K191236 · 3M Company · JOJ · Nov 22, 2019 · General Hospital
Device Facts
Record ID
K191236
Device Name
3M™ Attest™ Steam Chemical Integrators
Applicant
3M Company
Product Code
JOJ · General Hospital
Decision Date
Nov 22, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The 3M™ Attest™ Steam Chemical Integrators are designed to all critical parameters over a specified range of steam sterilization cycles. The integrating intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: Gravity 250°F/121°C 30 minutes; Gravity 270°F/132°C 3, 4, 10, 15, 25 minutes; Gravity 275°F/135°C 3, 10 minutes; Dynamic Air Removal 250°F/121°C 30 minutes; Dynamic Air Removal 270°F/132°C 4, 10 minutes; Dynamic Air Removal 273°F/134°C 3, 4 minutes; Dynamic Air Removal 275°F/135°C 3 minutes.
Device Story
Chemical indicator for steam sterilization monitoring; consists of paper wick and steam/temperature-sensitive chemical pellet inside paper/film/foil laminate. Pellet melts and migrates as dark color along wick upon exposure to steam, time, and temperature. User observes migration through window marked ACCEPT or REJECT. Placed in sterilization packs/trays by healthcare personnel. Provides visual confirmation of sterilization cycle parameters; assists in clinical decision-making regarding sterility of processed items. Benefits patient by ensuring medical devices used in procedures have undergone validated sterilization conditions.
Clinical Evidence
Bench testing only. Performance validated per ANSI/AAMI/ISO 11140-1:2014. Testing included Stated Value (SV) verification in saturated steam resistometers, simulated healthcare facility use testing, dry heat resistance, side-by-side comparison with biological indicators, and endpoint color stability. All tests passed.
Technological Characteristics
Chemical indicator; paper wick and proprietary chemical pellet in paper/film/foil laminate. Sensing principle: chemical pellet melting/migration dependent on steam, time, and temperature. Dimensions/form factor: strip. Connectivity: none. Sterilization: N/A (used to monitor sterilization).
Indications for Use
Indicated for use as an independent monitor of critical parameters in steam sterilization cycles (Gravity and Dynamic Air Removal) across specified temperatures (121°C–135°C) and exposure times. Intended for placement inside packs, pouches, containers, or trays.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M™ Comply™ SteriGage™ Chemical Integrator for Steam (K101249)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 22, 2019
3M Company Mary Fretland Senior Regulatory Affairs Associate 3M Center, Building 275-5W-06 St. Paul, Minnesota 55144
Re: K191236
Trade/Device Name: 3MTM Attest™ Steam Chemical Integrators Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: October 22, 2019 Received: October 23, 2019
Dear Mary Fretland:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Elizabeth Claverie, M.S. Assistant Director for THT4B2 Acting Assistant Director for THT4B1 DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K191236
### Device Name
3MTM Attest™ Steam Chemical Integrators
### Indications for Use (Describe)
The 3M™ Attest™ Steam Chemical Integrators are designed to all critical parameters over a specified range of steam sterilization cycles. The integrating intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
| Cycle Type | Temperature | Exposure Time |
|---------------------|-------------|--------------------------|
| Gravity | 250°F/121°C | 30 minutes |
| Gravity | 270°F/132°C | 3, 4, 10, 15, 25 minutes |
| Gravity | 275°F/135°C | 3, 10 minutes |
| Dynamic Air Removal | 250°F/121°C | 30 minutes |
| Dynamic Air Removal | 270°F/132°C | 4, 10 minutes |
| Dynamic Air Removal | 273°F/134°C | 3, 4 minutes |
| Dynamic Air Removal | 275°F/135°C | 3 minutes |
Minimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer: 250°F / 121°C: 16.5 minutes 270°F/ 132°C: 2.0 minutes 273°F/ 134°C: 1.4 minutes 275°F/ 135°C: 1.2 minutes
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" and the letter "M" in a bold, sans-serif font. The logo is simple and recognizable.
# 510(k) Summary for 3MTM Attest™ Steam Chemical Integrators
### Sponsor Information:
3M Health Care 3M Center, Bldg. 275-5W-06 St. Paul, MN 55144-1000
Contact: Mary Fretland Senior Regulatory Affairs Associate Phone Number: (651) 737-2296 Fax Number: (651) 737-5320 Email: mfretland@mmm.com
Date of Summary: November 21, 2019
Submission Number: K191236
{4}------------------------------------------------
#### 1. Device Name and Classification:
| Common or Usual Name: | Chemical Indicators |
|------------------------|----------------------------------------------------|
| Trade Name: | 3MTM AttestTM Steam Chemical Integrators |
| Classification Name: | Physical/chemical sterilization process indicators |
| Device Classification: | Class II, 21 CFR § 880.2800(b) |
| Product Code: | JOJ |
#### 2. Predicate Device:
3MTM ComplyTM SteriGage™ Chemical Integrator for Steam (K101249)
#### 3. Description of Device:
3M™ Attest™ Steam Chemical Integrators are chemical indicators consisting of a paper wick and a steam and temperature sensitive chemical pellet contained in a paper/film/foil laminate. The chemical pellet melts and migrates as a dark color along the paper wick. The migration is visible through a window marked ACCEPT or window marked REJECT; the extent of migration depends on steam, time, and temperature.
{5}------------------------------------------------
#### 4. Indications for Use
The 3M™ Attest™ Steam Chemical Integrators are designed to respond to all critical parameters over a specified range of steam sterilization cycles. The integrating indicator is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:
| Cycle Type | Temperature | Exposure Time |
|---------------------|-------------|---------------|
| Gravity | 250°F/121°C | 30 minutes |
| Gravity | 270°F/132°C | 3 minutes |
| Gravity | 270°F/132°C | 4 minutes |
| Gravity | 270°F/132°C | 10 minutes |
| Gravity | 270°F/132°C | 15 minutes |
| Gravity | 270°F/132°C | 25 minutes |
| Gravity | 275°F/135°C | 3 minutes |
| Gravity | 275°F/135°C | 10 minutes |
| Dynamic Air Removal | 250°F/121°C | 30 minutes |
| Dynamic Air Removal | 270°F/132°C | 4 minutes |
| Dynamic Air Removal | 270°F/132°C | 10 minutes |
| Dynamic Air Removal | 273°F/134°C | 3 minutes |
| Dynamic Air Removal | 273°F/134°C | 4 minutes |
| Dynamic Air Removal | 275°F/135°C | 3 minutes |
Minimum Stated Values for 3M™ Attest™ Steam Chemical Integrators as determined in a resistometer:
| 250°F/121°C | 270°F/132°C | 273°F/134°C | 275°F/135°C |
|--------------|-------------|-------------|-------------|
| 16.5 Minutes | 2.0 Minutes | 1.4 Minutes | 1.2 Minutes |
#### 5. Nonclinical Comparison to the Predicate Device
The design, fundamental technology and performance specifications for 3M™ Attest™ Steam Chemical Integrators are similar to the previously cleared device which is sold under the tradename 3MTM ComplyTM SteriGage™ Chemical Integrator for Steam (K101249).
There has been no change to the device's performance specifications or fundamental scientific technology. The intent of this submission is to expand the indications for use. re-brand the device, and modify the materials used to construct the device.
The differences between 3M™ Attest™ Steam Chemical Integrator and the predicate do not raise any new questions of safety and effectiveness as demonstrated by the performance testing and biocompatibility assessment.
{6}------------------------------------------------
#### 6. Technical Characteristics
### Technological Characteristics Comparison Between Subject and Predicate Device:
| Feature | Submission Device:<br>3MTM AttestTM Steam Chemical Integrators | Predicate Device (K101249):<br>3MTM ComplyTM SteriGageTM Chemical<br>Integrators for Steam |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Device Models | 1243RE and 1243RES | 1243RE |
| Device Design | 3MTM AttestTM Steam Chemical Integrators are<br>chemical indicators consisting of a paper wick and<br>steam and temperature sensitive chemical pellet<br>contained in a paper/film/foil laminate. The<br>chemical pellet melts and migrates as a dark color<br>along the paper wick. The migration is visible<br>through a window marked ACCEPT or window<br>marked REJECT; the extent of migration depends<br>on steam, time, and temperature. | Identical. |
| Indicator Agent | Proprietary formulation. | Identical. |
| Sterilization<br>method and cycles | Steam sterilization processes 250°F to 275°F<br>(121°C to 135°C) | Identical. |
| Shelf-life | One (1) year | Three (3) years |
## Indications for Use Comparison:
| Cycle Type | Temperature | Submission Device:<br>3MTM AttestTM Steam Chemical<br>Integrators<br>Exposure Time | Predicate Device (K101249):<br>3MTM Comply™ SteriGage™ Chemical<br>Integrators for Steam<br>Exposure Time |
|------------------------|-------------|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Gravity | 250°F/121°C | 30 minutes | ≥ 30 minutes |
| Gravity | 270°F/132°C | 3 minutes<br>4 minutes<br>10 minutes<br>15 minutes<br>25 minutes | ≥ 3 minutes (unwrapped) |
| Gravity | 275°F/135°C | 3 minutes<br>10 minutes | Not present. |
| Dynamic Air<br>Removal | 250°F/121°C | 30 minutes | Not present. |
| Dynamic Air<br>Removal | 270°F/132°C | 4 minutes<br>10 minutes | ≥ 3 minutes (unwrapped)<br>≥ 4 minutes (wrapped) |
| Dynamic Air<br>Removal | 273°F/134°C | 3 minutes<br>4 minutes | ≥ 3.5 minutes (unwrapped)<br>≥ 4 minutes (wrapped) |
| Dynamic Air<br>Removal | 275°F/135°C | 3 minutes | ≥ 3 minutes |
{7}------------------------------------------------
| Feature | Submission Device:<br>3MTM Attest™ Steam Chemical Integrators | | | | Predicate Device (K101249):<br>3MTM Comply™ SteriGage™ Chemical<br>Integrators for Steam | | | |
|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-----------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-----------------|-----------------|
| Endpoint<br>Specifications<br>(Minimum Stated<br>Values) | The minimum stated values for the 3MTM Attest™ Steam Chemical Integrators as determined using a resistometer are provided in the table below. | | | | The minimum stated values for the 3MTM Comply™ SteriGage™ Chemical Integrators for Steam as determined using a resistometer are provided in the table below. | | | |
| | Minimum Stated Values for 3MTM Attest™ Steam Chemical Integrators | | | | Minimum Stated Values for 3M™ ComplyTM SteriGage™ Steam Chemical Integrators | | | |
| | 250°F/<br>121°C | 270°F/<br>132°C | 273°F/<br>134°C | 275°F/<br>135°C | 250°F/<br>121°C | 270°F/<br>132°C | 273°F/<br>134°C | 275°F/<br>135°C |
| | 16.5<br>Minutes | 2.0<br>Minutes | 1.4<br>Minutes | 1.2<br>Minutes | 16.5<br>Minutes | 2.0<br>Minutes | 1.4<br>Minutes | 1.1<br>Minutes |
#### 7. Summary of Nonclinical Testing
Nonclinical testing of the 3M™ Attest™ Steam Chemical Integrators was conducted in accordance with the FDA Guidance for Industry and FDA Staff: Premarket Notification [510(k)] Submissions for Chemical Indicators, issued December 19, 2003 as well as ANSI/AAM/ISO 11140-1:2014.
The effectiveness of the 3MTM Attest™ Steam Chemical Integrator was demonstrated in the following tests:
| Test Name | Purpose | Acceptance Criteria | Results |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Stated Value (SV)<br>Testing | To identify the critical parameters<br>required to achieve a stated inactivation,<br>by referring to a stated test organism<br>with stated D and z values. The<br>integrator must turn to "ACCEPT" end<br>point at the stated value time and must<br>also remain "REJECT" when exposed to<br>conditions of -1°C/-15% set point of the<br>SV time. All testing is completed in a<br>saturated steam resistometer. Integrator<br>temperature coefficient and correlation<br>coefficient are calculated to confirm<br>alignment to biological indicator<br>performance. | SV at 121°C $\geq$ 16.5 minutes<br>SV at 132°C $\geq$ 2.0 minutes<br>SV at 134°C $\geq$ 1.4 minutes<br>SV at 135°C $\geq$ 1.2 minutes<br>Integrator<br>temperature<br>coefficient 10 - 27°C<br>Correlation<br>coefficient $\geq$ 0.9 | PASS |
| Health Care<br>Facility Simulated<br>Use Testing | Confirm integrators provide acceptable<br>performance in cleared customer use<br>sterilization cycles. | Device reaches "ACCEPT"<br>endpoint reaction when<br>exposed to customer use<br>cycles.<br>Device does not reach<br>endpoint ("REJECT") when<br>exposed to failing conditions<br>in customer use cycles. | PASS |
{8}------------------------------------------------
| Test Name | Purpose | Acceptance Criteria | Results |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------|
| Dry Heat Testing | Verify device requires the presence of saturated steam to turn to reach endpoint. | Endpoint must not be met following dry heat exposure at 140°C for 30 min. | PASS |
| Side-by-Side<br>Testing with<br>Biological<br>Indicator | Confirm integrators are parallel in performance to biological indicators (BI). | Chemical integrator parallels performance of BI and does not reach endpoint before BI is inactivated. | PASS |
| Endpoint Color<br>Stability | Confirm endpoint color stability for samples exposed to passing and failing conditions in a steam resistometer. | Endpoint decision must remain unchanged after 6 months. | PASS |
The results of performance testing on 3M™ Attest™ Steam Chemical Integrator demonstrate the device performs as intended in the claimed steam sterilization cycles.
#### 8. Conclusion
Based on the intended uses, technological characteristics and non-clinical performance data, the 3MTM AttestTM Steam Chemical Integrator is as safe, as effective and performs as well as or better than the legally marketed predicate, 3M™ Comply™ SteriGage™ Chemical Integrator for Steam cleared under K101249, Class II (CFR 880.2800, product code JOJ).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.