K180459 · Orthosensor, Inc. · ONN · Jun 7, 2018 · Neurology
Device Facts
Record ID
K180459
Device Name
VERASENSE for Zimmer Biomet Persona
Applicant
Orthosensor, Inc.
Product Code
ONN · Neurology
Decision Date
Jun 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The VERASENSE is indicated for any medical condition in which primary or revision Total Knee Arthroplasty (TKA) would be indicated. For use as a tool for adjustment of the implant to reduce instability from flexion gap asymmetry. The VERASENSE is sterile, for single patient use.
Device Story
VERASENSE is an intelligent disposable tibial insert for intraoperative knee balancing during TKA. It uses capacitive sensors to measure dynamic loads in medial and lateral knee compartments. Data is transmitted wirelessly to a LinkStation MINI or Evaluation Kit running the VERASENSE Software Application (VSA). The VSA displays real-time graphical and numerical load data and center of load (COL) location to the surgeon. Used in hospital operating rooms by surgeons to guide implant adjustment, reducing flexion gap asymmetry and improving joint stability. The device is sterile, single-use, and implant-system specific.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), electrical safety, EMC, software verification/validation, and performance testing (usability, sterilization, packaging, shelf life).
Technological Characteristics
Disposable tibial insert with capacitive sensors. Materials: PEEK (VITREX 381G) and stainless steel. Power: Two 3V Lithium batteries (parallel). Wireless: 2402–2480 MHz Bluetooth. Sterilization: Ethylene Oxide (10^-6 SAL). Software: VSA running on Windows 10 tablet. Connectivity: Bluetooth/USB. Electrical safety: AAMI ANSI ES60601-1. EMC: IEC 60601-1-2.
Indications for Use
Indicated for patients undergoing primary or revision Total Knee Arthroplasty (TKA) requiring adjustment of the femoral implant to reduce instability from flexion gap asymmetry.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white. Underneath the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
June 7, 2018
OrthoSensor, Inc. Deborah Johnson Senior Manager Regulatory Affairs 1855 Griffin Road, Suite A-310 Dania Beach, Florida 33004
Re: K180459
Trade/Device Name: VERASENSE for Zimmer Biomet Persona Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: ONN Dated: May 16, 2018 Received: May 17, 2018
Dear Deborah Johnson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Katherine D. Kavlock -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180459
Device Name VERASENSE for Zimmer Biomet Persona
### Indications for Use (Describe)
The VERASENSE is indicated for any medical condition in which primary or revision Total Knee Arthroplasty (TKA) would be indicated.
For use as a tool for adjustment of the implant to reduce instability from flexion gap asymmetry. The VERASENSE is sterile, for single patient use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Orthosensor. The logo consists of a stylized symbol above the company name. The symbol is made up of three curved shapes, one in dark blue, one in light blue, and one in gray. The company name is written in a simple, sans-serif font, with the "ORTHO" portion in gray and the "SENSOR" portion in blue.
K180459
### Traditional 510(k) Summary
### This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92
#### I. SUBMITTER
OrthoSensor, Inc. 1855 Griffin Road, Suite A-310 Dania Beach, FL 33004 Establishment Registration Number: 3008592715
Phone: (954) 577-7770 Fax: (954) 337-9222
Contact Person: Deborah Johnson, Senior Manager Regulatory Affairs
Date Prepared: May 15, 2018
#### II. DEVICE
Name of Device: VERASENSE for Zimmer Biomet Persona Common or Usual Name: Intraoperative Orthopedic Joint Assessment Aid Classification Name: Stereotaxic instrument Regulatory Class: II Regulation Number: 21 CFR 882.4560 Product Code: ONN
#### III. PREDICATE DEVICE
VERASENSE Knee System, K150372
#### IV. DEVICE DESCRIPTION
VERASENSE provides a means to dynamically balance the knee during primary or revision Total Knee Arthroplasty (TKA).
The VERASENSE consists of a knee sensor device and the following required accessories: LinkStation MINI or LinkStation Evaluation Kit and VERASENSE Software Application (VSA). The required accessories are intended to support the performance of the VERASENSE parent device. The LinkStation MINI / LinkStation MINI Evaluation Kit is capable of running the VSA and communicating wirelessly with the VERASENSE from outside the sterile field in the operating room. Individual VERASENSE devices are packaged sterile,
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Orthosensor. The logo consists of a stylized symbol above the word "ORTHOSENSOR". The symbol is made up of three curved shapes, two in blue and one in gray, arranged in a triangular formation. The word "ORTHOSENSOR" is written in a simple, sans-serif font, with the letters in gray.
for single patient use with a Shim Set for thickness adjustments. The required accessories are not packaged sterile and are reusable.
The VERASENSE device is an intelligent disposable tibial insert that measures dynamic loads in the medial and lateral compartments of the knee and wirelessly transmits the measured load data to the LinkStation MINI or LinkStation MINI Evaluation Kit. The VSA provides the surgeon with a graphical and numerical presentation of the loads in both the medial and lateral compartments of the knee. The VSA also provides location of the center of load (COL) from the femoral to the tibial component in each of the medial and lateral compartments for reference only.
VERASENSE devices are implant system specific due to variations in implant design. VERASENSE is compatible with the following implant systems:
- VERASENSE for Biomet Vanguard ●
- VERASENSE for Stryker Triathlon ●
- VERASENSE for Zimmer NexGen ●
- VERASENSE for Smith & Nephew Legion ●
- VERASENSE for Smith & Nephew Journey II ●
- . VERASENSE for Zimmer Biomet Persona
#### INDICATIONS FOR USE V.
The VERASENSE is indicated for any medical condition in which primary or revision Total Knee Arthroplasty (TKA) would be indicated.
For use as a tool for adjustment of the femoral knee implant to reduce instability from flexion gap asymmetry. The VERASENSE is sterile, for single patient use.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
At a high level, the subject and predicate devices are based on the following same technological elements:
- Capacitive sensors on the device measure the pressure exerted by the ● femoral component of the implant on the sensor.
- Required accessories intended to support the performance of the VERASENSE parent device are the LinkStation MINI or LinkStation MINI Evaluation Kit and VERASENSE Software Application (VSA)
- Patient contacting material of the top and bottom housings
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Orthosensor. The logo consists of a stylized symbol above the word "ORTHOSENSOR". The symbol is made up of three curved shapes, two in blue and one in gray, arranged in a triangular formation. The word "ORTHOSENSOR" is written in a simple, sans-serif font, with each letter in gray.
The following technological differences exist between the subject and predicate devices:
- the geometry / dimensions of the Zimmer Biomet Persona implant . system
- Articular surface is on shims, detached from the electronics ●
- Electronics module is reversible ●
- Patient contacting shim material changed to PolyEtherEtherKetone ● (PEEK) and stainless steel
- . Sensor wireless communication frequency changed to 2402 – 2480 MHz Bluetooth
- . New wireless connection mode for the LinkStation MINI / LinkStation MINI Evaluation Kit via tablet component integrated Bluetooth
- New version of the VSA with the addition of the VERASENSE for Zimmer Biomet Persona to the scope of the software application
- . Updated labeling to include compatibility with VERASENSE for Zimmer Biomet Persona variant and instructions for new wireless connection mode
| | Predicate Device K150372 | Subject Device |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification | 21 CFR 882.4560 | 21 CFR 882.4560 |
| Product Code | ONN | ONN |
| Indications for Use | The VERASENSE is indicated for any<br>medical condition in which primary or<br>revision Total Knee Arthroplasty (TKA)<br>would be indicated.<br>For use as a tool for adjustment of the<br>femoral knee implant to reduce<br>instability from flexion gap asymmetry.<br>The VERASENSE is sterile, for single<br>patient use. | The VERASENSE is indicated for any<br>medical condition in which primary or<br>revision Total Knee Arthroplasty (TKA)<br>would be indicated.<br>For use as a tool for adjustment of the<br>femoral knee implant to reduce<br>instability from flexion gap asymmetry.<br>The VERASENSE is sterile, for single<br>patient use. |
| Environment | Hospital | Hospital |
| VERASENSE (sensor) | | |
| Operating Principle | Capacitive sensors on the device<br>measure the pressure exerted by the<br>femoral component of the implant<br>on the sensor. | Capacitive sensors on the device<br>measure the pressure exerted by the<br>femoral component of the implant<br>on the sensor. |
| Power Requirements | Internally powered<br>Two 1.55V, 27mAh Silver Oxide<br>batteries (in series) | Internally powered<br>Two 3V, 30mAh Lithium batteries<br>(in parallel) |
| | Predicate Device K150372 | Subject Device |
| Form of Wireless technology | Radio Frequency | Radio Frequency |
| RF<br>Frequencies | VERASENSE<br>for Stryker<br>Triathlon<br>402.0 – 405.0 MHz | 2402 – 2480 MHz<br>Bluetooth |
| | VERASENSE<br>for Biomet<br>Vanguard | |
| | VERASENSE<br>for Zimmer<br>NexGen | |
| | VERASENSE<br>for Smith &<br>Nephew<br>Legion<br>404.3 MHz | |
| | VERASENSE<br>for Smith &<br>Nephew<br>Journey II | |
| | VERASENSE<br>for Zimmer<br>Biomet<br>Persona | |
| | | |
| Range | 2 m | 2 m |
| Packaging | Sterile, Double Tyvek /Film pouches,<br>chipboard box | Sterile, Double Tyvek /Film pouches,<br>chipboard box |
| Sterilization Method<br>(sensor only) | Ethylene Oxide 10-6 SAL | Ethylene Oxide 10-6 SAL |
| Materials<br>(patient<br>contacting) | Top<br>housing<br>Colored Polycarbonate Resin | Colorless Polycarbonate Resin |
| | Bottom<br>housing<br>Colorless Polycarbonate Resin | Colorless Polycarbonate Resin |
| | Adhesive<br>between<br>top and<br>bottom<br>housing<br>Loctite | Loctite |
| | Shims<br>Colored Polycarbonate Resin | 10, 11, 12, 13 mm<br>VITREX PolyEtherEtherKetone (PEEK)<br>381G & stainless steel metal plates<br><br>14 and 16 mm<br>VITREX PolyEtherEtherKetone (PEEK)<br>381G |
| Shelf Life | 24 months | 3 months |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Orthosensor. The logo consists of a stylized symbol above the company name. The symbol is composed of three curved shapes, two in blue and one in gray, arranged in a triangular formation. The text "ORTHOSENSOR" is written in a serif font, with the "S" in "SENSOR" being slightly larger and a different shade of blue than the rest of the letters.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Orthosensor. The logo consists of a stylized symbol above the company name. The symbol is made up of three curved lines, with the top line being blue, the middle line being gray, and the bottom line being blue. The company name is written in a sans-serif font, with the letters being gray.
| Predicate Device K150372 | | Subject Device | | |
|-------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Required Accessories (intended to support the performance of the VERASENSE parent device) | | | | |
| | VERASENSE Software Application (VSA) | | | |
| | VSA for Mac<br>OX | VSA for Microsoft<br>Windows | VSA for Microsoft Windows | |
| Operating Principle | The VERASENSE<br>Software<br>Application<br>calculates force<br>vectors and<br>positional data,<br>display both<br>numerically and<br>pictorially load<br>data versus<br>position,<br>according to the<br>description and<br>specifications. | The VERASENSE Software<br>Application calculates force<br>vectors and positional data,<br>display both numerically and<br>pictorially load data versus<br>position, according to the<br>description and<br>specifications. | The VERASENSE Software Application calculates<br>force vectors and positional data, display both<br>numerically and pictorially load data versus<br>position, according to the description and<br>specifications. | |
| Operating System<br>Compatibility | Mac OSX | Microsoft Windows | Microsoft Windows | |
| VSA version number(s) | 3.7 - 3.9.4.2 | 4.0.31.0 - 5.0.0.0 | 5.1.0.17 | |
| | | LinkStation | | |
| | | LinkStation | LinkStation<br>MINI | LinkStation MINI<br>Evaluation Kit |
| Operating Principle | Transceiver is connected to<br>LinkStation via a USB cable. Data<br>from the VERASENSE is wirelessly<br>received by the transceiver at a<br>frequency of 401.05-405.55 MHz.<br>The transceiver processes Gaussian<br>Frequency-Shift Keying (GFSK)<br>modulated data | | The transceiver<br>component of the<br>LinkStation<br>MINI/LinkStation MINI<br>Evaluation Kit can<br>receive and process<br>simultaneously 2<br>channels and the<br>carrier frequency band<br>for the wireless<br>communication is<br>401.05 - 405.55 MHz.<br>The integrated<br>Bluetooth transceiver<br>in the tablet computer<br>component of the<br>LinkStation<br>MINI/LinkStation MINI<br>Evaluation Kit carrier<br>frequency band for the<br>wireless<br>communication is 2402<br>– 2480 MHz Bluetooth<br>4.0 compatible. The<br>transceivers processes<br>Gaussian Frequency-<br>Shift Keying (GFSK)<br>modulated data. | The transceiver<br>component of the<br>LinkStation<br>MINI/LinkStation MINI<br>Evaluation Kit can<br>receive and process<br>simultaneously 2<br>channels and the carrier<br>frequency band for the<br>wireless communication<br>is 401.05 - 405.55 MHz.<br>The integrated Bluetooth<br>transceiver in the tablet<br>computer component of<br>the LinkStation<br>MINI/LinkStation MINI<br>Evaluation Kit carrier<br>frequency band for the<br>wireless communication<br>is 2402 — 2480 MHz<br>Bluetooth 4.0<br>compatible. The<br>transceivers processes<br>Gaussian Frequency-Shift<br>Keying (GFSK) modulated<br>data. |
| | | Predicate Device K150372 | Subject Device | |
| Sensor Compatibility | | VERASENSE for Stryker Triathlon | VERASENSE for Stryker Triathlon | VERASENSE for Stryker Triathlon |
| | | VERASENSE for Biomet Vanguard | VERASENSE for Biomet Vanguard | VERASENSE for Biomet Vanguard |
| | | VERASENSE for Zimmer NexGen | VERASENSE for Zimmer NexGen | VERASENSE for Zimmer NexGen |
| | | VERASENSE for Smith & Nephew Legion | VERASENSE for Smith & Nephew Legion | VERASENSE for Smith & Nephew Legion |
| | | VERASENSE for Smith & Nephew Journey II | VERASENSE for Smith & Nephew Journey II | VERASENSE for Smith & Nephew Journey II |
| Display Unit | | Apple iMac 27" | Tablet PC 12.5" | VERASENSE for Zimmer Biomet Persona<br>Tablet PC 12.5" |
| | Display Unit Connection Modes | WIFI connectivity, Bluetooth connectivity, USB ports, Ethernet, Thunderbolt, and FireWire | WIFI connectivity, Bluetooth connectivity, and a USB port | WIFI connectivity, Bluetooth connectivity, and a USB port |
| | Display Unit Power Requirements | 310W universal 3-pin jack, 100-240V, 50-60Hz | 65W universal 3-pin jack, 100-240V, 1.5A, 50-60Hz | 65W universal 3-pin jack, 100-240V, 1.5A, 50-60Hz |
| | Display Unit Operating System | Mac OSX | Windows 10 | Windows 10 |
| Stand Component | | Ergotron Neo-Flex Stand | Roll Stand | Tablet Kickstand |
| Magnet | | Neodymium magnet with a magnet to steel disc pull of greater than 2.75 lbs | Neodymium magnet with a magnet to steel disc pull of greater than 2.75 lbs | Neodymium magnet with a magnet to steel disc pull of greater than 2.75 lbs |
| Mount | | Mounted on roll stand | Mounted on roll stand | Mounted on tripod |
| Transceiver Component | Power requirements | USB powered and intended to be connected to the USB port of the LinkStation MINI display unit (5 V dc) | USB powered and intended to be connected to the USB port of the LinkStation MINI display unit (5 V dc) | USB powered and intended to be connected to the USB port of the LinkStation MINI display unit (5 V dc) |
| | Form of Wireless technology | Radio Frequency | Radio Frequency | Radio Frequency |
| | RF Frequencies | 401.05 - 405.55MHz | 401.05 – 405.55MHz | 401.05 – 405.55MHz |
| | Range | 2 m | 2 m | 2 m…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.