K221204 · Contego Medical, Inc. · QEW · Jun 23, 2022 · Cardiovascular
Device Facts
Record ID
K221204
Device Name
Excipio SV Thrombectomy Device
Applicant
Contego Medical, Inc.
Product Code
QEW · Cardiovascular
Decision Date
Jun 23, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5150
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Excipio SV Thrombectomy Device is indicated for the non-surgical removal of soft emboli and thrombi from peripheral blood vessels.
Device Story
Excipio SV Thrombectomy Device is a mechanical thrombus displacement catheter used in conjunction with an aspiration catheter to remove soft emboli and thrombi from peripheral blood vessels. The device features a distal nitinol (NiTi) braided basket component, which is expanded or collapsed by the operator via a proximal handle and activation wire. Radiopaque markers on the basket and catheter facilitate fluoroscopic visualization. The device is designed for 4-8mm diameter vessels and is compatible with 0.014" guidewires. It is intended for single-use in a clinical setting, operated by a physician. By mechanically displacing thrombus, the device facilitates aspiration, potentially restoring blood flow and benefiting patients with peripheral vascular occlusions.
Clinical Evidence
No human clinical data provided. Evidence consists of in vitro bench testing (visual/dimensional inspection, kink resistance, torsional strength, tensile testing, simulated use) and an acute GLP porcine animal study. Bench testing confirmed performance similar to predicate/reference devices. Animal study demonstrated device usability, non-thrombogenicity, and safety in peripheral vasculature.
Technological Characteristics
Mechanical thrombectomy catheter with nitinol (NiTi) braided basket. Features proximal handle for basket activation, 0.014" guidewire compatibility, and radiopaque markers. Dimensions: 40mm basket length, 4-8mm expanded diameter. Compatible with 7F or 8F aspiration catheters. Sterilization: Ethylene Oxide (EO). Single-use.
Indications for Use
Indicated for non-surgical removal of soft emboli and thrombi from peripheral blood vessels.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 23, 2022
Contego Medical Inc. Jim Clossick Vice President of Regulatory Affairs 3801 Lake Boone Trail, Suite 100 Raleigh, North Carolina 27607
Re: K221204
Trade/Device Name: Excipio SV Thrombectomy Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW Dated: April 25, 2022 Received: April 26, 2022
Dear Jim Clossick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221204
Device Name Excipio SV Thrombectomy Device
#### Indications for Use (Describe)
The Excipio SV Thrombectory Device is indicated for the non-surgical removal of soft emboli and thrombi from peripheral blood vessels
| Type of Use (Select one or both, as applicable) | <table><tr><td><div> <input type="checkbox"/> </div></td><td></td></tr></table> | <div> <input type="checkbox"/> </div> | |
|-------------------------------------------------|---------------------------------------------------------------------------------|---------------------------------------|--|
| <div> <input type="checkbox"/> </div> | | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Contego Medical. The logo features the company name in a combination of black and blue lettering, with a stylized blue arc above the word "Contego". Below the company name, the tagline "INTEGRATED EMBOLIC PROTECTION" is written in smaller, capitalized letters.
# 510(k) Summary [as required per 21 CFR 807.92]
### 510(K) K221204
| Date Prepared | April 22, 2022 |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Contego Medical, Inc.<br>3801 Lake Boone Trail, Suite 100<br>Raleigh, NC 27607 |
| FDA RegistrationNumber | 3011471056 |
| Contact Person: | Mr. Jim Clossick<br>Contego Medical, Inc.<br>3801 Lake Boone Trail, Suite 100<br>Raleigh, NC 27607<br>Phone: + 1 305 607 1708<br>Email: Jclossick@contegomedical.com |
| Proprietary Name: | Excipio SV Thrombectomy Device |
| Common Name: | Embolectomy Aspiration Device |
| Device Classification: | Class II per 21 CFR §870.5150 |
| Classification Name: | Peripheral Mechanical Thrombectomy With Aspiration |
| Product Code: | QEW |
| Predicate Device: | ReVive PV Thrombectomy Device (K132281) |
| Reference Device: | Penumbra Indigo Aspiration System (K142870)<br>Fogarty Arterial Embolectomy Catheter<br>(K193379) |
### Intended Use/Indications for Use:
The Excipio SV Thrombectomy Device is indicated for the non-surgical removal of soft emboli and thrombi from peripheral blood vessels.
### Device Description:
The Excipio SV Thrombectomy Device consists of a Thrombectomy Catheter to mechanically displace thrombus when used with an aspiration catheter. The Device
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Contego Medical. The word "Contego" is in black font, and the word "Medical" is in blue font. Above the word "Contego" is a blue swoosh. Below the words "Contego Medical" is the phrase "INTEGRATED EMBOLIC PROTECTION" in smaller, gray font.
K221204 Traditional 510(k) Summary Excipio SV Thrombectomy Device June 22, 2022
will be sold as a sterile, single use device.
### Thrombectomy Catheter
The Thrombectomy Catheter is a mechanical thrombus displacement device with a nitinol braided component (basket) at the distal end that can be opened and closed via an activation wire that attaches to the distal end of the braid and attaches to a proximal handle. The operator can open the braided component to a diameter that best matches the target vessel (4-8mm in diameter). Proximal and distal marker bands delineate the ends of the basket component, and 4 radiopaque markers delineate the circumference of the basket component to facilitate visualization under fluoroscopy.
#### Comparison with Predicate Devices
A comparison of the Excipio SV Thrombectomy Device and the predicate and reference devices show that the technological characteristics of the subject devices such as the design, sterilization method, and operating principles are similar to the currently marketed predicate and reference devices. The intended use of the subject device falls within the intended use of the predicate device.
| Device Name | Excipio SV<br>Thrombectomy<br>Device | ReVive PT<br>Thrombectomy<br>Device | Penumbra Indigo<br>Aspiration System | Fogarty Arterial<br>Embolectomy<br>Catheter |
|--------------------------------------|-------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Device | Subject Device | Predicate<br>Device | Reference Device | Reference Device |
| Manufacturer | Contego Medical | Codman & Shurtleff,<br>Inc. | Penumbra Inc. | Baxter Healthcare<br>Corp. |
| 510(k) Number | K221204 | K132281 | K142870 | K193379 |
| Class/Product Code | II/ QEW | II/QEW | II/QEW | II/DXE, GCA |
| Device Classification<br>Description | Peripheral<br>mechanical<br>thrombectomy with<br>aspiration | Peripheral mechanical<br>thrombectomy with<br>aspiration | Peripheral<br>mechanical<br>thrombectomy with<br>aspiration | Embolectomy<br>Catheter |
| Indications for Use | Non-surgical removal<br>of soft emboli and<br>thrombi from<br>peripheral blood<br>vessels | Indicated for:<br>non-surgical removal<br>of emboli and thrombi<br>from peripheral blood<br>vessels<br>non-surgical removal<br>of thrombi from<br>synthetic grafts | Removal of fresh,<br>soft emboli and<br>thrombi from vessels<br>in the arterial system | Removal of fresh,<br>soft emboli and<br>thrombi from vessels<br>in the arterial<br>system. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Contego Medical. The logo features the company name in a stylized font, with the word "Contego" in black and "Medical" in blue. Above the word "Contego" is a blue curved line. Below the company name is the tagline "INTEGRATED EMBOLIC PROTECTION" in smaller, black font.
#### K221204 Traditional 510(k) Summary Excipio SV Thrombectomy Device June 22, 2022
| Device Name | Excipio SV<br>Thrombectomy<br>Device | ReVive PT<br>Thrombectomy<br>Device | Penumbra Indigo<br>Aspiration System | Fogarty Arterial<br>Embolectomy<br>Catheter |
|-------------------------------------------------------|-----------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------------------------|
| Device | Subject Device | Predicate<br>Device | Reference Device | Reference Device |
| | | • temporary use in<br>peripheral vessel/graft<br>occlusion<br><br>• with aspiration and<br>with the injection or<br>infusion of contrast<br>media and other fluids | | |
| Principle of operation<br>- Thrombectomy<br>mechanism | Compressed NiTi<br>Basket | Compressed NiTi<br>Basket | Separator (for use in<br>catheter lumen) | Inflatable Balloon |
| Expanded diameter | 8 mm max | 4.5mm | 1.14 – 2.67 mm | 4 – 14 mm |
| Thrombectomy<br>mechanism length | 40 mm | 22-28mm | Unknown | Unknown |
| Artery diameters to<br>be treated | 4 – 8 mm | 1.5 – 5mm | Unknown | 4 – 14 mm |
| Guidewire<br>compatibility | 0.014" | 0.014" | 0.035" | N/A |
| Radiopaque markers | Proximal and distal to<br>basket and on basket<br>circumference | Proximal and distal to<br>basket | Unknown | Barium added to<br>catheter for<br>radiopacity |
| Aspiration Catheter<br>diameter | 7F or 8F | N/A | 3 – 8F | N/A |
| Sterilization | EO | E-beam | Unknown | Unknown |
## Non-Clinical Testing/Performance Data:
Non-clinical laboratory testing was performed on the Excipio SV Thrombectomy Device to determine substantial equivalence. The following testing/assessment were performed:
- Visual Inspection ●
- Dimensional Inspection .
- Kink Resistance .
- Torsional Strength ●
- Tensile Testing ●
- Simulated Use .
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for Contego Medical. The logo features the company name in a combination of black and blue text, with the word "Contego" in black and "Medical" in blue. Above the company name is a blue swoosh. Below the company name is the text "INTEGRATED EMBOLIC PROTECTION" in smaller, all-caps letters.
K221204 Traditional 510(k) Summary Excipio SV Thrombectomy Device June 22, 2022
The in vitro bench tests demonstrated that the Excipio SV Thrombectomy Devicemet all acceptance criteria and performed similarly to the predicate and referencedevices. Performance data demonstrate that the device functions as intended and has a safety and effectiveness profile that is similar to the predicate and reference devices.
### Biocompatibility:
Testing was performed to assess biocompatibility of the Excipio SVThrombectomy Device. The following biocompatibility tests were performed:
- Cytotoxicity ●
- Sensitization ●
- Irritation ●
- Acute System Toxicity ●
- Material Mediated Pyrogenicity
- Hemolysis
- Complement Activation ●
- Thrombogenicity ●
The results from the testing performed showed the Excipio SV Thrombectomy Device to be biocompatible.
#### GLP Animal Study:
An acute GLP animal study (porcine) was conducted to evaluate in vivo thrombogenicity and acute performance and safety of the Excipio Device.
The study provides evidence that the Excipio Device was usable, non-thrombogenic, and did not raise any new questions of safety in the peripheral vasculature of the porcine model. Therefore, the device is substantially equivalent to the predicate device.
#### Conclusion:
The Excipio SV Thrombectomy Device has a similar intended use and the same or similar technological characteristics such as design, sterilization method, and operating principles as the predicate and reference devices. Performance data demonstrates that the device functions as intended. The conclusions drawn from the nonclinical test demonstrate that the Excipio SV Thrombectomy Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.