the Temno Elite Biopsy system.
The Temno Elite Biopsy System is supplies sterile and is intended for single use only.
The Temno Elite Biopsy System is intended for use in obtaining biopsies from soft tissues such as liver, kidney, breast, prostate, spleen, lung, lymph nodes, thyroid, and various soft tissue masses. It is NOT intended for use in bone.
The Valved Coaxial Introducer Needle and Standard Coaxial Introducer Needles are intended for use as guiding needle in obtaining core biopsy Indications for samples from soft tissue such as liver, kidney, prostate, spleen, lung, Use lymph nodes, thyroid, and various soft tissue masses. It is NOT intended for use in bone.
> There is no change in the Indications for Use Statement from the predicate to the subject device.
The technological characteristics of the subject Temno Elite Biopsy System device are substantially equivalent to those of the predicate device #1 Bard® Mission® Disposable Core Biopsy and Reference Predicate Device #2, Achieve Programmable Automatic Biopsy.
At a high level, the subject and predicate devices are based on the following same technological elements:
- Clinical use
- Labeling ●
- Basic design ●
Comparison to Predicate Device
- Principle of operation ●
- Performance ●
- Soft tissues ●
- Adjustable throw lengths ●
The following technological differences exist between the subject and predicate devices:
- Needle gauges
- Materials .
- Biopsy chamber geometry
- Sample eject assist ●
{5}------------------------------------------------
FDA guidance and recognized performance standards have been established for Biopsv instrument under Section 514 of the Food. Drug and Cosmetic Act. A battery of tests was performed based on the requirements of the below recognized performance standards and quidance, as well as biocompatibility, sterilization, and labeling standards and quidance. Conformity to these standards demonstrates that the proposed Temno Elite Biopsy System met the standards' established acceptance criteria applicable to the safety and efficacy of the device. Performance testing was conducted based on the risk analysis and based on the requirements of the following documents:
- ANSI/AAMI/ISO 11135:2014 , Sterilization of health care products -● routine control of a sterilization process for medical devices
- ISO 10993-7: 2008, Biological evaluation of medical devices Part 7: ● Ethylene oxide residuals
- ISO 7864: 2016, Sterile hypodermic needles for single use -● Requirements and test methods
- ISO 9626:2016, Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- ISO 6009:2016, Hypodermic needles for single use - Color coding for identification
- ISO 2233:2000, Complete, filled transport packages and unit loads -● Conditioning for testing
- ISO 80369-7:2019, Small-bore connectors for liquids and gases in ● healthcare applications - Part 7: Connectors for intravascular or hypodermic applications
Performance Data
- . ISO 11607-1:2019, Packaging for terminally sterilized medical devices -Part 1: Requirements for materials, sterile barrier systems and packaging svstems
- . ISO 11607-2:2019, Packaging for terminally sterilized medical devices -Part 2: Validation requirements for forming, sealing and assembly processes
- . ASTM D4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems
- ASTM F1980:2016, Standard Guide for Accelerated Aging of Sterile ● Barrier Systems for Medical Devices
- ISO 10993-1:2018, Biological Evaluation of Medical Devices Part 1: ● Evaluation and Testing within a risk management process
- . ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-7:2008, Biological evaluation of medical devices -Part 7: Ethylene oxide sterilization residuals
- ISO 10993-10:2010, Biological evaluation of medical devices – Part10: Tests for irritation and skin sensitization
- . ISO 10993-11:2017, Biological evaluation of medical devices – Part 11: Tests for systemic toxicity
- United States Pharmacopeia 42, National Formulary 37 (USP) <151> Pyrogen Test, 2019
The following performance data were provided in support of the substantial equivalence determination.
Performance Testing- Bench
{6}------------------------------------------------
- Dimensional Verification ●
- Tensile of Joints ●
- Adjustable Throw Accuracy
- Multiple Samples ●
- Device Visibility ●
- Simulated Use ●
#### Biocompatibility testing
The biocompatibility evaluation for the Temno Elite Biopsy System was conducted in accordance with the FDA Guidance Document "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and testing within a risk management process," (2016), and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," (2018) as recognized by FDA. The battery of testing included the following test:
- Cytotoxicity
- Sensitization ●
- Irritation
- Acute Systemic Toxicity ●
- . Pyrogenicity
The result of biocompatibility testing demonstrate that the subject device Temno Elite Biopsy System is considered biocompatible for its intended use.
## Summary of Substantial Equivalence
Based on the indications for use, design, safety and performance testing, the subject Temno Elite Biopsy System meets the requirements that are considered essential for its intended use and is substantially equivalent to the Predicate device #1 Bard Mission Disposable Core Biopsy Instrument, K171953 and Reference Predicate device #2 Achieve Programmable Automatic Biopsy, K141552.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.