REDAPT Modular Acetabular Shell

K182109 · Smith & Nephew, Inc. · LPH · Nov 16, 2018 · Orthopedic

Device Facts

Record IDK182109
Device NameREDAPT Modular Acetabular Shell
ApplicantSmith & Nephew, Inc.
Product CodeLPH · Orthopedic
Decision DateNov 16, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3358
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity. The REDAPT Modular Acetabular Shell is for single use only and is intended for cementless use.

Device Story

REDAPT Modular Acetabular Shell is a cementless, single-use acetabular component for total hip arthroplasty; line extension of REDAPT Porous Acetabular Shell. Manufactured from Ti-6Al-4V powder via additive manufacturing; features porous structure identical to predicate. Inner surface machined to match R3 Multi-hole shell geometry and locking mechanism, enabling use of R3 poly liners. Used by orthopedic surgeons in clinical settings for primary or revision hip surgery. Device provides stable acetabular fixation; allows modularity with R3 liners to address patient-specific anatomical needs; benefits patients by restoring hip function and mobility.

Clinical Evidence

Bench testing only. Mechanical testing included acetabular shell and poly liner dissociation (push-in, push-out, lever-out, torque to failure), finite element analysis, fatigue testing, corrosion testing, and biocompatibility. Bacterial endotoxin testing met ANSI/AAMI ST72 standards. Porous structure characterization was previously cleared under K150790.

Technological Characteristics

Material: Ti-6Al-4V titanium alloy. Manufacturing: Additive manufacturing (porous structure). Form factor: Acetabular shell (48-80 mm). Fixation: Cementless. Connectivity: None. Sterilization: Not specified. Software: None.

Indications for Use

Indicated for individuals undergoing primary and revision hip surgery due to trauma, noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, pelvic fracture, diastrophic variant), inflammatory degenerative joint disease (rheumatoid arthritis, congenital dysplasia), or proximal femur fractures/nonunions unmanageable by other techniques.

Regulatory Classification

Identification

A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | Submitted by: | Smith & Nephew, Inc.<br>Advanced Surgical Devices Division<br>7135 Goodlett Farms Parkway<br>Cordova, Tennessee 38016 | |----------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Date of Submission: | August 2, 2018 | | Contact Person: | Meenakshi Gupta, Sr. Regulatory Affairs Specialist<br>T (901) 399-6139<br>F (901) 721-2748 | | Name of Device: | REDAPT Modular Acetabular Shell | | Common Name: | Total Hip Joint, Acetabular Component | | Device Classification Name<br>and Reference: | 21 CFR 888.3358 – Prosthesis, Hip, Semi-<br>Constrained, Metal/Polymer, Porous Uncemented | | Device Class: | Class II | | Panel Code: | Orthopaedics/87 | | Product Code: | LPH, MBL | | Predicate Device: | REDAPT Porous Acetabular Shell - K150790<br>R3 Shell Multi-Hole -K092386<br>REFLECTION Acetabular System-K920430<br>The predicate devices have not been subject to a<br>design related recall. | {1}------------------------------------------------ ### Device Description: The subject of this Traditional 510(k) is the REDAPT Modular Acetabular Shells. The REDAPT Modular Acetabular Shells are a line extension to the REDAPT Porous Acetabular Shells to allow the un-cemented use of R3 Poly liners. The REDAPT Porous Acetabular Shells were previously cleared for market via premarket notification K150790. The REDAPT Modular Acetabular Shells are available in same size range as REDAPT Porous Acetabular Shells from 48 mm to 80 mm. The REDAPT Modular Acetabular Shells are made from titanium alloy (Ti-6Al-4V) powder through an additive manufacturing process and has identical porous structure as the predicate REDAPT Porous Acetabular Shell. The inner surface of the subject REDAPT Modular Acetabular shell is modified by a machining process to have an identical inner geometry and locking mechanism as the R3 Multi-hole shells to allow the uncemented use of R3 poly liners that are cleared to use with R3 Multi-hole Shell. The R3 Multi-hole Shells are most recently cleared under premarket notification K092386.The R3 XLPE liners were cleared under 510(k) K070756 and K113848 and the R3 Constrained Liners were cleared under K111635, K122139 and K162641. ### Intended Use The REDAPT Modular Acetabular Shells are indicated for: Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture and trochanteric {2}------------------------------------------------ fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracturedislocation of the hip; and correction of deformity. The REDAPT Modular Acetabular Shell is for single use only and is intended for cementless use. The above indications are substantially equivalent to the indications cleared for the RADAPT Porous acetabular shell cleared under K150790 and R3 Multi-hole Shell Cleared under K092386. # Technological Characteristics Device comparisons described in this premarket notification demonstrate that the subject device REDAPT Modular Acetabular Shell are substantially equivalent to the below listed legally marketed predicate devices with regard to intended use, indications for use, design, material and performance characteristics. # Substantial Equivalence Information The overall design, materials, and indications for use for the REDAPT Modular Acetabular Shell are substantially equivalent to the following commercially available predicate devices. | Manufacturer | Description | Submission Number | Clearance Date | |----------------------|--------------------------------|-------------------|----------------| | Smith & Nephew, Inc. | REDAPT Porous Acetabular Shell | K150790 (Primary) | 11/16/2015 | | Smith & Nephew, Inc. | R3 Shell Multi-Hole | K092386 | 11/03/2009 | | Smith & Nephew, Inc. | REFLECTION Acetabular System | K920430 | 07/21/1992 | ## Table: Predicate Devices {3}------------------------------------------------ # Performance Testing To further support a determination of substantial equivalence, non-clinical bench (mechanical) testing were conducted on the REDAPT Modular Acetabular Shell. A review of the mechanical data and the technical memo, indicates that the REDAPT Modular Acetabular Shells are substantially equivalent to predicate devices listed in theTable above. The following tests were used as a basis for the determination of substantial equivalence. - . Acetabular Shell and Poly Liner Dissociation Testing-Push-in, Push-put, Lever-out and Torque to Failure testing - o Finite Element Analysis - Fatigue test - . Corrosion - . Biocompatibility Bacterial endotoxin testing was completed and met the acceptable endotoxin limits as stated in the FDA Guidance, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72. All tests which are in relation to the porous structure characterization (physical, chemical or mechanical) have been cleared in the REDAPT Porous Acetabular shell 510(k) K150790. ## Conclusion As previously noted, this 510(k) Premarket Notification is being submitted to request clearance for the REDAPT Modular Shell. Based on the similarities to the predicate devices and a review of the mechanical testing performed, the subject device is substantially equivalent to the commercially available predicate devices listed above. {4}------------------------------------------------ # Indications for Use 510(k) Number (if known) K182109 Device Name REDAPT Modular Acetabular Shells Indications for Use (Describe) The REDAPT Modular Acetabular Shells are indicated for: Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (ND)D) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, atthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracture-dislocation of the hip; and correction of deformity. The REDAPT Modular Acetabular Shell is for single use only and is intended for cementless use. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ | Submitted by: | Smith & Nephew, Inc.<br>Advanced Surgical Devices Division<br>7135 Goodlett Farms Parkway<br>Cordova, Tennessee 38016 | |----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------| | Date of Submission: | August 2, 2018 | | Contact Person: | Meenakshi Gupta, Sr. Regulatory Affairs Specialist<br>T (901) 399-6139<br>F (901) 721-2748 | | Name of Device: | REDAPT Modular Acetabular Shell | | Common Name: | Total Hip Joint, Acetabular Component | | Device Classification Name<br>and Reference: | 21 CFR 888.3358 – Prosthesis, Hip, Semi-<br>Constrained, Metal/Polymer, Porous Uncemented | | Device Class: | Class II | | Panel Code: | Orthopaedics/87 | | Product Code: | LPH, MBL | | Predicate Device: | REDAPT Porous Acetabular Shell - K150790<br>R3 Shell Multi-Hole -K092386<br>REFLECTION Acetabular System-K920430 | | | The predicate devices have not been subject to a<br>design related recall | {6}------------------------------------------------ ## Device Description: The subject of this Traditional 510(k) is the REDAPT Modular Acetabular Shells. The REDAPT Modular Acetabular Shells are a line extension to the REDAPT Porous Acetabular Shells to allow the un-cemented use of R3 Poly liners. The REDAPT Porous Acetabular Shells were previously cleared for market via premarket notification K150790. The REDAPT Modular Acetabular Shells are available in same size range as REDAPT Porous Acetabular Shells from 48 mm to 80 mm. The REDAPT Modular Acetabular Shells are made from titanium alloy (Ti-6Al-4V) powder through an additive manufacturing process and has identical porous structure as the predicate REDAPT Porous Acetabular Shell. The inner surface of the subject REDAPT Modular Acetabular shell is modified by a machining process to have an identical inner geometry and locking mechanism as the R3 Multi-hole shells to allow the uncemented use of R3 poly liners that are cleared to use with R3 Multi-hole Shell. The R3 Multi-hole Shells are most recently cleared under premarket notification K092386.The R3 XLPE liners were cleared under 510(k) K070756 and K113848 and the R3 Constrained Liners were cleared under K111635, K122139 and K162641. ### Intended Use The REDAPT Modular Acetabular Shells are indicated for: Hip components are indicated for individuals undergoing primary and revision surgery where other treatments or devices have failed in rehabilitating hips damaged as a result of trauma or noninflammatory degenerative joint disease (NIDJD) or any of its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant. Hip components are also indicated for inflammatory degenerative joint disease including rheumatoid arthritis, arthritis secondary to a variety of diseases and anomalies, and congenital dysplasia; treatments of nonunion, femoral neck fracture and trochanteric {7}------------------------------------------------ fractures of the proximal femur with head involvement that are unmanageable using other techniques; endoprosthesis, femoral osteotomy, or Girdlestone resection; fracturedislocation of the hip; and correction of deformity. The REDAPT Modular Acetabular Shell is for single use only and is intended for cementless use. The above indications are substantially equivalent to the indications cleared for the RADAPT Porous acetabular shell cleared under K150790 and R3 Multi-hole Shell Cleared under K092386. # Technological Characteristics Device comparisons described in this premarket notification demonstrate that the subject device REDAPT Modular Acetabular Shell are substantially equivalent to the below listed legally marketed predicate devices with regard to intended use, indications for use, design, material and performance characteristics. # Substantial Equivalence Information The overall design, materials, and indications for use for the REDAPT Modular Acetabular Shell are substantially equivalent to the following commercially available predicate devices. | Manufacturer | Description | Submission Number | Clearance Date | |----------------------|--------------------------------|-------------------|----------------| | Smith & Nephew, Inc. | REDAPT Porous Acetabular Shell | K150790 (Primary) | 11/16/2015 | | Smith & Nephew, Inc. | R3 Shell Multi-Hole | K092386 | 11/03/2009 | | Smith & Nephew, Inc. | REFLECTION Acetabular System | K920430 | 07/21/1992 | ## Table: Predicate Devices {8}------------------------------------------------ # Performance Testing To further support a determination of substantial equivalence, non-clinical bench (mechanical) testing were conducted on the REDAPT Modular Acetabular Shell. A review of the mechanical data and the technical memo, indicates that the REDAPT Modular Acetabular Shells are substantially equivalent to predicate devices listed in theTable above. The following tests were used as a basis for the determination of substantial equivalence. - Acetabular Shell and Poly Liner Dissociation Testing-Push-in, Push-put, Lever-out and Torque to Failure testing - o Finite Element Analysis - Fatigue test - Corrosion - . Biocompatibility Bacterial endotoxin testing was completed and met the acceptable endotoxin limits as stated in the FDA Guidance, "Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile," "Pyrogen and Endotoxins Testing: Questions and Answers," and ANSI/AAMI ST72. All tests which are in relation to the porous structure characterization (physical, chemical or mechanical) have been cleared in the REDAPT Porous Acetabular shell 510(k) K150790. ## Conclusion As previously noted, this 510(k) Premarket Notification is being submitted to request clearance for the REDAPT Modular Acetabular Shell. Based on the similarities to the predicate devices and a review of the mechanical testing performed, the subject device is substantially equivalent to the commercially available predicate devices listed above.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...