K220959 · Smith & Nephew, Inc. · LPH · Aug 18, 2022 · Orthopedic
Device Facts
Record ID
K220959
Device Name
OR3O Dual Mobility System
Applicant
Smith & Nephew, Inc.
Product Code
LPH · Orthopedic
Decision Date
Aug 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The OR3O Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.
Device Story
Modular dual mobility hip implant system; consists of diffusion hardened oxidized zirconium (OXINIUM DH) liners and highly crosslinked ultra-high-molecular-weight polyethylene (XLPE) inserts. Designed to mate with existing acetabular shells and femoral heads. Provides two articulating surfaces within the joint space to improve stability and reduce dislocation risk. Intended for use in primary and revision total hip arthroplasty. Implanted without bone cement. Used by orthopedic surgeons in clinical settings. Line extension adds smaller liner and insert sizes to accommodate wider patient anatomy. Benefits include increased range of motion and reduced risk of dislocation compared to traditional single-bearing implants.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility evaluation per ISO 10993, and mechanical verification testing (ASTM F1820, ASTM F1875, ISO 7206-12, ASTM F2582, ISO 21535-2009, ISO 14242).
Technological Characteristics
Modular hip prosthesis. Liner: OXINIUM DH (Zr-2.5Nb alloy, ASTM F2384-10R16). Insert: XLPE (ASTM F648-14, ISO 5834-2), 10 Mrad irradiated and remelted. Locking taper design. Uncemented fixation. Single-use. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring primary or revision total hip arthroplasty due to advanced hip joint degeneration (degenerative, post-traumatic, or rheumatoid arthritis), femoral head fracture or avascular necrosis, failure of previous hip surgery (reconstruction, arthrodesis, hemiarthroplasty, surface replacement, or total hip replacement), osteoarthritis, hips at risk of dislocation, or proximal hip joint/femoral neck fractures.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
August 18, 2022
Smith & Nephew, Inc. Rose Beifuss Senior Manager, Regulatory Affairs 7135 Goodlett Farms Parkway Cordova, Tennessee 38016
Re: K220959
Trade/Device Name: OR30 Dual Mobility System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: LPH Dated: July 14, 2022 Received: July 15, 2022
Dear Rose Beifuss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
510(k) Number (if known)
K220959
Device Name
OR30 Dual Mobility System
Indications for Use (Describe) Intended Use The OR3O Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.
Indications
- · Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or rheumatoid arthritis.
- · Fracture or avascular necrosis of the femoral head.
- · Failure of previous hip surgery: joint reconstruction, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
- · All forms of osteoarthritis.
- · Patients with hips at risk of dislocation.
- · Femoral neck fracture or proximal hip joint fracture.
The OR30 Dual Mobility System is intended for single use only. The modular OR30 Liners and Inserts are to be implanted without bone cement.
Mating components may be indicated for use without bone cement.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <span style="font-size: 10pt;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15" x="0" y="0"></rect> <path d="M2,2 L13,13 M13,2 L2,13" stroke="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <span style="font-size: 10pt;"> <svg height="15" width="15"> <rect fill="none" height="15" stroke="black" width="15" x="0" y="0"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary Smith & Nephew OR3O Dual Mobility System
| I. | SUBMITTER | Smith & Nephew, Inc.<br>Orthopaedics Division<br>7135 Goodlett Farms Parkway<br>Cordova, Tennessee 38016 |
|----|-----------|----------------------------------------------------------------------------------------------------------|
|----|-----------|----------------------------------------------------------------------------------------------------------|
Phone: (385) 253-2551
Contact Person: Rose Beifuss, Senior Manager, Regulatory Affairs Date Prepared: August 15, 2022
### II. DEVICE
| Name of Device: | OR3O Dual Mobility System |
|----------------------|------------------------------------------------------------------------------------------------------------|
| Common Name: | Hip Prosthesis |
| Regulatory Class: | Class II |
| Product Code: | LPH |
| Classification Name: | Hip joint metal/polymer/metal semi-constrained porous-coated<br>uncemented prosthesis,<br>(21 CFR888.3358) |
| PREDICATE DEVICE | Primary Predicate: OR3O Dual Mobility System - |
|------------------|------------------------------------------------|
| | K191002 (S.E. 10/31/2019) |
| | Predicate 2: POLARCUP Dual Mobility System - |
| | K110135 (S.E. 10/14/2011) |
| | Predicate 3: REFLECTION Hip System - |
| | K071160 (S.E. 10/05/2007) |
| | Predicate 4: REFLECTION Hip System - |
| | K932755 (S.E. 05/06/1994) |
| | Predicate 5: BH Dual Mobility System - |
| | K171934 (S.E. 11/30/2017) |
The predicate devices have not been subject to a design related recall.
{4}------------------------------------------------
#### III. DEVICE DESCRIPTION
The OR30 Hip System is a modular dual mobility implant system consists of diffusion hardened oxidized zirconium (OXINIUM DH) alloy liners with a highly polished zirconium oxide inner surface and a machined locking taper and backside of Zr2.5Nb alloy (F2384-10R16). The locking taper and outside profile is designed to mate with a dedicated Ti6Al4V R3 acetabular shell with OD sizes 44mm to 74mm or REDAPT Modular Press-fit acetabular shell with OD sizes 48mm to 74mm. For each assembled OR30 liner and R3 shell size from 44mm to 74mm and REDAPT Modular shell size from 48mm to 74mm, a dedicated insert is available. Inserts are made of highly crosslinked ultra-high-molecularweight polyethylene (ISO 5834-2/ASTM F648) that is irradiated with a 10 Mrad dose and remelted . These can be combined with oxidized zirconium or CoCr alloy femoral heads of sizes 22 mm (for Size 44mm-52mm) and 28 mm (for size 50mm-74mm). The final OR30 Dual Mobility construct will include an acetabular shell (with optional screws and optional screw hole covers), an OXINIUM DH Liner, a cross linked polyethylene (XLPE) Insert and a femoral head.
The purpose of this Traditional 510(k) submission is to add additional components, namely additional sizes of the OR3O Dual Mobility OXINIUM DH Liners and OR3O Dual Mobility XLPE Inserts to the existing OR30 Dual Mobility System.
This line extension will consist of two smaller size OXINIUM DH Liners and four additional XLPE inserts to allow for a wider range of implant options for physicians to choose from. The materials and manufacturing processes will be the same, with the exception of accounting for different sized implants.
{5}------------------------------------------------
#### IV. INDICATIONS FOR USE
## Intended Use
The OR3O Dual Mobility System is intended for use in primary and revision total hip arthroplasty in skeletally mature patients.
## Indications
- Advanced degeneration of the hip joint as a result of degenerative, post-traumatic, or . rheumatoid arthritis.
- Fracture or avascular necrosis of the femoral head. .
- Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, . hemiarthroplasty, surface replacement arthroplasty, or total hip replacement.
- All forms of osteoarthritis. .
- Patients with hips at risk of dislocation. .
- Femoral neck fracture or proximal hip joint fracture. •
The OR3O Dual Mobility System is intended for single use only. The modular OR3O Liners and Inserts are to be implanted without bone cement.
Mating components may be indicated for use without bone cement.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE V. PREDICATE DEVICE
The subject OR3O Dual Mobility System implants have the same or similar intended use, fundamental scientific technology, materials, and indications for use as the following FDA cleared predicates: OR3O Dual Mobility System - K191002 (S.E 10/31/2019), POLARCUP Dual Mobility System - K110135 (S.E. 10/14/2011), REFLECTION Hip System - K071160 (S.E. 10/05/2007) - REFLECTION Hip System (K932755 S.E. 05/06/1994), BH Dual Mobility System - K171934 (S.E. 11/30/2017)
The subject device includes two implant components to form a dual mobility concept where there are two articulating surfaces in the same joint space:
{6}------------------------------------------------
## LINER
- The subject OR30 Dual Mobility System Liner uses an OXIDIZED DH Zirconium . (OXINIUM DH) material with a highly polished inner surface identical to the predicate OR3O Dual Mobility System - K191002 (S.E 10/31/2019). The OR3O Dual Mobility Liners have a locking taper and are designed to mate with existing Smith & Nephew R3 (K092386 – S.E. 11/03/2009) or REDAPT Modular Shell - K182109 (S.E. 11/16/2018).
## INSERT
- The subject OR30 Dual Mobility XLPE material is identical to the predicate OR3O ● Dual Mobility XLPE material - OR3O Dual Mobility K191002 (S.E 10/31/2019), with only minor design iterations made to allow for smaller size implants. Both the subject and predicate OR30 Dual Mobility XLPE Inserts are intended to be used as an articulating component of the OR30 Dual Mobility System.
The OR30 Dual Mobility Liners and OR30 Dual Mobility Inserts are intended to be used with existing Smith & Nephew Acetabular Shells and Femoral Heads:
- Existing Acetabular Shells The OR30 Dual Mobility System uses existing Smith & . Nephew R3 Shells - K092386 (S.E. 11/03/2009) or Smith & Nephew REDAPT Modular Press-fit acetabular shells - K182109 (S.E. 11/16/2018).
- Existing Femoral Heads The OR3O Dual Mobility System uses existing Smith & . Nephew OXINIUM Femoral Heads K110101 (S.E. 04/11/2011) or Cobalt Chrome Femoral Heads – K963509 (S.E. 01/27/1997) and K963486 (S.E. 11/27/1996).
#### PERFORMANCE DATA VI.
The following performance data are provided in support of the substantial equivalence determination.
## Biocomnatibility
The biocompatibility evaluation for the OR3O Dual Mobility Implants was conducted in accordance with FDA's Guidance for Industry and FDA Staff "Use of International Standard
{7}------------------------------------------------
ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing Within a Risk Management Process".
The subject OR30 Dual Mobility Implants are permanent implants and will be classified as permanent, >30 day body contact according to ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process".
The subject OR30 Dual Mobility Liners are manufactured from Zr-2.5Nb alloy materials in accordance with the following ASTM standard: F2384- 10R16 Standard Specification for Wrought Zirconium-2.5Niobium Alloy for Surgical Implant Applications (UNS R60901).
The subject OR3O Dual Mobility Inserts are manufactured from identical polyethylene XLPE materials as the predicate device, in accordance with the following standards: ASTM F648-14 (Standard Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants) and ISO 5834-2 (Implants for surgery. Ultra-highmolecular-weight polyethylene. Moulded forms).
A biocompatibility evaluation has been completed and a Declaration of Conformity's has been provided for the OR3O Dual Mobility System line extension.
## Mechanical testing
Smith & Nephew has evaluated the subject OR30 Dual Mobility System to demonstrate substantial equivalence to the predicate devices: OR3O Dual Mobility System - K191002 (S.E 10/31/2019), POLARCUP Dual Mobility System - K110135 (S.E. 10/14/2011), REFLECTION Hip System – K071160 (S.E. 10/05/2007), REFLECTION Hip System (K932755 S.E. 05/06/1994) and BH Dual Mobility System - K171934 (S.E. 11/30/2017), and determined that the subject devices do not represent a new worst case.
{8}------------------------------------------------
| Standard | Testing |
|----------------|----------------------------------------------------------------|
| ASTM F1820 | Push Out Testing Torque to Failure Testing Lever-Out Testing |
| ASTM F1875 | Environmental Fatigue and Corrosion Assessment |
| ISO 7206-12 | Deformation Testing |
| ASTM F2582 | Neck Impingement Testing |
| ISO 21535-2009 | Range of Motion |
| ISO 14242 | Wear Testing |
| N/A | Jump Distance Insert Pull-Out Testing Insert Lever-out Testing |
Biomechanical testing and/ or rationales were completed on the subject devices as follows:
{9}------------------------------------------------
The subject OR3O Dual Mobility devices met the pre-determined acceptance criteria for each intended output. Therefore, design verification testing determined that the subject OR3O Dual Mobility System is substantially equivalent to the identified predicate devices.
## Non-Pvrogenicity Endotoxin Testing
Bacterial endotoxin testing was completed and met the acceptable endotoxin limit as stated in the FDA Guidance, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72 "Bacterial endotoxins – Test methods, routine monitoring and alternatives to batch testing".
### CONCLUSIONS VII.
Based on the verification evidence activities provided in this pre-market notification application, the subject OR3O Dual Mobility System is substantially equivalent to the legally marketed primary predicate device - OR3O Dual Mobility System - K191002 (S.E 10/31/2019), and the other predicates: POLARCUP Dual Mobility System - K110135 (S.E. 10/14/2011), REFLECTION Hip System K071160 (S.E. 10/05/2007), REFLECTION Hip System -K932755 (S.E. 05/06/1994) and BH Dual Mobility System - K171934 (S.E. 11/30/2017).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.