K220830 · These Three Medical, LLC · ONN · Sep 15, 2022 · Neurology
Device Facts
Record ID
K220830
Device Name
TrueBalance™ Surgical System
Applicant
These Three Medical, LLC
Product Code
ONN · Neurology
Decision Date
Sep 15, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
TrueBalance™ Surgical System is indicated for any medical condition in which primary or revision Total Knee Arthroplasty (TKA) would be indicated. For use as a tool to measure implant alignment and for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. TrueBalance™ Surgical System is completely reusable with the exception of the Ultra-Thin Force Sensor which is single use.
Device Story
TrueBalance™ Surgical System is an intraoperative orthopedic joint assessment tool used by surgeons during TKA. It measures dynamic loads in medial/lateral knee compartments and static angular alignment of bone cuts. Inputs include force data from an Ultra-Thin Force Sensor and orientation data from an embedded accelerometer. The system transforms these inputs into numerical and pictorial load/alignment data, transmitted via Bluetooth to an iPad running the TrueBalance™ Software Application (TSA). The surgeon uses this real-time feedback to adjust femoral implant positioning and balance flexion gaps, aiming to reduce post-operative instability. The system is implant-agnostic and reusable, except for the single-use force sensor. All hardware components (excluding the iPad) are steam-sterilizable. The device provides quantifiable data to assist in surgical decision-making, potentially improving implant alignment and joint stability.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including software lifecycle validation (IEC 62304), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), sterilization validation, biocompatibility (ISO 10993), shelf-life/packaging testing, and performance testing for load/force and alignment accuracy.
Technological Characteristics
Materials: VICTREX 450G PEEK, silicone gasket. Sensing: Capacitive load sensors, accelerometer. Energy: Internal 3.6V lithium battery. Connectivity: Bluetooth 4.2 to iPad. Form factor: Modular, autoclavable components. Software: TSA application on iOS. Sterilization: Steam autoclave. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993, ISO 17665, IEC 62304.
Indications for Use
Indicated for patients undergoing primary or revision Total Knee Arthroplasty (TKA).
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
September 15, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
These Three Medical, LLC Moj Eram Regulatory Affairs Correspondent 20860 N. Tatum Blvd.. Suite 300 Phoenix, Arizona 85050
Re: K220830
Trade/Device Name: TrueBalance™ Surgical System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: ONN Dated: August 15, 2022 Received: August 16, 2022
Dear Moj Eram:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220830
Device Name TrueBalance™ Surgical System
Indications for Use (Describe)
TrueBalance™ Surgical System is indicated for any medical condition in which primary or revision Total Knee Arthroplasty (TKA) would be indicated.
For use as a tool to measure implant alignment of the femoral knee implant to reduce instability from flexion gap asymmetry. TrueBalance™ Surgical System is completely reusable with the exception of the Ultra-Thin Force Sensor which is single use.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for These Three Medical, LLC. The logo features a blue circle with three curved lines emanating from it, resembling a signal or sound wave. To the right of the symbol, the words "THESE THREE" are written in a bold, sans-serif font, with "MEDICAL, LLC" written in a smaller font below.
# 510(k) Summary
# 21 CFR 807.92(a)
#### 1 General Provisions
| Submitter Name: | These Three Medical, LLC |
|----------------------|------------------------------------------------------|
| Address: | 20860 N. Tatum Blvd., Suite 300<br>Phoenix, AZ 85050 |
| Contact Person: | Jay Pierce |
| Telephone Number: | 480.771.4133 |
| Date of Preparation: | 15 August 2022 |
#### Principle and Regulatory Contacts 2
### Submitted By/Principal Contact:
| Contact Name: | Jay Pierce |
|---------------|------------------------------------------------------|
| Title: | President/CEO<br>These Three Medical, LLC |
| Address: | 20860 N. Tatum Blvd., Suite 300<br>Phoenix, AZ 85050 |
| Phone Number: | 480.771.4133 |
| Email: | huan@generalvibronics.com |
### Regulatory Affairs Correspondance:
| Name: | Moj Eram, PhD |
|---------------|-----------------------------------|
| Title: | Regulatory Affairs Representative |
| | Sage BioPartners, LLC |
| Phone Number: | 801.230.8611 (Mobile) |
| Email: | moj.eram@sagebiopartners.com |
#### 3 Subject Device
| Trade Name: | TrueBalance™ Surgical System |
|-------------------------------|------------------------------|
| Classification Name: | Stereotaxic Instrument |
| Classification: | Class II |
| Product Code: | ONN |
| Regulation Number: | 21 CFR 822.4560 |
| Regulation Medical Specialty: | Neurology |
| Review Panel: | Orthopedic |
{4}------------------------------------------------
#### Predicate Devices ব
- 1. VERASENSE For Zimmer Biomet Persona CR C-D/3-9 Left, VERASENSE For Zimmer Biomet Persona CR C-D/3-9 Right, VERASENSE For Zimmer Biomet Persona CR E-F/3-11 Left, VERASENSE For Zimmer Biomet Persona CR E-F/3-11 Right, VERASENSE For Zimmer Biomet Persona CR G-H/7-12 Left (hereafter VERASENSE For Zimmer Biomet Persona)
| Manufacturer: | OrthoSensor, Inc. |
|-------------------------------|------------------------------------------------|
| 510(k) Number: | K193580 |
| Classification: | Class II |
| FDA Product Code: | ONN (Subsequent OLO) |
| Regulation Number: | 21 CFR 882.4560 |
| Regulation Medical Specialty: | Neurology |
| Review Panel: | Orthopedic |
| Device Classification Name: | Intraoperative Orthopedic Joint Assessment Aid |
#### VERASENSE For Zimmer Biomet Persona 2.
| Manufacturer: | OrthoSensor, Inc. |
|----------------------------------------------------------------------------|-------------------|
| 510(k) Number: | K180459 |
| Classification: | Class II |
| FDA Product Code: | ONN |
| Regulation Number: | 21 CFR 882.4560 |
| Regulation Medical Specialty: | Neurology |
| Review Panel: | Orthopedic |
| Device Classification Name: Intraoperative Orthopedic Joint Assessment Aid | |
#### 5 Device Description
TrueBalance™ Surgical System provides a means to measure implant alignment and dynamically balance the knee during primary or revision Total Knee Arthroplasty (TKA).
TrueBalance™ Surgical System consists of seven components and one optional component:
- Electronic Module, 1.
- 2. Ultra-Thin Force Sensor,
- Multifunction Alignment Handle, 3.
- 4. Sensor Covers (8 sizes),
- 5. Alignment Rod,
- Sterilization Tray, and 6.
- 7. TrueBalance™ Software Application (TSA) and Display (iPad)
- 8. (Optional) Off-the-shelf Tablet Floor Stand for iPad
The TrueBalance™ Surgical System Electronic Module and hardware components are steam sterilizable via autoclave and reusable via a Sterilization Tray. TrueBalance™ Surgical System
{5}------------------------------------------------
Ultra-Thin Force Sensor is single use, packaged non-sterilizable via autoclaye. The TrueBalance™ Surgical System Electronic Module is steam sterilizable via autoclave in either the Sterilization Tray or standard autoclavable pouches.
TrueBalance™ Surgical System measures dynamic loads in the medial and lateral compartments of the knee and static angular measurements for the alignment functionality. The data from the Ultra-Thin Force Sensor and embedded accelerometer is wirelessly transmitted to the TrueBalance™ Software Application (TSA) and Display (iPad) which is located outside the sterile field in the operating room.
#### Indications for Use 6
TrueBalance™ Surgical System is indicated for any medical condition in which primary or revision Total Knee Arthroplasty (TKA) would be indicated.
For use as a tool to measure implant alignment and for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. TrueBalance™ Surgical System is completely reusable with the exception of the Ultra-Thin Force Sensor which is single use.
#### 7 Comparison of Technological Characteristics with the Predicate Devices
A comparison of the subject device. TrueBalance™ Surgical System and predicate devices. VERASENSE For Zimmer Biomet Persona (K193580) & VERASENSE For Zimmer Biomet Persona (K180459), is based on the following same technologies:
- . Electronics and sensors to measure load/force pressure between the femur and tibia in TKA procedures.
- . Electronics and accelerometer to measure coronal plane alignment of the tibial resection relative to the patient specific tibia reference axis defined by the system registered landmarks, and numerical value of the varus/valgus tibial mechanical axis.
- Required components to support the system includes a device that communicates via Bluetooth communications and a software application.
- Algorithms to calculate medial and lateral load pressures and present to a user interface. ●
- Sensor covers (shims) for thickness. ●
In addition to the similarities stated above, the following technological differences exist between the subject device, TrueBalance™ Surgical System and Predicate devices, VERASENSE For Zimmer Biomet Persona (K193580) & (K180459):
- . The Predicate device measures condylar load/force only when the trial hardware components are placed while the TrueBalance™ Surgical System measures condylar loading during gap balancing and trialling phase.
- TrueBalance Surgical System utilizes a Multifunctional Alignment Handle to measure ● load/force in the medial and lateral compartments, along with load positions in the anterior and posterior compartments.
- In addition to measuring the resected the tibia, TrueBalance Surgical System utilizes an accelerometer and 21inch Alignment Rod to measure coronal plane alignment of the femoral resection relative to the patient specific femur reference axis defined by the
{6}------------------------------------------------
system registered landmarks, and numerical value of the varus/valgus femoral mechanical axis. Note that most conventional instrumentation for TKA procedures includes a spacer block and alignment rod in order to provide the surgeon a tool for visual interpretation of post cut limb alignment, albeit without quantifiable data.
- TrueBalance™ Software Application (TSA) and Display (iPad) presents tibia slope and . mechanical axis of the limb.
- . The Predicate devices enclose all electronics and sensors in a plastic, hermetically sealed, single use component. The TrueBalance Surgical System electronics module is decoupled from the sensor and the multifunction alignment handle. The system requires assembly in the operating room.
- . The Ultra-Thin Force Sensor fits directly between the Multifunction Alignment Handle and numerous size Sensor Covers and is placed between the Femur and resected Tibia with or without trial hardware.
- . The entire TrueBalance Surgical System (excluding TSA) is autoclavable.
- TrueBalance Surgical System is reusable. (Exception is the Ultra-thin Force Sensor which . single use)
- The TrueBalance™ Surgical System is implant company agnostic system and not dependent on the dimensions of OEM implant components.
- . The Predicate devices provide a separate computer (link station) that houses the Software Application. TrueBalance™ Software Application (TSA) and Display (iPad) is provided to the hospital as part of a complete system.
| Table 5.1 - Subject and Predicate devices Comparison | | | | |
|----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| | Subject Device<br>TrueBalance™ Surgical System | Predicate Device<br>VERASENSE for Zimmer<br>Biomet Persona<br>(K193580) | Predicate Device<br>VERASENSE for Zimmer<br>Biomet Persona<br>(K180459) | |
| Sponsor / Owner | These Three Medical, LLC | OrthoSensor, Inc. | OrthoSensor, Inc. | |
| Regulation<br>Number | 21 CFR 882.4560 | 21 CFR 882.4560 | 21 CFR 882.4560 | |
| Product Code | ONN | ONN<br>Subsequent OLO | ONN | |
| Environment for<br>Use | Hospital (operating room) | Hospital (operating room) | Hospital (operating room) | |
| Intended User | Surgeon | Surgeon | Surgeon | |
| General Device<br>Description | A measurement tool to provide the<br>surgeon with quantifiable data<br>relative to load and angular<br>measurements. | A measurement tool to<br>provide the surgeon with<br>quantifiable data relative to<br>load and angular<br>measurements. | A measurement tool to<br>provide the surgeon with<br>quantifiable data relative to<br>load measurements. | |
| Table 5.1 – Subject and Predicate devices Comparison | | | | |
| | Subject Device<br>TrueBalance™ Surgical System | Predicate Device<br>VERASENSE for Zimmer<br>Biomet Persona<br>(K193580) | Predicate Device<br>VERASENSE for Zimmer<br>Biomet Persona<br>(K180459) | |
| Principle of<br>Operation | A measurement system used to<br>collect quantifiable data that is<br>useful to the surgeon performing<br>Total Knee Arthroplasty (TKA)<br>procedures.<br>Capacitive sensors on the device<br>measure the pressure exerted by the<br>femoral component of the implant<br>on the sensor.<br>The components include an<br>Electronic Module, Ultra-Thin<br>Force Sensor, Sensor Covers<br>(various thicknesses), Multifunction<br>Alignment Handle, Alignment Rod,<br>and an iPad with a Software<br>Application. The load sensors on the<br>device, measure the pressures<br>exerted between the cut femur and<br>tibia and between the femoral<br>component of the implant on the<br>sensor. The embedded<br>accelerometer works in conjunction<br>with the Alignment Rod and is<br>placed in specified procedural<br>positions, in order to measure angles<br>associated with bone cuts.<br>The Electronic Module is connected<br>via Bluetooth to an iPad hosting the<br>TrueBalance™ Software<br>Application (TSA). The TSA<br>displays both numerically and<br>pictorially load data along with<br>angular measurements (alignment<br>data), according to the description<br>and specifications. | A measurement system used<br>to collect quantifiable data<br>that is useful to the surgeon<br>performing Total Knee<br>Arthroplasty (TKA)<br>procedures.<br>Capacitive sensors on the<br>device measure the pressure<br>exerted by the femoral<br>component of the implant on<br>the sensor.<br>The VERASENSE Software<br>Application calculates force<br>vectors and positional data,<br>display both numerically and<br>pictorially load data versus<br>position, according to the<br>description and<br>specifications.<br>Transceiver is connected to<br>LinkStation via a USB cable.<br>Data from the VERASENSE<br>is wirelessly received by the<br>transceiver at a frequency of<br>401.05-405.55 MHz. The<br>transceiver processes<br>Gaussian Frequency-Shift<br>Keying (GFSK) modulated<br>data. | A measurement system used<br>to collect quantifiable data<br>that is useful to the surgeon<br>performing Total Knee<br>Arthroplasty (TKA)<br>procedures.<br>Capacitive sensors on the<br>device measure the pressure<br>exerted by the femoral<br>component of the implant on<br>the sensor.<br>The VERASENSE Software<br>Application calculates force<br>vectors and positional data,<br>display both numerically and<br>pictorially load data versus<br>position, according to the<br>description and<br>specifications.<br>Transceiver is connected to<br>LinkStation via a USB cable.<br>Data from the VERASENSE<br>is wirelessly received by the<br>transceiver at a frequency of<br>401.05-405.55 MHz. The<br>transceiver processes<br>Gaussian Frequency-Shift<br>Keying (GFSK) modulated<br>data. | |
| Control<br>Mechanism | Microelectronics, sensors and<br>accelerometer to measure the<br>femoral component load and angles<br>of bone cuts. | Microelectronics, capacitive<br>sensors and accelerometer to<br>measure the femoral<br>component load and angles<br>of bone cuts. | Microelectronics, capacitive<br>sensors and accelerometer to<br>measure the femoral<br>component load. | |
| Table 5.1 - Subject and Predicate devices Comparison | | | | |
| | Subject Device<br>TrueBalance™ Surgical System | Predicate Device<br>VERASENSE for Zimmer<br>Biomet Persona<br>(K193580) | Predicate Device<br>VERASENSE for Zimmer<br>Biomet Persona<br>(K180459) | |
| | | Materials | | |
| Electronic Module<br>Top Housing | VICTREX450G Natural PEEK -<br>Material Polyetheretherketone<br>(Electronic Module) | Colored Polycarbonate Resin | Colored Polycarbonate Resin | |
| Electronic Module<br>Bottom Housing | VICTREX450G Natural PEEK -<br>Material Polyetheretherketone<br>(Electronic Module) | Colored Polycarbonate Resin | Colored Polycarbonate Resin | |
| Electronic Module<br>Seal Between Top<br>and Bottom<br>Housing | Silicone Gasket | Loctite | Loctite | |
| Sensor Covers | VICTREX450G Natural PEEK -<br>Material Polyetheretherketone…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.