Fogarty Arterial Embolectomy Catheter: The Fogarty Arterial Embolectorny Catheter is indicated for the removal and thrombi from vessels in the arterial system. To remove fibrous or adherent material, alternative devices such as the Fogarty Adherent Clot and Graft Thrombectomy Catheter are recommended. Fogarty Biliary Balloon Probe: The Fogarty Biliary Balloon Probe is indicated for the removal of stones and ductal debris and exploration.
Device Story
Fogarty Arterial Embolectomy Catheter and Fogarty Biliary Balloon Probe are manual surgical instruments used in clinical environments. The Arterial Embolectomy Catheter features a PVC or nylon body (40-80 cm) with a distal latex balloon and proximal hub for syringe-based inflation; a soft, rounded tip facilitates arterial insertion. The Biliary Balloon Probe consists of a single-lumen PVC catheter with a distal latex balloon and a flexible stylet for navigation through sharp angulations. Both devices are single-use, sterile, and manually operated by clinicians. The balloon is inflated via syringe to engage and remove thrombi or stones from the target vessel or duct. These devices provide mechanical assistance for clot or stone extraction, aiding in the restoration of flow or clearance of obstructions.
Clinical Evidence
Bench testing only. Evaluations included product performance testing, balloon inflation testing, packaging validation (ISO 11607-1), sterilization validation (ISO 11135:2014), and ethylene oxide/ethylene chlorohydrin residuals evaluation (ISO 10993-7:2008).
Technological Characteristics
Catheter body: PVC or nylon (3F only). Balloon: Latex. Stylet: Flexible material for stiffness/navigation. Energy source: Manual (syringe-based inflation). Connectivity: None. Sterilization: Ethylene oxide (EO). Form factor: 40, 60, or 80 cm lengths.
Indications for Use
Indicated for removal of fresh, soft emboli and thrombi from arterial vessels (Arterial Embolectomy Catheter) and removal of stones and ductal debris and exploration (Biliary Balloon Probe).
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
Predicate Devices
Pre-amendment devices
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 12, 2020
Edwards Lifesciences, LLC Wendy Gonzalez Associate Manager, Regulatory Affairs One Edwards Way Irvine, California 92614
# Re: K193379
Trade/Device Name: Fogarty Arterial Embolectomy Catheter, Fogarty Biliary Balloon Probe Regulation Number: 21 CFR 870.5150 and 21 CFR 876.5010 Regulation Name: Embolectomy Catheter and Biliary Catheter and Accessories Regulatory Class: Class II Product Code: DXE, GCA Dated: May 18, 2020 Received: May 20, 2020
Dear Ms. Gonzalez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known) K193379
#### Device Name
Fogarty Arterial Embolectomy Catheter and Fogarty Biliary Balloon Probe
Indications for Use (Describe)
Fogarty Arterial Embolectomy Catheter:
The Fogarty Arterial Embolectorny Catheter is indicated for the removal and thrombi from vessels in the arterial system. To remove fibrous or adherent material, alternative devices such as the Fogarty Adherent Clot and Graft Thrombectomy Catheter are recommended.
Fogarty Biliary Balloon Probe:
The Fogarty Biliary Balloon Probe is indicated for the removal of stones and ductal debris and exploration.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Fogarty Arterial Embolectomy Catheter and Fogarty Biliary Balloon Probe | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| 510(k) Submitter | Edwards Lifesciences, LLC | |
| Contact Person | Wendy Vera González<br>Associate Manager, Regulatory Affairs<br>Edwards Lifesciences<br>(949) 250- 2500 | |
| Date Prepared | June 9, 2020 | |
| Trade Name | Fogarty Arterial Embolectomy<br>Catheter | Fogarty Biliary Balloon Probe |
| Regulation<br>Description | Embolectomy Catheter | Biliary catheter and accessories |
| Regulation Class /<br>Product Code | Class II<br>DXE | Class II<br>GCA |
| Predicate<br>Device(s) | Pre-amendment | Pre-amendment |
| Regulation<br>Number | 21 CFR 870.5150 | 21 CFR 876.5010 |
| Device<br>Description | Fogarty Arterial Embolectomy Catheter<br>The Fogarty Arterial Embolectomy Catheter consists of a catheter<br>body tube made out of polyvinyl chloride (PVC) or nylon (3F only)<br>of 40, 60, or 80 cm lengths.<br><br>The catheter contains a latex balloon at the distal end and a PVC<br>hub at the proximal end for connection to a syringe for balloon<br>inflation. They contain a soft, rounded tip, which promotes easy<br>arterial insertion.<br><br>The model with an “S” has a stylet at the proximal end to support<br>the inflation lumen during shipping and storage.<br><br>These devices are provided sterile and labeled for single use only. | |
| Fogarty Biliary Balloon Probe | | |
| The Fogarty Biliary Balloon Probe consists of a single-lumen<br>polyvinyl chloride (PVC) catheter body with a latex balloon at the<br>distal end. A hub at the proximal end is used for balloon inflation. | | |
| A flexible stylet is provided with Fogarty Biliary Balloon Probe.<br>The stylet is of a designed stiffness intended to avoid perforation<br>and yet maintain enough shape to facilitate introduction of the<br>probe into areas of sharp angulation. These devices are provided<br>sterile and labeled for single use only. The Fogarty Biliary Balloon<br>Probe is not intended to be used with other devices. | | |
| Indications for<br>Use/Intended Use | The Fogarty Arterial Embolectomy Catheter is indicated for the<br>removal of fresh, soft emboli and thrombi from vessels in the<br>arterial system. To remove fibrous or adherent material, alternative<br>devices such as the Fogarty Adherent Clot and Graft<br>Thrombectomy Catheter are recommended. | |
| | The Fogarty Biliary Balloon Probe is indicated for the removal of<br>stones and ductal debris and exploration. | |
| Comparative<br>Analysis | The subject device is identical to the predicate devices in terms of<br>intended use, indications for use, and technology. The proposed<br>changes involve the creation of new model numbers for the existing<br>legally marketed device with a new packaging configuration, new<br>labels and updates to the IFU as a result of the new model numbers. | |
| | Packaging design verification, sterilization (EO resistance and<br>residuals), product performance and biocompatibility evaluations<br>were conducted to ensure that the change in final packaging did not<br>alter the performance of the subject devices. | |
| | The subject Fogarty Arterial Embolectomy Catheter and Fogarty<br>Biliary Balloon Probe, have been shown to be substantially<br>equivalent to the predicate devices for its intended use in hospitals<br>and other appropriate clinical environments. | |
| Functional/ Safety<br>Testing | The Fogarty Arterial Embolectomy Catheter and Fogarty Biliary<br>Balloon Probe with new packaging configuration are identical to the<br>predicate device in terms of final device and indications for use. To<br>ensure that the Fogarty Arterial Embolectomy Catheter and Fogarty<br>Biliary Balloon Probe with new packaging configuration did not<br>raise any new concerns of safety and effectiveness, and are<br>substantially equivalent to the predicate devices, the following<br>evaluation and testing were performed:<br>Product Performance testing evaluation and Product Test<br>/Balloon Inflation test to demonstrate final device is not<br>affected. | |
| | > Packaging testing in accordance to ISO 11607-1 – Packaging<br>for Terminally Sterilized Medical Devices.<br>> Sterilization testing in accordance to ISO 11135:2014 -<br>Sterilization of health-care products- Ethylene oxide-<br>Requirements for the development, validation and routing<br>control of a sterilization process for medical devices.<br>> EO/ ECH residuals evaluation in accordance to ISO10993-<br>7:2008 - Biological evaluation of medical devices – Part 7:<br>Ethylene oxide sterilization residuals. | |
| | The Fogarty Arterial Embolectomy Catheter and Fogarty Biliary<br>Balloon Probe models have successfully passed performance,<br>packaging, sterility assurance and EO/ECH residuals testing<br>demonstrating that the subject devices are substantially equivalent to<br>the predicate devices. | |
| Conclusion | The Fogarty Arterial Embolectomy Catheter and Fogarty Biliary<br>Balloon Probe are substantially equivalent to the predicate device. | |
{4}------------------------------------------------
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.