K220838 · Howmedica Osteonics, Dba Stryker Orthopaedics · LZN · May 20, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220838
Device Name
Artisan Bone Plug, Universal Cement Restrictor
Applicant
Howmedica Osteonics, Dba Stryker Orthopaedics
Product Code
LZN · General, Plastic Surgery
Decision Date
May 20, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Indications for Use
Artisan Bone Plug: These bone plugs are intended to be placed in the femoral canal prior to the introduction of bone cement in a cemented hip procedure. The plug is placed distally to the femoral stem to help allow cement pressurization and to help prevent cement migration further down the femoral canal. Universal Cement Restrictor: For cement spacers, mid-shaft restrictors and Cement Plugs: · In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be advantageous.
Device Story
Artisan Bone Plug and Universal Cement Restrictor are orthopedic implants used during cemented hip arthroplasty. Devices are placed distally within the femoral canal by a surgeon to act as a barrier; this facilitates bone cement pressurization and prevents distal cement migration. The current submission updates labeling to include MR Conditional status. No changes were made to device design, materials, or function. Devices are used in clinical settings by orthopedic surgeons.
Clinical Evidence
No clinical data. Substantial equivalence is supported by engineering analysis and non-clinical bench testing, including MR safety testing (ASTM F2052-15, F2213-17, F2119-07, F2182-19) and bacterial endotoxin testing (ANSI/AAMI ST72:2019).
Technological Characteristics
Passive orthopedic implants. Materials and design remain unchanged from predicates. MR safety testing performed per ASTM F2052-15, ASTM F2213-17, ASTM F2119-07, and ASTM F2182-19. Pyrogenicity testing per ANSI/AAMI ST72:2019.
Indications for Use
Indicated for patients undergoing cemented hip arthroplasty requiring a bone plug, cement spacer, or mid-shaft restrictor to facilitate cement pressurization and prevent distal cement migration.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
May 20, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Howmedica Osteonics, dba Stryker Orthopaedics Margaret Klippel Chief Specialist 325 Corporate Drive Mahwah, New Jersey 07430
# Re: K220838
Trade/Device Name: Artisan Bone Plug, Universal Cement Restrictor Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: LZN, JDI Dated: March 18, 2022 Received: March 22, 2022
# Dear Margaret Klippel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220838
Device Name Artisan Bone Plug
Indications for Use (Describe)
These bone plugs are intended to be placed in the femoral canal prior to the introduction of bone cemented hip procedure. The plug is placed distally to the femoral stem to help allow cement pressurization and to help prevent cement migration further down the femoral canal.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220838
Device Name Universal Cement Restrictor
Indications for Use (Describe)
For cement spacers, mid-shaft restrictors and Cement Plugs:
· In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be advantageous.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
| Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Margaret Klippel<br>Chief Specialist, Regulatory Affairs<br>Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Telephone: 201-831-5559 |
| Date Prepared: | May 20, 2022 |
| Proprietary Name: | Labeling Update for Artisan Bone Plug and Universal Cement Restrictor |
| Common Name: | Artificial Hip Replacement Components |
| Regulatory Class: | Class II |
| Regulation: | Hip joint metal/polymer semi-constrained cemented prosthesis 21 CFR<br>§888.3350 |
| | Surgical Mesh 21 CFR §878.3300 |
| Product Codes: | LZN – cement obturator |
| | JDI - Prosthesis, hip, semi-constrained, metal/polymer, cemented |
Legally Marketed Device to Which Substantial Equivalence is Claimed: Artisan Bone Plug - K951860 Universal Cement Restrictor - K924323
Legally Marketed Additional Predicate Devices Used to Support Substantial Equivalence: Universal Distal Spacer - K914406, K153345
## Device Description:
The Artisan Bone Plug and Universal Cement Restrictor are commercially available devices that have been determined substantially equivalent in previous 510(k) premarket notifications. The purpose of this submission is to modify the labeling of these devices to add MR Conditional labeling and make other minor labeling updates.
# Indication for Use:
{5}------------------------------------------------
There are no changes to the previously cleared indications for use. The indications for the subject components are as follows.
#### Artisan Bone Plug:
These bone plugs are intended to be placed in the femoral canal prior to the introduction of bone cement in a cemented hip procedure. The plug is placed distally to the femoral stem to help allow cement pressurization and to help prevent cement migration further down the femoral canal.
## Universal Cement Restrictor:
For cement spacers, mid-shaft restrictors and Cement Plugs:
- . In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be advantageous.
## Summary of Technological Characteristics:
There have been no changes to the technological characteristics of the subject devices as a result of the revision to the labeling. The subject devices have the same design and are manufactured from the same materials as the predicate devices.
## Non-Clinical Testing:
In previous premarket notification K153345 non-clinical testing was presented to characterize the compatibility of Stryker Orthopaedics Hip System devices in the MR environment. Testing was performed according to the standards listed below:
- Magnetically Inducted Displacement Force performed per ASTM F2052-15, Standard Test ● Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the MR Environment
- Magnetically Induced Torque performed per ASTM F2213-17, Standard Test Method for ● Measurement of Magnetically Induced Torque on Medical Devices in the MR Environment
- Image Artifact performed per ASTM F2119-07 (Reapproved 2013), Standard Test Method ● for Evaluation of MR Image Artifacts from passive Implants
- Heating by RF Fields per ASTM F2182-19, Standard Test Method for Measurement of Radio Frequency Induced Heating near Passive Implants during MR Imaging
An engineering analysis was conducted to determine if the Artisan Bone Plug and Universal Cement Restrictors pose a new worst case when compared to the devices previously tested in K153345. The results of this analysis indicated that the subject devices do not create a new worst case.
The labeling has been modified to include the MR conditional symbol, and to provide the parameters under which a patient who has the device can be safely scanned.
{6}------------------------------------------------
Bacterial endotoxin testing (BET) as specified in ANSI/AAMI ST72:2019 was used for pyrogenicity testing to achieve an endotoxin limit of < 20EU/Device.
Clinical Testing: Clinical testing was not required as a basis for substantial equivalence.
Conclusion: The subject Artisan Bone Plug and Universal Cement Restrictor components are substantially equivalent to the predicate devices identified in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.