K221139 · Feliks Plastik Laminasyon VE Ambalaj Malzemeleri · FYA · Nov 10, 2022 · General, Plastic Surgery
Device Facts
Record ID
K221139
Device Name
Bodygard SFS Surgical Gown Level 4
Applicant
Feliks Plastik Laminasyon VE Ambalaj Malzemeleri
Product Code
FYA · General, Plastic Surgery
Decision Date
Nov 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The Bodygard SFS Surgical Gown Level 4 is intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, Bodygard SFS Surgical Gown Level 4 met the requirements for Level 4 classification.
Device Story
Bodygard SFS Surgical Gown Level 4 is a single-use, poly-reinforced surgical gown; constructed from SMS nonwoven/film/SMS nonwoven (SFS) material; manufactured via ultrasonic bonding. Provides AAMI Level 4 liquid barrier protection in critical (arms/chest) and non-critical zones. Used in operating rooms by surgical personnel to prevent cross-contamination of microorganisms, body fluids, and particulates between patient and staff. Supplied sterile via Ethylene Oxide (EO).
Clinical Evidence
No clinical data. Bench testing only. Performance verified via AATCC 127 (hydrostatic pressure), AATCC 42 (water impact), ASTM D5034-09 (breaking strength), ASTM D5587-14 (tear strength), ASTM D1683-17 (seam strength), ASTM F1671 (viral penetration), and ISO 10993 series (biocompatibility/cytotoxicity/irritation/sensitization).
Indicated for operating room personnel during surgical procedures to protect patient and personnel from transfer of microorganisms, body fluids, and particulate material. No specific age or gender restrictions.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 10, 2022
Feliks Plastik Laminasyon Ve Ambalaj Malzemeleri % Darren Reeves President DP Distribution & Consulting, LLC 12240 Hunting Horn Lane Rockville, Virginia 23146
Re: K221139
Trade/Device Name: Bodygard SFS Surgical Gown Level 4 Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: October 7, 2022 Received: October 11, 2022
Dear Darren Reeves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Bifeng Qian -S
Bifeng Qian, M.D., Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K221139
Device Name Bodygard SFS Surgical Gown Level 4
#### Indications for Use (Describe)
The Bodygard SFS Surgical Gown Level 4 is intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, Bodygard SFS Surgical Gown Level 4 met the requirements for Level 4 classification.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PSC Publishing Services (301) 443-6740
FF
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
# K221139 510(k) Summary Traditional 510(k)
In accordance with 210 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Bodygard SFS Surgical Gown Level 4 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Feliks Plastik Lam Ve Amb Mal San Ve Tic A. S.<br>Eskishir Organize Sanayi Bolgesi 26. Cad No. 9 26110<br>Eskisehir |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Ali Serdar Serteser<br>Felix@felixplastic.com<br>+90 533 209 72 95<br>Eskisehir Industrial Zone 26st No 9<br>Odunpazari, Eskisehir Turkey |
| Contact Person/<br>Prepared by: | Darren Reeves<br>President<br>DP Distribution & Consulting, LLC ®<br>(804) 307-7706<br>dreeves@dpdconline.com<br>12240 Hunting Horn Lane<br>Rockville, Virginia 23146 |
| Preparation Date: | 11/08/2022 |
| Subject Device: | Trade Name: Bodygard SFS Surgical Gown Level 4<br>510(k) #: K221139<br>Common Name: Surgical Gown<br>Classification Name: Surgical Gown (21 CFR 878.4040,<br>Product Code FYA) |
| Predicate Device: | Disposable Surgical Gown, Disposable Reinforced Surgical<br>Gown (K212869) |
{4}------------------------------------------------
### Device Description:
The Bodygard SFS Surgical Gown Level 4 is a single use poly reinforced surgical gown, SMS Nonwoven/Film/SMS Nonwoven (SFS) that provides AAMI Level 4 liquid barrier protection in the critical zones (arms and chest) and non-critical zones of the gown. The Bodygard SFS Surgical Gown Level 4 is manufactured using ultrasonic bonding technique are available in the color blue and four different sizes (M, L, XL, and XXL).
Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities. The Bodygard SFS Surgical Gown Level 4, meets the requirements for Level 4 classification, are disposable medical devices and provided in sterile.
### Intended Use and Indication for Use:
The Bodygard SFS Surgical Gown Level 4 is intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, Bodygard SFS Surgical Gown Level 4 met the requirements for Level 4 classification.
## Comparison of Predicate Device:
| Elements of<br>Comparison | Proposed Device<br>Bodygard SFS Surgical<br>Gown Level 4 | Predicate Device K212869<br>Disposable Surgical Gown,<br>Disposable Reinforced Surgical<br>Gown (Sterile) | Remark |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| 510 (k) Number | K221139 | K212869 | N/A |
| Product Code | FYA | FYA | Same |
| Regulation<br>Number | 21CFR 878.4040 | 21CFR 878.4040 | Same |
| Indication for Use | | | |
| | The Bodygard SFS Surgical<br>Gown Level 4 is intended<br>to be worn by operating<br>room personnel during<br>surgical procedure to pro-<br>tect both the surgical pa-<br>tient and the operating<br>room personnel from<br>transfer of microorgan-<br>isms, body fluids,<br>and particulate material.<br><br>Per ANSI/AAMI PB70:2012<br>Liquid barrier performance<br>and classification of pro-<br>tective apparel and drapes<br>intended for use in health<br>care facilities, Bodygard<br>SFS Surgical Gown Level 4<br>met the requirements for<br>Level 4 classification. | Disposable Surgical Gown and<br>Disposable Reinforced Surgical<br>Gown are intended to be worn by<br>operating room personnel during<br>surgical procedure to protect<br>both the surgical patient and the<br>operating room personnel from<br>transfer of microorganisms, body<br>fluids, and particulate material.<br><br>Per ANSI/AAMI PB70:2012 Liquid<br>barrier performance and classifi-<br>cation of protective apparel and<br>drapes intended for use in health<br>care facilities, Disposable Surgical<br>Gown ML515M45U met the re-<br>quirements for Level 3 classifica-<br>tion, Disposable Surgical Gown<br>GD524ME65 and Disposable Re-<br>inforced Surgical Gown met the<br>requirements of Level 4 classifi-<br>cation. | Similar |
| Style | Poly Reinforced | Non-reinforced/Reinforced | Similar |
| Durability | Disposable | Disposable | Same |
| Color | Blue | Blue | Same |
| Material | Complete gown has 4 com-<br>ponents/Component A<br>Fabric SMS Nonwoven/<br>Lamination Reinforcement<br>Film/SMS Nonwoven<br>(SFS)/Component B Knit<br>Cuff Component C<br>Velcro/Component D<br>Sewing thread SMS | Level 4 Standard Surgical Gown:<br>SMS nonwoven, PE film,<br>Polyester and blue masterbatch;<br>Level 4 Reinforced Surgical<br>Gown: SMS nonwoven Polyester,<br>PE film reinforced film and Blue<br>masterbatch | Similar |
| Weight per<br>square | $60 g/m^2$ | $65 g/m^2$ | Similar |
| Label and<br>Labeling | Conforms with 21 CFR Part<br>801 | Conforms with 21 CFR Part 801 | Same |
| Size | M, L, XL, and XXL | S, M, L, XL, XXL, and XXXL | Similar |
| Break Strength<br>(ASTM D 5034-<br>09) | >30 N | >30 N | Same |
| Tear Strength<br>(ASTM D5587-14) | >20 N | >20 N | Same |
| Seam Strength<br>(ASTM D1683-17) | >50 N | >50 N | Same |
| Linting (ISO 9073-10:2003) | Log10 < 4 | Log10 < 4 | Same |
| Flammability (16 CFR 1610) | Class I | Class I | Same |
| Hydrostatic pressure (AATCC 127) | >50 cm | >50 cm | Same |
| Water impact<br>(AATC Test<br>Method 42) | ≤1.0 g | ≤1.0 g | Same |
| Barrier Penetration<br>(ASTM F1671) | No detectable transfer of the Phi-X174 Bacteriophage | No detectable transfer of the Phi-X174 Bacteriophage | Same |
| Barrier Protection Level | Level 4 per AAMI PB 70 | Level 4 per AAMI PB 70 | Same |
| Water Vapor Transmission<br>(ASTM D6701) | Average transmission/per-<br>meation rate 4.360 g/m² day | Unknown | Unknown |
| Elements of Comparison | Proposed Device | Predicate Device | Remark |
| Biocompatibility | | | |
| Cytotoxicity (ISO 10993-5) | No Cytotoxicity | No Cytotoxicity | Same |
| Skin Irritation (ISO 10993-10) | No Irritation | No Irritation | Same |
| Sensitization (ISO 10993, 10993-1, and 10993-10) | No Sensitization | No Sensitization | Same |
| Sterile | Ethylene Oxide (EO),<br>SAL=10-6 Sterile | Ethylene Oxide (EO), SAL=10-6<br>Sterile | Same |
#### Table 2 Technological Characteristic Comparison
{5}------------------------------------------------
{6}------------------------------------------------
# Summary of Non-Clinical Test:
## Non-Clinical Tests:
The product was tested in alignment with "Guidance on Premarket Notification 510(k) Submissions for Surgical Gowns and Surgical" Guidance Document
{7}------------------------------------------------
| Table 3: | | | |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test Method | Purpose | Acceptance Criteria | Results |
| AATCC 127 | Water resistance<br>Hydrostatic<br>Pressure | >50 cm H2O (AQL<br>4%, RQL=20%) | All test samples were<br>> 50 cm H2O |
| AATCC 42 | Water Resistance<br>impact<br>penetration | <1.0 g penetration<br>(AQL 4%, RQL=20%) | All test samples were<br>< 1 g penetration |
| ASTM D<br>5034-09 | Breaking<br>Strength | >30 N (AQL 4%,<br>RQL=20%) | All test samples were<br>>30N |
| ASTM D5587-<br>14 | Tearing Strength | >20 N (AQL 4%,<br>RQL=20%) | All test samples were<br>>20 N |
| 16 CFR 1610 | Flammability<br>testing | Class I | Meets Class I |
| ASTM D1683-<br>17 | Seam Strength | >50 N (AQL 4%,<br>RQL=20%) | All test samples were<br>>50 N |
| D6701-16 | Water vapor<br>transmission of<br>Nonwoven and<br>plastic Barriers | >500 g/m² Day<br>WVTR | Passed |
| ASTM<br>D3776/D | Mass Per Area<br>(Weight) of fab-<br>ric | Has met acceptance<br>criteria according to<br>ASTM F2407 | Has met acceptance<br>criteria according to<br>ASTM F2407 |
| ASTM F1670 | Resistance by<br>synthetic Blood | No Penetration at 2<br>psi (13.8 kPA) | Passed |
| ASTM F1671 | Resistance of<br>Materials Used in<br>Protective Cloth-<br>ing to Penetra-<br>tion by<br>Blood-Borne<br>Pathogens | No detectable trans-<br>fer of the Phi-X174<br>Bacteriophage | Passed |
| ISO 9073-<br>10:2003 | Lint and Other<br>particles genera-<br>tion in the dry<br>state | Log 10<4 | Below Log10<4<br>Passed |
| ASTM D4169-<br>16 | Performance<br>testing of ship-<br>ping containers<br>and systems | Products must with-<br>stand the distribu-<br>tion environment | Passed |
| | | | |
| ASTM F88-<br>07A | Seal strength of<br>Flexible Barrier<br>Materials | Package Seal in-<br>tegrity must be in-<br>tact. | Passed |
| ASTM F2096-<br>04 | Detecting Gross<br>Leaks in medical<br>packaging by in-<br>ternal pressuriza-<br>tion (Bubble test) | Package integrity<br>must be intact with-<br>out failed seal loca-<br>tions. | Passed |
| ASTM F1980-<br>07 | Accelerated Ag-<br>ing of Sterile Bar-<br>rier Systems for<br>medical devices | Package integrity<br>must be intact after<br>accelerated aging | Passed |
| ISO 10993-5 | Biological Evalua-<br>tion of medical<br>devices - Part 5:<br>tests for In vitro<br>cytotoxicity of<br>medical devices | Device must not be<br>cytotoxic | Under the condition<br>of the testing, the de-<br>vice is non-cytotoxic |
| ISO 10993-10 | Biological Evalua-<br>tion of medical<br>devices - Part<br>10: Tests for irri-<br>tation and skin<br>sensitization / Ir-<br>ritation | Device must not be<br>irritant | Under the condition<br>of the testing, the de-<br>vice is not an irritant |
| ISO 10993-10 | Biological Evalua-<br>tion of medical<br>devices - Part<br>10: Tests for irri-<br>tation and skin<br>sensitization/<br>sensitization | Device must not be<br>sensitizer | Under the condition<br>of the testing, the de-<br>vice is not a sensitizer |
| ISO 10993-7 | Biological Evalua-<br>tion of medical<br>devices- Part 7:<br>Ethylene Oxide<br>Sterilization<br>Residuals | Residual Ethylene<br>oxide levels must be<br>below limits EO <4<br>mg/device. The de-<br>vice met require-<br>ments of ISO 1099-<br>7:2008 | Ethylene Oxide resid-<br>ual levels are below<br>limitations. Data not<br>available, required to<br>meet the require-<br>ments of ISO 10993-<br>7:2008 |
{8}------------------------------------------------
{9}------------------------------------------------
## Summary of Clinical Tests:
No clinical tests were performed.
#### Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the device Bodygard SFS Surgical Gown Level 4 is as safe, as effective, and performs as well as or better than the legally marketed predicate device Disposable Surgical Gown, Disposable Reinforced Surgical Gown.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.