3M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)
K220942 · 3M Company · JOJ · Jun 24, 2022 · General Hospital
Device Facts
Record ID
K220942
Device Name
3M Attest Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)
Applicant
3M Company
Product Code
JOJ · General Hospital
Decision Date
Jun 24, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: Gravity 250°F/121°C 30 minutes; Gravity 270°F/132°C 3 minutes; Gravity 270°F/132°C 4 minutes; Gravity 270°F/132°C 10 minutes; Gravity 270°F/132°C 15 minutes; Gravity 270°F/132°C 25 minutes; Gravity 275°F/135°C 3 minutes; Gravity 275°F/135°C 10 minutes; Dynamic Air Removal 250°F/121°C 30 minutes; Dynamic Air Removal 270°F/132°C 4 minutes; Dynamic Air Removal 270°F/132°C 5 minutes; Dynamic Air Removal 270°F/132°C 6 minutes; Dynamic Air Removal 270°F/132°C 7 minutes; Dynamic Air Removal 270°F/132°C 8 minutes; Dynamic Air Removal 270°F/132°C 9 minutes; Dynamic Air Removal 270°F/132°C 10 minutes; Dynamic Air Removal 273°F/134°C 3 minutes; Dynamic Air Removal 273°F/134°C 4 minutes; Dynamic Air Removal 275°F/135°C 3 minutes.
Device Story
Chemical indicator for steam sterilization monitoring; consists of paper wick and steam/temperature-sensitive chemical pellet in paper/film/foil laminate. Placed inside sterilization containment devices (packs, trays, etc.) by healthcare personnel. During steam sterilization, chemical pellet melts and migrates as dark color along wick. Migration extent depends on steam, time, and temperature. Visual output window indicates 'ACCEPT' or 'REJECT' based on whether critical sterilization parameters were met. Provides immediate visual confirmation of sterilization cycle performance to healthcare providers; ensures sterility assurance of processed items; benefits patient by reducing risk of using non-sterile instruments.
Clinical Evidence
Bench testing only. Performance verified via resistometer testing for stated values at various temperatures (250°F-275°F), dry heat testing, side-by-side testing with biological indicators, and endpoint color stability testing. All tests passed.
Technological Characteristics
Chemical indicator; paper wick and steam/temperature-sensitive chemical pellet; paper/film/foil laminate packaging. Operates via chemical pellet melting and migration in response to saturated steam. Shelf life: 3 years.
Indications for Use
Indicated for use as an independent monitor of critical parameters in steam sterilization cycles (Gravity and Dynamic Air Removal) for packs, pouches, containers, or trays. No specific patient population or contraindications defined.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
June 24, 2022
3M Company Yumi Wackerfuss Senior Regulatory Affairs Associate 2510 Conway Avenue, Bldg. 275-5W-06 Saint Paul, Minnesota 55144-1000
Re: K220942
Trade/Device Name: 3M Attest Steam Chemical Integrators (1243A, 1243RE, 1243RES) Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: May 25, 2022 Received: May 27, 2022
### Dear Yumi Wackerfuss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220942
Device Name
3MTM Attest™ Steam Chemical Integrators (1243A, 1243B, 1243RE, 1243RES)
Indications for Use (Describe)
The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles.
The integrating indicator is intended to be placed in each pack, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
| Cycle Type | Temperature | Exposure Time |
|---------------------------------------------------------------------------------------------------|----------------------------|----------------------------|
| Gravity | 250°F/121°C | 30 minutes |
| Gravity | 270°F/132°C | 3 minutes |
| Gravity | 270°F/132°C | 4 minutes |
| Gravity | 270°F/132°C | 10 minutes |
| Gravity | 270°F/132°C | 15 minutes |
| Gravity | 270°F/132°C | 25 minutes |
| Gravity | 275°F/135°C | 3 minutes |
| Gravity | 275°F/135°C | 10 minutes |
| Dynamic Air Removal | 250°F/121°C | 30 minutes |
| Dynamic Air Removal | 270°F/132°C | 4 minutes |
| Dynamic Air Removal | 270°F/132°C | 5 minutes |
| Dynamic Air Removal | 270°F/132°C | 6 minutes |
| Dynamic Air Removal | 270°F/132°C | 7 minutes |
| Dynamic Air Removal | 270°F/132°C | 8 minutes |
| Dynamic Air Removal | 270°F/132°C | 9 minutes |
| Dynamic Air Removal | 270°F/132°C | 10 minutes |
| Dynamic Air Removal | 273°F/134°C | 3 minutes |
| Dynamic Air Removal | 273°F/134°C | 4 minutes |
| Dynamic Air Removal | 275°F/135°C | 3 minutes |
| Minimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer: | | |
| 250°F/121°C<br>16.5 Minutes | 270°F/132°C<br>2.0 Minutes | 273°F/134°C<br>1.4 Minutes |
| | | 275°F/135°C<br>1.2 Minutes |
| Type of Use (Select one or both, as applicable) | | |
| ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the 3M logo. The logo consists of the number "3" and the letter "M" in a bold, sans-serif font. The color of the text is red, and the background is white.
# 510(k) Summary for 3M™ Attest™ Steam Chemical Integrators 1243A, 1243B, 1243RE and 1243RES
#### Sponsor Information:
3M Company 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
Contact: Yumi Wackerfuss Senior Regulatory Affairs Associate Phone Number: (651) 737-3556 Fax Number: (651) 737-5320
Date of Summary: March 29th, 2022
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the 3M logo. The logo is red and consists of the number "3" followed by the letter "M". The logo is simple and recognizable.
#### 1. Device Name and Classification:
| Common or Usual Name | Chemical Indicators |
|------------------------|---------------------------------------------------|
| Proprietary Name: | 3M™ Attest™ Steam Chemical Integrators |
| Classification Name: | Physical/chemical sterilization process indicator |
| Device Classification: | Class II, 21 CFR § 880.2800(b) |
| Product Code: | JOJ |
#### 2. Predicate Device:
K193254, 3MTM Attest™ Steam Chemical Integrators
#### 3. Description of Device:
3M™ Attest™ Steam Chemical Integrators are chemical indicators consisting of a paper wick and a steam and temperature sensitive chemical pellet contained in a paper/film/foil laminate. The chemical pellet melts and migrates as a dark color along the paper wick. The migration is visible through a window marked ACCEPT or window marked REJECT; the extent of migration depends on steam, time, and temperature.
#### 4. Indications for Use
The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
| Cycle Type | Temperature | Exposure Time |
|---------------------|-------------|---------------|
| Gravity | 250°F/121°C | 30 minutes |
| Gravity | 270°F/132°C | 3 minutes |
| Gravity | 270°F/132°C | 4 minutes |
| Gravity | 270°F/132°C | 10 minutes |
| Gravity | 270°F/132°C | 15 minutes |
| Gravity | 270°F/132°C | 25 minutes |
| Gravity | 275°F/135°C | 3 minutes |
| Gravity | 275°F/135°C | 10 minutes |
| Dynamic Air Removal | 250°F/121°C | 30 minutes |
| Dynamic Air Removal | 270°F/132°C | 4 minutes |
| Dynamic Air Removal | 270°F/132°C | 5 minutes |
| Dynamic Air Removal | 270°F/132°C | 6 minutes |
| Dynamic Air Removal | 270°F/132°C | 7 minutes |
| Dynamic Air Removal | 270°F/132°C | 8 minutes |
| Dynamic Air Removal | 270°F/132°C | 9 minutes |
| Dynamic Air Removal | 270°F/132°C | 10 minutes |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The number and letter are in a bold, sans-serif font. The logo is simple and recognizable.
### 3M™ Attest™ Steam Chemical Integrators 1243A, 1243B, 1243RE and 1243RES
| Cycle Type | Temperature | Exposure Time |
|---------------------|-------------|---------------|
| Dynamic Air Removal | 273°F/134°C | 3 minutes |
| Dynamic Air Removal | 273°F/134°C | 4 minutes |
| Dynamic Air Removal | 275°F/135°C | 3 minutes |
Minimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer:
| 250°F/121°C | 270°F/132°C | 273°F/134°C | 275°F/135°C |
|--------------|-------------|-------------|-------------|
| 16.5 Minutes | 2.0 Minutes | 1.4 Minutes | 1.2 Minutes |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the 3M logo. The logo is in red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable.
#### 510(k) Summary 3M™ Attest™ Steam Chemical Integrators 1243A, 1243B, 1243RE and 1243RES
#### 5. Summary of Technological Characteristics compared to Predicate Device
| Element | Subject Devices: (this submission)<br>3M™ Attest™ Steam Chemical Integrators | Predicate Device (K193254: Cleared on Dec 23rd, 2019)<br>3M™ Attest™ Steam Chemical Integrators | Comparison | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Device Models | 1243A, 1243B, 1243RE and 1243RES | 1243A and 1243B | 1243A and 1243B are substantially equivalent as 1243RE and 1243RES (K191236). | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| Indications for use | The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: | The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: | Similar. This submission proposes the additional exposure times for 5, 6, 7, 8 and 9 mintues for Dynamic Air Removal, 270°F/132° C. | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
| | Cycle Type Temperature Exposure Time Gravity 250°F/121°C 30 minutes Gravity 270°F/132°C 3 minutes Gravity 270°F/132°C 4 minutes Gravity 270°F/132°C 10 minutes Gravity 270°F/132°C 15 minutes Gravity 270°F/132°C 25 minutes Gravity 275°F/135°C 3 minutes Gravity 275°F/135°C 10 minutes Dynamic Air Removal 250°F/121°C 30 minutes Dynamic Air Removal 270°F/132°C 4 minutes Dynamic Air Removal 270°F/132°C 5 minutes Dynamic Air Removal 270°F/132°C 6 minutes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Cycle Type Temperature Exposure Time Gravity 250°F/121°C 30 minutes Gravity 270°F/132°C 3 minutes Gravity 270°F/132°C 4 minutes Gravity 270°F/132°C 10 minutes Gravity 270°F/132°C 15 minutes Gravity 270°F/132°C 25 minutes Gravity 275°F/135°C 3 minutes Gravity 275°F/135°C 10 minutes Dynamic Air Removal 250°F/121°C 30 minutes Dynamic Air Removal 270°F/132°C 4 minutes Dynamic Air Removal 270°F/132°C 10 minutes | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the 3M logo. The logo consists of the number "3" and the letter "M" in a bold, sans-serif font. The color of the text is red. The background is white.
#### 510(k) Summary 3M™ Attest™ Steam Chemical Integrators 1243A, 1243B, 1243RE and 1243RES
| Element | Subject Devices: (this submission)<br>3MTM Attest™ Steam Chemical Integrators | | | Predicate Device (K193254: Cleared on Dec 23rd, 2019)<br>3MTM Attest™ Steam Chemical Integrators | | | | Comparison | |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-------------|-------------|--------------------------------------------------|--|
| | Dynamic Air<br>Removal | 270°F/132°C | 7 minutes | Dynamic Air<br>Removal | 273°F/134°C | 3 minutes | | | |
| | Dynamic Air<br>Removal | 270°F/132°C | 8 minutes | Dynamic Air<br>Removal | 273°F/134°C | 4 minutes | | | |
| | Dynamic Air<br>Removal | 270°F/132°C | 9 minutes | Dynamic Air<br>Removal | 275°F/135°C | 3 minutes | | | |
| | Dynamic Air<br>Removal | 270°F/132°C | 10 minutes | | | | | | |
| | Dynamic Air<br>Removal | 273°F/134°C | 3 minutes | | | | | | |
| | Dynamic Air<br>Removal | 273°F/134°C | 4 minutes | | | | | | |
| | Dynamic Air<br>Removal | 275°F/135°C | 3 minutes | | | | | | |
| Indicator<br>Agent | Proprietary formulation. | | | Proprietary formulation. | | | | Identical | |
| Sterilization<br>method and<br>cycles | Steam sterilization processes 250°F to 275°F (121°C to 135°C) | | | Steam sterilization processes 250°F to 275°F (121°C to 135°C) | | | | Identical | |
| Endpoint<br>Specificatio<br>ns<br>(Minimum<br>Stated<br>Values) | The minimum stated values for the 3M™ Attest™ Steam<br>Chemical Integrators as determined using a resistometer are<br>provided in the table below. | | | The minimum stated values for the 3MTM Attest™ Steam<br>Chemical Integrators as determined using a resistometer<br>are provided in the table below. | | | | Identical | |
| | 250°F/121°C | 270°F/132°C | 273°F/134°C | 275°F/135°C | 250°F/121°C | 270°F/132°C | 273°F/134°C | 275°F/135°C | |
| | 16.5<br>Minutes | 2.0 Minutes | 1.4 Minutes | 1.2 Minutes | 16.5<br>Minutes | 2.0 Minutes | 1.4 Minutes | 1.2 Minutes | |
| Shelf life | Three (3) years | | | One (1) year for 1243A and 1243B | | | | Extended<br>via<br>identical<br>protocol.<br>The | |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the 3M logo. The logo is red and consists of the number "3" and the letter "M" in a bold, sans-serif font. The logo is simple and recognizable.
#### 510(k) Summary 3M™ Attest™ Steam Chemical Integrators 1243A, 1243B, 1243RE and 1243RES
| Element | Subject Devices: (this submission) | Predicate Device (K193254: Cleared on Dec 23rd, 2019) | Comparison |
|---------|----------------------------------------|-------------------------------------------------------|----------------------------------------------------------------------------------|
| | 3M™ Attest™ Steam Chemical Integrators | 3M™ Attest™ Steam Chemical Integrators | 1243RE<br>and RES<br>was cleared<br>for 6<br>months<br>shelf life<br>originally. |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable. The red color is bold and eye-catching.
### 510(k) Summary 3MTM Attest™ Steam Chemical Integrators 1243A, 1243B, 1243RE and 1243RES
### 6. Nonclinical Comparison to the Predicate Device
3MTM Attest™ Steam Chemical Integrators are identical to the previously cleared devices which are sold under the same tradename 3M™ Attest™ Steam Chemical Integrators (K193254) and was tested using the identified test methodology shown below.
The results of performance testing on 3MTM Attest™ Steam Chemical Integrators demonstrate the device performs and meet the acceptance criteria shown below.
#### Results Test Name Acceptance Criteria Purpose SV at 250°F/121°C Stated Value To identify the critical > 16.5 minutes Pass (SV) Testing parameters required to achieve a stated inactivation, by SV at 270°F/132°C > 2.0 minutes referring to a stated test organism with stated D and z SV at 273°F/134°C ≥ 1.4 minutes values. The integrator must turn to "ACCEPT" end point SV at 275°F/135°C > 1.2 minutes at the stated value time and must also remain "REJECT" when exposed to conditions of Integrator 10 - 27°C -1°C/-15% set point of the SV temperature time. All testing is completed coefficient in a saturated steam Correlation > 0.9 resistometer. Integrator coefficient temperature coefficient and correlation coefficient are calculated to confirm alignment to biological indicator performance. Health Care Confirm integrators provide Device reaches "ACCEPT" endpoint Pass Facility reaction when exposed to customer acceptable performance in Simulated Use cleared customer use use cvcles. Testing sterilization cycles. Device does not reach endpoint ("REJECT") when exposed to failing conditions in customer use cycles. Dry Heat Verify device requires the Endpoint must not be met following Pass Testing presence of saturated steam to dry heat exposure at 140°C for 30 turn to reach endpoint. min. Side-by-Side Confirm integrators are Chemical integrator parallels Pass Testing with parallel in performance to performance of BI and does not reach Biological endpoint before BI is inactivated. biological indicators (BI). Indicator Endpoint Confirm endpoint color Endpoint decision must remain Pass unchanged after 6 months. Color Stability stability for samples exposed to passing and failing
#### Table 6.1: Summary of Nonclinical Testing
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the 3M logo. The logo consists of the number "3" and the letter "M" in a bold, red font. The two characters are connected to each other. The logo is simple and recognizable.
### 510(k) Summary 3MTM Attest™ Steam Chemical Integrators 1243A, 1243B, 1243RE and 1243RES
| Test Name | Purpose | Acceptance Criteria | Results |
|-----------|----------------------------------------|---------------------|---------|
| | conditions in a steam<br>resistometer. | | |
### 7. Conclusion
Based on the intended uses, technological characteristics and non-clinical performance data, 3M™ Attest™ Steam Chemical Integrators are as safe, as effective, and performs as well as or better than the legally marketed predicate device, Attest™ Steam Chemical Integrators cleared under K193254, Class II (21 CFR 880.2800(b)), product code JOJ.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.