Disposable Surgical Gown and Disposable Reinforced Surgical Gown are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, Disposable Surgical Gown ML515M45U met the requirements for Level 3 classification, Disposable Surgical Gown GD524ME65 and Disposable Reinforced Surgical Gown met the requirements of Level 4 classification.
Device Story
Disposable surgical gowns (reinforced and non-reinforced) designed for use by operating room personnel; intended to provide a physical barrier against microorganisms, body fluids, and particulates during surgery. Devices are single-use, sterile, and available in six sizes (S-XXXL). Principle of operation relies on material barrier properties (SMS nonwoven, PE film) to meet ANSI/AAMI PB70 Level 3 or Level 4 liquid barrier performance standards. Healthcare providers wear the gown as personal protective equipment; output is the physical protection of the wearer and patient. Benefits include reduced risk of cross-contamination and infection transmission in the surgical environment.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via AATCC 127 (hydrostatic pressure), AATCC 42 (water impact), ISO 9073-10 (linting), ASTM D5034 (breaking strength), ASTM D5587 (tearing strength), ASTM D737 (air permeability), and ASTM F1671 (blood-borne pathogen resistance). Biocompatibility confirmed via ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
Materials: SMS nonwoven, PE film, polyester, blue masterbatch. Barrier levels: AAMI PB70 Level 3 and Level 4. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Physical properties: Class I flammability, >50cm hydrostatic pressure, <1.0g water impact. Sizes: S-XXXL.
Indications for Use
Indicated for operating room personnel during surgical procedures to protect the patient and personnel from the transfer of microorganisms, body fluids, and particulate material.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
Predicate Devices
Level 2 Standard Surgical Gown, Level 3 Standard Surgical Gown (K211422)
SMS Standard Surgical Gown; SMS Surgical Gown with Reinforcement; BVB Surgical Gown; BVB Splicing Surgical Gown (K192290)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
December 21, 2021 Fugou County Shenxiang Manufacturing Co., LTD. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.box 120-119 Shanghai, 200120 China
Re: K212869
Trade/Device Name: Disposable Surgical Gown, Disposable Reinforced Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical apparel Regulatory Class: Class II Product Code: FYA
Dear Diana Hong:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter for your device cleared on December 13, 2021. Specifically, FDA is updating this SE letter due to the clearance date not appearing on the original letter.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Clarence W. Murray III, Assistant Director of Office of Surgical and Infection Control Devices, at Tel: 301-796-0270 or Email: Clarence.Murrav@fda.hhs.gov.
Sincerely,
# Bifeng Qian -S
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Fugou County Shenxiang Manufacturing Co., LTD. % Diana Hong General Manager Mid-Link Consulting Co.,Ltd P.O.box 120-119 Shanghai, 200120 China
Re: K212869
Trade/Device Name: Disposable Surgical Gown, Disposable Reinforced Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FYA Dated: August 19, 2021 Received: September 9, 2021
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bifeng Qian -S
For Clarence W. Murray, III, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K212869
Device Name
Disposable Surgical Gown, Disposable Reinforced Surgical Gown
#### Indications for Use (Describe)
Disposable Surgical Gown and Disposable Reinforced Surgical Gown are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, Disposable Surgical Gown ML515M45U met the requirements for Level 3 classification, Disposable Surgical Gown GD524ME65 and Disposable Reinforced Surgical Gown met the requirements of Level 4 classification.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|× Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{4}------------------------------------------------
# K212869
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
- 1. Date of Preparation: 12/13/2021
- 2. Sponsor Identification
Fugou County Shenxiang Manufacturing Co., LTD. Industrial Clusters, No.2 Road, Fugou County, Henan Province, China
Establishment Registration Number: 3017149307
Contact Person: Carina Si Position: Business Manager Tel: +86- 21-54889822 Fax: +86- 21-54889180 Email: mike.xu(@fgshenxiang.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Jinlei Tang (Alternative Contact Person)
#### Mid-Link Consulting Co., Ltd.
P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 360-925-3199 Email: info@mid-link.net
{5}------------------------------------------------
#### Identification of Proposed Device 4.
Trade Name: Disposable Surgical Gown, Disposable Reinforced Surgical Gown Common Name: Surgical Gown
#### Regulatory Information
Classification Name: Gown, Surgical Classification: II; Product Code: FYA; Regulation Number: 21 CFR 878.4040 Review Panel: General Hospital;
#### Indication for use:
Disposable Surgical Gown and Disposable Reinforced Surgical Gown are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.
Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, Disposable Surgical Gown ML515M45U met the requirements for Level 3 classification, Disposable Surgical Gown GD524ME65 and Disposable Reinforced Surgical Gown met the requirements of Level 4 classification.
Device Description:
The proposed devices are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The proposed devices are single use, disposable medical devices and are provided in sterile.
There are two types of surgical gown: Disposable Surgical Gown and Disposable Reinforced Surgical Gown. And each type of surgical gown is available in six product sizes, including S, M, L, XL, XXL and XXXL. Disposable Surgical Gown have two models: ML515M45U and GD524ME65. Disposable Reinforced Surgical Gown have one model: PA528ME45R. Disposable Surgical Gown ML515M45U met the requirements for Level 3 classification, Disposable Surgical Gown GD524ME65 and Disposable Reinforced Surgical Gown met the requirements of Level 4 classification.
Table 1. Surgical Gowns Description
| Proposed | Model | Size | Color | Style | AAMI | Material |
|----------|-------|------|-------|-------|------|----------|
|----------|-------|------|-------|-------|------|----------|
2 / 13
{6}------------------------------------------------
| device | | | | Level | | |
|--------------------------------|----------------------------------------------|------------------------------|------------------------------|----------------|------------|---------------------------------------------------------------------------|
| Disposable<br>Surgical<br>Gown | ML515M45U | S, M, L, XL,<br>XXL,<br>XXXL | Blue | Non-reinforced | 3 | 45g/m² PP SMS<br>nonwoven, Polyester,<br>Blue masterbatch |
| | GD524ME65 | S, M, L, XL,<br>XXL,<br>XXXL | Blue | Non-reinforced | 4 | 45g/m² SMS<br>nonwoven, 20g/m² PE<br>film, Polyester, Blue<br>masterbatch |
| | Disposable<br>Reinforced<br>Surgical<br>Gown | PA528ME45R | S, M, L, XL,<br>XXL,<br>XXXL | Blue | Reinforced | 4 |
#### 5. Identification of Predicate Devices
Predicate Device 1 510(k) Number: K211422 Product Name: Level 2 Standard Surgical Gown, Level 3 Standard Surgical Gown
Predicate Device 2 510(k) Number: K192290 Product Name: SMS Standard Surgical Gown; SMS Surgical Gown with Reinforcement; BVB Surgical Gown; BVB Splicing Surgical Gown
Predicate Device 3 510(k) Number: K121152 Product Name: Surgical Gown
- 6. Summary of Clinical Testing
No clinical study is included in this submission.
{7}------------------------------------------------
#### 7. Summary of Technological characteristics
| Item | Proposed Device | Predicate Device K211422 | Remark |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code | FYA | FYA | Same |
| Regulation No. | 21CFR 878.4040 | 21CFR 878.4040 | Same |
| Class | II | II | Same |
| Indication for Use | Disposable Surgical Gown and<br>Disposable Reinforced Surgical<br>Gown are intended to be worn by<br>operating room personnel during<br>surgical procedure to protect both<br>the surgical patient and the<br>operating room personnel from<br>transfer of microorganisms, body<br>fluids, and particulate material.<br>Per ANSI/AAMI PB70:2012<br>Liquid barrier performance and<br>classification of protective apparel<br>and drapes intended for use in<br>health care facilities, Disposable<br>Surgical Gown ML515M45U met<br>the requirements for Level 3<br>classification, Disposable Surgical<br>Gown GD524ME65 and<br>Disposable Reinforced Surgical<br>Gown met the requirements of<br>Level 4 classification. | Surgical gown is intended to be<br>worn by operating room personnel<br>during surgical procedure to<br>protect both the surgical patient<br>and the operating room personnel<br>from transfer of microorganisms,<br>body fluids, and particulate<br>material.<br>Per ANSI/AAMI PB70:2012<br>Liquid barrier performance and<br>classification of protective apparel<br>and drapes intended for use in<br>health care facilities, the Level 2<br>standard surgical gowns met the<br>requirements for Level 2<br>classification, the Level 3 standard<br>surgical gowns and Level 3<br>reinforced surgical gowns met the<br>requirements for Level 3<br>classification. | Same |
| Style | Non-reinforced | Non-reinforced | Same |
| Durability | Disposable | Disposable | Same |
| Color | Blue | Blue | Same |
| Labeling | Conform with 21CFR Part 801 | Conform with 21CFR Part 801 | Same |
| Weight per square (g) | 45g/m2 | Level 3 Standard Surgical Gown:<br>43g/m² | Different |
| Size | S, M, L, XL, XXL, XXXL | XS, S, M, L, XL, XXL, XXXL | Different |
| Flammability | Class I | Class I | Same |
| Hydrostatic pressure | >50 cm | Level 3 Standard Surgical<br>Gown: >50 cm | Same |
| Water impact | <1.0 g | <1.0 g | Same |
| Table 1 Comparison of Technology Characteristics for Level 3 Surgical Gown | |
|----------------------------------------------------------------------------|--|
| | |
4 / 13
{8}------------------------------------------------
| Breaking strength | >20N | >20N | Same |
|-----------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-----------|
| Tearing strength | >20N | >20N | Same |
| Linting | $Log_{10}<4$ | $Log_{10}(particle count) <4$ | Same |
| Air permeability | >30 ft³/min/ft² | >30 ft³/min/ft² | Same |
| Barrier protection<br>level | Level 3 per AAMI PB 70 | Level 3 per AAMI PB 70 | Same |
| Material | SMS, Polyester, Blue masterbatch | Level 3 Standard Surgical Gown:<br>SMS nonwoven, Polyester and<br>Polyamide | Different |
| Biocompatibility | | | |
| Cytotoxicity | Under the conditions of the study, | Under the conditions of the study, | Same |
| Irritation | the device is non-toxic,<br>non-irritating, and | the device is non-toxic,<br>non-irritating, and non-sensitizing. | |
| Sensitization | non-sensitizing. | | |
| Sterilization | Sterile<br>Method: Ethylene Oxide (EO);<br>Sterilization Assurance Level<br>(SAL): 10-6 | Sterile<br>Method: Ethylene Oxide (EO);<br>Sterilization Assurance Level<br>(SAL): 10-6 | Same |
#### Different - Weight per square
The weight per square for the proposed surgical gowns is different from the predicate device. However, the difference in the weight per square will not affect the intended use. In addition, the performance testing results demonstrated that the proposed surgical gowns can meet the barrier protection level 3 requirement as required by PB70. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Size
The size for the proposed surgical gowns is different from the predicate device. The proposed surgical gowns are available in 6 product sizes, including S, M, L, XL, XXL and XXXL. However, the difference in the size will not affect the device performance. And the specifications of the proposed device can be covered by the predicated products. Different size can be selected by surgeon's condition. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
#### Different- Material
The material for the proposed surgical gowns is different from the predicated device. However, the biocompatibility test for proposed device was performed and the results showed no adverse effect. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
{9}------------------------------------------------
| Item | Proposed Device | Predicate Device K192290 | Predicate Device K121152 | Remark |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| Product Code | FYA | FYA | FYA | Same |
| Regulation No. | 21CFR 878.4040 | 21CFR 878.4040 | 21CFR 878.4040 | Same |
| Class | II | II | II | Same |
| Indication for Use | Disposable Surgical Gown and Disposable Reinforced Surgical Gown are intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. | Surgical gown is intended to be worn by operating room personnel during surgical procedure to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Per ANSI/AAMI PB70:2012 Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities, SMS Standard Surgical Gown and SMS Surgical Gown with Reinforcement met the requirements for Level 3 classification; BVB Surgical Gown and BVB Splicing Surgical Gown met the requirements for Level 4 classification. | Jiangsu Guangda's Reinforced Surgical Gowns, Model Number GD-SG-01, are non-sterile, single use surgical gowns intended to protect surgical patients and operating room personnel from the transfer of microorganisms, body fluids, and particulate material. This product may be sterilized using Ethylene Oxide (EO) following the Validation and routine control under ANSI/AAMI/ISO 11135. | Same |
| Style | Non-reinforced/Reinforced | Non-reinforced | Reinforced | Different |
| Durability | Disposable | Disposable | Disposable | Same |
| Color | Blue | Blue | Blue | Same |
| Labeling | Conform with 21CFR Part 801 | Conform with 21CFR Part 801 | Conform with 21CFR Part 801 | Same |
| Weight per | Level 4 Standard Surgical | 68 g/m² | 45 g/m² | Different |
| square (g) | Gown: 45g/m² and 20g/m²<br>PE film<br>Level 4 Reinforced Surgical<br>Gown: 45g/m² and 25g/m²<br>PE film reinforced film | | | |
| Size | S, M, L, XL, XXL, XXXL | M, L, XL, XXL, XXXL,<br>XXXL-XLONG | XL | Different |
| Flammability | Class I | Class I | Class I | Same |
| Hydrostatic<br>pressure | Level 4 Standard Surgical<br>Gown: >50 cm;<br>Level 4 Reinforced Surgical<br>Gown: >50 cm | >50 cm | >20 | Same |
| Water impact | ≤1.0 g | ≤1.0 g | ≤1.0 g | Same |
| Breaking strength | Level 4 Standard Surgical<br>Gown:<br>latitude: 111 N<br>longitude: 184 N<br>Level 4 Reinforced Surgical<br>Gown:<br>latitude: 109 N<br>longitude: 122 N | latitude: 92.3N<br>longitude: 177.11N | Passed | Different |
| Tearing strength | Level 4 Standard Surgical<br>Gown:<br>latitude: 90 N<br>longitude: 137 N<br>Level 4 Reinforced Surgical<br>Gown:<br>latitude: 80 N<br>longitude: 137 N | latitude: 35.30N<br>longitude: 56.46N | Passed | Different |
| Linting | Log10<4 | Log10<4 | Log10<4 | Same |
| Bacterial<br>Penetration | No detectable transfer of the<br>Phi-X174 Bacteriophage | No detectable transfer of the<br>Phi-X174 Bacteriophage | No detectable transfer of the<br>Phi-X174 Bacteriophage | Same |
| Barrier protection<br>level | Level 4 per AAMI PB 70 | Level 4 per AAMI PB 70 | Level 4 per AAMI PB 70 | Same |
| Material | Level 4 Standard Surgical<br>Gown: SMS nonwoven, PE<br>film, Polyester and blue<br>masterbatch;<br>Level 4 Reinforced Surgical<br>Gown: SMS nonwoven | SMS nonwoven, white<br>knitted cuff, white<br>spunbond, and BVB | SMS and PE+PP two layer<br>compound | Different |
| | Polyester, PE film reinforced<br>film and Blue masterbatch | | | |
| Biocompatibility | | | | |
| Cytotoxicity | Under the conditions of the<br>study, the device is<br>non-toxic, non-irritating, and<br>non-sensitizing. | No Cytotoxicity<br>No Irritation<br>No Sensitization | Passed | Same |
| Irritation | | | Passed | |
| Sensitization | | | Passed | |
| Sterilization | Sterile<br>Method: Ethylene Oxide (EO);<br>Sterilization Assurance Level<br>(SAL): 10-6 | Ethylene Oxide (EO) | Although sold non-sterile,<br>gowns can be EO Sterilized | Same |
# Table 2 Comparison of Technology Characteristics for Level 4 Surgical Gown
6 / 13
{10}------------------------------------------------
{11}------------------------------------------------
#### Different - Style
The style of the proposed device is different from the predicate device is available in two styles, non-reinforced and reinforced, which can be covered by predicate device K192290 and reference device K121152. The proposed device offers more options that physician can choose based on clinical conditions. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Weight per square
Although the weight per square of the proposed device and the weight per square of the predicate device are different, the difference in the weight per square will not affect the intended use. In addition, the performance testing results demonstrated that the proposed surgical gowns can meet the barrier protection level 4 requirement as required by PB70. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Size
The size for the proposed surgical gowns is different from the predicate device. The proposed surgical gowns are available in 6 product sizes, including S, M, L, XL, XXL and XXXL. However, the difference in the size will not affect the device performance. Different size can be selected by surgeon's condition. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Breaking strength
The longitude and latitude breaking strength of the Level 4 Standard Surgical Gown is larger than that of the predicate device K192290, so it is better than the predicate device. Although the Breaking strength
{12}------------------------------------------------
of the longitude of the Level 4 Reinforced Surgical Gown is smaller than that of the predicate device K192290, according to EN 13795-1:2019 Surgical clothing and drapes - Requirements and test methods, the acceptance criteria of breaking strength shall be more than 20N and the test result for Level 4 Reinforced Surgical Gown can meet this acceptance criteria. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
## Different - Tearing strength
Although the longitude and latitude tear strength of the proposed device is different from that of the predicate device K192290, the longitude and latitude tear strength of the proposed device is larger than that of predicate device K192290, so the tearing strength is better than that of the predicate device K192290. Therefore, although the data are different, this difference will not affect the safety and effectiveness of the proposed device.
#### Different - Material
The material for the proposed surgical gowns is different from the predicated device. However, the biocompatibility test for proposed device was performed and the results showed no adverse effect. Therefore, this difference will not affect the safety and effectiveness of the proposed device.
#### Summary of Non-Clinical Tests 8.
Non clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
- > 16 CFR Part 1610 Standard for the Flammability of Clothing Textiles;
- > AATCC 127: 2017 Water Resistance: Hydrostatic Pressure Test;
- > AATCC 42: 2017 Water Resistance: Impact Penetration Test;
- > ISO 9073-10: 2003 Textiles-Test Methods for Nonwovens-Part 10: Lint and Other Particles Generation in the Dry State;
- > ASTM D1683/D1683M: 2017(2018) Standard Test Method for Failure in Sewn Seams of Woven Fabrics;
- > ASTM D5587: 2015(2019) Standard Test Method for Tearing Strength of Fabrics by Trapezoid Procedure:
- > ASTM D5034: 2009(2017) Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test);
- > ASTM D737: 2018 Standard Test Method for Air Permeability of Textile Fabrics;
- > ASTM F1886/F1886M: 2016 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection;
- > ASTM F88/F88M: 2015 Standard Test Method for Seal Strength of Flexible Barrier Materials;
{13}------------------------------------------------
- > ASTM F…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.